K Number
K093143
Date Cleared
2010-01-28

(115 days)

Product Code
Regulation Number
862.3050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Breath Alcohol Detector is for the semi-quantitative rapid detection of the presence of alcohol in the exhaled breath. The Breath Alcohol Detector indicates relative Blood Alcohol Concentration (BAC) at 0.02%, 0.04%, 0.05%, 0.06%, 0.08% or 0.10% cut-off levels. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.

Device Description

The Breath Alcohol Detector is a visual semi-quantitative test for the detection of alcohol in the exhaled breath. The Breath Alcohol Detector consists of a plastic tube, two plastic plugs, a blow bag (optional), a tube label and a glass vial encased with reaction crystals. The crystals employ a solid-phase chemistry system based on chemically chromogenic reaction. Alcohol, if present in the exhaled breath, reacts with the chemically coated crystals and produces a color change. This color change is proportional to the concentration of alcohol in the breath, which is an approximation of relative Blood Alcohol Concentration (BAC). The Breath Alcohol Detector is available with or without blow bags in six cut-off levels: 0.02%, 0.04%, 0.05%, 0.06%, 0.08% or 0.10% cut-off levels.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Mission® Breath Alcohol Detector, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria in a table format. However, it implies that the device had to demonstrate properties equivalent to the predicate device and meet certain performance characteristics. It indicates that the device achieved "comparable readings" to a DOT/NHTSA approved evidential breath test device.

Acceptance Criteria (Implied)Reported Device Performance
Accuracy / Comparability to Evidential Breath Test Device: Device readings should be comparable to a recognized standard."Study results indicate that nonprofessional, inexperienced laypersons were able to obtain comparable readings when using the Mission® Breath Alcohol Detector as compared to the results obtained by the trained technicians [using] an evidentiary breath test, Alco-Sensor IV, manufactured by Intoximeters Inc. which is a DOT/NHTSA approved device." Additionally, "The laboratory testing and clinical study results demonstrate that the Mission® Breath Alcohol Detector is safe, accurate and easy-to-use."
Ease of Use for Laypersons: Non-professional users should be able to operate the device effectively."nonprofessional, inexperienced laypersons were able to obtain comparable readings" and "the participating laypersons were questioned and responded as satisfied with the ease of operation by following the Instructions for Use and the overall performance of the Mission® Breath Alcohol Detector."
Substantial Equivalence to Predicate Device: The device should be substantially equivalent to the Akers Bioscience BreathScan Alcohol Detector.The entire submission and the FDA's final determination confirm substantial equivalence. The document includes a comparison table highlighting highly similar features and minor differences (measurement range and blow bag option). The conclusion explicitly states: "It also demonstrates that the Mission® Breath Alcohol Detector is substantially equivalent to the Akers Biosciences BreathScan Alcohol Detector, currently sold on the U.S. market."
Precision: The device should provide consistent results.Evaluated through "precision" studies. No specific metrics or results are provided, but its inclusion implies that satisfactory precision was demonstrated.
Analytical Specificity (Blank, Cigarette Smoke, Volatile Substances): The device should not react to common interfering substances.Evaluated through "analytical specificity-blank reading, analytical specificity-cigarette smoke, analytical specificity-volatile substances" studies. No specific results are provided, but its inclusion implies that satisfactory specificity was demonstrated.
Environmental Robustness (Temperature, Vibration, Lighting): The device should perform reliably under varying environmental conditions.Evaluated through "temperature flexibility, vibration effect, lighting effect" studies. No specific results are provided, but its inclusion implies that satisfactory robustness was demonstrated.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: The document does not specify the exact number of individuals (laypersons and trained laboratory technicians) who participated in the clinical study. It states that "Clinical studies were conducted with laypersons and trained laboratory technicians."
  • Data Provenance: The data was gathered from a prospective clinical study specifically conducted for this device. The country of origin for the data is not explicitly stated, but the submitting company (ACON Laboratories, Inc.) is based in San Diego, California, USA, and the predicate device is from New Jersey, USA, suggesting the study was likely conducted in the USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: Not explicitly stated. The ground truth was established by "trained laboratory technicians" using an "evidentiary breath test, Alco-Sensor IV."
  • Qualifications of Experts: The experts were described as "trained laboratory technicians." Their specific experience (e.g., years of experience) is not provided, but their use of a DOT/NHTSA approved device implies they are qualified operators of such equipment for accurate breath alcohol measurement.

