(293 days)
The TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH IBLOCK is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. It is specifically indicated for: Total Knee Arthroplasty
As the equivalent TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES, the TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH IBLOCK consists of the following major components and subsystems: - The Station (Surgetics or NanoStation), consisting of a mobile computer system and an optical localizer - . Ancillary instruments, a specific motorized cutting block and reflective markers used for reference and registration - TOTAL KNEE SURGETICS software application with iBlock option . The main modification to the predicate device K081232 concerns the modification of the motorized cutting block PRAXITELES®.
The provided text describes a 510(k) summary for the TOTAL KNEE SURGETICS Navigation System with iBlock. Based on the available information, here is an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance | Study Type |
|---|---|---|
| Accuracy of the system for intended use | "Analyses show that the accuracy and performance of the system are adequate for its intended use" | Non-clinical setting (bench testing, specimen) |
| Performance of the system for intended use | "Analyses show that the accuracy and performance of the system are adequate for its intended use" | Non-clinical setting (bench testing, specimen) |
| Device does not raise new safety issues | "tested in a non clinical setting (bench testing, specimen) to assess that no new safety and efficiency issues were raised with this device." | Non-clinical setting (bench testing, specimen) |
| Device does not raise new efficiency issues | "tested in a non clinical setting (bench testing, specimen) to assess that no new safety and efficiency issues were raised with this device." | Non-clinical setting (bench testing, specimen) |
| Performance not reduced compared to predicate device | "and not reduced in comparison to the predicate device." | Comparison to predicate device (K081232) performance via non-clinical testing |
2. Sample Size Used for the Test Set and Data Provenance
The document states, "The TOTAL KNEE SURGETICS Navigation System with iBlock was tested in a non clinical setting (bench testing, specimen)". While it specifies "specimen," it does not provide a specific sample size for the test set or specify the data provenance (e.g., country of origin, retrospective or prospective).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not specify the number of experts used to establish the ground truth for the test set or their qualifications. The testing appears to be non-clinical (bench testing, specimen), which typically relies on established physical standards or measurements rather than expert consensus on images.
4. Adjudication Method for the Test Set
The document does not mention an adjudication method (such as 2+1 or 3+1). Given it's non-clinical testing on specimens, human adjudication in the typical sense for image interpretation would not be applicable.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted for this device. The study described is a non-clinical performance evaluation, not a study involving human readers or AI assistance. The device is a surgical navigation system, not an AI-based diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
The performance testing was a standalone evaluation of the device in a non-clinical setting. The statement "tested in a non clinical setting (bench testing, specimen)" implies testing the system's inherent performance characteristics (accuracy, efficiency) without direct human-in-the-loop clinical scenarios.
7. The Type of Ground Truth Used
The ground truth for the non-clinical testing (bench testing, specimen) would likely be established through physical measurements and engineering specifications to determine the accuracy of the system's guidance and its motorized cutting block. This is not explicitly stated as "expert consensus, pathology, or outcomes data," but rather implied by the nature of non-clinical, benchtop testing for a navigation system.
8. The Sample Size for the Training Set
The document does not provide information about a training set size. This device is a navigation system, and the testing described is for its performance, not for an AI model that requires a training set in the conventional sense. The "TOTAL KNEE SURGETICS software application with iBlock option" likely refers to the operational software, not a machine learning algorithm that needs a training set.
9. How the Ground Truth for the Training Set Was Established
Since there is no mention of a training set in the context of an AI/machine learning model, there is no information about how its ground truth would have been established.
{0}------------------------------------------------
Page 1 of (3)
510(k) Summary
for
TOTAL KNEE SURGETICS Navigation System with iBlock
Submitter Name and Address: 1.
