K Number
K090236
Date Cleared
2009-04-27

(84 days)

Product Code
Regulation Number
866.6010
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AFP method is an in vitro diagnostic test for the quantitative measurement of alphafetoprotein in human serum and lithium heparinized plasma on the Dimension Vista® System. Measurements of alpha-fetoprotein are used as an aid in managing nonseminomatous testicular cancer when used in conjunction with physical examination, histology/pathology, and other clinical evaluation procedures.

Device Description

The Dimension Vista® AFP method is a homogeneous, sandwich chemiluminescent immunoassay based on LOCI® technology. The LOCI® reagents include two synthetic bead reagents and a biotinylated anti-AFP monoclonal antibody fragment. The first bead reagent (Chemibeads) is coated with an anti-AFP monoclonal antibody and contains chemiluminescent dye. The second bead reagent (Sensibeads) is coated with streptavidin and contains a photosensitizer dye. Sample is incubated with biotinylated antibody and Chemibeads to form bead-AFP-biotinylated antibody sandwiches. Sensibeads are added and bind to the biotin to form bead-pair immunocomplexes. Illumination of the complex at 680 nm generates singlet oxygen from Sensibeads which diffuses into the Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 mm and is a direct function of the AFP concentration in the sample.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Siemens Healthcare Diagnostics Inc. Dimension Vista® AFP Flex® reagent cartridge, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance:

The primary objective of this submission is to add lithium heparinized plasma as an approved sample type. Therefore, the "acceptance criteria" are implied by the comparison to the existing, approved serum sample type. The study demonstrates analytical and clinical agreement between serum and lithium heparinized plasma samples.

ParameterAcceptance Criteria (Implied by Predicate Performance)Reported Device Performance (Lithium Heparin Plasma vs. Serum)
SlopeClose to 1.000.99
Intercept (ng/mL)Close to 0.00-0.05
Correlation CoefficientClose to 1.0000.997

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: 70
  • Data Provenance: The document states "human serum versus lithium heparinized plasma samples," implying human-derived samples. There is no information regarding the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This submission does not involve clinical experts establishing a ground truth for diagnostic accuracy in the traditional sense (e.g., radiologists interpreting images). Instead, it's a comparative analytical study assessing the equivalence of different sample types for the same assay. The ground truth for AFP levels would be the measurement obtained from the predicate device using serum samples. Therefore, this section is not directly applicable.

4. Adjudication Method for the Test Set:

Not applicable. This was an analytical comparison study between two sample types using instrument measurements, not a diagnostic interpretation study requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic device for quantitative measurement, not an imaging or interpretation device that would typically involve human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, a standalone performance assessment was conducted in the sense that the device (Dimension Vista® AFP Flex® reagent cartridge) was used to measure AFP levels in both serum and lithium heparin plasma samples. The reported performance refers to the analytical agreement of these measurements, without human intervention in the result generation itself.

7. The Type of Ground Truth Used:

The "ground truth" in this context is the analytical measurement obtained from the predicate Dimension Vista® AFP method using serum samples. The study aims to demonstrate that lithium heparin plasma samples provide equivalent measurements to these established serum values.

8. The Sample Size for the Training Set:

The document does not explicitly mention a "training set" in the context of an algorithm or machine learning model. This is an update to a reagent cartridge for an established immunoassay. The method itself (LOCI® technology) is a well-defined chemical reaction, not a machine learning algorithm that requires a separate training set for model development.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as explained in point 8.

{0}------------------------------------------------

Siemens Healthcare Diagnostics Inc.
Modification to Dimension Vista® AFP Flex® reagent cartridge 510(k) Premarket Notification

APR 2 7 2009

510(k) Summary for Dimension Vista® AFP Flex® reagent cartridge

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K090236

A. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation:

Manufacturer:Siemens Healthcare Diagnostics Inc.500 GBC Drive, M/S 514P.O. Box 6101Newark, Delaware 19714-6101
Contact Information:Siemens Healthcare Diagnostics Inc.500 GBC Drive, M/S 514P.O. Box 6101Attn: Janet M. FoseTel: 302-631-8826Fax: 302-631-6299

January 30, 2009 Preparation date:

B. Name of Device:

Dimension Vista® AFP Flex® reagent cartridge

C. Regulatory Information:

CFR Section: 866.6010 - Tumor-Associated Antigen Immunological Test System Classification: Class II Classification Panel: Immunology (82) Product Code: LOJ

D. Predicate Device:

Dimension Vista® AFP Flex® reagent cartridge (K071597)

E. Device Description:

The Dimension Vista® AFP method is a homogeneous, sandwich chemiluminescent immunoassay based on LOCI® technology. The LOCI® reagents include two synthetic bead reagents and a biotinylated anti-AFP monoclonal antibody fragment. The first bead reagent (Chemibeads) is coated with an anti-AFP monoclonal antibody and contains chemiluminescent dye. The second bead reagent (Sensibeads) is coated with streptavidin and contains a photosensitizer dye. Sample is incubated with biotinylated antibody and Chemibeads to form

{1}------------------------------------------------

Siemens Healthcare Diagnostics Inc. Modification to Dimension Vista AFP Flex® reagent cartridge 510(k) Premarket Notification

bead-AFP-biotinylated antibody sandwiches. Sensibeads are added and bind to the biotin to form bead-pair immunocomplexes. Illumination of the complex at 680 nm generates singlet oxygen from Sensibeads which diffuses into the Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 mm and is a direct function of the AFP concentration in the sample.

