K Number
K091061

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-11-13

(213 days)

Product Code
Regulation Number
878.5000
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Implicitguide™ Surgical Suture System is indicated for use in soft tissue approximation and elevation of subdermis and underlying muscle.

Device Description

The ImplicitCare Implicitguide™ Surgical Suture System combines USP size 4-0 braided nonabsorbable poly(ethylene terephthalate), or polyester, surgical suture with a set of manual instruments designed to efficiently and accurately place the suture for the purpose of performing approximation of tissue and elevation of subdermis/underlying muscle procedure. All components of the Implicitguide™ Surgical Suture System are provided sterile, and are intended for single use only. The manual instruments consist of the following: Tape (marking tape) to aid in marking the skin for needle (Suturod) point of entry locations)- if needed, Lancet to create a point of entry for needle (Suturod) and suture, Clearing device to clear subcutaneous ligamentous attachments around the puncture site, Two (2) stainless steel Suturods™ which, like suture needles, guide the suture through the tissue. Each Suturod™ has an internal light guide which, when connected to the Light Handle, illuminates the distal tip of the Suturod™M, Light Handle which contains an LED light source to connect to Suturod™.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the Implicitguide™ Surgical Suture System based on the provided text. It's important to note that the document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against novel acceptance criteria in a clinical trial. Therefore, some requested information (like effect size of AI, standalone algorithm performance, or detailed ground truth for training) is not applicable or not present in this type of submission.

1. Table of Acceptance Criteria and Reported Device Performance

Test CategoryAcceptance Criteria (Stated or Implied)Reported Device Performance
BiocompatibilityCompliance with AAMI/ISO 10993-1 requirements and FDA's Blue Book memorandum (G95-1) for an Implant Device contacting Tissue, Bone, or Blood permanently. Specific tests: Physico-chemical (Plastics, USP), Cytotoxicity (ISO MEM Elution), Sensitization (ISO Murine Lymph Node Assay), Irritation (ISO Intracutaneous Reactivity Test), Systemic Toxicity (ISO Acute Systemic Injection Test), Genotoxicity (ISO Salmonella typhimurium Reverse Mutation), Hemocompatibility (NIH Method), Pyrogen (ISO Materials Mediated Rabbit Pyrogen), Implantation (7 day and 90 day Muscle implant)."All results are acceptable." (for the TEVDEK® II polyester suture, which is stated to be exactly the same suture used in the Implicitguide™ system). "These tests demonstrate that the suture contained in the Implicitguide™ Surgical Suture System is as safe as the predicated devices."
Suture MechanicalMeets all requirements of USP Monograph for Nonabsorbable Surgical Suture. Specifically USP XXXI <871> Needle attachment - Sutures."The subject device will meet all requirements of USP XXXI <871> Needle attachment - Sutures." (due to being the same as TEVDEK® II).
Device Functionality (Cadaver Tests)Performance of brow elevation and shaping. Trans-illumination and color of light for Suturod. Tensile strength of medial platysmal border muscles and facial retaining ligaments. Demonstrated substantial equivalence to predicate device with added advantage of Suturod light trans-illumination."Test results are acceptable and test reports are on file."
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2002/A1:2004 and EN60601-1-2:2007 for Radio Frequency Emissions and Electromagnetic Immunity Tests for medical electrical equipment."All test required passed requirements." "This report demonstrates the safety of the Suturod."
Design VerificationDesign and USP requirements for various components, including: Clearing Tool Torque, Lancet Blade Puncture, Suture Length, Suture Attachment Force, Suturod Bending, Light Intensity, Light On-Time."All test results are acceptable." "These tests demonstrate that the device meets design and USP requirements."