Praxim "Le Grand Sablon" 4, Avenue de l'Obiou 38 700 La Tronche France
心
JAN 2 1 2010
Contact Name: Sebastien Burtin Telephone: (+33) 4 76.54 95 03
Date Prepared: March 13th 2009
2. Device Name:
TOTAL KNEE SURGETICS Navigation System with Proprietary Name: iBlock Common/Usual Name: Image guided surgical navigation system Classification Name: Stereotaxic instrument
3. Predicate Device:
PRAXIM - TOTAL KNEE SURGETICS Navigation System with Praxiteles (K081232)
్లు
{1}------------------------------------------------
4. Intended Use:
Page 2 of 3
The TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH IBLOCK is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures.
It is specifically indicated for:
- Total Knee Arthroplasty o
5. Device Description:
As the equivalent TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES, the TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH IBLOCK consists of the following major components and subsystems:
- The Station (Surgetics or NanoStation), consisting of a mobile computer ● system and an optical localizer
- . Ancillary instruments, a specific motorized cutting block and reflective markers used for reference and registration
- TOTAL KNEE SURGETICS software application with iBlock option .
The main modification to the predicate device K081232 concerns the modification of the motorized cutting block PRAXITELES®.
Technological Characteristics and Substantial Equivalence 6.
The underlying technology of the TOTAL KNEE SURGETICS Navigation System with iBlock is the same as for the predicate device K081232. The system is based on the same operating principle and control mechanism to provide the user with the same kind of information and guidance for the same surgery. The main changes with respect to the predicate device concern the modification of the motorized cutting block Praxiteles. The iBlock® is an improved version of the Praxiteles motorized cutting block and is also controlled by the navigation system. Once the cutting block has been positioned to the desired cut, the surgeon can proceed to the cut as it is done in the conventional way using a traditional saw.
{2}------------------------------------------------
K 0 90 953
್ಗಳ
7. Performance Testing
The TOTAL KNEE SURGETICS Navigation System with iBlock was tested in a non clinical setting (bench testing, specimen) to assess that no new safety and efficiency issues were raised with this device. Analyses show that the accuracy and performance of the system are adequate for its intended use and not reduced in comparison to the predicate device.
In conclusion the modified device TOTAL KNEE SURGETICS Navigation System with iBlock is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0605 Silver Spring, MD 20993-0002
Praxim S.A. % Mr. Sébastien Burtin Regulatory Affairs and Quality Assurance Director Le Grand Sablon 4, Avenue De L'Obiou 38700 La Tronche, France
JAN 2 1 2010
Re: K090953
Trade/Device Name: TOTAL KNEE SURGETICS Navigation System with iBlock Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: December 16, 2009 Received: December 23, 2009
Dear Mr. Burtin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of the Pederal Food, Drug,
You may therefore, market the donice approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misting and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. Sébastien Burtin
comply with all the Act's requirements, including, but not limited to: registration and ligting (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (2) CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolling the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark V. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "BRAHIM" in a stylized, blocky font. Each letter is formed with thick lines, giving it a bold and somewhat industrial appearance. The letters are evenly spaced and aligned horizontally, creating a clear and legible word.
PRAXIM S.A. LE GRAND SABLON 4, AVENUE DE L'OBIQU 38700 Fr TRONCHE France
PHONE +33 4 76 54 95.03 になる +33 × 76 × 1.93 20
હે. ને. ਕਪ CAPITAL DE 3.448.888.800 NO SIRET 400 491 304 00020 CODE APE 331 B
ww.praxim.com
Praxim 510(k)
STATEMENT OF INDICATIONS FOR USE
TOTAL KNEE SURGETICS Navigation System with iBlock
510(k) Number (if known): KO90953
Device Name: TOTAL KNEE SURGETICS Navigation System with iBlock
The TOTAL KNEE SURGETICS Navigation System with iBlock is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures.
It is specifically indicated for :
- Total Knee Arthroplasty
Prescription Use X AND / OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Praxim 510(k) March 13ʰ 2009
Concurrence of CDRH, Office of Device Evaluation (ODE)
TOTA KNEE SURGETICS Navigation System with Division Sign-Off)
CONFIDENTIAL
Page 1/1
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K090952
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).