K Number
K091978

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-03-25

(267 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DBB-06 Hemodialysis Delivery System is indicated for hemodialysis prescribed by physicians for adult patients with acute and chronic renal failure, treated in hospitals and dialysis clinics by qualified operators. The DBB-06 Hemodialysis Delivery System is not indicated for pediatric patients. It is not for home use.

Device Description

The DBB-06 HEMODIALYSIS DELIVERY SYSTEM is composed of a hydraulic unit for the delivery of dialysate and extracorporeal blood circuitry. The permeate is heated and deaerated in the hydraulic section, which is then mixed with concentrate and fed into the dialyzer through the dialysate fluid feeder. The closed balancing system assures the amount of dialysate infused corresponds to the amount of dialysate extracted. The interior pressure of the dialyzer is controlled automatically by adjustment of the ultra filtration amount and UF rate by the dialyzer. Heparinization of the external circulating blood is accomplished with the heparin pump either by continuous or bolus injection before it is passed on to the dialyzer. The DBB-06 HEMODIALYSIS DELIVERY SYSTEM uses both acetate dialysis and bicarbonate dialysis. Utilizing the various functions of the device, the conductivity and UF profile can be programmed. In addition, the device incorporates all functions necessary for double-needle dialysis as well as single-needle dialysis procedures. The hydraulic unit is cleaned and disinfected using selectable cleaning programs and is equipped with protective systems for patient safety and proper operation. A Blood Volume Monitor (BVM) is an optional accessory available for the DBB-06 HEMODIALYSIS DELIVERY SYSTEM, and was not available with the predicate device, the DBB-05 Hemodialysis Delivery System. The BVM monitors blood volume to assist in the prevention of excessive removal of fluid.

AI/ML Overview

The Acceptance Criteria and the study proving the device meets these criteria for the NIKKISO DBB-06 HEMODIALYSIS DELIVERY SYSTEM, specifically focusing on its new optional accessories, the Blood Volume Monitor (BVM) and the Non-Invasive Blood Pressure (NIBP) measurement module, are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

Feature/ParameterAcceptance Criteria (Predicate)Reported Device Performance (DBB-06 BVM / M2500 NIBP)
Blood Volume Monitor (BVM)
Intended UseNon-invasive hematocrit, oxygen saturation, and blood volume monitoring for dialysis patients.Non-invasive blood volume monitoring (percent change in blood volume in real-time) for dialysis patients.
Principle of operationLight absorption through blood to calculate relative blood volume.Light absorption through blood to calculate relative blood volume.
Microprocessor controlYesYes
Measurement via ADCYesYes
AccuracyApproximately +/-2% (compared with other methods like centrifuge).Approximately +/-2% (compared with other methods like centrifuge).
NIBP Measurement Module (M2500)
Cuff (Arm circumference range)1240cm (M2100) / 1066cm (Press-Mate Advantage)12~40cm
Cuff BiocompatibilityCompliantCompliant
IEC60601-2-30Compliant (1999 for M2100) / (1995 for Press-Mate Advantage)Compliant (1999)
EN1060-1CompliantCompliant
EN1060-3CompliantCompliant
ANSI/AAMI SP-10CompliantCompliant
Measuring range and Accuracy0 to 300mmHg, Within +/-3mmHg0 to 300mmHg, Within +/-3mmHg
Power supplyDC 12 VDC 12 V
Pin assigns3 Signal Pins and 7 Power supply pins3 Signal Pins and 7 Power supply pins
Technical Alarms12 types (C11-21)12 types (C11-21)
Fatal Alarms4 types (E03, E07, E08, E09)4 types (E03, E07, E08, E09)
Software algorithmsSameSame

2. Sample Size Used for the Test Set and Data Provenance

  • Blood Volume Monitor (BVM):

