(141 days)
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner. These gloves are not intended to be used as a chemical barrier.
Sunmax Enterprise Shanghai Co Powder free blue Nitrile Examination Glove is a class II device having product code 80LZA. It is a disposable device that meets all requirements of ASTM D 631900a-05 and is tested with chemotherapy drugs.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The device is a patient examination glove, and its performance is assessed against established ASTM standards and specific physical characteristics. The document doesn't explicitly state "acceptance criteria" as a distinct section but lists characteristics and test results that imply acceptance based on meeting these values or passing these tests.
| Characteristic/Test | Acceptance Criteria (Implied by Predicate/Standard) | Reported Device Performance (Sunmax Device) |
|---|---|---|
| Physical Dimensions (Size Medium) | ||
| Width (size medium) | e.g., ~92mm (based on predicates) | 89mm |
| Overall length | 240mm (based on predicates) | 240mm |
| Palm thickness | e.g., ~0.17mm (based on predicates) | 0.12mm |
| Finger thickness | e.g., ~0.18mm (based on predicates) | 0.12mm |
| Mechanical Properties | ||
| Tensile strength pre-aging min | 21mpa (based on predicates for similar devices) | 18mpa |
| Tensile strength after-aging min | 16mpa (based on predicates for similar devices) | 18mpa |
| Ultimate elongation pre-aging min | 500 (based on predicates for similar devices) | 600 |
| Ultimate elongation after-aging min | 500 (based on predicates for similar devices) | 570 |
| Biocompatibility | ||
| Meets Biocompatibility | Yes (based on predicates) | Yes |
| Duration of biocompatibility | Limited (based on predicates) | Limited |
| Skin irritation test | Passes | Passes (Result of Primary Skin Irritation Test) |
| Dermal sensitization | Passes | Passes (Result of Dermal Sensitization Test) |
| Barrier Performance | ||
| Residual powder test | Passes | Passes |
| Permeation testing per ASTM D 6978-05 | Passes (Performance against specific chemotherapy drugs) | Passes (Table of breakthrough times for 9 chemo drugs) |
| Iodine Test | Passes | Passes |
| Barrier strength | Meets requirements of ASTM D6319-00a | Meets the requirements of ASTM D6319-00a |
| General Glove Requirements | ||
| Meets ASTM D 6319-00a-05 (overall standard) | Meets requirements | Gloves meets the requirements of ASTM D6319-00a (for tensile strength) and overall testing meets requirements |
| Inspection level, AQL (pinholes) | S-2, AQL 4.0, AQL 2.5 (pinholes) | (Implied by adherence to ASTM D6319-00a standards) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a separate "test set" sample size for the various tests. It generally refers to testing being conducted to meet ASTM standards. For pinholes and general inspection, it mentions "Inspection level S-2, AQL 4.0, pinholes at AQL 2.5," which are statistical sampling plans defined within the ASTM standard, but not explicit sample numbers.
- Data Provenance: Not explicitly stated. The manufacturer is in Shanghai, China, so it's most likely that the testing was performed there. It's retrospective in the sense that the data is presented as a summary of completed tests for the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This device is a medical glove, and "ground truth" as typically defined for diagnostic or AI-driven devices (e.g., expert consensus on images, pathology reports) is not relevant. The performance is assessed through standardized physical, chemical, and biological tests.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like "2+1" are used for expert review where there might be disagreement, usually for image interpretation or similar qualitative assessments. For material property testing, the results are quantitative and objective measurements or pass/fail criteria against a standard.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-driven device or a diagnostic device where human readers would be involved in interpretation.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device (glove), not an algorithm or software.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by:
- ASTM Standard Specifications: Specifically, ASTM D 6319-00a-05 for physical properties and dimensions.
- Permeation Testing per ASTM D 6978-05: For chemotherapy drug resistance, the "ground truth" is whether the glove prevents breakthrough for a specified duration according to this standard.
- Biocompatibility Testing: "Passes" results for dermal sensitization and primary skin irritation tests establish the ground truth for biocompatibility.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable. No training set is involved.
{0}------------------------------------------------
EXHIBIT #3
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
1. Applicant:
Sunmax Enterprise Shanghai Co., Ltd No.2 New Industrial Area, Zhu Hang Zhen Jin Shan Xian, Shanghai, China
JUL - 1 2009
- Manufacturer: Sunmax Enterprise Shanghai Co., Ltd No.2 New Industrial Area, Zhu Hang Zhen Jin Shan Xian, Shanghai, China
3. Submitter:
Mr. Jigar Shah Official Correspondent for Sunmax Enterprise Shanghai Co., Ltd
4. Address:
mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021 Tel: 516-482-9001 Fax: 516-482-0186 jigar@mdiconsultants.com
5. Trade/proprietary Name:
Sunmax Enterprise Shanghai Co Powder free Blue Nitrile Patient Examination Glove tested with chemotherapy drugs.
