K Number
K091190
Manufacturer
Date Cleared
2009-07-30

(98 days)

Product Code
Regulation Number
888.3010
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Universal Clamp Stainless Steel System is a temporary implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone factures. The indications for use include the following applications:

  • Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring techniques;
  • Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as scoliosis, kyphosis, spondylolisthesis;
  • Spinal degenerative surgery, as an adjunct to spinal fusions;
    The Universal Clamp Stainless Steel System may also be used in conjunction with other medical implants made of stainless steel whenever "wiring" may help secure the attachment of other implants.
Device Description

The Universal Clamp Stainless Steel System is a temporary orthopedic implant intended to provide stabilization during the development of solid bony fusion and aid in the repair of bone fractures. The device system is designed to be incorporated into constructs and used in conjunction with other medical implants made of stainless steel whenever "wiring" may help secure the attachment of other implants.

AI/ML Overview

The provided text describes the Zimmer Spine Universal Clamp® Stainless Steel System. However, it does not include information about acceptance criteria or a study proving the device meets them in the way a medical AI/software device submission would.

Instead, the document is a 510(k) premarket notification for a physical medical device (a bone fixation cerclage system). For such devices, "acceptance criteria" and "study that proves the device meets the acceptance criteria" are typically addressed through bench testing and engineering evaluations demonstrating substantial equivalence to a legally marketed predicate device, rather than clinical performance metrics in the context of sensitivity, specificity, or reader studies.

Based on the provided text, here's the breakdown of what is available:

1. A table of acceptance criteria and the reported device performance:

This information is not present in the provided text. The document refers to "engineering evaluations and bench testing" but does not detail specific acceptance criteria (e.g., tensile strength, fatigue life thresholds) or specific quantitative performance values.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

This information is not applicable in the context of an AI/software device, as this is a physical implant. For the engineering evaluations and bench testing, sample sizes for the mechanical tests would typically be defined by engineering standards, but these details are not provided in the summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not applicable as this is a physical implant, not a diagnostic or interpretive device requiring expert-established ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable as this is a physical implant.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable as this is a physical implant, not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable as this is a physical implant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

This information is not applicable as this is a physical implant. Performance for such devices is based on mechanical properties and material science.

8. The sample size for the training set:

This information is not applicable as this is a physical implant. There is no "training set" in the context of device approval for a physical implant.

9. How the ground truth for the training set was established:

This information is not applicable as this is a physical implant.


Summary of available information regarding performance:

The document states:

  • "Engineering evaluations and bench testing were conducted to assess the physical and mechanical properties of the subject device."
  • "These results demonstrate that the performance of the Abbott Spine Universal Clamp® Stainless Steel System is substantially equivalent to the predicate stainless steel version cleared under premarket notification K081622."

This indicates that the Zimmer Spine Universal Clamp® Stainless Steel System achieved regulatory acceptance by demonstrating substantial equivalence to a predicate device (Abbott Spine Universal Clamp® Stainless Steel System, cleared under K081622) through bench testing and engineering evaluations of its physical and mechanical properties. The specific details of these tests, including acceptance criteria and quantitative results, are not included in this summary document but would have been part of the full 510(k) submission.

{0}------------------------------------------------

Zimmer Spine SUMMARY OF SAFETY AND EFFECTIVENESS

SUBMITTER:

Zimmer Spine

JUL 30 2009

3003853072

David Padgett, RAC (US) Senior Specialist, Regulatory Affairs Telephone: (512) 533-1998 Fax: (512) 258-0995

June 30, 2009

Universal Clamp® Stainless Steel System

JDQ

Bone Fixation Cerclage

21 CFR § 888.3010

Abbott Spine Universal Clamp Stainless Steel System

The Universal Clamp Stainless Steel System is a temporary orthopedic implant intended to provide stabilization during the development of solid bony fusion and aid in the repair of bone fractures. The device system is designed to be incorporated into constructs and used in conjunction with other medical implants made of stainless steel whenever "wiring" may help secure the attachment of other implants.

The Universal Clamp Stainless Steel System is a temporary orthopedic implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications:

  • -Spinal trauma surgery, used in sublaminar. interspinous, or facet wiring techniques.
  • Spinal reconstructive surgery, incorporated into 2. constructs for the purpose of correction of spinal deformities such કર scoliosis, kvohosis. spondylolisthesis.
    1. Spinal degenerative surgery, as an adjunct to spinal fusions;

The Universal Clamp Stainless Steel System may also be used in conjunction with other medical implants made

CONTACT PERSON:

TRADE NAME:

Date:

Product Code:

CLASSIFICATION NAME:

Classification Reference:

MANUFACTURER ESTABLISHMENT

REGISTRATION NUMBER:

PREDICATE DEVICE:

DEVICE DESCRIPTION:

INDICATIONS:

{1}------------------------------------------------

of stainless steel whenever "wiring" may help secure the attachment of other implants.

COMPARISON TO PREDICATE DEVICE:

Engineering evaluations and bench testing were conducted to assess the physical and mechanical properties of the subject device. These results demonstrate that the performance of the Abbott Spine Universal Clamp® Stainless Steel System is substantially equivalent to the predicate stainless steel version cleared under premarket notification K081622.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a seal or emblem.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 30 2009

Zimmer Spine % Mr. David Padgett, RAC Senior Specialist, Regulatory Affairs 5301 Riata Park Court, Building F Austin, TX, 78727

Re: K091190

Trade/Device Name: Clamp® Stainless Steel System Regulation Number: 21 CFR 888.3010 · Regulation Name: Bone fixation cerclage Regulatory Class: II Product Code: JDQ Dated: June 30, 2009 Received: July 2, 2009

Dear Mr. Padgett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. David Padgett, RAC

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Harbara Buchus

fris

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K091190

Device Name:

Universal Clamp® Stainless Steel System

Indications for Use:

The Universal Clamp Stainless Steel System is a temporary implant for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone factures. The indications for use include the following applications:

  • Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring 1 techniques;
  • Spinal reconstructive surgery, incorporated into constructs for the purpose of 2. correction of spinal deformities such as scoliosis, kyphosis, spondylolisthesis;
  • റ്റ് Spinal degenerative surgery, as an adjunct to spinal fusions;

The Universal Clamp Stainless Steel System may also be used in conjunction with other medical implants made of stainless steel whenever "wiring" may help secure the attachment of other implants.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDBH. Office of Device Evaluation (ODE) (ExT forma) (Division Sign-Off) Division of Surgical, Orthopedic,

and Restorative Devices

510(k) Number K091190

§ 888.3010 Bone fixation cerclage.

(a)
Identification. A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.(b)
Classification. Class II.