    • Test Set Sample Size: Not explicitly stated as a number of subjects. The document mentions "Clinical Test was performed." for the DBB-06 BVM. For the predicate CRIT-LINE Monitor III (K972470), "Comparison testing was performed on blood bank blood under simulated conditions" without specifying a sample size.
    • Data Provenance: Not specified for the DBB-06 BVM clinical test (e.g., country, specific demographics). For the predicate, "blood bank blood" suggests in-vitro testing. Whether the BVM clinical test was prospective or retrospective is not explicitly stated.
  • Non-Invasive Blood Pressure (NIBP) measurement module (M2500):

    • Test Set Sample Size: Not specified. The comparison relies on compliance with international standards (IEC, EN, ANSI/AAMI) and documented specifications that match the predicate devices. Performance tests of these blood pressure monitors were conducted, but the sample size for these tests is not provided in the summary.
    • Data Provenance: Not specified. The manufacturer for both the M2100 and M2500 is "Omron Colin Co., Ltd., 1-12-14 Koishikawa, Bunkyo-ku, Tokyo Japan," suggesting Japanese origin for the device's development and potentially testing data. The nature of these "performance tests" (e.g., clinical, bench, simulated) and their provenance is not detailed.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Blood Volume Monitor (BVM): Not applicable. Ground truth for BVM accuracy is stated as "compare well with other methods such as centrifuge." This implies a gold standard measurement method rather than expert consensus on interpretation.
  • Non-Invasive Blood Pressure (NIBP) measurement module (M2500): Not applicable. The ground truth for blood pressure measurement accuracy is typically established against validated reference devices (e.g., intra-arterial measurements) or through compliance with recognized standards (e.g., ANSI/AAMI SP-10). Expert interpretation of NIBP readings against a gold standard is generally not how NIBP accuracy is assessed in regulatory submissions.

4. Adjudication Method for the Test Set

Not applicable for either the BVM or NIBP module, as the assessments are based on objective measurements and compliance with technical specifications and standards rather than subjective expert review requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. The devices are objective measurement tools, not systems that rely on human interpretation of complex data (like diagnostic imaging). Therefore, an MRMC study to measure improvement in human reader performance with AI assistance is not relevant to this type of device.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

  • Blood Volume Monitor (BVM): Yes. The BVM's performance, including its accuracy of approximately +/-2%, operates independently to calculate blood volume based on light absorption. The "Clinical Test was performed" for the BVM would assess this standalone performance.
  • Non-Invasive Blood Pressure (NIBP) measurement module (M2500): Yes. The M2500 NIBP module's performance, including its measuring range and accuracy (0 to 300mmHg, within +/-3mmHg), is an inherent function of the device itself. Performance tests (referenced but not detailed) would evaluate its standalone accuracy against an appropriate reference.

7. Type of Ground Truth Used

  • Blood Volume Monitor (BVM): "Comparison well with other methods such as centrifuge." This refers to established laboratory methods for measuring hematocrit, which is then used to calculate blood volume.
  • Non-Invasive Blood Pressure (NIBP) measurement module (M2500): Compliance with international standards (IEC60601-2-30, EN1060-1, EN1060-3, ANSI/AAMI SP-10) which define accuracy requirements, often referencing invasive blood pressure measurements or other validated non-invasive methods as the "ground truth."

8. Sample Size for the Training Set

No training set information is provided as the devices described (BVM and NIBP module) are based on established physical principles and algorithms, not machine learning models that require training data.

9. How Ground Truth for the Training Set Was Established

Not applicable, as no training set is described for these devices.

{0}------------------------------------------------

NIKKISO CO., LTD

Section 6- 510(k) Summary

MAR 2 5 2010

  • a. Company name, address NIKKISO CO., LTD. Medical Equipment Unit 498-1 Shizutani, Makinohara-shi Shizuoka, 421-0496, Japan
  • b. Contact Hachiro Edamura Manager of Quality Control
  • Date prepared ن

June 19, 2009

d. Name of device.

