(22 days)
The KOWA VX-10a is intended for taking pictures of fundus images with mydriatic or without mydriatic.
The KOWA VX-10a is a simplified version of the predicate device [KOWA VX-10i (K062021)] by removing the indocyanine green (ICG) angiography mode, removing the use of Polaroid film as a recording media, and changing the availability of the data card for writing patient information as an optional accessory. In addition the lens coating was changed.
The provided text describes a 510(k) premarket notification for the KOWA VX-10α fundus camera. This document is a submission to the FDA demonstrating that a new device is substantially equivalent to a legally marketed predicate device. As such, it does not contain acceptance criteria for device performance based on a study, nor does it present results from a clinical or standalone effectiveness study.
The primary "study" proving the device meets acceptance criteria in this context is a comparison to a predicate device and a series of safety tests.
Here's a breakdown of the requested information based on the provided text, with explanations for what is missing or not applicable to this type of submission:
1. A table of acceptance criteria and the reported device performance
This document does not present acceptance criteria in the form of performance metrics for a clinical study or functional performance against specific thresholds. Instead, the "acceptance criteria" are implied by the substantial equivalence argument, meaning the new device must perform at least as well as (or equivalently to) the predicate device for its intended use, and demonstrate compliance with relevant safety standards.
Therefore, the "reported device performance" is essentially that the KOWA VX-10α is comparable to the KOWA VX-10i.
| Acceptance Criteria | Reported Device Performance (vs. Predicate Device KOWA VX-10i) |
|---|---|
| Equivalent Indications For Use | Same |
| Equivalent Picture Magnifications | Same |
| Acceptable Photography Modes | Non-mydriatic, Mydriatic color, Fluorescein angiography (ICG angiography removed as a deliberate simplification) |
| Equivalent Working Distance | Same |
| Equivalent CCD Camera for Observation | Same |
| Acceptable Record Media | 35mm film (Polaroid film removed as a deliberate simplification) |
| Equivalent Video Camera Connectability | Yes |
| Equivalent Observation System | Same |
| Equivalent Dioptric Compensation | Same |
| Equivalent Focusing | Same |
| Equivalent Filter for FA | Present |
| Filter for ICG | Not applicable (deliberately removed) |
| Equivalent Observation Light Source | Same |
| Equivalent Photographing Light Source | Same |
| Equivalent Power Consumption | Same |
| Equivalent Dimension | Same |
| Equivalent Weight | Same |
| Safety Compliance (due to lens coating change) | Conforms to ISO10940, ISO 15004-1, ISO 15004-2 (no safety concerns raised) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not applicable. This submission relies on a comparison to a predicate device and engineering/safety tests, not a clinical test set of patient images or data.
- Data Provenance: Not applicable for a traditional clinical test set. The safety test reports (ISO standards) would have their own data provenance, likely from internal testing by Kowa Company, Ltd. in Japan.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. No clinical test set requiring expert ground truth was evaluated in this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. No clinical test set requiring adjudication was evaluated.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a fundus camera, an imaging acquisition device, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to AI assistance is not relevant or described.
6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done
Not applicable. This device is an imaging hardware device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For the purpose of this 510(k) submission, the "ground truth" for the device's functionality is its equivalency to the predicate device and compliance with established international safety standards for ophthalmic instruments.
8. The sample size for the training set
Not applicable. This is hardware, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is hardware, not an AI/ML algorithm that requires a training set.
Summary of the Study Proving Acceptance Criteria:
The "study" conducted for the KOWA VX-10α to meet acceptance criteria (in the context of a 510(k) submission) primarily involved:
- Predicate Device Comparison: A detailed feature-by-feature comparison demonstrating that the KOWA VX-10α is substantially equivalent to the KOWA VX-10i (K062021), with any differences being simplifications that do not raise new questions of safety or effectiveness.
- Safety Testing: Specific tests were performed to address the impact of a change in lens coating:
- Test Report for ISO10940, 1998: Ophthalmic instruments - Fundus cameras
- Test Report for ISO 15004-1, 2006: Ophthalmic instruments-Fundamental requirements and test methods Part 1: General requirements applicable to all ophthalmic instruments
- Test report for ISO 15004-2:2007 Ophthalmic instruments-Fundamental requirements and test methods Part 2: Light hazard protection
The results of these safety tests demonstrated conformity to the listed ISO standards, confirming that the lens coating change did not introduce any new optical hazards or safety concerns. This, combined with the substantial equivalence argument, formed the basis for the FDA's clearance.
