K Number
K090963

Validate with FDA (Live)

Date Cleared
2009-04-30

(24 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Description

KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers. The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 2×60 times. If any irregular heartbeat is detected, it can be shown on the LCD. It also has the function of averaging the last three measurements.

AI/ML Overview

The KD-556 Fully Automatic Electronic Blood Pressure Monitor is designed and manufactured according to the ANSI/AAMI SP10 standard for manual, electronic, or automated sphygmomanometers. This standard sets forth the accuracy requirements for such devices.

Here's a breakdown of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from ANSI/AAMI SP10)Reported Device Performance (KD-556)
Blood Pressure Accuracy:Conforms to AAMI SP10:2002, A1:2003, and A2:2006
Mean difference between device and reference measurementNot explicitly stated with specific numerical values in the provided document, but implicitly meets the standard's requirements.
Standard deviation of the differencesNot explicitly stated with specific numerical values in the provided document, but implicitly meets the standard's requirements.
Pulse Rate Accuracy:Conforms to AAMI SP10:2002, A1:2003, and A2:2006
Accuracy within specified limitsNot explicitly stated with specific numerical values in the provided document, but implicitly meets the standard's requirements.
Other Performance:
Cuff pressure rangeNot explicitly stated with specific numerical values for KD-556, but noted to be different from the predicate device KD-591.
Overpressure limitNot explicitly stated with specific numerical values for KD-556, but noted to be different from the predicate device KD-591.
Pulse rate range40-180 times/min (changed from predicate device's 30-180 times/min)

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It simply states that the KD-556 conforms to the AAMI SP10 standard, which implies that a study was conducted according to the standard's requirements. These requirements would typically specify sample sizes for clinical validation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide information on the number or qualifications of experts used to establish ground truth for the test set. For blood pressure monitors, ground truth is typically established by trained technicians using a reference sphygmomanometer (e.g., mercury sphygmomanometer) following a specific protocol outlined in standards like AAMI SP10.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method. In the context of blood pressure monitor validation, "adjudication" in the sense of expert consensus for image interpretation is not directly applicable. Instead, the validation involves direct comparison of the device's readings against a reference standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done, and it is not applicable for this type of device. The KD-556 is a standalone blood pressure monitoring device; its performance is measured against a reference standard, not against human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance study was implicitly done. The document states that the device "conforms to the following standards: AAMI SP10:2002, A1:2003, and A2:2006." The AAMI SP10 standard outlines rigorous clinical validation protocols for automated sphygmomanometers, which inherently involve testing the device's accuracy without human intervention influencing the measurement process itself beyond proper application.

7. The Type of Ground Truth Used

The ground truth used for blood pressure monitors following the AAMI SP10 standard is typically a reference measurement obtained by trained observers using a directly auscultated sphygmomanometer (e.g., mercury sphygmomanometer) or an equivalent validated reference device. This involves simultaneous or near-simultaneous measurements by the test device and the reference device on research subjects.

8. The Sample Size for the Training Set

The document does not mention a training set or its sample size. For a device like a blood pressure monitor, the "training set" concept is usually associated with machine learning algorithms. While the device uses an "oscillometric and silicon integrate pressure sensor technology," it's likely pre-calibrated and validated against established standards rather than being "trained" in the typical AI sense.

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned in the provided text, the method for establishing its ground truth is not applicable or provided. The accuracy of this device is primarily established through its adherence to the AAMI SP10 standard, which governs the clinical validation process.

{0}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

APR 3.0 2009

1.0 submitter's information

Name: Andon Health Co., Ltd. Address: No 31, Changjiang Road, Nankai District, Tianjin, P.R. China Phone number: 86-22-6052 6161 Fax number: 86-22-6052 6162 Contact: Liu Yi Date of Application: 03/30/2009

2.0 Device information

Trade name:Fully Automatic Electronic Blood Pressure Monitor
Common name:Noninvasive blood pressure measurement system
Classification name:Noninvasive blood pressure measurement system

3.0 Classification

Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular

4.0 Predict device information

Manufacturer: Andon Health Co., Ltd. KD-591 Fully Automatic Electronic Blood Pressure Monitor Device: 510(k) number: K080319

5.0 Device description

KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

{1}------------------------------------------------

K090963
$2/3

It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.

The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 2×60 times. If any irregular heartbeat is detected, it can be shown on the LCD. It also has the function of averaging the last three measurements.

6.0 Intended use

KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

The intended use and the indication for use of KD-556, as described in its labeling are the same as the predict device KD-591.

7.0 Summary comparing technological characteristics with predicate device

Technological CharacteristicsComparison result
Design principleIdentical
AppearanceSimilar
Patients contact MaterialsIdentical
PerformanceSimilar
BiocompatibilityIdentical
Mechanical safetyIdentical
Energy sourceIdentical
Standards metIdentical
Electrical safetyIdentical
EMCIdentical
FunctionSimilar

{2}------------------------------------------------

23/3

8.0 Performance summary

KD-556 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:

  • · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
  • · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
  • · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
  • · AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
  • AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.

9.0 Comparison to the predict device and the conclusion

Our device KD-556 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-591 whose 510(k) number is K080319.

The two devices are very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Only their appearance and some functions such as 2×60 times memory and averaging the last three measurements are different. The performance parameter of cuff pressure range and overpressure limit are different from the predicted device KD-591. The Environmental parameters is changed, the operational range for humidity (<90%) are changed from the predict device whose operational range for humidity is <80%. The pulse rate range is changed from 30-180 times/min to 40-180 times/min. The MCU is also changed.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the bird symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 3 0 2009

Andon Health Co., Ltd. c/o Mr. Liu Yi Principal Consultant, Regulatory Devices No 31, Changjiang Road Nankai District, Tianjin P.R. China

Re: K090963

KD-556 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870. 1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: March 30, 2009 Received: April 06, 2009

Dear Mr. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Liu Yi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Signature

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Statement of Indications for Use

510(k) Number :

Device name: KD-556 Fully Automatic Electronic Blood Pressure Monitor

Indications for use:

KD-556 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Prescription use Part 21 CFR 801 Subpart D)

Over-The-Counter Use YES AND/OR (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices4/28/09
510(k) NumberK090963Page 1 of 1
-------------------------------------

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).