K Number
K093142

Validate with FDA (Live)

Date Cleared
2010-02-04

(122 days)

Product Code
Regulation Number
882.5950
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Target Detachable Coils are intended for use in the treatment of intracranial aneurysms and other neuro and peripheral vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.

Target Coils are indicated for endovascular embolization of:
• Intracranial aneurysms
• Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
• Arterial and venous embolizations in the peripheral vasculature

Boston Scientific's InZone Detachment System is intended for use with all Boston Scientific Detachable Coils in the embolization of intracranial aneurysms and other vascular malformations of the neuro and peripheral vasculature.

Device Description

Boston Scientific Corporation's Target Detachable Coils are comprised of four coil types: Target Coil 360 STANDARD, Target Coil 360 SOFT, Target Coil 360 ULTRA and Target Coil HELICAL ULTRA. All Target Coils are stretch resistant coils. Target Coils incorporate a length of multi-strand material through the center of the coil designed to help resist stretching. Target Coils are designed for use with Boston Scientific's InZone™ Detachment System (sold separately).

Each Target Coil type consists of a platinum-tungsten alloy coil attached to a stainless steel delivery wire. For Target Coil 360 STANDARD, Target Coil 360 SOFT and Target Coil 360 ULTRA coils the distal end of the main coil is formed such that there is a smaller distal loop at the end of the main coil to facilitate placement of the coil. The diameter of the distal loop is 75% that of the rest of the main coil loops.

Boston Scientific's InZone Detachment System is intended for use with all Boston Scientific Detachable Coils in the embolization of intracranial aneurysms and other vascular malformations of the neuro and peripheral vasculature.

AI/ML Overview

The provided text describes Boston Scientific Corporation's Target Detachable Coils and InZone Detachment System. The submission is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information regarding acceptance criteria, specific device performance metrics, sample sizes for test/training sets, expert qualifications, and adjudication methods is not present in the provided document.

However, based on the available information, here's what can be inferred:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not specify quantitative acceptance criteria or detailed performance metrics. Instead, it states that "Verification testing has demonstrated Target Detachable Coils are substantially equivalent to the current legally marketed predicate devices" and similarly for the InZone Detachment System, electrical safety testing "has demonstrated the devices to be substantially equivalent to the current legally marketed predicate device."

The "performance" reported is that the devices are substantially equivalent to their respective predicate devices. This implies that their performance, in terms of safety and effectiveness for their intended use, is considered comparable to already approved devices.

2. Sample Sizes Used for the Test Set and Data Provenance:

The document states "Verification testing has demonstrated..." and "Verification testing of the InZone Detachment System, including electrical safety testing in accordance with applicable parts of the EN 60601-series of standards, has demonstrated..."

This indicates that pre-clinical verification testing was performed. However, specific sample sizes for these tests are not provided. The data provenance is also not specified, but given this is pre-clinical bench and electrical safety testing, it would largely be laboratory-generated data, not human patient data in the context of geographical origin or retrospective/prospective nature.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided and is generally not applicable to a 510(k) submission for this type of device, which relies heavily on bench testing and comparison to predicate devices, rather than human expert evaluation of clinical cases or images.

4. Adjudication Method for the Test Set:

This information is not provided and is not applicable in the context of the reported verification testing.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device (embolization coils and their detachment system) is a physical medical device, not an AI or imaging diagnostic tool that would typically involve human readers or AI assistance in interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is a physical medical device.

7. The Type of Ground Truth Used:

For the "verification testing," the ground truth implicitly used would be established engineering specifications, performance standards (like EN 60601 for electrical safety), and the known characteristics/performance of the predicate devices. For example, if testing tensile strength, the "ground truth" would be the specified strength range for safe performance.

8. The Sample Size for the Training Set:

This information is not provided and is not applicable as this is a physical medical device and not an AI/machine learning algorithm requiring a training set.

9. How the Ground Truth for the Training Set was Established:

This information is not provided and is not applicable for the same reason as point 8.

{0}------------------------------------------------

K093142

FEB - 4 2010

510(k) Summary Of Safety And Effectiveness

Summary DateOctober 1, 2009
Submitter Name andAddressBoston Scientific Corporation47900 Bayside ParkwayFremont, CA. 94538
Contact Person:Jim LeathleyRegulatory Affairs Project ManagerPhone: 510 440 7836Fax: 510 440 7752Email: leathlej@bsci.com
Trade Name:Target® Detachable CoilsInZone Detachment System
Common Name:Occlusion Coil, Vascular Occlusion Coil, Neurovascular OcclusionCoil
Power Supply
Classification Name:Target Detachable Coils are classed as vascular and neurovascularembolization devices under 21 CFR 870.3300 (KRD) and 21 CFR882.5950 (HCG), respectively, and are Class II devices (specialcontrols).
The special control for the devices is FDA's guidance document,Class II Special Controls Guidance Document: Vascular andNeurovascular Embolization Devices (issued 29 Dec 2004).
The InZone Detachment System is intended for use with all currentlymarketed Boston Scientific Detachable Coils in the embolization ofintracranial aneurysms and other vascular malformations of theneuro and peripheral vasculature.

