K Number
K090922

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-06-05

(65 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Empi Active Transcutaneous Nerve Stimulator Device is used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. It is also used as an adjunctive treatment for post-surgical and post-trauma-acute pain

Device Description

The Empi Active Transcutaneous Nerve Stimulator is a battery powered single-function device intended to provide clinicians with a TENS Simple Modulated Pulse (SMP) therapy. This device is intended to be easy to use without the hassle of configuration. Simply turn the device on and increase or decrease the intensity. It is a single-function electrotherapy device with one mode of treatment that allows for Transcutaneous Electrical Nerve Stimulation (TENS).

AI/ML Overview

The provided text describes a 510(k) summary for the "Empi Active Transcutaneous Nerve Stimulator." This document confirms that no clinical studies were required or performed to demonstrate the safety and efficacy of this device.

The submission asserts that the device is substantially equivalent to a predicate device (Empi SELECT Model 4600, K061650) because it has the same intended use, technological characteristics, and does not significantly influence clinical endpoints where prospective clinical studies would be necessary. Therefore, the device meets the criteria for exemption from clinical testing as per FDA guidance document 95-4158.

As no clinical study was conducted, the following information cannot be provided from the given document:

  1. Table of acceptance criteria and reported device performance: No clinical performance data is presented.
  2. Sample size used for the test set and the data provenance: No test set was used for clinical evaluation.
  3. Number of experts used to establish the ground truth for the test set and their qualifications: No ground truth was established from clinical data.
  4. Adjudication method for the test set: Not applicable as no clinical test set.
  5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not conducted.
  6. Standalone (algorithm only without human-in-the-loop performance) study: Not conducted, as this is a medical device, not an algorithm.
  7. Type of ground truth used: Not applicable as no clinical ground truth was established.
  8. Sample size for the training set: Not applicable as no clinical training set was used.
  9. How the ground truth for the training set was established: Not applicable.

Summary of Non-Clinical Testing and Acceptance Criteria (as described in the document):

The Empi Active Transcutaneous Nerve Stimulator underwent non-clinical testing to demonstrate compliance with product specifications and user needs.

Acceptance CriteriaReported Device Performance
Electrical testsMeets product specifications over a range of operating and storage conditions.
Mechanical testsMeets product specifications over a range of operating and storage conditions.
Software testsMeets product specifications over a range of operating and storage conditions.
User needsMeets user needs according to marketing requirements.
Voluntary Standards ConformanceConforms to: - UL 60601-01 Medical electrical equipment Part 1: General requirements for safety. 2003 - CAN/CSA C22.2 No.601.1-M90 with Amendment A2. 2005 - IEC60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety – Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004)

Study Proving Device Meets Acceptance Criteria:

The study proving the device meets the acceptance criteria was non-clinical verification and validation testing. This included:

  • Electrical testing: To ensure electrical safety and performance within specified parameters.
  • Mechanical testing: To assess the durability and physical integrity of the device.
  • Software testing: To verify the functionality and reliability of the device's embedded software.
  • Validation testing: To confirm the device meets user needs and marketing requirements.

The document explicitly states: "No prospective clinical studies are required to demonstrate safety and efficacy of the device in support of an application for premarket clearance in the target markets." The justification for this is that the device "does not differ from the predicate device in technological characteristics or intended use, where the device has a significant influence on clinical endpoints, and where prospective clinical studies would be necessary to determine safety and efficacy equivalence."

{0}------------------------------------------------

6090922 510 上:

JUN - 5 2009

Image /page/0/Picture/3 description: The image shows the word "Empi" in a stylized font. The letters are all lowercase except for the "E", which is uppercase. There is a registered trademark symbol to the right of the "i".

510(k) Summary

510(k) Owner's Name:Address:Empi, Inc.Clear Lake Industrial ParkClear Lake, SD 57226
Phone number:Fax number:651-415-7344651-415-7497
Email:Contact person:Date prepared:swalrod@empi.comSandra Walrod, Regulatory Affairs AssociateMarch 30, 2009
Trade name:Common name:Empi Active Transcutaneous Nerve StimulatorTENS
Classification name:Transcutaneous Electrical Nerve Stimulator (21 CFR 882.5890)
Product Code:GZJ, NYN
Classification:II
Predicate device(s):Empi SELECT Model 4600 Transcutaneous Nerve StimulatorK061650

Device Description:

The Empi Active Transcutaneous Nerve Stimulator is a battery powered single-function device intended to provide clinicians with a TENS Simple Modulated Pulse (SMP) therapy. This device is intended to be easy to use without the hassle of configuration. Simply turn the device on and increase or decrease the intensity. It is a single-function electrotherapy device with one mode of treatment that allows for Transcutaneous Electrical Nerve Stimulation (TENS).

