K Number
K091042

Validate with FDA (Live)

Device Name
CRONO S-PID 50
Manufacturer
Date Cleared
2009-05-12

(29 days)

Product Code
Regulation Number
880.5725
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

the Crono S-PID 50 has been designed for use in subcutaneous infusion of prescribed liquid medicines.

Device Description

Canè s.r.l., a company that specializes in the production of ambulatory pumps, has now produced a new generation of compact pumps: Crono, a perfect combination of high technology and innovative design.

Crono S-PID 50 is an ambulatory syringe infusion pump intended for the controlled administration of liquids into the patients. This means that the pump can be worn during the infusion.

Crono S-PID 50 combines high technology with innovative design. Thanks to the small size of the pump the patients can administer the drug any time during the day without interrupting daily life or leisure activities.

Crono S-PID 50 pump uses 50 ml dedicated syringes.

Crono S-PID 50 has a particular mechanism which pushes directly on the rubber syringe piston, which makes it possible to reach a significant thrust force and high accuracy of administration.

Crono S-PID 50 administers 20 ul per impulse.

In case of occlusion, an innovative infusion control system makes it possible to proceed with the infusion automatically and, after the occlusion is eliminated, to complete it.

Crono S-PID 50 is fitted with a liquid cristal display which shows the time it takes to complete the delivery, the syringe size and the battery charge status.

AI/ML Overview

Here's an analysis of the provided information to describe the acceptance criteria and the study proving the device meets them:

Device: Crono S-PID 50 Ambulatory Infusion Pump

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Predicate Device K052217)Reported Device Performance (Crono S-PID 50)
Intended UseSubcutaneous infusion of prescribed liquid medicines.SAME to Crono Super PID - K052217
Infusion per impulse22 μl20 μl (SAME to Crono PCA 50 - K052218)
Flow rate accuracy+/-2%+/-3% (SAME to Crono PCA 50 - K052218)
Max. Occlusion pressure4.5 bar +/-1 bar3 bar +/-1.5 bar
Capacity10 or 20 ml50 ml (SAME to Crono PCA 50 - K052218)
Physical Size3" x 1.85" x 1.14" (77 x 48x 29 mm)3.3" x 2.15" x 1.65" (84.5 x 55 x 42 mm) (SAME to Crono PCA 50 - K052218)
Weight4.0 oz (115 g)4.85 oz (140 g) (battery included) (SAME to Crono PCA 50 - K052218)
BatteryPower Source Lithium battery (3V) of the 123 A type.SAME

Note regarding acceptance criteria: The document specifically states that the Crono S-PID 50 is "SAME to Crono Super PID - K052217" for its intended use, implying that the acceptance criterion for intended use is alignment with that predicate device. For other characteristics, the comparison is often made to Crono PCA 50 (K052218), indicating that the acceptance criteria for the new device are based on either of the two predicate devices, with the new device demonstrating equivalence or improvement in performance. The table above reflects these comparisons.

2. Sample size used for the test set and the data provenance

The document states: "The results of bench, EMC, and user testing indicates that the new device is as safe and effective as the predicate device." However, no specific sample sizes for the test set (number of devices, number of tests, specific test parameters or conditions) are provided. The data provenance is implied to be from bench testing, EMC (Electromagnetic Compatibility) testing, and user testing, conducted by Canè S.p.A. in Italy. The study appears to be prospective in the sense that these tests were performed explicitly for the submission of the Crono S-PID 50.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided. The document refers to "bench, EMC, and user testing" but does not detail the nature of "user testing" or if any experts were involved in establishing "ground truth" beyond standard engineering and testing procedures. For a medical device, ground truth for performance metrics would typically be established through validated testing methodologies and measurements, not necessarily expert consensus in the way a diagnostic imaging study would.

4. Adjudication method for the test set

This information is not provided. Adjudication methods are typically relevant for studies involving subjective assessments or interpretation by multiple human readers, which is not clearly described as part of the Crono S-PID 50's testing. For bench and EMC testing, objective measurements would be compared against predefined specifications.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was performed or mentioned. The device is an ambulatory infusion pump, not a diagnostic AI system, so an MRMC study with human readers assisting AI or vice-versa is not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a medical pump, not an algorithm or AI system for diagnostic or interpretive tasks. The "standalone" performance would refer to the pump's mechanical and electronic functioning as a device. The "bench, EMC, and user testing" described would represent the standalone performance evaluation of the device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance would be established through objective engineering measurements and validated test methodologies against specified performance parameters (e.g., flow rate accuracy, occlusion pressure, infusion per impulse). For instance, flow rate accuracy would be tested by precisely measuring the volume delivered over a set period and comparing it to the programmed delivery, with the true value being the precisely measured volume.

8. The sample size for the training set

This information is not applicable and not provided. The Crono S-PID 50 is a medical device (infusion pump), not a machine learning or AI model that requires a "training set."

9. How the ground truth for the training set was established

This information is not applicable and not provided as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "CANE" next to a plus sign. The plus sign is made up of two thick black lines that intersect in the middle. The word "CANE" is written in a bold, sans-serif font. The letters are all capitalized and are black in color.

Canė S.p.A. - Socio Unico
Direzione e coordinamento CMF S.r.l.
C.F.-P.IVA - Reg. Impr. TO: 04384410017
R.E.A. ΤΟ Ν. 629783
Cap. Soc. Euro 120.000,00 i.v.

K091042

ea Qualità /

Image /page/0/Picture/5 description: The image shows a logo with a stylized "C" at the top, followed by the word "CERMET" in bold, sans-serif font. Below "CERMET" is a rectangular box containing the words "SISTEMA DI GESTIONE CERTIFICATO" stacked on top of each other. The text "REG N. 3506 UNI EN ISO 9001:2000 ISO 13485:2004" is at the bottom of the logo.