Trade Name:DBB-06 HEMODIALYSIS DELIVERY SYSTEM
Common Name:High Permeability Hemodialyzer
Classification Name:Dialyzer, High Permeability With or Without Sealed Dialysate System

e. Predicate devices

The DBB-06 HEMODIALYSIS DELIVERY SYSTEM is substantially equivalent to:

510(k):K023509
Trade name:DBB-05 Hemodialysis Delivery System
Product code:KDI
510(k):K061519
Trade name:Modification to DBB-05 Hemodialysis Delivery System
Product code:KDI

Description of the device ﮨﮯ

The DBB-06 HEMODIALYSIS DELIVERY SYSTEM is composed of a hydraulic unit for the delivery of dialysate and extracorporeal blood circuitry. The permeate is heated and deaerated in the hydraulic section, which is then mixed with concentrate and fed into the dialyzer through the dialysate fluid feeder. The closed balancing system assures the amount of dialysate infused corresponds to the amount of dialysate extracted. The interior pressure of the dialyzer is controlled automatically by adjustment of the ultra filtration amount and UF rate by the dialyzer. Heparinization of the external circulating blood is accomplished with the heparin pump either by continuous or bolus injection before it is passed on to the dialyzer.

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)

SECTION 6 6-2 OF 6-11

{1}------------------------------------------------

K091978 Page 2 of 10

NIKKISO CO., LTD

The DBB-06 HEMODIALYSIS DELIVERY SYSTEM uses both acetate dialysis and bicarbonate dialysis. Utilizing the various functions of the device, the conductivity and UF profile can be programmed. In addition, the device incorporates all functions necessary for double-needle dialysis as well as single-needle dialysis procedures. The hydraulic unit is cleaned and disinfected using selectable cleaning programs and is equipped with protective systems for patient safety and proper operation.

A Blood Volume Monitor (BVM) is an optional accessory available for the DBB-06 HEMODIALYSIS DELIVERY SYSTEM, and was not available with the predicate device, the DBB-05 Hemodialysis Delivery System. The BVM monitors blood volume to assist in the prevention of excessive removal of fluid.

g. Indications for Use

Indication for Use

The DBB-06 Hemodialysis Delivery System is indicated for hemodialysis prescribed by physicians for adult patients with acute and chronic renal failure, treated in hospitals and dialysis clinics by qualified operators. The DBB-06 Hemodialysis Delivery System is not indicated for pediatric patients. It is not for home use.

h. Statement of substantial equivalence

The DBB-06 HEMODIALYSIS DELIVERY SYSTEM is substantially equivalent to the DBB-05 Hemodialysis Delivery System (K023509) and the Modification to DBB-05 Hemodialysis Delivery System (K061519).

Table 1 below compares the DBB-06 HEMODIALYSIS DELIVERY SYSTEM to the DBB-05 Hemodialysis Delivery System (K023509) and the Modification to DBB-05 Hemodialysis Delivery System (K061519). All three devices are totally self-contained machines and include:

  • · the same Indication for Use
  • · use the same Operating Principle
  • · incorporate the same Basic System Design
  • · automatic priming of the extracorporeal circuit
  • · prepare dialysate
  • · monitors for the dialysate and blood
  • · pumps for blood and anticoagulant (heparin) at predetermined rates
  • · controls fluid removal
  • · automatically cleans, disinfects, and rinses the dialysate flow path
  • · incorporate the same materials

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)

SECTION 6 6-3 OF 6-11

{2}------------------------------------------------

NIKKISO CO., LTD

esponse to fda letter dated september 4, 200

Table 1. Comparison table of DBB-05 and DBB-06 Device Characteris

Device CharacteristicsPREDICATESPROPOSED
Nikkiso DBB-05Hemodialysis Delivery System(K023509)Modification toNikkiso DBB-05Hemodialysis Delivery System(K061519)Nikkiso DBB-06Hemodialysis Delivery System
Product CodeKDISameSameSame
Indications for UseThe DBB-05 HemodialysisDelivery System is indicatedfor hemodialysis prescribed byphysicians for adult andpediatric patients with acute orchronic renal failure. TheDBB-05 is intended forhemodialysis performed inhospitals and dialysis clinicsby a qualified operator.SameSameSame adult indication foruse
Proportioning systemContinuous volumetric dilutionwith duplex pumpSameSameSame
Temperature control (°C)34 to 40SameSameSame
Temperature alarm limit (°C)Fixed: 33, 41Auto: ±1 from set value¹SameSameSame
Bicarbonate conductivity range(mS/cm)2.30 to 7.00SameSameSame
Total conductivity range (mS/cm)12.5 to 15.5SameSameSame
Flow (mL/min)0, 300 to 800SameSameSame
Transmembrane pressure (mmHg)-100 to +500SameSameSame
Sodium therapyYes, ProfiledSameSameSame
Ultrafiltration removal rate (L/h)0.00; 0.10 to 4.00, Profiled UFSameSameSame
pH monitorNoneSameSameSame