2. Sample Size Used for the Test Set and Data Provenance

  • Biocompatibility/Suture Mechanical Tests: The tests were conducted on the TEVDEK® II polyester suture, which is described as being "exactly the same suture" as used in the Implicitguide™ system. The sample size for these specific tests is not explicitly stated, but it would have followed the requirements of the standards (e.g., ISO, USP). The data provenance is implied to be from Teleflex Medical (formerly Genzyme Corp.), the manufacturer of TEVDEK® II. These are likely retrospective, pre-existing test results.
  • Cadaver Tests: Four cadavers (two females aged 92 and 84, two males aged 58). Data provenance: University of Maryland, State Anatomy Board, Baltimore, MD, USA. This was a prospective, specifically designed pre-clinical test.
  • EMC Evaluation: Not specified as a "test set" in terms of biological samples. The "test required" refers to the device itself being subjected to EMC testing in a laboratory environment.
  • Design Verification Tests: Sample sizes for each specific design verification test (e.g., number of sutures for attachment force, number of Suturods for bending) are not explicitly stated, but are implied by the nature of engineering design verification. The data provenance is ImplicitCare, LLC's internal testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Biocompatibility/Suture Mechanical Tests: Ground truth is established by compliance with international standards (ISO, AAMI) and USP monographs, which are developed by expert consensus in those fields. No individual experts for a specific test set are mentioned as establishing "ground truth" in the way one would for clinical imaging studies.
  • Cadaver Tests: Two Board Certified plastic surgeons, Dr. Gregory Mueller and Dr. Sherrell Aston, performed the procedures and evaluated the results. Their CVs were on file, indicating their qualification as experts.
  • EMC Evaluation & Design Verification Tests: Ground truth is defined by the technical specifications of the standards (IEC, EN) and internal design requirements. Expert involvement would be in the design and interpretation of these tests, but not in establishing "truth" for a clinical test set.

4. Adjudication Method for the Test Set

  • Cadaver Tests: No explicit adjudication method (like 2+1 or 3+1) is mentioned. The description states the procedures were "conducted by" the two plastic surgeons. It implies they jointly or individually evaluated the results, possibly through consensus, but this is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "No clinical testing was conducted to support this submission." This type of study is typically associated with clinical trials evaluating diagnostic or prognostic devices with human readers interpreting cases.

6. Standalone (Algorithm Only) Performance Study

  • No, a standalone (algorithm only) performance study was not done. This device is a surgical suture system, not an AI algorithm.

7. Type of Ground Truth Used

  • Biocompatibility/Suture Mechanical: Established by compliance with recognized standards and monographs (e.g., ISO, USP) and prior regulatory approval (510(k) for predicate suture). This is a form of regulatory and scientific consensus.
  • Cadaver Tests: Expert evaluation/observation by Board Certified plastic surgeons during a pre-clinical cadaver study.
  • EMC Evaluation & Design Verification: Technical specifications and engineering requirements for the device components, validated against international standards.

8. Sample Size for the Training Set

  • This information is not applicable. The device is a surgical suture system and does not involve machine learning algorithms that require a "training set." The submission details non-clinical tests demonstrating safety and equivalence.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

page 1/5

October 9, 2009

રે.0 510(k) SUMMARY

In accordance with Title 21 Code of Federal Regulations (21 CFR), Part 807, and in particular, §807.92, the following 510(k) summary is provided for ImplicitguideTM Surgical Suture System:

5.1 Submitted By:

NOV 1 3 2009

Implicit Care, LLC 5920 Friars Road, Suite 102 San Diego, California 92108

Contact: Dionicia B. Reblando, Vice President, RA/QA

Date Prepared: 10/9/2009

5.2 Device Name

Trade or Proprietary Name: Implicitguide™ Surgical Suture System

Common or Usual Name: Nonabsorbable Polyester Suture

Nonabsorbable poly(ethylene terephthalate) surgical suture Classification Name:

5.3 · Predicate Devices

The subject device is substantially equivalent, in whole or in part, to the following commercially available predicate device:

  • TEVDEK® II Polyester Suture K001434, K021019; Genzyme Corp. ●
  • Featherlift Extended Length Aptos Threads (Contour Threads) K041593; . Surgical Specialties Corporation

