(44 days)
The SoftLase Pro™/OrthoLase™/HygieneLase™ is to provide the ability to perform intraoral soft tissue maxilla-facial dental. general, oral and i cosmetic surgery. The SoftLase ProTM/OrthoLase™/HygieneLase™ is intended for ablating, incising, excising, vaporizing and coagulation of soft tissues using a contact fiber optic delivery system.
The device will be used in the following areas: general and cosmetic dentistry otolaryngology, arthroscopy, gastroenterology, general surgery, dermatology & plastic surgery, neurosurgery, genecology, urology, opthamology and pulmonary surgery. The following are the oralpharneeal indications for use for which the device will be marketed:
-Excision and Incision Biopsies
-Hemostatic assistance
-Treatment of Apthous Ulcers
-Frenectomy
-Frenotomy
-Gingival Incision and Excision
-Gingivectomy
-Gingivoplasty
-Incising and Draining of Abscesses
-Operculectomy
-Oral Papillectomy
-Removal of Fibromas
-Soft Tissue Crown Lengthening
-Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
-Tissue retraction for Impression
-Vestibuloplasty
-Light activation of bleaching materials for teeth whitening
-Laser-assisted bleaching/whitening of teeth
Not Found
The provided text is a 510(k) summary for the SoftLase Pro/OrthoLase/HygieneLase family of products, which are surgical laser systems. It does not contain information about acceptance criteria or a study proving the device meets those criteria in the way a typical diagnostic or AI device submission would.
Instead, the document focuses on demonstrating substantial equivalence to predicate devices already on the market. This means the manufacturer is asserting that their new device is as safe and effective as existing legally marketed devices, rather than conducting new clinical trials against specific performance metrics for novel technology.
Here's a breakdown of why the requested information cannot be fully provided based on the input and what the document does say:
-
A table of acceptance criteria and the reported device performance:
- No specific acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds) are mentioned.
- No reported device performance metrics (e.g., specific clinical outcomes, diagnostic accuracy) are presented.
- The document states: "Safety and performance testing" was conducted, but no details of this testing or its results are provided. The "performance" in this context refers to the device's functional characteristics (e.g., laser output values, safety features) being comparable to predicate devices, not clinical effectiveness in a statistical sense.
-
Sample size used for the test set and the data provenance:
- Not applicable. There is no "test set" in the context of diagnostic performance evaluation or AI model validation described here. The evaluation is based on demonstrating technical equivalence and safety to existing devices.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth for a test set is not established as there's no diagnostic or AI performance being measured against a gold standard.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. No test set requiring adjudication is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a surgical laser, not a diagnostic imaging device with an AI component for human reader assistance. No MRMC study is mentioned.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical surgical device, not a standalone algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not applicable. The "ground truth" here is the established safety and effectiveness of the predicate devices based on their prior clearances and market use. The new device is deemed equivalent to these known-safe-and-effective devices.
-
The sample size for the training set:
- Not applicable. This document describes a physical medical device, not an AI or machine learning model that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable. No training set is involved.
Summary of what the document does state regarding "proof" of meeting acceptance criteria (implicitly, suitability for market):
The "acceptance criteria" for this 510(k) submission are implicitly tied to the concept of substantial equivalence to predicate devices and compliance with relevant standards.
-
Acceptance Criteria (Implicit): The device must be as safe and effective as legally marketed predicate devices and comply with applicable performance standards and regulations.
- Compliance with 21 CFR 1010 & 1020 (mandated performance standards for laser products).
- Compliance with IEC60601-1:1995+A1+A2, IEC60601-2-22:1995, and IEC60825-1:1993+A1+A2 (international electrical and laser safety standards).
- Compliance with Quality Systems Regulations, ISO materials standards, cGMP & ISO 9000 series quality regulations.
- Technical characteristics (e.g., laser output values) similar to predicate devices.
- Intended use and indications for use previously cleared for predicate devices.
-
Study/Evidence Proving Device Meets Criteria:
- The "study" is a demonstration of substantial equivalence (not a traditional clinical trial comparing two intervention groups).
- Evidence presented:
- Comparison of intended use and indications for use with cleared predicate devices (SoftLase G2, ODYSSEY NAVIGATOR, STYLA MICROLASER/STYLAORTHO Diode Laser System).
- Comparison of technical characteristics with predicate devices.
- Verification that laser output values are "well within previously cleared values" of predicate devices.
- Reference to the "proven safety and effectiveness" of predicate devices for the claimed indications.
- Statement that "Safety and performance testing" was conducted (without providing details of the testing or results in this summary).
In essence, the "study" is a regulatory submission arguing that the new device shares fundamental characteristics and performance safety/effectiveness profiles with already approved devices, therefore requiring no new extensive clinical trials to prove efficacy or safety.
{0}------------------------------------------------
SoftLase Pro OrthoLase HygieneLase 510K
CONFIDENTIAL
KC091796
JUL 81 2009
Appendix D PREMARKET NOTIFICATION |510(k)] Summary
510(K) Summary of Safety and Effectiveness Premarket Notification 510(k)
Zap Lasers, LLC 2621-B Pleasant Hill Road Pleasant Hill, Ca. 94523
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
1. Device Name:
SoftLase Pro/OrthoLase/HygieneLase family of products Trade Name: Common Name(s): Surgical Laser System Classification Name(s): Laser, Surgical
- Establishment Name & Registration Number: Name: Zap Lasers, LLC Number: 3005212363
3. Classification(s):
$878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology. (a) Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide. (2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b) Classification. Class II.
| Device Class: | Class II for all requested indications |
|---|---|
| Classification Panel: | General and Plastic Surgery & Others |
| Product Code(s): | GEX |
.