4. Adjudication Method for the Test Set:

  • The document does not describe an adjudication method in the typical sense of expert review for discrepancies. Instead, the study directly compared the device's readings obtained by laypersons to the results obtained by trained technicians using a reference device (Alco-Sensor IV). The 'ground truth' was assumed to be the Alco-Sensor IV reading.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, a formal MRMC comparative effectiveness study was not reported. The study focused on comparing the performance of the device (operated by laypersons) against a reference device (operated by trained technicians), not on comparing multiple human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • This device is a physical, chromogenic reaction test. It actively involves a human user blowing into it and visually interpreting the results. Therefore, a "standalone algorithm only" performance evaluation is not applicable and was not performed. The performance of the device without a human "in the loop" doesn't make sense for this type of product.

7. The Type of Ground Truth Used:

  • The ground truth was established by an evidentiary breath test (Alco-Sensor IV) which is a DOT/NHTSA approved device. This represents a highly reliable, accepted method for measuring breath alcohol concentration, essentially acting as the gold standard for comparison in this context.

8. The Sample Size for the Training Set:

  • The document does not provide information regarding a separate "training set" or its sample size. This type of chemical-based, visual read device typically undergoes analytical validation and clinical validation, rather than a machine learning model's training and testing split. The "analytical performance characteristics" and "clinical study" sections describe the validation process.

9. How the Ground Truth for the Training Set Was Established:

  • Given that no machine learning training set is mentioned or implied, this question is not applicable. The "ground truth" for the overall validation was established by the Alco-Sensor IV as described above.

{0}------------------------------------------------

K093143, ACON Reply, Dated 11-Dec-09 Attachment F, 510(k) Summary for Breath Alcohol Detector

5. 510(k) SUMMARY

JAN 2 8 2010

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The Assigned 510(k) number is _ K 09 31 4 3

Submitter's Identification:

ACON Laboratories, Inc. 10125 Mesa Rim Road San Diego, California 92121

Tel.: 858-875-8019 Fax: 858-875-8099

Date Prepared: October 2009

Contact Person:

Richard Lenart Regulatory Affairs Manager

Proprietary Name of the Device:

Mission® Breath Alcohol Detector

Common Name:

Breath-alcohol test system

Regulation Section and Classification:

Class I, Breath-alcohol test system 21 CFR § 862.3050

F-1 of F-4

{1}------------------------------------------------

Product Code:

DJZ Devices, Breath trapping, alcohol

Medical Specialty:

Toxicology

Predicate Device:

BreathScan Alcohol Detector Akers Biosciences Inc., Thorofare, New Jersey, USA 08086. 510(k) Number: K060761

Description:

The Breath Alcohol Detector is a visual semi-quantitative test for the detection of alcohol in the exhaled breath. The Breath Alcohol Detector consists of a plastic tube, two plastic plugs, a blow bag (optional), a tube label and a glass vial encased with reaction crystals. The crystals employ a solid-phase chemistry system based on chemically chromogenic reaction. Alcohol, if present in the exhaled breath, reacts with the chemically coated crystals and produces a color change. This color change is proportional to the concentration of alcohol in the breath, which is an approximation of relative Blood Alcohol Concentration (BAC). The Breath Alcohol Detector is available with or without blow bags in six cut-off levels: 0.02%, 0.04%, 0.05%, 0.06%, 0.08% or 0.10% cut-off levels.

Intended Use:

The Breath Alcohol Detector is for the semi-quantitative rapid detection of the presence of alcohol in the exhaled breath. The Breath Alcohol Detector indicates relative Blood Alcohol Concentration (BAC) at 0.02%, 0.04%, 0.05%, 0.06%, 0.08% or 0.10% cut-off levels.

{2}------------------------------------------------

K093143, ACON Reply, Dated 11-Dec-09 Attachment F, 510(k) Summary for Breath Alcohol Detector

.

Technological Characteristics:

FeatureSpecifications
MethodologyChromogenic reaction
SpecimenExhaled breath
Measurement Range0.02%, 0.04%, 0.05%, 0.06, 0.08% and 0.10%
Measuring UnitsBAC %
Reading Time2 minutes
Reading Stability5 minutes
Storage Temperature2-30°C (36-86°F)
Shelf Life3 years
Dimensions0.94 x 8.0 cm (0.37 x 3.15 inches)
Weight2.0 g (0.07 oz)

.

