(288 days)
The Merits Model N Series Nebulizer Compressors are AC-powered air compressors intended to provide a source of compressed air for medical purposes for use in home health care. These devices are used in conjunction with a pneumatic nebulizers to produce fine aerosol mists of medication for respiratory therapy, for both children and adults suffering from respiratory disorders such as asthma, allergies, etc.
The Merits Model N Series Nebulizer Compressors consist of AC powered piston-type compressors that generates compressed air. They are small, lightweight and designed for convenience. The piston-type compressor is housed in a plastic cabinet with rubber bumpers. The motor driven piston compressor is connected with a power cord and a switch. The devices contain no microprocessors or other electronic components. They operate from 115VAC/60Hz. Each of them is supplied with tubing, a nebulizer kit and several replaceable filters.
Model N281 and N282 are identical in specifications and components. The only difference is they use different cabinets which are different in size and shape.
When using, the compressor is placed on a flat surface. Air delivery tubing and a nebulizer kit are connected to the outlet hose barb. The air passes the replaceable filter is compressed and delivered to the nebulizer kit through the tubing.
This 510(k) summary describes a Nebulizer Compressor, a device that provides a source of compressed air for pneumatic nebulizers. The context provided is a regulatory submission for a medical device trying to demonstrate "substantial equivalence" to a predicate device. This type of submission generally focuses on comparing the new device's specifications and performance to an already legally marketed device rather than proving novel effectiveness through a new study with strict "acceptance criteria" and "ground truth" as might be found for AI/ML or diagnostic devices.
Given the nature of the device (a nebulizer compressor) and the available document, the questions provided in the prompt are not directly applicable. This document is a 510(k) summary for a relatively simple mechanical device, not a sophisticated diagnostic or AI-driven system. Therefore, concepts like "ground truth," "expert adjudication," "MRMC studies," "training set," and "acceptance criteria" in the way they are typically applied to AI/ML or complex diagnostic devices are not detailed here.
However, I can interpret the document's content to extract what could be considered analogous to acceptance criteria and the "study" that proves it, within the framework of a substantial equivalence submission for this type of device.
Acceptance Criteria and Device Performance for Merits Model N Series Nebulizer Compressor
The "acceptance criteria" for this device are implicitly defined by its performance comparison to a legally marketed predicate device, the SAN UP S.A. Nebulizer Compressor Model: 3050 (K002468). The goal is to demonstrate "substantial equivalence," meaning its characteristics and performance are "basically the same" and "bring forth no new questions of safety and effectiveness."
1. Table of Acceptance Criteria (Implicit) and Reported Device Performance:
Acceptance Criteria Category (Implicit) | Specific Criteria (Implicitly based on Predicate) | Reported Device Performance | Study that Proves it Meets Criteria |
---|---|---|---|
AC-Powered Operation | Should operate from 115VAC/60Hz, similar to predicate. | Operates from 115VAC/60Hz. | Comparative Performance Test of Merits N281/N282 Nebulizer Compressor & SanUp Model 3050 Compressor |
Filter Material | Use the same or equivalent filter material as predicate. | Contains the same filter material. | Comparative Performance Test |
Environmental/Safety/EMC | Meet relevant environmental, safety, and electromagnetic compatibility (EMC) requirements similar to predicate. | Meets Environmental, Safety and EMC requirements. | IEC/UL 60601-1, Medical Electrical Equipment Part1: General Requirements for Safety; IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility |
Compressor Operating Pressure Range | Similar compressor operating pressure range as predicate. | Within similar compressor operating pressure range. | Comparative Performance Test |
Compressor Flow Range | Similar compressor flow range as predicate. | Within similar compressor flow range. | Comparative Performance Test |
Performance Characteristics | Performance characteristics are "basically the same" as the predicate. | Performance characteristics are "basically the same." | Comparative Performance Test |
Particle Size Analysis (Nebulization Efficacy) | Produce fine aerosol mists for respiratory therapy, comparable to predicate. | Data from Cascade Impaction Testing for Particle Size Analysis. (Specific values not provided but implied to be acceptable) | Cascade Impaction Testing for Particle Size Analysis |
Emission of Harmful Substances | No harmful emissions (VOCs, CO, O3, PM2.5) above acceptable levels. | Emissions (VOCs, CO, O3 & PM2.5) evaluated. (Specific values not provided but implied to be acceptable) | Emission of VOCs, CO, O3 & PM2.5 |
Intended Use | Aligns with the predicate's intended use for respiratory therapy in home health care. | Intended use matches description for home health care, pneumatic nebulizers, respiratory therapy for children/adults with asthma, allergies, etc. | (Implicit, based on device design and intended use statement) |
The study that proves the device meets the acceptance criteria is detailed in Section 9.10, "Performance Data." This section lists various tests performed to demonstrate the device's safety and performance characteristics are substantially equivalent to the predicate.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for the Test Set: Not explicitly stated. For a mechanical device like a nebulizer compressor, "sample size" usually refers to the number of units tested. It's common in such submissions to test a limited number of production units (e.g., 3-5 units) to confirm specifications. The document refers to "Merits N281/N282 Nebulizer Compressor," implying at least one of each model was tested, but does not specify the number of individual units.
- Data Provenance (Country of origin, retrospective/prospective): Not specified. The submitter is Merits Health Products Co., LTD. in Taichung, Taiwan R.O.C. It is highly probable that the tests were conducted at their facilities or by a contracted lab, likely in Taiwan or a region that adheres to international testing standards (e.g., IEC/UL). These would be prospective tests performed on the new device designs.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- This concept is not applicable here. For a nebulizer compressor, "ground truth" is established by direct measurement of physical properties (e.g., pressure, flow, electrical safety, particle size) against established engineering standards and the predicate device's performance, not by expert consensus on interpretations of data (like medical images).
- The
IEC/UL 60601-1
andIEC 60601-1-2
standards are objective engineering and safety standards.Cascade Impaction Testing
andEmission Testing
also rely on objective laboratory measurements.
4. Adjudication Method for the Test Set:
- Not applicable. As described above, the evaluation relies on objective measurements against engineering standards and comparison to the predicate device's physical specifications and performance, not on subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, an MRMC study was not done. MRMC studies are typically performed for diagnostic devices where human readers (e.g., radiologists) interpret cases (e.g., medical images) and their performance is compared with and without AI assistance. This device is a mechanical air compressor, not a diagnostic tool or an AI-driven system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device does not have an "algorithm" in the sense of a standalone AI/ML component that performs a function subject to such evaluation. It is a mechanical device.
7. The Type of Ground Truth Used:
- For this device, the "ground truth" is objective physical and performance measurements obtained through standardized engineering and safety tests (e.g., IEC/UL standards, comparative performance tests for pressure/flow, cascade impaction for particle size, emission testing). The "truth" is that the device's measured parameters meet safety standards and are "basically the same" as those of the predicate device.
8. The Sample Size for the Training Set:
- Not applicable. This device does not use an "algorithm" based on AI/ML that requires a training set. Its design and operation are based on established mechanical engineering principles.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there is no training set for an AI/ML algorithm.
§ 868.5630 Nebulizer.
(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).