(296 days)
Not Found
No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the device description focuses on mechanical and electrical components for oxygen pressurization and filling, with no mention of data processing or algorithmic decision-making.
No.
The device is described as an oxygen filling machine that pressurizes oxygen from an oxygen concentrator to fill gas cylinders for ambulatory use, and it is not intended for life support or patient monitoring. While the oxygen it processes is for therapeutic use, this specific device (Profill Q601 series) is for filling cylinders, not directly administering therapy itself.
No
The device is an oxygen concentrator with an oxygen filling accessory, which provides supplemental oxygen and fills gas cylinders. It explicitly states it "is not intended for life support nor does it provide any patient monitoring capabilities," indicating it does not diagnose medical conditions.
No
The device description explicitly details hardware components like a compressor module, buffer tank, and cylinder fitting, indicating it is a physical medical device, not software-only.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to provide supplemental oxygen to patients and fill oxygen cylinders for personal ambulatory use. This is a therapeutic and delivery function, not a diagnostic one.
- Device Description: The description details a compressor module and cylinder for pressurizing and filling oxygen. This aligns with oxygen delivery, not in vitro testing of biological samples.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), detecting analytes, or providing diagnostic information about a patient's health status.
IVD devices are used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This device's function is entirely focused on providing and managing oxygen for breathing.
N/A
Intended Use / Indications for Use
The Oxygen Concentrators with the Oxygen Filling accessory are intended to provide supplemental oxygen to patients in the home and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. They are not intended to sustain or support life.
The Oxygen Concentrators with the Oxygen Filling accessory are indicated to provide supplemental oxygen to patients in the home and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. The device is not intended for life support nor does it provide any patient monitoring capabilities.
Product codes
CAW
Device Description
The Profill Q601 series oxygen filling machines are prescription devices designed for use in home by patients that require supplemental oxygen. It is intended to pressurize oxygen from an oxygen concentrator to fill gas cylinders for the patient's personal ambulatory use. It is not intended for life support nor does it provide any patient monitoring capabilities.
The device consists of a compressor module and a portable oxygen cylinder with a specially adapted cylinder fitting. It is compatible with available standard oxygen concentrators as the source of oxygen. The oxygen generated by the oxygen concentrator is inducted into a buffer tank. Then it flows to the compressor where the oxygen is pressurized and filled into the cylinder. The devices are not sold or labeled as sterile.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The Safety tests under conduct are summarized in the following.
- a. Leakage Currents Test
- b. Dielectric Strenath
- c. Rigidity Test
- d. Stability Test
- e. Heating Test
- f. Abnormal operation and fault condition
- g. Power cord Test
- h. Transformer dielectric strength
- Ball pressure Test -
The regarding electromagnetic compatibility (EMC) tests has been finished are listed in the following.
- a. EN60601-1-2
- b. EN61000-3-2
- c. EN61000-3-3
Verification testing has confirmed the product meets its specifications.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 868.5440 Portable oxygen generator.
(a)
Identification. A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).(b)
Classification. Class II (performance standards).
0
!
9.0 SUMMARY OF SAFETY AND EFFECTIVENESS
APR - 2 2010
"510(k) SUMMARY" | ||
---|---|---|
9.1 | Manufacturer: | Merits Health Products Co., LTD. |
9, Road 36, | ||
Taichung Industrial Park | ||
Taichung, Taiwan R.O.C | ||
9.2 | Submitted By: | Steve Chao |
Merits Health Products Co., LTD. | ||
9, Road 36, | ||
Taichung Industrial Park | ||
Taichung, Taiwan R.O.C. | ||
Tel: 886-4-2359-4985 ext. 200 | ||
Fax: 886-4-2359-4992 | ||
E-mail: stevechao@merits.com.tw | ||
9.3 | Trade/Proprietary Name: | Merits Health Products Oxygen Filling |
Accessory | ||
9.4 | Common/Usual Name: | Oxygen Concentrator (Accessory) |
9.5 | Classification Name: | Portable Oxygen Generator |
21 CFR 868.5440 | ||
73 CAW |
9.6 Comparison to Currently Marketed Devices
The modified Merits Health Products Profill Q601 Series Oxygen Filling Accessory is substantially equivalent to the currently marketed Merits Oxygen Filling Accessory (K050430).
9.7 Device Description
The Profill Q601 series oxygen filling machines are prescription devices designed for use in home by patients that require supplemental oxygen. It is intended to pressurize oxygen from an oxygen concentrator to fill gas cylinders for the patient's personal ambulatory use. It is not intended for life support nor does it provide any patient monitoring capabilities.
The device consists of a compressor module and a portable oxygen cylinder with a specially adapted cylinder fitting. It is compatible with available standard oxygen concentrators as the source of oxygen. The oxygen generated by the oxygen concentrator is inducted into a buffer tank. Then it flows to the compressor where the oxygen is pressurized and filled into the cylinder. The devices are not sold or labeled as sterile.
9 ก
ﻴﺔ
1
SUMMARY OF SAFETY AND EFFECTIVENESS (Con't)
Indications for Use 9.8
The Oxygen Concentrators with the Oxygen Filling accessory are intended to provide supplemental oxygen to patients in the home and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. They are not intended to sustain or support life.
g.g Technological Characteristics
The subject and the predicate devices both use the oxygen concentrator as its oxygen source, similar AC-Powered Motors to drive the oxygen compressors. Both the oxygen compressors use cylinders of the same size and material. The oxygen flows in is compressed in the cylinders and then fill into a medical cylinder through the filling connector. The filling time and filling pressure are the same. Both meet the same Safety and EMC requirements. The technological characteristics of the Profill 0601 series oxygen filling machines are the same as the predicate devices.
9.10 Performance Data
The Safety tests under conduct are summarized in the following.
- a. Leakage Currents Test
- b. Dielectric Strenath
- c. Rigidity Test
- d. Stability Test
- e. Heating Test
- f. Abnormal operation and fault condition
- g. Power cord Test
- h. Transformer dielectric strength
- Ball pressure Test -
The regarding electromagnetic compatibility (EMC) tests has been finished are listed in the following.
- a. EN60601-1-2
- b. EN61000-3-2
- c. EN61000-3-3
Verification testing has confirmed the product meets its specifications.
9.11 Conclusion
Based on the design, performance specifications, testing and intended use, the Oxygen Filling Accessories are substantially equivalent to the currently marketed devices.
2
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle, represented by three curved lines that suggest the bird's head, body, and tail feathers.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR - 2 2010
Mr. Steve Chao Merits Health Products Company, Limited 9, Road 36, Taichung Industrial Park Taichung, Taiwan 407
Re: K091716
Trade/Device Name: The Merits Health Products Profill Q601 Series Oxygen Filling Accessory Regulation Number: 21CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: March 10, 2010 Received: March 11, 2010
Dear Mr. Chao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
. Page 2- Mr. Chao
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. · Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
INDICATIONS FOR USE STATEMENT
510(k) File Number:
Device Name:
The Merits Health Products Profill Q601 Series Oxygen Filling Accessory
Indications For Use:
The Oxygen Concentrators with the Oxygen Filling accessory are indicated to provide supplemental oxygen to patients in the home and to supply pressurized oxygen to fill gas cylinders for the patient's personal ambulatory use. The device is not intended for life support nor does it provide any patient monitoring capabilities.
Prescription Use ✓ (Per 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
A
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _ Co91716
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
. . . . . . . . . . . . .
Concurrence of CDRH, Office of Device Evaluation (ODE)