K091978

Page 3 of 10

ACHMENT QUESTION 1, 2, and 3

SECTION 6
6-4 OF 6-11

DBB-06 HEMODIALYSIS DELIVERY SYSTEI
(K091978)

{3}------------------------------------------------

(t
A State Base1Call of Children
Device CharacteristicsPREDICATESNikkiso DBB-05Hemodialysis Delivery System(K023509)Modification toNikkiso DBB-05Hemodialysis DeliverySystem(K061519)PROPOSEDNikkiso DBB-06-Hemodialysis DeliverySystem
BYPASS INDICATORVisualSameSame
BLOOD CIRCUIT
Arterial pressure (mmHg)-300 to +300SameSame
Venous pressure (mmHg)-200 to +500SameSame
Blood pump range (mL/min)40 to 600SameSame
Heparin pump range (mL/h)0.0 to 9.9 (10, 20, 30 mL syringe)SameSame
DISINFECTIONChemical, thermo-chemical, hot rinseSameSame
DISPLAY TYPELCD, 12.1" color, SVGASameSame
DISPLAY PARAMETERS
Dialysate pressureYesSameSame
Transmembrane pressureYesSameSame
Bicarbonate conductivityYes2SameSame
Total conductivityYesSameSame
Flow rateYesSameSame
Elapsed timeYesSameSame
Remaining timeYesSameSame
Complete timeYesSameSame
Kt/V ratio calculation displayNoYesYes
Blood pressure value historyNumericNumeric or GraphicalNumeric or Graphical
Blood volumeNoNoYes
MICROPROCESSOR
Type3 microprocessor system, TX1941AF Toshiba, 32bit.SameSame
StorageTreatment dataSameSame

DBB-06 HEMODIALYSIS DELIVERY SYSTER

(K091978)

ACHMENT QUESTION 1, 2, and

SECTION 6
6-5 OF 6-11

Page 4 of 10

  1. 1

esponse to fda letter dated september 4, 200

{4}------------------------------------------------

RESPONSE TO FDA LETTER DATED SEPTEMBER 4, 2009
--------------------------------------------------

NIKKISO CO., LTD

Device CharacteristicsPREDICATESNikkiso DBB-05Hemodialysis Delivery System(K023509)Modification toNikkiso DBB-05Hemodialysis Delivery System(K061519)PROPOSEDNikkiso DBB-06Hemodialysis DeliverySystem
InterfaceBuilt-in RS-232 for technician,Optional TCP/IP for networksystem.SameSame
LOSS-OF-WATER ALARMYesSameSame
OTHER SPECIFICATIONYesSameSame
Single needle Click-ClackNoYesYes
Arterial clamp for use during Singleneedle Click-ClackLiquidLiquid or Dry PowderLiquid or Dry Powder
Bicarbonate concentrate typeYesSameSame
Built in BP monitoringYesSameSame
Isolation UFYes, Continuously3SameSame
Online UF control testNoSameSame
Built in Blood Volume Monitor (BVM)Notes:
  • The dialysate temperature alam limit is set to ±1°C from the temperature target value automatically. The alam wind
    #1°C can be changed from 0 to ±3°C.
    The maximum concentrate "I ನ ಗ
    • -

{5}------------------------------------------------

New optional accessories

Blood Volume Monitor (BVM)

A Blood Volume Monitor (BVM) is a new optional accessory available with the DBB-06 HEMODIALYSIS DELIVERY SYSTEM, which was not available in the predicate devices; the DBB-05 Hemodialysis Delivery System (K023509) and the Modification to DBB-05 Hemodialysis Delivery System (K061519).

