K Number
K093210
Date Cleared
2010-03-04

(142 days)

Product Code
Regulation Number
886.4670
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AquaLase® Capsule Wash Tip is indicated to assist in the removal of residual cortical material and lens epithelial cells.

Device Description

The AquaLase® Capsule Wash Tip is a sterile, single-use product. It is an accessory to the AquaLase® Handpiece (K980292), which produces the heated intraocular irrigating solution (Balanced Salt Solution or equivalent fluid) that washes the capsular surface. The AquaLase® Capsule Wash Tip and the AquaLase® Handpiece are accessories to the INFINITI® Vision System (K021566). The AquaLase® Capsule Wash Tip utilizes existing packaging configurations and has the same shelf life as existing AquaLase® tips.

To use the AquaLase® Capsule Wash Tip as intended, no modification to the existing INFINITI® System is required.

AI/ML Overview

The document describes the acceptance criteria and performance testing for the AquaLase® Capsule Wash Tip, an accessory to a phacoemulsification device.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria DescriptionReported Device Performance
Nozzle and Tube Retention (Minimum Retention Strength)Tips passed the minimum retention strength.
Passive Flow (Requirements)Tips passed the passive flow requirements.
Wrench/Tip Detachment (Force)Tips passed the wrench/tip detachment force.
Fluid Output Profile (Collimated Requirement)Tips passed the fluid output profile requirement of being collimated.
Handpiece Compatibility (Requirement)Tips passed the handpiece compatibility requirement.
Irrigation Free Flow (Minimum Rate)Tips passed the minimum irrigation free flow rate requirement.
Injection Fluid Active Flow (Average Pulse Flow Rate)Tips passed the injection fluid active flow.
Injection Pulse Volume (Maximum)Tips passed the maximum injection pulse volume requirement.
Pulse Fluid Exit Temperature (Maximum)Tips passed the maximum temperature requirement.
Leakage (Leak-free during use)Tips passed the requirement to remain leak free during use.
Particulates (Metallic Particulate Requirements)Tips passed the metallic particulate requirements.
Endurance (Injection Fluid Active Flow & Leakage)Tips passed the endurance requirement for Injection Fluid Active Flow and the endurance requirement for Leakage

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify a distinct "test set" in terms of a clinical study with human patients. The performance testing appears to be primarily non-clinical bench testing of the device itself. Therefore, sample sizes would refer to the number of tips or tests conducted on the device components. This information (e.g., number of tips tested for each criterion) is not provided in the summary.

Data provenance is not explicitly stated as country of origin, but given the submission is to the FDA, it's implied that the testing was conducted to meet US regulatory requirements. The testing is retrospective in the sense that it's conducted on the manufactured device prior to market introduction, but not in the context of analyzing previously collected patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not applicable as the "ground truth" for the non-clinical performance testing of this medical device is based on predefined engineering specifications, regulatory standards, and comparison to predicate devices, not on expert consensus from clinical cases. There is no mention of experts establishing ground truth for the device's functional performance.

4. Adjudication Method for the Test Set:

This information is not applicable as the testing described is objective, non-clinical performance testing against pre-defined specifications. There is no mention of subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document focuses on the non-clinical performance of the device itself and its equivalence to predicate devices, not on the clinical effectiveness or improvement of human readers with or without AI assistance. The device is a surgical accessory, not an AI software.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

This question is not applicable as the device is a physical surgical accessory (Capsule Wash Tip), not an algorithm or AI software. There is no "algorithm only" performance to evaluate.

7. The Type of Ground Truth Used:

The "ground truth" for this device's performance testing is based on:

  • Engineering specifications and requirements: Such as minimum retention strength, passive flow requirements, collimation of fluid output profile, maximum temperature, and leak-free operation.
  • Regulatory standards: Adherence to standards like ISO 10993 (biocompatibility) and ISO 11135-1 (sterilization).
  • Comparison to predicate devices: "Salient tip performance characteristics such as fluid output volume and fluid output temperature are analogous to the 1.1mm Liquefaction Tip."

8. The Sample Size for the Training Set:

This information is not applicable. The device is a physical medical device, not an AI or software algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as there is no "training set" for this type of medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date "MAR - 4 2010" at the top. Below the date is the word "Alcon" in a bold, sans-serif font. The word "Alcon" is the primary focus of the image, with the date appearing as a secondary element.

K093210

PREMARKET NOTIFICATION 510(k) SUMMARY

This summary document is being prepared in accordance with section 21 CFR 807.92(c).

The submitter of the 510(k) is:

Martin A. Kaufman Director, Regulatory Affairs Alcon Research, Ltd. 15800 Alton Parkway Irvine, CA 92618 Phone: (949) 753-6250 Fax: (949) 753-6237

Device Subject to this 510(k):

Trade Name:AquaLase® Capsule Wash Tip
Common Name:Accessory to a Phacoemulsification Device
Classification Name:Accessory to a Phacofragmentation Device (per 21CFR 886.4670)

1. Predicate Devices

The legally marketed devices(s) to which we are claiming equivalence to are:

510(k) NumberDevice
K980292/K021566AquaLase ® 1.1 mm Liquefaction Tip
K911808/K910245ALCON® Silicone I/A Tip
K030957Milvella Perfect capsule

{1}------------------------------------------------

2. Device Description

ﻣﺮ ﻣﺮ ﻣﺮﺍﺟﻊ

The AquaLase® Capsule Wash Tip is a sterile, single-use product. It is an accessory to the AquaLase® Handpiece (K980292), which produces the heated intraocular irrigating solution (Balanced Salt Solution or equivalent fluid) that washes the capsular surface. The AquaLase® Capsule Wash Tip and the AquaLase® Handpiece are accessories to the INFINITI® Vision System (K021566). The AquaLase® Capsule Wash Tip utilizes existing packaging configurations and has the same shelf life as existing AquaLase® tips.