MAY 12 2009

EXHIBIT 2

CANĖ S.r.l. Via Pavia, 105/l 10090 Rivoli-Cascine Vica (Torino) Italy Tel .: ++39-011-957.48.72 Fax ++39-011-959.88.80 Contact: Mario Cané, President April 1, 2009

510(k) Summary

  1. Identification of the Device: Proprietary - Trade Name: Crono S-PID 50. Classification Name: 80 FRN. Common/Usual Name: Ambulatory Infusion Pump.

2. Equivalent legally marketed devices

This product is similar in function and design to the pumps cleared under 510(k) number K052217 (Crono Super PID) and K052218 (Crono PCA 50).

3. Indications for Use (intended use)

The portable Crono S-PID 50 infusion devices have been designed for use in subcutaneous infusion of prescribed liquid medicines.

4. Description of the Device

Canè s.r.l., a company that specializes in the production of ambulatory pumps, has now produced a new generation of compact pumps: Crono, a perfect combination of high technology and innovative design.

Crono S-PID 50 is an ambulatory syringe infusion pump intended for the controlled administration of liquids into the patients. This means that the pump can be worn during the infusion.

Crono S-PID 50 combines high technology with innovative design. Thanks to the small size of the pump the patients can administer the drug any time during the day without interrupting daily life or leisure activities.

Crono S-PID 50 pump uses 50 ml dedicated syringes.

Crono S-PID 50 has a particular mechanism which pushes directly on the rubber syringe piston, which makes it possible to reach a significant thrust force and high accuracy of administration.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "CANE" in bold, block letters next to a symbol that resembles a plus sign. The plus sign is also made of bold, block shapes. The text and symbol are black against a white background.

MEDICAL TECHNOLOGY

Rep. Imar. TO: 0438 TO N 629783 Soc. Euro 120,000.00 f. Area Qualità

Image /page/1/Picture/4 description: The image shows a logo with the text "SINCERT" at the top. Below that is a circular logo with the letters "C" and "Cermet" inside. The text "Sistema di Gestione Certificato" is below the logo. The bottom of the image contains the text "REG N. 3506, UNI EN ISO 9001:2000, ISO 13485:2004".

Crono S-PID 50 administers 20 ul per impulse.

In case of occlusion, an innovative infusion control system makes it possible to proceed with the infusion automatically and, after the occlusion is eliminated, to complete it.

Crono S-PID 50 is fitted with a liquid cristal display which shows the time it takes to complete the delivery, the syringe size and the battery charge status.

5. Safety and Effectiveness, comparison to predicate device.

The results of bench, EMC, and user testing indicates that the new device is as safe and effective as the predicate device.

6. Substantial Equivalence Chart

CharacteristicCrono PCA 50 - K052218Crono Super PID - K052217Crono S-PID 50
Intended UseSubcutaneous, intravenousand epidural infusion ofprescribed liquid medicines.Subcutaneous infusion ofprescribed liquid medicines.SAME to CronoSuper PID -K052217
Physical characteristicsCrono PCA 50 - K052218Crono Super PID- K052217Crono S-PID 50
Size3.3" x 2.15" x 1.65"(84.5 x 55 x 42 mm).3" x 1.85" x 1.14"(77 x 48x 29 mm).SAME to Crono PCA50 - K052218
Weight4.85 oz (140 g)(battery included).4.0 oz (115 g)(battery included).SAME to Crono PCA50 - K052218
BatteryPower Source Lithium battery(3V) of the 123 A type.Power Source Lithium battery(3V) of the 123 A type.SAME
Infusion per impulse20 μl.22 μl.SAME to Crono PCA50 - K052218
Flow rate accuracy+/-3%.+/-2%.SAME to Crono PCA50 - K052218
Max. Occlusion pressure2.2 bar +/-0.8 bar.4.5 bar +/-1 bar.3 bar +/-1.5 bar.
Capacity50 ml.10 or 20 ml.SAME to Crono PCA50 - K052218

7. Conclusion

After analyzing both bench and user testing data, it is the conclusion of CANÈ S.p.A. that Crono pumps are as safe and effective as the predicate device, has few software differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black against a white background.

Public Health Service

MAY 12 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Carlo Musso Quality Manager Cane S.p.A. Via Pavia, 105/I 10090 Rivoli-Cascine Vica (Torino) ITALY

Re: K091042

Trade/Device Name: Crono S-PID 50 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: April 9, 2009 Received: April 15, 2009

Dear Mr. Musso:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Musso

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Anthony D. Minter Sr.

Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "CANE" in a stylized font. The letters are bold and black, and the word is horizontally oriented. The letters are spaced closely together, and the overall impression is one of a simple, graphic design.

ne e coordinamento CMF 5 Reg. Impr. TO: 04384410017 Area Qualità

Image /page/4/Picture/3 description: The image shows a logo with the text "SINCERT" at the top. Below that is a circular logo with the text "CERMET" below it. Further down, the text "SISTEMA DI GESTIONE CERTIFICATO" is present, followed by "REG. N. 3506", "UNI EN ISO 9001:2000", and "ISO 13485:2004".

j) Indications for Use

510(k) Number (if known): _

Device Name: Crono S-PID 50.

Indications For Use: the Crono S-PID 50 has been designed for use in subcutaneous infusion of prescribed liquid medicines.

Prescription Use X . (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Light Illum. for LCDR, Colburn 05/14/09

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: KO9 10 42

Canè S.p.A. Via Pavia, 105/110090 RIVOLI Cascine Vica (TO) Tel. +39.011.9598880 +39.011.9598880 +39.011.9575089 www.canespa.it - carlo.musso@canespa.it

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).