{0}------------------------------------------------
-- СНУО КН. ТОКУО 193 8432 ИРА ਕ ਨ
K091683
4-14, NIHONBASHHONCHO 3-CHOME
CHUO-KU, TOKYO 103-8433 JAPAN
| Applicant: | Kowa Company, Ltd. |
|---|---|
| 4-14, Nihonbashi-honcho 3-chome | |
| Chuo-ku, Tokyo 103-8433 Japan | |
| Contact: | Akihiro Fujita |
| Date Summary Prepared: | May 15, 2009 |
| Device Trade Name: | KOWA VX-10α |
| Classification name: | Camera, Ophthalmic, AC-powered |
| Product code: | HKI |
Intended use:
イ
The KOWA VX-10a is intended for taking pictures of fundus images with mydriatic or without mydriatic.
{1}------------------------------------------------
Comparison:
The KOWA VX-10i was chosen as the predicate device.
The KOWA VX-10a is a simplified version of the predicate device [KOWA VX-10i (K062021)] by removing the indocyanine green (ICG) angiography mode, removing the use of Polaroid film as a recording media, and changing the availability of the data card for writing patient information as an optional accessory. In addition the lens coating was changed.
In order to evaluate the effects of the change of lens coating for any optical hazards, the following tests were performed:
-
Test Report for ISO10940, 1998: Ophthalmic instruments - Fundus cameras
-
Test Report for ISO 15004-1, 2006: Ophthalmic instruments-Fundamental requirements and test methods Part 1: General requirements applicable to all ophthalmic instruments
-
Test report for ISO 15004-2:2007 Ophthalmic instruments-Fundamental requirements and test methods Part 2: Light hazard protection
The test results demonstrated that the KOWA VX-10α conforms to the above listed standards. which confirmed that the change in the lens coating did not raise any safety concerns. The test results are included in Appendix D.
A comparison of the KOWA VX-10a and the predicate device is shown below in Table B.
Table A. Predicate device
| Predicate Device | Manufacturer | 510(k) Number | Date Cleared |
|---|---|---|---|
| KOWA VX-10i | KOWA Company Ltd. | K062021 | Nov. 1, 2006 |
Conclusion:
The KOWA VX-10a is equipped with the same fundamental technology features equivalent to the predicate devices, and also delivers the equivalent level safety and effectiveness. Therefore since there are no significant difference in the basic function, safety and effectiveness between the KOWA VX-10a and the predicate device, KOWA concludes that the KOWA VX-10a is substantially equivalent to the KOWA VX-10i (K062021).
{2}------------------------------------------------
Table B. Comparison table
| Proposed Device | Predicate Device | |
|---|---|---|
| KOWA VX-10a | KOWA VX-10i (K062021) | |
| IndicationsFor Use | Same | Take pictures of fundus imageswith or without mydriatic. |
| Picturemagnifications | Mydriatic : 50°/ 30°Non-mydriatic : 45°/ 27° | Mydriatic : 50°/ 30°Non-mydriatic : 45°/ 27° |
| Photography mode | Non-mydriaticMydriatic colorFluorescein angiography | Non-mydriaticMydriatic colorFluorescein angiographyIndocyanine Green angiography |
| Working distance | Same | 39 mm |
| CCD camera forobservation | Same | Monochrome CCD |
| Record media | 35mm film | 35mm film/ Polaroid film |
| Video cameraconnect ability | Yes | Yes |
| Observation system | Same | Mydriatic : FinderNon-mydriatic : LCD |
| Dioptriccompensation | Same | -32D ~ +35D |
| Focusing | Same | By aligning the split lines |
| Filter for FA | Present | Present |
| Filter for ICG | Not applicable | Present |
| Observation LightSource | Same | Halogen lamp100W |
| Photographing LightSource | Same | Xenon flash lamp300WS |
| Power consumption | Same | 180VA1800VA (Instantaneous) |
| Dimension | Same | 400(W) x 520(D) x 752(H) mm |
| Weight | 35.5 kgwith power unit | 35.5 kgwith power unit |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in a simple, sans-serif font.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kowa Company, Ltd. c/o Akihiro Fujita, General Manager Electronics and Optics Division 4-14. Nihonbashi-honcho 3-chrome Chuo-ku, Tokyo Japan 103-8433
JUL - 2 2009
Re: K091683
Trade/Device Name: Kowa VX-10a Fundus Camera Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: II Product Code: HKI Dated: May 15, 2009 Received: June 10, 2009
Dear Mr. Fujita:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indication for Use
510(k) Number (if known):
Device Name: KOWA VX-10α
Indication For Use:
The KOWA VX-10a is intended for taking pictures of fundus images with mydriatic or without mydriatic.
V Prescription Use (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(División Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
§ 886.1120 Ophthalmic camera.
(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.