K Number
K092261
Manufacturer
Date Cleared
2010-07-23

(360 days)

Product Code
Regulation Number
872.3640
Panel
DE
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DEFCON TSA/TSH implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients.

DEFCON TSA/TSH implants are for single-stage or two-stage surgical procedures and cement or screw retained restorations.

Implants are intended for immediate loading on single-tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function if the requirement detailed in the surgical manual is satisfied. When placing implants in the posterior region, we recommend using only large diameter (Ø6.0 mm and above) implants.

Specific indications for small diameter (Ø3.3 mm) implants: Because of their reduced mechanical stability, small diameter implants are only used in cases with a low mechanical load. Placement in the molar region is not recommended.

Device Description

DEFCON TSA / TSH dental implant systems are threaded, root-form endosseous implants of various diameters and lengths and corresponding abutments. Implants are composed of Titanium commercially pure, feature different implant to abutment connection options and are available with modified surfaces to promote improved osseointegration. Implant abutments are composed of Titanium commercially pure, Titanium alloy 6Al 4V and POM-C.

Primary stability, adequate osseointegration and mechanical performance of the implant connection are fundamental to implant success. Device design characteristics including the thread profile, geometry, surface finish and crestal characteristics are based on these concepts.

AI/ML Overview

The provided text describes the "IMPLADENT DEFCON TSA/TSH Dental Implant Systems," a root-form endosseous dental implant. The document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to already marketed predicate devices rather than providing detailed acceptance criteria and a study to prove performance against those criteria in the way a clinical trial might.

Therefore, the information requested regarding acceptance criteria, device performance, study characteristics (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance), and training set details for an AI/CAD device is not available in the provided text.

The document explicitly states:

  • "No clinical studies are submitted." (Section 5: 510(k) SUMMARY)
  • The device was subject to "bench testing to determine conformance to performance [requirements]... specifically including mechanical performance... and biological testing in accordance with the level and duration of contact with the human body, surface finish evaluation including chemical analyses, and sterilization process validation." (Section 5: 510(k) SUMMARY - NON-CLINICAL TEST SUMMARY)

This indicates that compliance was demonstrated through non-clinical, laboratory bench testing, not through human clinical studies where acceptance criteria for diagnostic accuracy (common in AI/CAD systems) would typically be evaluated.

The "acceptance criteria" in this context are related to the mechanical, biological, and material properties of the implant, which were likely evaluated against industry standards or internal specifications during bench testing. However, the specific metrics and comparison to reported performance are not detailed in the summary provided.

In summary, as this is a 510(k) for a physical medical device (dental implant) and not an AI/CAD system, the requested information on acceptance criteria, clinical study design, and performance metrics (like sensitivity, specificity, reader improvement, etc.) is not applicable or present in the provided document.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.