ﻴﺔ - ﺗﻮ

{1}------------------------------------------------

Legally Marketed

Predicate Devices:

Reference (Clearance Date)Device
K002181 (11 Aug 2000)GDC 10-UltraSoft® Coils(introduction of UltraSoft coils)
K021494 (6 June 2002)GDC Power Supply and Detachable Coil Connecting Cables
K031049 (3 June 2003)Clearance of ISAT indication for all GDC devices
K042539 (19 Oct 2004)GDC 360 Detachable Coils(introduction of 360 shape coils)
K050700 (15 April 2005)Matrix2® Detachable Coils

Device Description:

Boston Scientific Corporation's Target Detachable Coils are comprised of four coil types: Target Coil 360 STANDARD, Target Coil 360 SOFT, Target Coil 360 ULTRA and Target Coil HELICAL ULTRA. All Target Coils are stretch resistant coils. Target Coils incorporate a length of multi-strand material through the center of the coil designed to help resist stretching. Target Coils are designed for use with Boston Scientific's InZone™ Detachment System (sold separately).

Each Target Coil type consists of a platinum-tungsten alloy coil attached to a stainless steel delivery wire. For Target Coil 360 STANDARD, Target Coil 360 SOFT and Target Coil 360 ULTRA coils the distal end of the main coil is formed such that there is a smaller distal loop at the end of the main coil to facilitate placement of the coil. The diameter of the distal loop is 75% that of the rest of the main coil loops.

Boston Scientific's InZone Detachment System is intended for use with all Boston Scientific Detachable Coils in the embolization of intracranial aneurysms and other vascular malformations of the neuro and peripheral vasculature.

. ﺍ

{2}------------------------------------------------

510(k) Summary Of Safety And Effectiveness (cont.)

Accessories:Target Detachable Coils are packaged within a flushing dispensercoil assembly. The dispenser coil is an accessory item with anattached flushport used to hydrate the coil prior to use.There are no accessories to the InZone Detachment System.
Indications for Use /Intended Use:Target Detachable Coils are intended for use in the treatment ofintracranial aneurysms and other neuro and peripheral vascularabnormalities such as arteriovenous malformations and arteriovenousfistulae.
Target Coils are indicated for endovascular embolization of:• Intracranial aneurysms• Other neurovascular abnormalities such as arteriovenousmalformations and arteriovenous fistulae• Arterial and venous embolizations in the peripheral vasculatureBoston Scientific's InZone Detachment System is intended for usewith all Boston Scientific Detachable Coils in the embolization ofintracranial aneurysms and other vascular malformations of theneuro and peripheral vasculature.
Comparison to Predicate Device:
Target Detachable Coils
Boston Scientific Corporation's Target Detachable Coils have thesame intended use/indications for use as the predicate devices.
Although the coils incorporate modifications in design, materials,packaging, and instructions for use, the modifications do not alter thefundamental scientific technology of the predicate devices.

Risk assessment of the modifications in the form of design and use failure modes and effects analysis (design and use FMEAs) has been conducted in accordance with EN ISO 14971 +A1:2003. Boston Scientific has determined the modifications to the predicate devices raise no new questions of safety or effectiveness.

Verification testing has demonstrated Target Detachable Coils are substantially equivalent to the current legally marketed predicate devices.

.

Page 3 of 4

{3}------------------------------------------------

Comparison to Predicate Device (cont.):

InZone Detachment System

Boston Scientific Corporation's InZone Detachment System has the same intended use and indications for use as the current legally marketed predicate device, Boston Scientific's Detachable Coil Power Supply cleared under premarket notification K021494 (cleared 6 June 2002).

Although the InZone Detachment System incorporates modifications in materials, firmware, packaging, and instructions for use, the modifications do not alter the fundamental scientific technology of the predicate device.

Risk assessment of the modifications, in the form of design and use failure modes and effects analysis (design and use FMEAs), has been conducted in accordance with EN ISO 14971 +A1:2003. Boston Scientific has determined the modifications to the predicate device raise no new questions of safety or effectiveness.

Verification testing of the InZone Detachment System, including electrical safety testing in accordance with applicable parts of the EN 60601-series of standards, has demonstrated the devices to be substantially equivalent to the current legally marketed predicate device.

Conclusion:

Because the subject modifications do not alter the intended use or indications for use of the predicate devices, or the fundamental scientific technology of the predicate devices; and because risk assessment of the modifications and successful verification testing raise no new questions of safety and effectivenss, Boston Scientific has determined the Target Detachable Coils and InZone Detachment System to be substantially equivalent to the current legally marketed predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

James Leathley Regulatory Affairs Project Manager Boston Scientific Neurovascular 47900 Bayside Parkway Fremont, CA 94538-6515

FEB - 4 2010

Re: K093142

Trade/Device Name: Target Detachable Coils and InZone Detachment System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: II Product Code: HCG and KRD Dated: January 22, 2010 Received: January 25, 2010

Dear Mr. Leathley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you active oper.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number: _ Ko93142_

Device Name: Target Detachable Coils

Indications for Use:

Target Detachable Coils are indicated for the endovascular embolization of:

  • · Intracranial aneurysms
  • · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae

  • · Arterial and venous embolizations in the peripheral vasculature
    Device Name: InZone Detachment System

Intended Use:

The InZone Detachment System is intended for use with all Boston Scientific Detachable Coils in the embolization of intracranial aneurysms and other vascular malformations of the newro and oeriohenal vasculature.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)


K 093142

Prescription Use (Per 21 CFR 801.109)X
OR. Over The Counter Use

KRISTEN BOWSHER

(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear,

Nose and Throat Devices

510(k) Number

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).