To use the AquaLase® Capsule Wash Tip as intended, no modification to the existing INFINITI® System is required.

3. Indications for Use

The AquaLase® Capsule Wash Tip is indicated to assist in the removal of residual cortical material and lens epithelial cells.

Brief Summary of Non-clinical test and Results 4.

Biocompatibility evaluations of materials coming in contact with the patient fluid path have been performed to the following standards:

Standard #Title
10993-1: 2003 AAMI/ANSI/ISOBiological Evaluation of Medical Devices --Part 1: Evaluation and testing
10993-5: 1999 AAMI/ANSI/ISOBiological Evaluation of medical devices --Part 5: Tests for In Vitro cytotoxicity
10993-7:2008 AAMI/ANSI/ISOBiological Evaluation of Medical Devices --Part 7: Ethylene Oxide Sterilization Residuals
10993-10:2002/A1:2006 AAMI/ANSI/ISOBiological Evaluation of Medical Devices --Part 10: Tests for irritation and delayed-typehypersensitivity - Including A1:2006
10993-11:2006 AAMI/ANSI/ISOBiological Evaluation of Medical Devices --Part 11: Tests for systemic toxicity
10993-12:2007 AAMI/ANSI/ISOBiological Evaluation of Medical Devices --Part 12: Sample Preparation and ReferenceMaterials

{2}------------------------------------------------

The AquaLase® Capsule Wash tip is provided sterile and intended for single use only. This product is Ethylene Oxide sterilized and the process has been validated to a SAL of 10t per FDA Recognized Consensus Standard - ""ISO 11135-1:2007, Sterilization of health care products — Ethylene oxide — Requirements for development, validation and routine control of a sterilization process for medical devices."

Technological characteristics affecting clinical performance are similar to those of predicate devices previously listed. The AquaLase® Capsule Wash Tip has been developed and will be manufactured in compliance with section 21 CFR 820 and ISO 14971:2003. Non-clinical testing has demonstrated that the functional requirements have been met and that the device is equivalent to the predicate devices.

Summary of Performance Testing 5.

The AquaLase® Capsule Wash Tip has been developed and manufactured in compliance with recognized international consensus standards and FDA requlations/ guidance documents. Performance testing demonstrates that the AquaLase® Capsule Wash Tip operates as intended on the INFINITI® Vision System and is substantially equivalent to the predicate device. Salient tip performance characteristics such as fluid output volume and fluid output temperature are analogous to the 1.1mm Liquefaction Tip. The fluid output profile is modified from the 1.1mm Liquefaction Tip; the Capsule Wash Tip's profile is to better serve the intended use.

Performance test results are provided in the following table.

{3}------------------------------------------------

DescriptionResults
Nozzle and Tube RetentionTips passed the minimum retentionstrength.
Passive FlowTips passed the passive flowrequirements.
Wrench/Tip DetachmentTips passed the wrench/tip detachmentforce.
Fluid Output ProfileTips passed the fluid output profilerequirement of being collimated.
Handpiece CompatibilityTips passed the handpiece compatibilityrequirement.
Irrigation Free FlowTips passed the minimum irrigation freeflow rate requirement.
Injection Fluid Active Flow (AveragePulse Flow Rate)Tips passed the injection fluid active flow.
Injection Pulse VolumeTips passed the maximum injection pulsevolume requirement.
Pulse Fluid Exit TemperatureTips passed the maximum temperaturerequirement.
LeakageTips passed the requirement to remainleak free during use.
ParticulatesTips passed the metallic particulaterequirements.
EnduranceTips passed the endurance requirementfor Injection Fluid Active Flow and theendurance requirement for Leakage

Results Summary for AquaLase® Capsule Wash Tip (CW Tip)

For this table, the predicate device is the 1.1mm AquaLase® Liquefaction Tip (K980292/ K021566).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping, curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

MAR - 4 2010

Alcon Research, Ltd. c/o Mr. Martin A. Kaufman, RAC Director, Regulatory Affairs 15800 Alton Parkway Irvine, CA 92816-3818

Re: K093210

Trade/Device Name:AquaLase® Capsule Wash Tip Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System (Accessory) Regulatory Class: Class II Product Code: HQC Dated: January 28, 2010 Received: January 29, 2010

Dear Mr. Kaufman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Martin A. Kaufman, RAC

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Kesia Alexander for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

4. Indications for Use Statement

The Indications for Use statement is provided here and is also included in Attachment A. The AquaLase® Capsule Wash Tip is intended for prescription use only.

510(k) Number (if known): K 0932/0 Device Name: AquaLase® Capsule Wash Tip Indications for Use:

The AquaLase® Capsule Wash Tip is indicated to assist in the removal of residual cortical material and lens epithelial cells.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Callaway
(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K093210

Page 8 of 51

§ 886.4670 Phacofragmentation system.

(a)
Identification. A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.(b)
Classification. Class II.