K Number
K092847
Date Cleared
2010-03-11

(176 days)

Product Code
Regulation Number
868.5895
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 840 Ventilator System with Expanded NeoMode Option provides continuous ventilation to patients requiring respiratory support. The 840 Ventilator System with Expanded NeoMode Option, is intended for patients with an Ideal Body Weight (IBW) as low as 0.3 kg and for use in a wide variety of clinical conditions.

The 840 Ventilator System with Expanded NeoMode Option is intended for use in hospitals and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.

Device Description

The PB 840 Ventilator is a dual-microprocessor-based, touch-screen controlled, critical care ventilator intended to provide continuous ventilation for neonate to adult patients (with expanded NeoMode Option) or infant to adult patients (with no Option) who require either invasive ventilation or non-invasive ventilation.

The 840 Ventilator Expanded Neomode Option includes the software enhancements that lower the ideal body weight (IBW) from 0.5kg to 0.3kg and provides the user an option to lower the tidal volume to 2mL. The ventilator determines values for operational variables and allowable settings based on breathing circuit type and IBW. The software controls prevent inadvertent mismatching of patient size and breathing circuit type.

The 840 Ventilator Expanded Neomode Option is available as an integrated part of the 840 Ventilator or as separate software upgrade kit.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the 840 Ventilator System with Expanded NeoMode Option:

It's important to note that this document is a 510(k) summary for a ventilator, not an AI-powered diagnostic device. Therefore, many of the typical questions regarding AI device studies (like MRMC, number of experts for ground truth, sample sizes for test/training sets in the context of image analysis or similar AI tasks) are not applicable here. The "device" in this context is a medical hardware and software system, and the "study" refers to engineering verification and validation testing, not clinical trials in the way AI diagnostics are evaluated.

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Software validation for expanded ranges of IBW (down to 0.3kg) and Tidal Volume (down to 2mL) in the Patient Settings subsystem, in accordance with "General Principles of Software Validation, Final Guidance for Industry and FDA Staff" (2002) and "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (2005)."Test results met the required acceptance criteria."
Verification of 2mL breath delivery performance in accordance with applicable sections of the FDA Draft Reviewer Guidance for Ventilators."The 840 ventilator 2mL breath delivery performance was verified in accordance with applicable sections of the FDA Draft Reviewer Guidance for Ventilators."
Control systems performance verification tests to ensure breath delivery response and accuracy met Product and Controls specification requirements."Control systems performance verification tests verified the 840 ventilator breath delivery response and accuracy met Product and Controls specification requirements."
Ensure no new hazards were introduced compared to predicate devices due to software changes."There are no new hazards introduced by the 840 Ventilator System with Expanded Neomode Option as compared with the predicate devices." (This is a conclusion from the absence of new hazards, supported by the nature of the changes being software-only and the testing performed).
Continued safety and effectiveness for intended use."We conclude the results of this testing support the changes to the indications for use and support continued safety and effectiveness of the device."

Study Details (as applicable to a medical device modification, not an AI study)

  1. Sample size used for the test set and the data provenance:

    • The document does not specify a sample size for the test set in terms of patient data or clinical cases. The testing described is primarily engineering verification and validation of software and hardware performance against specifications. These tests would involve various simulations, bench testing, and potentially some animal or human user testing, but not a "test set" in the sense of a dataset for an AI algorithm.
    • Data provenance: Not applicable in the context of a typical "AI study." The context here is product development and regulatory submission for a physical medical device with software components.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable for this type of device submission. Ground truth for a ventilator's performance is established by engineering specifications, regulatory standards (like FDA guidances for ventilators), and physiological principles, not by expert consensus on a test set of data points in the same way it would be for an AI diagnostic. The "experts" would be the engineers and compliance officers designing and testing the device, ensuring it meets these established criteria.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable. Adjudication methods like 2+1 are typically used in clinical studies or for establishing ground truth from multiple human readers in an AI development context. For a ventilator, performance is objectively measured against predefined engineering and regulatory specifications.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is specific to AI systems that assist human readers (e.g., radiologists interpreting images). The 840 Ventilator is a life-support device, and its expanded NeoMode option is a software enhancement to its functionality, not an AI-assisted diagnostic tool.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device's core functionality is a "standalone" system in that it delivers ventilation based on its internal algorithms and control systems. However, this is not an "AI algorithm only" in the sense of an AI diagnostic that might be compared to human performance. The ventilator's performance was validated, which inherently evaluates its "standalone" function in delivering breaths.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance is based on engineering specifications, established physiological parameters for ventilation, and regulatory guidance documents (e.g., "FDA Draft Reviewer Guidance for Ventilators"). These define the acceptable ranges and accuracy for breath delivery, tidal volume, and other operational variables.
  7. The sample size for the training set:

    • This is not applicable as the document describes a software upgrade for an existing medical device, not the development of a machine learning model where a "training set" would be used to teach an algorithm. The software was developed and then validated against specifications.
  8. How the ground truth for the training set was established:

    • Not applicable for the reasons stated above.

{0}------------------------------------------------

K092847 Page 1 of 3

Image /page/0/Picture/1 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white rectangle in the middle and some other design elements. The word is written in all capital letters and is the main focus of the image. The image is simple and clear, with a focus on the company name.