The predicate device used to determine substantial equivalence to the Blood Volume Monitor (BVM) was the CRIT-LINE Monitor III (CLMII) (K972470). Both the Blood Volume Monitor (BVM) and the CRIT-LINE Monitor III (CLMIII) (K972470) are identical in function and operate under the same principle of light absorption passing through the blood under test, to measure hematocrit and calculate the related value of blood volume.

Following is comparison table for the Blood Volume Monitor of DBB-06 and the CRIT-LINE Monitor III (K972470);

Table 2 BVM Comparison Table
ItemPredicate Device:CRIT-LINE Monitor III (K972470)DBB-06 Blood Volume Monitor
Intended UseUsed for non-invasive hematocrit, oxygen saturation and blood volume monitoring. Percent change in blood volume in real time application in the treatment of dialysis patients with intended purpose of providing a more effective treatment for the dialysis patient. Based on the data that the monitors provide, the dialysis technician increases or decreases the rate at which fluid is removed from the blood in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.Used for non-invasive blood volume monitoring. This monitoring provides the percent change in blood volume in real time. The healthcare provider may increase or decrease the rate at which fluid is removed from the blood based on the data that blood volume monitor provides.
TechnologicalCharacteristics
Principle of bloodvolume monitoringThe device functions and operates under the principles of light absorption through the blood to calculate the relative value of Blood Volume (BV).The device functions and operates under the principles of light absorption through the blood to calculate the relative value of Blood Volume (BV).
MicroprocessorcontrolYesYes

Table 2 BVM Comparison Table

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)

SECTION 6 6-7 OF 6-11

{6}------------------------------------------------

RESPONSE TO FDA LETTER DATED SEPTEMBER 4, 2009

K091978

NIKKISO CO., LTD

The measurement is accomplished by calibrated Analog to Digital Converters (ADC).YesYes
AccuracyApproximately +/-2% and compare well with other methods such as centrifuge.Approximately +/-2% and compare well with other methods such as centrifuge.
Clinical testingNo clinical testing was performed. Comparison testing was performed on blood bank blood under simulated conditions.Clinical Test was performed

Based on the technical characteristics, performance and failure simulation testing of the Blood Volume Monitor (BVM), NIKKISO CO., LTD. concludes that the Blood Volume Monitor (BVM) performed as well as and is substantially equivalent to the CRIT-LINE Monitor III (CLMIII) (K972470) and does not raise any new questions regarding safety or effectiveness.

Non-Invasive Blood Pressure (NIBP) measurement module

A Non-Invasive Blood Pressure (NIBP) measurement module is an optional accessory used to monitor the patient's blood pressure during hemodialysis treatment. The DBB-05 Hemodialysis Delivery System (K023509) and the Modification to DBB-05 Hemodialysis Delivery System (K061519) both use the Omron Colin Model M2100 NIBP. The DBB-06 HEMODIALYSIS DELIVERY SYSTEM uses a newer model, the Omron Colin Model M2500 NIBP. The manufacturer for both models is Omron Colin Co., Ltd., 1-12-14 Koishikawa, Bunkyo-ku, Tokyo Japan.

The M2500 is integrated into a variety of patient monitors worldwide and is used to measure patient blood pressure using the oscillometric method with a standard blood pressure cuff wrapped on the patient's upper arm. The M2500 consists of the following four major hardware and software blocks:

  • · Cuff pressure control hardware block
  • · Cuff pressure detection and pulse detection hardware block
  • · Cuff pressure and pulse amplitude data set generation block
  • · Blood pressure determination block

The M2100 was developed using similar technology of the predicate device, the Press-Mate Advantage (K973637). All three (3) NIBPs, the M2500, M2100 and the Press-Mate Advantage are manufactured by Colin Corporation, which merged with Omron Group creating the new company Omron Colin Co., Ltd.

