K Number
K092804

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-10-08

(27 days)

Product Code
Regulation Number
892.5700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MicroSelectron HDR Version 2 is intended to enable an operator to apply, by remote control, a radionuclide source into the body (including Interstitial, Intracavitary, Intraluminal, Bronchial, Endovascular and Intra-operative) or to the surface of the body for radiation therapy.

Device Description

The microSelectron-HDR is a remote afterloading system for high dose rate brachytherapy treatment using a single iridium-192 radioactive source. The MicroSelectron-HDR delivers a radiation dose distribution conforming to treatment data, r he Miorooolouren . Alle workstation or imported from a treatment planning system. The modifications to the cleared device K953946 are: - Increase of maximum source strength for treatment of patients from 10 Ci (aprox. 40.000 . μGy.m²/h) to 12Ci (aprox. 48.000 μGy.m²/h).

AI/ML Overview

The provided text is a 510(k) summary for the MicroSelectron HDR Version 2, a remote afterloading system for high dose rate brachytherapy. It details the device's technical specifications, intended use, and its substantial equivalence to previously cleared devices. However, this document does not contain information on acceptance criteria, device performance studies, sample sizes, expert qualifications, ground truth establishment, or any details about multi-reader multi-case studies or standalone algorithm performance.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and the reported device performance
  2. Sample sized used for the test set and the data provenance
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
  4. Adjudication method for the test set
  5. MRMC comparative effectiveness study results (effect size)
  6. Standalone (algorithm only) performance
  7. The type of ground truth used
  8. The sample size for the training set
  9. How the ground truth for the training set was established

The document primarily focuses on regulatory approval based on substantial equivalence to an existing device, rather than detailed performance studies against specific acceptance criteria.

{0}------------------------------------------------

Korssoy

NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 +31 318 550485 Fax

Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Special 510(k) section

OCT - 8 2009

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

as required by section 807.92(c)

Submitter of 510(k):

Company name: Registration number: Address:

Phone: Fax: Correspondent:

Nucletron Corporation 1121753 8671 Robert Fulton Drive Columbia, MD 21046 410-312-4100 410-312-4197 Elaina Colby Manager Quality Assurance & Regulatory Affairs

Modified Device Name:

Trade/Proprietary Name:MicroSelectron HDR Version 2
Common/Usual Name:Afterloader
Classification Name:Remote controlled radionuclide applicator system
Classification:21Cfr892.5700 Class II
Product CodeJAQ

Legally Marketed Device(s)

Our device is substantially equivalent to the legally marketed predicate device cited in the table below:

. Manufacturer Same Calif. Device La Device La 200 10 510 K) 1
Nucletron BVComments of the control of the later the large of the comments of the comments of the comments of the contribution of the comments of the contribution of the contribution ofMicroSelectron HDR Version 2K953946

Description:

The microSelectron-HDR is a remote afterloading system for high dose rate brachytherapy treatment using a single iridium-192 radioactive source.

{1}------------------------------------------------

The MicroSelectron-HDR delivers a radiation dose distribution conforming to treatment data, r he Miorooolouren . Alle workstation or imported from a treatment planning system.

The modifications to the cleared device K953946 are:

  • Increase of maximum source strength for treatment of patients from 10 Ci (aprox. 40.000 . μGy.m²/h) to 12Ci (aprox. 48.000 μGy.m²/h).

Intended use:

The modified device has the same intended use as the legally marketed predicate device cited:

The MicroSelectron HDR Version 2 is intended to enable an operator to apply, by remote control, a radionuclide source into the body (including Interstitial, Intracavitary, Intraluminal, Bronchial, Endovascular and Intra-operative) or to the surface of the body for radiation therapy.

Summary of technological considerations:

MicroSelectron HDR Version 2 is substantially equivalent to the previously cleared device (K953946) and to the microSelectron_V3 (K061354).

Name: John Lapre

Aug 28, 2009
Date

Title: Vice President R&D Nucletron BAV. Veenendaal, The Netherlands

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms. Elaina M. Colby Manager, Quality Assurance & Regulatory Affairs Nucletron Corporation 8671 Robert Fulton Drive COLUMBIA MD 21046-2133

OCT = 8 2009

Re: K092804

Trade/Device Name: MicroSelectron HDR Version 2 Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radio-nuclide applicator system Regulatory Class: II Product Code: JAO Dated: September 10, 2009 Received: September 11, 2009

Dear Ms. Colby:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{3}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office i of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm

Sincerely yours,

Janine M. Morris

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

MicroSelectron HDR Version 2

510(k) Number

K092804

Device Name

Indications for Use

The MicroSelectron HDR Version 2 is intended to enable an operator to apply, by remote control, a radionuclide source into the body (including Interstitial, Intracavitary, Intraluminal, Bronchial, Endovascular and Intra-operative) or to the surface of the body for radiation therapy.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

logmthty

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number.

§ 892.5700 Remote controlled radionuclide applicator system.

(a)
Identification. A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.(b)
Classification. Class II.