(90 days)
Not Found
Not Found
No
The summary describes software for data management and analysis from an oximeter, including a desaturation detection algorithm. There is no mention of AI, ML, or related concepts, and the description of the algorithm and performance studies does not suggest the use of such technologies.
No
The description states the device is intended for viewing, downloading, analyzing, and printing data from oximeters, and is not a central station or patient monitoring system, indicating it's a data management tool, not for direct therapy.
Yes
Explanation: The "Intended Use / Indications for Use" section states: "It can be used to view the data being recorded in real time; download the recorded data to the PC from the device; analyze and print the recorded data and can be used in sleep screening." Additionally, the "Device Description" section notes it can "view event logs and self diagnostic information from the device and can be used in sleep screening." The phrase "sleep screening" directly implies a diagnostic purpose.
Yes
The device description explicitly states it is "BCI® WW1021" and describes its functions as software interacting with a separate oximeter/device via a USB connection. It does not include or describe any hardware components of its own.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are devices used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information about a person's health.
- Device Function: The description clearly states that the WW1021 is software used with an oximeter/device to view, download, analyze, and print data recorded by that device. It interacts with the oximeter, which measures physiological parameters (likely oxygen saturation and pulse rate) in vivo (within the living body).
- No Specimen Analysis: There is no mention of the WW1021 analyzing any biological specimens. Its function is to manage and process data acquired by another device that is directly connected to a patient.
Therefore, the WW1021 is a data management and analysis software for an oximeter, not a device that performs in vitro diagnostic testing.
N/A
Intended Use / Indications for Use
The WW1021 is intended for use with any oximeter/device which supports the BCI® Communication Protocol 1030 (BCICP 1030), if the device is connected to the PC through the USB port. It can be used to view the data being recorded in real time; download the recorded data to the PC from the device; analyze and print the recorded data and can be used in sleep screening. It can show only visual alarms and cannot annunciate audio alarms. It is not intended to be used as a central station monitoring system or as a patient monitoring system. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, sleep technicians and other qualified health care providers. The WW1021 may be used in the hospital or clinical environment.
Product codes (comma separated list FDA assigned to the subject device)
DQA
Device Description
The aim of this submission is to obtain market clearance for the BCI® WW1021 PC Application Oximetry Software.
The BCI® WW1021 is intended for use with any device which supports the BCI® Communication Protocol (BCICP), if the device is connected to the PC through the USB port. It can be used to configure the device; view the data being recorded in real time; download the recorded data to the PC from the device; analyze and print the recorded data; view event logs and self diagnostic information from the device and can be used in sleep screening. It can show only visual alarms and cannot annunciate audio alarms. It is not intended to be used as a central station monitoring system or as a patient monitoring system. It cannot be used to configure or download data to the PC from devices that are wirelessly connected. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians and other qualified health care providers. BCI® WW1021 may be used in the hospital or clinical environment. Communication between the device and BCI® WW1021 is possible using any standard USB cable.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians and other qualified health care providers. BCI® WW1021 may be used in the hospital or clinical environment.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The BCI WW1021 PC Application Software passed software verification and validation testing.
Clinical Testing: The intent of this study was to re-analyzed clinical data previously gathered for the investigation of the BCI® 1621 Software to determine the clinical accuracy and functionality of the BCI® WW1021 software's desaturation detection algorithm. There were four subjects enrolled. One subject (Subject A) was studied twice. Subjects included those not suspected of having obstructive sleep apnea, those with the diagnosis of sleep apnea or those suspected of having sleep apnea determined by self disclosure. This study was a descriptive, cross-sectional investigation of adult subjects including those not suspected of having obstructive sleep apnea, those with the diagnosis of sleep apnea or those suspected of having sleep apnea determined by self disclosure. This study involved the comparison to clinician scoring of desaturation events using the definition used by the BCI® WW1021 Software. The Clinical Investigation results were 7.49% false negative and 0.17% false positive events. This constitutes a pass condition for both the false negative (less than or equal to 10%) and false positive (less than or equal to 10%) criteria.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
false negative: 7.49%
false positive: 0.17%
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
BCI® WW1621 Oximetry Data Management Oximetry Software
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
APR 9 8 2009
510(k) Summary of Safety and Effectiveness
Manufacturer: | Smiths Medical |
---|---|
Address: | N7 W22025 Johnson Drive |
Waukesha, WI 53186 | |
Telephone Number: | (262) 542-3100 |
Fax Number: | (262) 542-3325 |
Contact Person: | Donald Alexander |
Director Regulatory Affairs | |
Date Prepared: | January 23, 2009 |
Proprietary Name: | BCI® WW1021 PC Application Software |
Common/Classification Name: | Oximetry PC Software |
Product Code: | DQA |
Regulation Number: | 870.2700 (Class II) |
Predicate Devices: | BCI® WW1621 Oximetry Data Management Oximetry |
Software |
Device Description:
The aim of this submission is to obtain market clearance for the BCI® WW1021 PC Application Oximetry Software.