Device Description 5.4

The ImplicitCare Implicitguide™ Surgical Suture System combines USP size 4-0 braided nonabsorbable poly(ethylene terephthalate), or polyester, surgical suture with a set of manual instruments designed to efficiently and accurately place the suture for the purpose of performing approximation of tissue and elevation of subdermis/underlying muscle procedure. All components of the Implicitguide™ Surgical Suture System are provided sterile, and are intended for single use only

The manual instruments consist of the following:

  • Tape (marking tape) to aid in marking the skin for needle (Suturod) point of . entry locations)- if needed

{1}------------------------------------------------

  • . Lancet to create a point of entry for needle (Suturod) and suture
  • Clearing device to clear subcutaneous ligamentous attachments around the . puncture site
  • Two (2) stainless steel Suturods™ which, like suture needles, guide the . suture through the tissue. Each Suturod™ has an internal light guide which, when connected to the Light Handle, illuminates the distal tip of the Suturod™M
    • Light Handle which contains an LED light source to connect to Suturod™ .

ર્સ્ડ Intended Use

The Implicitguide™ Surgical Suture System is intended for use in soft tissue approximation and elevation of subdermis and underlying muscle.

Comparison to Predicate Devices ર્ડ.6

  • The Implicitguide™ Surgical Suture System is equivalent in the intended use to a. the predicate device, the Featherlift Extended Length Aptos Threads (Contour Threads) except for the surgical technique. The Featherlift Aptos Threads tissue elevation is achieved by tacking to the temporal fascia. The Implicitguide™ Surgical Suture System tissue elevation is achieved by the placement of the suture over the muscle and/or other soft tissues, with subsequent tensioning and knotting of the suture.
  • b. Suture- The nonabsorbable polyester surgical suture contained in the Implicitguide™ Surgical Suture System is exactly the same suture as the TEVDEK® II polyester suture currently distributed commercially by Teleflex (510[k] nos. K001434 and K021019 by Genzyme Corp.), and meets all requirements of the USP Monograph for Nonabsorbable Surgical Suture
  • c. Device design, physical configuration and materials-Comparisons of design characteristics and physical configurations of the predicate devices have established that Implicitguide™ Surgical Suture System is substantially equivalent in design, physical configuration and materials of composition. The predicate device has two needles, one needle attached at each end of suture. The Implicitguide™ Surgical Suture System has two needles called Suturods, one attached to each end of the suture at the midpoint of the needle. Materials of construction are both stainless steel and materials that have long history of successful use in medical devices. See attachment 1, or Addendum 3, Section 10.4 Device Comparison table.

{2}------------------------------------------------

091061
page 3/5

§510(k) PreMarket Notification, K091061 ImplicitCare, LLC

October 9. 2009

5.7 Summary of Non-Clinical Test

  • The nonabsorbable polyester surgical suture contained in the Implicitguide™ a. Surgical Suture System is exactly the same suture as the TEVDEK® II polyester suture currently distributed commercially by Teleflex (510[k] nos. K001434 and K021019 by Genzyme Corp.).
    Teleflex Medical has issued a certification to ImplicitCare, that tests on the Teleflex suture have been completed, test profile and results comply with AAMI/ISO 10993-1 requirements and FDA's Blue Book memorandum (G95-1) to assure safe patient contact of an Implant Device contacting Tissue, Bone, or Blood permanently. All results are acceptable. See attachment 2.

The following are the tests conducted:

  • Physico-chemical: Plastics, USP ●
  • Cytotoxicity:-ISO MEM Elution .
  • Sensitization:-ISO Murine Lympth Node Asay ●
  • Irritation: ISO Intracutaneous Reactivity Test .
  • Systemic Toxicity: ISO Acute Systemic Injection Test �
  • Genotoxicity: ISO Salmonella typhyimurium Reverse Mutation .
  • . Hemocompatibility: Hemolysis test (NIH Method)
  • Pyrogen: ISO Materials Mediated Rabbit Pyrogen .
  • . Implantation: 7 day and 90 day Muscle implant

These tests demonstrate that the suture contained in the Implicitguide™ Surgical Suture System is as safe as the predicated devices.