MAR 1 1 2010

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

A. SUBMITTER INFORMATION:

Submitter's Name:

Covidien, LP, formerly Nellcor Puritan-Bennett Corporation

Address:

6135 Gunbarrel Avenue Boulder, CO 80301

Contact: Phone: Fax:

Jean Simon 303-305-2435 303-305-2212

Date of Preparation:

February 26, 2010

B. DEVICE NAME:

Trade Name(s):

Common/Usual Name: Continuous Ventilator Classification Names: CFR Reference: Product Code:

840 Ventilator System with Expanded NeoMode Option Class II; Anesthesiology 8868.5895 СВК

C. PREDICATE DEVICE:

510(k) # / Name of Predicate Device(s):

K001646 / PB 840 Ventilator with Neomode Option

K081837 / Cardinal Health / Viasys AVEA Comprehensive Ventilator

D. DEVICE DESCRIPTION:

The PB 840 Ventilator is a dual-microprocessor-based, touch-screen controlled, critical care ventilator intended to provide continuous

{1}------------------------------------------------

ventilation for neonate to adult patients (with expanded NeoMode Option) or infant to adult patients (with no Option) who require either invasive ventilation or non-invasive ventilation.

The 840 Ventilator Expanded Neomode Option includes the software enhancements that lower the ideal body weight (IBW) from 0.5kg to 0.3kg and provides the user an option to lower the tidal volume to 2mL. The ventilator determines values for operational variables and allowable settings based on breathing circuit type and IBW. The software controls prevent inadvertent mismatching of patient size and breathing circuit type.

The 840 Ventilator Expanded Neomode Option is available as an integrated part of the 840 Ventilator or as separate software upgrade kit.

E. INDICATIONS FOR USE:

The 840 Ventilator System with Expanded NeoMode Option provides continuous ventilation to patients requiring respiratory support. The 840 Ventilator System with Expanded NeoMode Option, is intended for patients with an Ideal Body Weight (IBW) as low as 0.3 kg and for use in a wide variety of clinical conditions.

The 840 Ventilator System with Expanded NeoMode Option is intended for use in hospitals and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.

F. TECHNOLOGICAL CHARACTERISTICS SUMMARY & SUBSTANTIAL EQUIVALENCE STATEMENT:

The subject device, the 840 Ventilator System with Expanded NeoMode Option, has the same intended use, general design and fundamental scientific technology as the predicate devices (K001646 and K081837).

The 840 Ventilator System with Expanded NeoMode Option uses technology substantially equivalent to the CBK Family of Systems (K001646 and K081837). There are no new hazards introduced by the 840 Ventilator System with Expanded NeoMode Option as compared with the predicate devices.

{2}------------------------------------------------

PERFORMANCE DATA SUMMARY: G.

As the changes to the device were made to the software and graphic user interface only and no changes were made to the hardware or firmware, repeat electrical safety testing (IEC testing) was not warranted. Electrical safety was previously established in the PB 840 Ventilator predicate device.

The PB 840 ventilator software change had no changes to patient contacting materials. Therefore, new biocompatibility testing was not warranted.

Performance testing was conducted to ensure that the specifications of the 840 Ventilator System with Expanded NeoMode Option were met thereby demonstrating the safety and effectiveness of the subject device for its intended use.

Software validation evaluating the software changes in the 840 ventilator's Patient Settings subsystem for the expanded ranges for IBW and Tidal Volume was completed in accordance with the Guidance entitled, "General Principles of Software Validation, Final Guidance for Industry and FDA Staff', issued on January 11, 2002 and "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued on May 11, 2005. Test results met the required acceptance criteria.

The 840 ventilator 2mL breath delivery performance was verified in accordance with applicable sections of the FDA Draft Reviewer Guidance for Ventilators. Control systems performance verification tests verified the 840 ventilator breath delivery response and accuracy met Product and Controls specification requirements.

We conclude the results of this testing support the changes to the indications for use and support continued safety and effectiveness of the device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with three curved lines representing its wings or body. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

.

Ms. Jean Simon Senior Director, Regulatory Affairs Covidien, Formerly Nellcor Puritan Bennett, Incorporated 6135 Gunbarrel Avenue Boulder, Colorado 80301

MAR 1 1 2010

Re: K092847

Trade/Device Name: 840 Ventilator System with Expanded Neomode Option Regulation Number: 21CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: March 1, 2010 Received: March 8, 2010

Dear Ms. Simon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

Device Name: 840 Ventilator System with Expanded Neomode Option

Indications for Use:

The 840 Ventilator System with Expanded NeoMode Option provides continuous ventilation to patients requiring respiratory support. The 840 Ventilator System with Expanded NeoMode Option, is intended for patients with an Ideal Body Weight (IBW) as low as 0.3 kg and for use in a wide variety of clinical conditions.

The 840 Ventilator System with Expanded NeoMode Option is intended for use in hospitals and hospital-type facilities. It may be used during hospital and hospitaltype facility transport provided that electrical power and compressed gas are supplied.

L. Schutter

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: KC092847

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).