Intended Use:

The Empi Active Transcutaneous Nerve Stimulator is indicated for use as a Transcutaneous Electrical Nerve Stimulation (TENS) device. This is the same indication as for the Empi SELECT device. The Empi Active Transcutaneous Nerve Stimulator incorporates a TENS

Empi

i

{1}------------------------------------------------

Simple Modulated Pulse mode for relief and management of chronic and intractable pain. The Empi SELECT incorporates the very same TENS Simple Modulated Pulse stimulation treatment mode for relief and management of chronic and intractable pain. Both devices are prescription devices to be used under the direction of a medical professional. Both devices are designed to be used in a clinic setting or by the patient at home. Therefore, the intended use of the Empi Active Transcutaneous Nerve Stimulator are the same as those for the Empi SELECT.

Comparison to predicate:

This submission is intended to demonstrate that the Empi Active Transcutaneous Nerve Stimulator has the same intended use as the predicate device and that there are no changes to the fundamental scientific technology and the basic considerations described in the guidance.

The Empi Active Transcutaneous Nerve Stimulator is indicated for use as a Transcutaneous Electrical Nerve Stimulation (TENS) device. This is the same indication as for the Empi SELECT device. The Empi Active Transcutaneous Nerve Stimulator incorporates a TENS Simple Modulated Pulse mode for relief and management of chronic and intractable pain. The Empi SELECT incorporates the very same TENS Simple Modulated Pulse stimulation treatment mode for relief and management of chronic and intractable pain. Therefore, the indications for use of the Empi Active Transcutaneous Nerve Stimulator are the same as those for the Empi SELECT. Both devices are prescription devices to be used under the direction of a medical professional. Both devices are designed to be used in a clinic setting or by the patient at home. Therefore, the intended use of the Empi Active Transcutaneous Nerve Stimulator are the same as those for the Empi SELECT.

Non-clinical Testing:

Verification of the Empi Active Transcutaneous Nerve Stimulator includes electrical, mechanical, and software tests to show that the device meets its product specifications over a range of operating and storage conditions. Validation testing for the Empi Active Transcutaneous Nerve Stimulator includes testing to show the device meets user needs according to marketing requirements.

The Empi Active device conforms to the following voluntary standards:

  • UL 60601-01 Medical electrical equipment Part 1: General requirements for safety . 2003
  • CAN/CSA C22.2 No.601.1-M90 with Amendment A2. 2005 �
  • IEC60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety � - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004)

Clinical Testing:

1

No prospective clinical studies are required to demonstrate safety and efficacy of the device in support of an application for premarket clearance in the target markets. The Empi Active

{2}------------------------------------------------

Transcutaneous Nerve Stimulator does not differ from the predicate device in technological characteristics or intended use, where the device has a significant influence on clinical endpoints, and where prospective clinical studies would be necessary to determine safety and efficacy equivalence. The Empi Active Transcutaneous Nerve Stimulator does not fit the profile of devices that might require clinical data per FDA guidance document 95-4158.

Conclusion:

The Empi Active Transcutaneous Nerve Stimulator is substantially equivalent the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Empi, Inc. c/o Virginia L. Conger Director of Quality and Regulatory 205 Hwy 22 East Clear Lake, SD 57226-0709

JUN - 5 2009

Re: K090922

Trade/Device Name: Empi Active Transcutaneous Nerve Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: May 11, 2009 Received: May 15, 2009

Dear Ms. Conger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Virginia L. Conger

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Kevin Alexander Jr.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Empi

Empi SELECT Special 510(k) Submission

Indications for Use

K090922 510(k) Number (if known):

Device Name: Empi Active Transcutaneous Nerve Stimulator

Indications for Use:

The Empi Active Transcutaneous Nerve Stimulator Device is used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. It is also used as an adjunctive treatment for post-surgical and post-trauma-acute pain

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K090922

(Division Sign-Off) Division of Ophthalmic and Ear, Jose and Throat Devices

510(k) Number K090522

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).