Comparison to Predicate Devices:

The Mission® Breath Alcohol Detector is substantially equivalent to the Akers Biosciences BreathScan Alcohol Detector.

FeatureMission® Breath AlcoholDetectorAkers BioscienceBreathScan AlcoholDetector
Similarities
Intended UseDetect presence of alcohol inexhaled breathSame
Target PopulationOver the counterSame
CalibrationNone requiredSame
MethodologyChromogenic reactionSame
Anatomical SiteMouthSame
Test SampleExhaled breathSame
ResultSemi-quantitativeSame
InterpretationVisual color changeSame
Measuring UnitsBAC %Same
MouthpieceNone requiredSame
Blowing Time12 secondsSame
Dimensions0.94 x 8.0 cm (0.37 x 3.15inches)Same
Weight2.0 g (0.07 oz)Same
Differences
Measurement RangeSeparate devices available atdifferent cut-off levels:0.02%, 0.04%, 0.05%, 0.06%,0.08% and 0.10%Separate devices available atdifferent cut-off levels:0.02%, 0.04%, 0.05%, and0.08%
Blow BagMay be used with or withoutblow bagNone

{3}------------------------------------------------

Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Guidance documents included the "NHTSA/DOT Highway Safety Programs; Model Specifications for Screening Devices to Measure Alcohol in Bodily Fluids, (Federal Register/Vol.59, No.147, August 2, 1994/Notices/39382)," and NHTSA/DOT Highway Safety Programs; Model Specifications for Screening Devices to Measure Alcohol in Bodily Fluids, (Federal Register/Vol.73, No.62, March 3, 2008/Notices)"

Laboratory Testing:

The performance characteristics of the Mission® Breath Alcohol Detector were evaluated by the following studies: precision, analytical specificity-blank reading, analytical specificity-cigarette smoke, analytical specificity-volatile substances, temperature flexibility, vibration effect, lighting effect and device comparison study.

Discussion of Clinical Tests Performed:

Clinical studies were conducted with laypersons and trained laboratory technicians using the Mission Breath Alcohol Detector. The study data were presented evaluating the accuracy of the Mission® Breath Alcohol Detector compared to an evidentiary breath test, Alco-Sensor IV, manufactured by Intoximeters Inc. which is a DOT/NHTSA approved device (Conforming Product List of Evidential Breath Alcohol Measurement Devices - FR/Vol 72, No 241/December 2007), per the ACON Clinical Study Protocol for the Breath Alcohol Detector. Study results indicate that nonprofessional, inexperienced laypersons were able to obtain comparable readings when using the Mission® Breath Alcohol Detector as compared to the results obtained by the trained technicians. In addition, the participating laypersons were questioned and responded as satisfied with the ease of operation by following the Instructions for Use and the overall performance of the Mission® Breath Alcohol Detector.

Conclusion:

The laboratory testing and clinical study results demonstrate that the Mission® Breath Alcohol Detector is safe, accurate and easy-to-use. It also demonstrates that the Mission® Breath Alcohol Detector is substantially equivalent to the Akers Biosciences BreathScan Alcohol Detector, currently sold on the U.S. market.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Acon Laboratories, Inc. c/o Mr. Richard Lenart Regulatory Affairs Manager 10125 Mesa Rim Road San Diego, CA 92121

Re: K093143

Trade Name: Mission® Breath Alcohol Detector Regulation Number: 21 CFR §862.3050 Regulation Name: Breath-Alcohol Test System Regulatory Class: Class I, reserved Product Codes: DJZ Dated: December 11, 2009 Received: December 15, 2009

JAN 2 8 2010

Dear Mr Lenart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Acts requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

CA

Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT 4.

510(k) Number (if known): K o 9 3 l 4 3

Mission® Breath Alcohol Detector Device Name:

Indications for Use:

The Breath Alcohol Detector is for the semi-quantitative rapid detection of the presence of alcohol in the exhaled breath. The Breath Alcohol Detector indicates relative Blood Alcoho! Concentration (BAC) at 0.02%, 0.04%, 0.05%, 0.06%, 0.08% or 0.10% cut-off levels. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1

E-1 of E-1

§ 862.3050 Breath-alcohol test system.

(a)
Identification. A breath-alcohol test system is a device intened to measure alcohol in the human breath. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.