-
- Common Names: POWDER-FREE Patient Examination Glove
7. Classification name: Patient Examination Glove
-
- Classification number: 21 CFR 880.6250
{1}------------------------------------------------
9. Device Description:
Sunmax Enterprise Shanghai Co Powder free blue Nitrile Examination Glove is a class II device having product code 80LZA. It is a disposable device that meets all requirements of ASTM D 631900a-05 and is tested with chemotherapy drugs.
10. Intended Use:
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner. These gloves are not intended to be used as a chemical barrier.
11. Substantial Equivalence Discussion:
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is substantially equivalent to the predicate devices.
| Characteristicand parameters | SunmaxEnterpriseShanghai Co.,LTD (NewDevice) | Shanghai ChinaStar Corp(K) 071072 | MEDLINEINDUSTRIES,INC(K) 040841 | SubstantialEquivalence(SE) |
|---|---|---|---|---|
| Product Code | LZA | LZA | LZA | |
| Intended Use | A powder freepatientexaminationglove (Testedfor Use withChemotherapyDrugs) is adisposabledevice intendedfor medicalpurposes that isworn on theexaminers handor finger topreventcontaminationbetween patientand examiner.These glovesare not intendedto be used as achemical barrier | A powder freepatientexaminationglove (Tested forUse withChemotherapyDrugs) is adisposable deviceintended formedical purposesthat is worn onthe examinershand or finger topreventcontaminationbetween patientand examiner.These gloves arenot intended to beused as achemical barrier | MedlinePowder-FreeBlue NitrileExaminationGloves (Testedfor Use withChemotherapyDrugs) is adisposabledevice intendedfor medicalpurposes that isworn on theexaminer's handor finger topreventcontaminationbetween patientand examiner. | SE |
| Width (sizemedium) | 89mm | 92mm | 92mm | Minordifference |
2
{2}------------------------------------------------
| Overall length | 240mm | 240mm | 240mm | SE |
|---|---|---|---|---|
| Palm thickness | 0.12mm | 0.17mm | 0.17mm | |
| Finger thickness | 0.12mm | 0.18mm | 0.18mm | |
| Tensile strengthpre aging min | 18mpa | 21mpa | 21mpa | |
| Tensile strengthafter aging min | 18mpa | 16mpa | 16mpa | |
| Ultimateelongation preaging min | 600 | 500 | 500 | Minordifference |
| Ultimateelongation afteraging min | 570 | 500 | 500 | |
| MeetsBiocompatibility | Yes | Yes | Yes | |
| Duration of bio-compatibility | Limited | Limited | Limited | |
| Skin irritation test | Passes | Passes | Passes | SE |
| Dermalsensitization | Passes | Passes | Passes | |
| Residual powdertest | Passes | Passes | Passes |
- Summary of Testing:
| Test | Results | |
|---|---|---|
| a. | Dermal Sensitization Test | Passes |
| b. | Primary Skin irritation | Passes |
| c. | Permeation testing per ASTM D 6978-05 | Passes |
| d. | Iodine Test | Passes |
| e. | Tensile strength | Gloves meets the requirements ofASTM D63 19-00a. |
| f. | Barrier strength | Gloves meets the requirements ofASTM D63 19-00a. |
The standards used by Sunmax Enterprise Shanghai Co to determine substantial equivalence are based on ASTM D 631900a-2005. All testing meets requirements for physical specifications and dimensions conducted on gloves, Inspection level S-2, AQL 4.0, pinholes at AQL 2.5
There are no special labeling claims and we do not claim our gloves to be hypoallergenic.
{3}------------------------------------------------
13. Conclusion:
Powder free Blue Nitrile Patient Examination Glove tested with chemotherapy drugs performance was equivalent to any other conventional method evaluated. Our evaluation concluded that our device raises no new issues of Safety and effectiveness.
.
4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Public Health Service
JUL - 1 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sunmax Enterprise Shanghai Company, Limited C/O Mr. Jigar Shah mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K090336
Trade/Device Name: Powder-Free, Blue, Nitrile Examination Gloves Tested with
Chemotherapy Drugs
Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: June 25, 2009 Received: June 30, 2009
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Shah
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections $31-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Anthony D.m
Susan Runner, D.D.S., M.A.
Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
EXHIBIT #1
Indications for Use
510(k) Number (if known): K090336
Applicant: Sunmax Enterprise Shanghai Co Ltd
Device Name: Powder free Blue Nitrile Patient Examination Glove tested with chemotherapy drugs.
Indications for Use:
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.
Chemotherapy drugs with their breakthrough times:
| Dacarbazine | 240 minutes |
|---|---|
| Cyclophosphamide(Cytoxan) | 240 minutes |
| Doxorubicin Hydrochloride | 240 minutes |
| Fluorouracil | 240 minutes |
| Cisplatin | 240 minutes |
| Etoposide (Toposar) | 240 minutes |
| Paclitaxel (Taxol) | 240 minutes |
| Thio-Tepa | 23.3 minutes |
| Carmustine | 18.7 minutes |
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shule A Murphy, MD
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090336
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.