4. Section 514 Compliance
Zap Lasers, LLC intends to comply fully with the general controls authorized under Sections 501, 502, 510, 516, 518, and 520 of the Food, Drug, and Cosmetic Act.
{1}------------------------------------------------
5. Performance Standards
United States Food and Drug Administration mandated performance standards for this device exist and are provided under Sections 21 CFR 1010 & 1020, with permissible deviations relative to Laser Notice 50, dated July 26, 2001. The device also complies with IEC60601-1:1995+A1+A2, IEC60601-2-22:1995, and IEC60825-1:1993+A1+A2. In addition, various voluntary performance standards are utilized. Voluntary standards utilized include Standard Operating Procedures, vendor & process certification and qualification procedures, Quality Systems Regulations, ISO materials standards and cGMP & ISO 9000 series quality regulations.
6. Special Controls:
All Class II devices are subject to Special Controls.
7. Labeling:
.
The laser system discussed in this premarket notification will be manufactured by Zap Lasers, LLC and labeled as such. Zap Lasers, LLC will market the system exclusively to healthcare facilities, physicians and dentists. In addition to the usual package and identification labeling, the following additional Warnings, Cautions & Precautions statements are displayed as appropriate on or within the device packaging. They are repeated here for ease of use.
Warning: Federal (United States) Law restricts this device to sale by or on the order of a physician or dentist only.
9. Predicate Device (legally marketed comparison device)
Zap Lasers, Inc. believes that the following surgical laser systems are substantially equivalent to the SoftLase Pro/OrthoLase/HygieneLase Family of products surgical diode system.
-
- SoftLase G2 (K021227, Zap Lasers, LLC)
-
- ODYSSEY NAVIGATOR (K062258, Ivoclar Vivadent, Inc.)
-
- STYLA MICROLASER/STYLAORTHO Diode Laser System (K081214, Zap Lasers LLC.)
10. Summary of Equivalence:
There are no unique applications, material or specifications presented herein. Evidence of equivalence has been demonstrated through:
- The SoftLase Pro ™/OrthoLase™/HygieneLase™ family of products intended use and indications . for use were previously cleared by FDA for the predicate devices.
- The technical characteristics of the SoftLase Pro ™/OrthoLase™/HygieneLase™ family of products . are similar to those of the cleared SoftLase G2, Odyssey Navigator and the Styla MicroLase.
- Laser output values of the SoftLase Pro ™/OrthoLase™/HygieneLase™ family of products are well . within previously cleared values of the predicate dental laser system as described.
- The predicate devices and other previously cleared laser systems with similar power outputs have a . proven safety and effectiveness in the treatment of the claimed indications.
- . Safety and performance testing.
Therefore, the SoftLase Pro/OrthoLase Family of products is substantially equivalent to its predicate devices cited above and raises no new safety and/or effectiveness issues.
Page 46 of 72
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ZAP LASERS, LLC % Mr. Jay Goble, DDS President 2621-B Pleasant Hill Road Pleasant Hill, California 94523
JUL 3 1 2009
Re: K091796
Trade/Device Name: Softlase Pro™/OrthoLase™/HygieneLase™ Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery
and in dermatology
Regulatory Class: II Product Code: GEX Dated: July 15, 2009 Received: July 22, 2009
Dear Mr. Goble:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Jay Goble, DDS
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours, Mark N.Melkerso Director Division of Surgical, Orthopedic and Restorative Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
CONFIDENTIAL
Tab 2 Indications for Use Statement
Device Name: SoftLase Pro™/OrthoLase™/HygieneLase™
Indication for Use:
The SoftLase Pro™/OrthoLase™/HygieneLase™ is to provide the ability to perform intraoral soft tissue maxilla-facial dental. general, oral and i cosmetic surgery. The SoftLase ProTM/OrthoLase™/HygieneLase™ is intended for ablating, incising, excising, vaporizing and coagulation of soft tissues using a contact fiber optic delivery system.
The device will be used in the following areas: general and cosmetic dentistry otolaryngology, arthroscopy, gastroenterology, general surgery, dermatology & plastic surgery, neurosurgery, genecology, urology, opthamology and pulmonary surgery. The following are the oralpharneeal indications for use for which the device will be marketed:
-Excision and Incision Biopsies
-Hemostatic assistance
-Treatment of Apthous Ulcers
-Frenectomy
-Frenotomy
-Gingival Incision and Excision
-Gingivectomy
-Gingivoplasty
-Incising and Draining of Abscesses
-Operculectomy
-Oral Papillectomy
-Removal of Fibromas
-Soft Tissue Crown Lengthening
-Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
-Tissue retraction for Impression
-Vestibuloplasty
-Light activation of bleaching materials for teeth whitening
-Laser-assisted bleaching/whitening of teeth
| Prescription Use | X |
|---|---|
| (21 CFR Part 801 Subpart D) | |
| And/Or | |
| Over the Counter Use | . |
| (21 CFR Part 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
| (Division Sign-Off) |
|---|
Division of Surgical, Orthopedic, and Restorative Devices
Page 6 of 72
6/15/09
510(k) Number K091796
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.