(24 days)
Not Found
No
The summary describes a standard oscillometric blood pressure monitor and does not mention any AI or ML capabilities. The operational principle is based on traditional pressure sensing and signal processing.
No
The device is a blood pressure monitor, which is a diagnostic tool used to measure blood pressure and pulse rate, not to treat a condition.
Yes
The device is intended to measure diastolic and systolic blood pressures and pulse rate, providing information that can be used to assess an individual's health status, which falls under the definition of diagnostics.
No
The device description explicitly mentions hardware components such as an inflatable cuff, LCD, electronic interface module, and silicon integrate pressure sensor technology, indicating it is not a software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The KD-556 Fully Automatic Electronic Blood Pressure Monitor measures blood pressure and pulse rate using a non-invasive technique (a cuff on the upper arm). It does not analyze samples taken from the body.
- Intended Use: The intended use is to measure blood pressure and pulse rate, not to diagnose or monitor conditions based on in vitro analysis.
Therefore, the device falls under the category of a non-invasive medical device, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Product codes
DXN
Device Description
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.
The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 2×60 times. If any irregular heartbeat is detected, it can be shown on the LCD. It also has the function of averaging the last three measurements.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
upper arm
Indicated Patient Age Range
adult individual
Intended User / Care Setting
medical professionals or at home
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
KD-556 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:
- · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
- · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
- · AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
Results: "the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness."
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
APR 3.0 2009
1.0 submitter's information
Name: Andon Health Co., Ltd. Address: No 31, Changjiang Road, Nankai District, Tianjin, P.R. China Phone number: 86-22-6052 6161 Fax number: 86-22-6052 6162 Contact: Liu Yi Date of Application: 03/30/2009
2.0 Device information
Trade name: | Fully Automatic Electronic Blood Pressure Monitor |
---|---|
Common name: | Noninvasive blood pressure measurement system |
Classification name: | Noninvasive blood pressure measurement system |
3.0 Classification
Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular
4.0 Predict device information
Manufacturer: Andon Health Co., Ltd. KD-591 Fully Automatic Electronic Blood Pressure Monitor Device: 510(k) number: K080319
5.0 Device description
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
1
K090963
$2/3
It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.
The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 2×60 times. If any irregular heartbeat is detected, it can be shown on the LCD. It also has the function of averaging the last three measurements.
6.0 Intended use
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
The intended use and the indication for use of KD-556, as described in its labeling are the same as the predict device KD-591.
7.0 Summary comparing technological characteristics with predicate device
Technological Characteristics | Comparison result |
---|---|
Design principle | Identical |
Appearance | Similar |
Patients contact Materials | Identical |
Performance | Similar |
Biocompatibility | Identical |
Mechanical safety | Identical |
Energy source | Identical |
Standards met | Identical |
Electrical safety | Identical |
EMC | Identical |
Function | Similar |
2
23/3
8.0 Performance summary
KD-556 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:
- · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
- · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
- · AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
- AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
9.0 Comparison to the predict device and the conclusion
Our device KD-556 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-591 whose 510(k) number is K080319.
The two devices are very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Only their appearance and some functions such as 2×60 times memory and averaging the last three measurements are different. The performance parameter of cuff pressure range and overpressure limit are different from the predicted device KD-591. The Environmental parameters is changed, the operational range for humidity ( KD-556 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870. 1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: March 30, 2009 Received: April 06, 2009
Dear Mr. Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Mr. Liu Yi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Statement of Indications for Use
510(k) Number :
Device name: KD-556 Fully Automatic Electronic Blood Pressure Monitor
Indications for use:
KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Prescription use Part 21 CFR 801 Subpart D)
Over-The-Counter Use YES AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) | |
---|---|
Division of Cardiovascular Devices | 4/28/09 |
510(k) Number | K090963 | Page 1 of 1 |
---|---|---|
--------------- | --------- | ------------- |