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)


SECTION 6 6-8 OF 6-11

{7}------------------------------------------------

RESPONSE TO FDA LETTER DATED SEPTEMBER 4, 2009

Following is comparison table of the M2500, M2100 and the Press-Mate Advantage.

Table 3 comparison table of the M2500, M2100 and the Press-Mate Advantage.

No.ItemM2500NIBP for DBB-06(K091978) underapplicationM2100NIBP for DBB-05(K023509) andmodified DBB-05(K061519)Press-Mate Advantage(K973637)
1Cuff(Arm circumferencerange)12~40cmSame10~66cm
2Cuff BiocompatibilityCompliantCompliantCompliant
3IEC60601-2-30:1999CompliantCompliantIEC60601-2-30:1995
4EN1060-1:1995CompliantCompliantCompliant
5EN1060-3:1997CompliantCompliantCompliant
6ANSI/AAMI SP-10for NIBP MeasurementCompliantCompliantCompliant
7Measuring range andAccuracy0 to 300mmHWithin +/-3mmHg0 to 300mmHWithin +/-3mmHg0 to 300mmHWithin +/-3mmHg
8Power supplyDC 12 VDC 12 V*
9Pin assigns3 Signal Pins and 7Power supply pins3 Signal Pins and 7Power supply pins-*
10Technical Alarms12 types of alarms(C11-21)12 types of alarms(C11-21)12 types of alarms(C11-21)
11Fatal Alarms4 types of alarms(E03,E07, E08,E09)4 types of alarms(E03,E07, E08,E09)4 types of alarms(E03,E07, E08,E09)
12Software algorithmsSameSameSame

*: Device is an all-in-one vital signs monitor with NIBP, and is not connected to another device.

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)

SECTION 6 6-9 OF 6-11

{8}------------------------------------------------

RESPONSE TO FDA LETTER DATED SEPTEMBER 4, 2009

NIKKISO CO., LTD

As shown in the above comparison table as well as in the performance test in those blood pressure monitors, it is demonstrated that M2500 is substantially equivalent to the M2100 used in the predicate devices, the DBB-05 Hemodialysis Delivery System (K023509) and the Modification to DBB-05 Hemodialysis Delivery System (K061519) and to Press-Mate Advantage, and does not raise any new questions regarding safety or effectiveness.

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)

SECTION 6 6-10 Ob 6-11

{9}------------------------------------------------

NIKKISO CO., LTD

i. Conclusion

Based on the above discussion and enclosed sections regarding substantial equivalence to predicate devices, NIKKISO CO., LTD. concludes that the DBB-06 HEMODIALYSIS DELIVERY SYSTEM is substantially equivalent to the DBB-05 Hemodialysis Delivery System (K023509) and the Modification to DBB-05 Hemodialysis Delivery System (K061519), and does not raise any new questions regarding safety or effectiveness.

DBB-06 HEMODIALYSIS DELIVERY SYSTEM (K091978)

SECTION 6 6-11 OF 6-11

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Nikkiso Co., Ltd. % Fumiaki Kanai, Ph.D. President and CEO MIC International 4-1-17 Hongo Bunkyo-ku Tokyo, 113-0035 JAPAN

MAR 2 5 2010

Re: K091978

Trade/Device Name: DBB-06 Hemodialysis Delivery System Regulation Number: 21 CFR $876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Codes: KDI Dated: March 19, 2010 Received: March 22, 2010

Dear Dr. Kanai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{11}------------------------------------------------

Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indications for Use

1091978 510(k) Number (If known):

Device Name: DBB-06 HEMODIALYSIS SYSTEM

Indication for Use

The DBB-06 Hemodialysis Delivery System is indicated for hemodialysis prescribed by physicians for adult patients with acute and chronic renal failure, treated in hospitals and dialysis clinics by qualified operators.

The DBB-06 Hemodialysis Delivery System is not indicated for pediatric patients. It is not for home use.

Prescription UseX
(Per 21 CFR 801 Subpart D)

AND/OR

Over-the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Ott) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”