F. Device Intended Use:

The AFP method is an in vitro diagnostic test for the quantitative measurement of alphafetoprotein in human serum and lithium heparin plasma on the Dimension Vista® System. Measurements of alpha-fetoprotein are used as an aid in managing non-seminomatous testicular cancer when used in conjunction with physical examination, histology/pathology, and other clinical evaluation procedures.

G. Substantial equivalence information:

The Siemens Dimension Vista® AFP method modification to add lithium heparinized plasma as a sample type and the predicate Dimension Vista® AFP method are both in vitro diagnostic immunoassays intended for the measurement of alpha-fetoprotein (AFP) as an aid in managing non-seminomatous testicular cancer when used in conjunction with physical examination, histology/pathology, and other clinical evaluation procedures.

Similarities
FeatureDevicePredicate
Intended UseThe AFP method is an in vitrodiagnostic test for the quantitativemeasurement of alpha-fetoprotein inhuman serum and lithium heparinplasma on the Dimension Vista®System. Measurements of alpha-fetoprotein are used as an aid inmanaging non-seminomatoustesticular cancer when used inconjunction with physicalexamination, histology/pathology,and other clinical evaluationprocedures.The AFP method is an in vitrodiagnostic test for the quantitativemeasurement of alpha-fetoproteinin human serum on the DimensionVista® System. Measurements ofalpha-fetoprotein are used as an aidin managing non-seminomatoustesticular cancer when used inconjunction with physicalexamination, histology/pathology,and other clinical evaluationprocedures
StandardizationTraceable to the World HealthOrganization (WHO) Referencepreparation for human AFP(72/225)Traceable to the World HealthOrganization (WHO) Referencepreparation for human AFP(72/225)
Measurement methodChemiluminescent:Homogenous sandwichimmunoassay based on LOCI®technologyChemiluminescent:Homogenous sandwichimmunoassay based on LOCI®technology
Measuring Range$0.5- 1000.0$ ng/mL$0.5- 1000.0$ ng/mL
Sample2 uL2 uL

A comparison summary of the features of the products is included in the following table:

{2}------------------------------------------------

Siemens Healthcare Diagnostics Inc.
Modification to Dimension Vista® AFP Flex® reagent cartridge 510(k) Premarket Notification

Differences
FeatureDevicePredicate
Intended UseThe AFP method is an in vitrodiagnostic test for the quantitativemeasurement of alpha-fetoprotein inhuman serum and lithium heparinplasma on the Dimension Vista®System. Measurements of alpha-fetoprotein are used as an aid inmanaging non-seminomatoustesticular cancer when used inconjunction with physicalexamination, histology/pathology.and other clinical evaluationproceduresThe AFP method is an in vitrodiagnostic test for thequantitative measurement ofalpha-fetoprotein in humanserum on the Dimension Vista®System. Measurements of alpha-fetoprotein are used as an aid inmanaging non-seminomatoustesticular cancer when used inconjunction with physicalexamination,histology/pathology, and otherclinical evaluation procedures
Sample TypesSerum and plasma (lithium heparin)Serum

H. Comments about Substantial Equivalence:

The same scientific technology and manufacturing processes are used for both devices. No formulation or design changes were made for the method modification. Comparative data for equivalency studies on human serum versus lithium heparinized plasma samples demonstrates good analytical and clinical agreement between the sample types. Summary results for the matched pairs study are described below:

Comparative MethodSlopeInterceptng/mLCorrelationCoefficientn
Dimension Vista® AFP0.99-0.050.99770

I. Conclusion:

The Siemens Dimension Vista® AFP method modification to add lithium heparinized plasma as a sample type and the predicate Dimension Vista® AFP method (K071597) are substantially equivalent based on their intended use, device features and performance characteristics as described above.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Siemens Healthcare Diagnostics Inc. c/o Ms Janet M. Fose Regulatory Affairs Specialist 500 GBC Drive, M/S 514 P.O. Box 6101 Newark, DE 19714-6101

APR 9 7 2009

Re: K090236

Trade/Device Name: Dimension Vista® AFP Flex® reagent cartridge Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-associated antigen immunological test system Regulatory Class: II Product Code: LOJ Dated: January 30, 2009 Received: February 2, 2009

Dear Ms. Fose:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the

{4}------------------------------------------------

Page 2 – Ms Janet M. Fose

quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mana M Clan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Dimension Vista® AFP Flex® reagent cartridge

Indication for Use:

The AFP method is an in vitro diagnostic test for the quantitative measurement of alphafetoprotein in human serum and lithium heparinized plasma on the Dimension Vista® System. Measurements of alpha-fetoprotein are used as an aid in managing nonseminomatous testicular cancer when used in conjunction with physical examination, histology/pathology, and other clinical evaluation procedures.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

ia M Chan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K090236

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.