The BCI® WW1021 is intended for use with any device which supports the BCI® Communication Protocol (BCICP), if the device is connected to the PC through the USB port. It can be used to configure the device; view the data being recorded in real time; download the recorded data to the PC from the device; analyze and print the recorded data; view event logs and self diagnostic information from the device and can be used in sleep screening. It can show only visual alarms and cannot annunciate audio alarms. It is not intended to be used as a central station monitoring system or as a patient monitoring system. It cannot be used to configure or download data to the PC from devices that are wirelessly connected. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians and other qualified health care providers. BCI® WW1021 may be used in the hospital or clinical environment. Communication between the device and BCI® WW1021 is possible using any standard USB cable.
Indications for Use
The WW1021 is intended for use with any oximeter/device which supports the BCI® Communication Protocol 1030 (BCICP 1030), if the device is connected to the PC through the USB port. It can be used to view the data being recorded in real time; download the recorded data to the PC from the device; analyze and print the recorded data and can be used in sleep screening. It can show only visual alarms and cannot annunciate audio alarms. It is not intended to be used as a central station monitoring
1
1
system or as a patient monitoring system. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, sleep technicians and other qualified health care providers. The WW1021 may be used in the hospital or clinical environment.
Risk Mitigation Table
Below is a summary of risks common to pulse oximeters and how this submission addresses those risks.
Identified Risk | Mitigation Measures |
---|---|
Inadequate Device Performance | Software Validation (Volume 8) |
Clinical Testing (Volume 10) | |
Improper Use | Proposed Labeling (Volume 7) |
An in-depth risk management analysis, including mitigation measures, was performed on the BCI WW1021 software. The results are provided under Volume 5.
Performance Testing
The BCI WW1021 PC Application Software passed software verification and validation testing. Test reports, including test protocols, pass/fail criteria, results and conclusions, are provided under Volume 8.
Clinical Testing
The intent of this study was to re-analyzed clinical data previously gathered for the investigation of the BCI® 1621 Software to determine the clinical accuracy and functionality of the BCI® WW1021 software's desaturation detection algorithm. There were four subjects enrolled. One subject (Subject A) was studied twice. Subjects included those not suspected of having obstructive sleep apnea, those with the diagnosis of sleep apnea or those suspected of having sleep apnea determined by self disclosure. This study was a descriptive, cross-sectional investigation of adult subjects including those not suspected of having obstructive sleep apnea, those with the diagnosis of sleep apnea or those suspected of having sleep apnea determined by self disclosure. This study involved the comparison to clinician scoring of desaturation events using the definition used by the BCI® WW1021 Software. The Clinical Investigation results were 7.49% false negative and 0.17% false positive events. This constitutes a pass condition for both the false negative (less than or equal to 10%) and false positive (less than or equal to 10%) criteria. A complete test protocol and results can be found under Volume 10.
Conclusion
Supporting information per this premarket submission confirms that the BCI® WW (030 Pulse Oximeter is substantially equivalent to its predicate devices.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
2
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 8 2009
Mr. Donald Alexander Director Regulatory Affairs Smiths Medical PM. Incorporated N7 W22025 Johnson Drive Waukesha, Wisconsin 53186-1856
Re: K090205
Trade/Device Name: BCI WW1021 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA Dated: January 23, 2009 Received: January 28, 2009
Dear Mr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
3
Page 2- Mr. Alexander
and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Cooper
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Indications for Use
510(k) Number (if known):
Device Name: _BC1® WW1021 PC Application Software
Indications for Use:
WW1021 is intended for use with any oximeter/device which supports the BCI® Communication Protocol 1030 (BCICP 1030), if the device is connected to the PC through the USB port. It can be used to view the data being recorded in real time; download the recorded data to the PC from the device; and print the recorded data and can be used in sleep screening. It can show only visual alarms and cannot annunciate audio alarms. It is not intended to be used as a central station monitoring system or as a patient monitoring system. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, sleep technicians and other qualified health care providers. WW1021 may be used in the hospital or clinical environment.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susan Rose
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: KC90005