Since the nonabsorbable polyester surgical suture contained in the Implicitguide™ Surgical Suture System is exactly the same suture as the TEVDEK® II, and, the subject device will meet all requirements of USP XXXI <871> Needle attachment - Sutures. As such, no additional non-clinical testing is necessary to further demonstrate substantial equivalence to other polyester sutures marketed in the U.S.

  • b. Cadaver Tests. The study was conducted on four cadavers, two females and two males. The two males are both 58 years old and the females are 92 and 84 years old. The procedure was conducted by Board Certified plastic surgeons, Dr. Gregory Mueller and Dr. Sherrell Aston. Both of their CV's are on file. Procedure was conducted on August 31, 2009, in University of Maryland, State Anatomy Board, 655 West Baltimore Street Room B-023, Baltimore, MD 21201

{3}------------------------------------------------

K091061 Page 4/5

ADDENDUM NUMBER 2-510(k) SUMMARY

October 9, 2009

Tests completed:

    1. Brow elevation and shaping.
    1. Suturod: Trans-Illumination and color of light.
  • Tensile strength of medial platysmal border muscles, and facial retaining 3. ligaments.

Test results are acceptable and test reports are on file. See attachment 3 for summary. These tests were conducted to demonstrate substantial equivalence to predicate device, Contour Threads, with added advantage of the Suturod light transillumination.

  • EMC Evaluation for the (Surgical) Light tool of Suturod (Needle). ﻦ ﺗ
    This evaluation was conducted in accordance with the specifications for medical electrical equipment as specified by International Electrotechnical Commission (IEC) and the European Committee for Electrotechnical Standardization (CENELEC).

Test specification: Radio Frequency Emissions and Electromagnetic Immunity Tests in accordance with requirements of IEC 60601-1-2:2002/A1:2004 and EN60601-1-2:2007

Test Results: All test required passed requirements. See attachment 4 for Certificate and summary of test results. Complete test report on file.

This report demonstrates the safety of the Suturod.

  • d. Design Verification Tests
    Design Verification Tests have been completed. All test results are acceptable.

The following test report are on file. See attachment 5 for summary of Design validation

  • -TR090109-01 Design Verification Clearing Tool Torque Test
  • -TR090109-02 Design Verification Lancet Blade Puncture Test
  • -TR090109-03 Design Verification Suture Length Test
  • -TR090109-04 Design Verification Suture Attachment force Test
  • Design Verification Suturod Bending Test -TR090109-05
  • -TR090109-06 Design Verification Light Intensity Test
  • -TR090109-07 Design Verification Light on- time Test

These tests demonstrate that the device meets design and USP requirements

{4}------------------------------------------------

K09 1061
page 5/5

October 9, 2009

5.8 Summary of Clinical Tests

No clinical testing was conducted to support this submission.

5.9 Conclusions

The results of our comparison demonstrated the substantial equivalence of the subject device to the identified predicate device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with outstretched wings, rendered in a simple, abstract design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

NOV 1:3 2009

Implicit Care, LLC % Ms. Dionicia B. Reblando Vice President, RA/QA 5920 Friars Road, Suite 102 San Diego, California 92108

Re: K091061

Trade/Device Name: Implicitguide™ Surgical Suture System Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: October 23, 2009 Received: October 28, 2009

Dear Ms. Reblando:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{6}------------------------------------------------

Page 2 - Ms. Dionicia B. Reblando

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

N. Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):

K091061

Device Name:

Implicitguide™ Surgical Suture System

Indications for Use:

The Implicitguide™ Surgical Suture System is indicated for use in soft tissue approximation and elevation of subdermis and underlying muscle.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Krone for MXM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091061

2

§ 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture.

(a)
Identification. Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.