K Number
K090178
Date Cleared
2009-04-24

(88 days)

Product Code
Regulation Number
892.1200
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PET scanner is intended to be used for whole body, multi-slice, Positron Emission Tomography diagnostic imaging.

Device Description

The Positron Emission Tomography (PET) scanners from Neusoft Positron Medical Systems Corporation image the distribution of positron-emitting radionuclides in the human body. These software-controlled medical devices use positron emitting radioactive isotopes to create cross-sectional tomographic images that show function or metabolism, rather than anatomy, as in conventional scanning techniques.

The PET scanner's software generates quantitative images through Attenuation Correction, which corrects for loss of information due to absorption and scatter by overlying soft tissue, and through theoretical, measured, or hybrid (segmented attenuation correction) Image Reconstruction, which generates transaxial images from acquired data using a filtered back projection or iterative reconstruction process.

The PET scanner consists of Detector System, Gantry, Front-end Electronics, Patient Table, Transmission Source, Operator's Control Workstation, Image Processing Workstation and optional Remote Display Workstation.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study conducted for Neusoft Positron Medical Systems Co., Ltd.'s PET scanner (Attrius™, Attrius L™, Truesight).

Important Note: The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It primarily describes comparisons in technological characteristics and lists various tests performed to ensure safety and effectiveness. It does not present a detailed clinical study with specific acceptance criteria metrics, sample sizes for test/training sets, expert qualifications, or MRMC study results (like an AI improvement effect size) in the way that would typically be found in a later clinical trial report for AI-powered diagnostic devices. The performance evaluation here is largely based on phantom studies (NEMA NU2) and compliance with established standards.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the 510(k) summary, explicit "acceptance criteria" for diagnostic performance in terms of sensitivity, specificity, or AUC against a clinical ground truth are not provided. Instead, the performance is demonstrated through compliance with established industry standards for PET scanners using phantoms.

Acceptance Criteria Category (Implied by FDA submission)Reported Device Performance (Summary from text)
Technological Equivalence / Performance Standards Compliance (NEMA NU2)"The performance test data of the candidate device have been obtained using NEMA NU2 phantoms, under the test procedure of NEMA NU2 standards."
Safety TestingIEC 60601-1 Safety tests, IEC 60601-1-1 Safety tests performed.
Electromagnetic Compatibility (EMC)IEC 60601-1-2 Electromagnetic Compatibility test performed.
Software Validation"Software validation, US FDA Software Used in Medical Devices Guidance" performed.
Radiation Measurement"Radiation measurement" tests performed.
Substantial Equivalence (Overall Conclusion)"All the above differences have been verified and determined to not affect safety and effectiveness of the device. We submit that the candidate device and the predicate device are substantially equivalent."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): The document does not specify a sample size of patient cases for a clinical test set. The primary performance evaluation cited is a phantom study. Phantom studies use physical objects designed to mimic human tissue for reproducible testing, not patient data.
  • Data Provenance: Not applicable for a phantom study. The NEMA NU2 phantoms are standardized physical objects.
  • Retrospective or Prospective: Not applicable for a phantom study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable. The ground truth for NEMA NU2 phantom studies is inherently known by the phantom's design and physical properties (e.g., known radioactivity concentration, dimensions of objects). No human experts are used to establish this kind of ground truth.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. As discussed, the test set is based on NEMA NU2 phantoms, where the "ground truth" is objectively defined by the phantom itself and its preparation, not by expert consensus or adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No, an MRMC comparative effectiveness study is not mentioned or implied in this 510(k) summary. The summary focuses on the device's technical performance and safety, primarily through phantom testing and comparisons to a predicate device's technical specifications. This type of study would typically be performed for AI algorithms designed to assist human readers in interpretation.
  • Effect Size of Human Reader Improvement: Not applicable, as no MRMC study was performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Standalone Performance: The performance testing reported (NEMA NU2) is essentially a standalone evaluation of the PET scanner's image acquisition and reconstruction capabilities. However, this is for the imaging device itself, not for a separate AI algorithm designed for interpretation. The phrase "algorithm only" or "standalone" typically refers to the diagnostic accuracy of an AI model in performing a task (e.g., detecting lesions) without human intervention. This document describes the performance of the PET scanner as a whole system. Therefore, while technically an "algorithm" (reconstruction software) is evaluated in a standalone manner, it's not in the context of an AI-driven interpretive tool.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the performance evaluation cited is primarily physical phantom properties (NEMA NU2 phantoms). These phantoms have known geometric configurations and radioactivity distributions, allowing for objective measurement of parameters like spatial resolution, sensitivity, and image uniformity.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable/Not provided. This document does not describe the development or training of an AI algorithm in the contemporary sense. The PET scanner uses "3rd party OSEM image reconstruction software" and other 3rd party software (Linux, RTAI). These are established software components for image reconstruction and operating systems, not typically "trained" on a dataset in the way a modern deep learning AI model would be. Therefore, there is no mention of a dedicated training set.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as no training set for an AI algorithm is mentioned. The software components (OSEM, Linux, RTAI) would have been developed and validated by their respective creators through their own, unspecified methods, prior to being integrated into this PET scanner.

{0}------------------------------------------------

K090178
Page 1 of 4
510(k) (Traditional) Submission
Section 5 510(k) Summary

NEUSOFT POSITRON MEDICAL SYSTEMS CO., LTD.

510(k) Summary

Date: _ 01/21/2009

APR 2 4 2009

Company:

1

510(k) ownerNeusoft Positron Medical Systems Co., Ltd.
AddressNo.16 Shiji Road, Hunnan New District, Shenyang, P.R. China
Telephone86-24-83662197
Fax86-24-83661796
ContactSun Yan, Chief Quality Engineer, Quality Management Department
U.S. AgentDelphi Consulting Group
ContactJ. Harvey Knauss
Telephone713-723-4080
Fax713-723-0786

Device:

Trade NameAttrius™, Attrius L™, Truesight
Common NamePositron Emission Tomography (PET) Scanner
Classification NameSystem, tomography, computed, emission
Device Class21 CFR 892.1200 Class II
Product CodeKPS
Classification panelRadiology
Predicate DeviceK022001, mPower, Model 712, Positron Corp.

Device description summary:

The Positron Emission Tomography (PET) scanners from Neusoft Positron Medical Systems Corporation image the distribution of positron-emitting radionuclides in the human body. These software-controlled medical devices use positron emitting radioactive isotopes to create cross-sectional tomographic images that show function or metabolism, rather than anatomy, as in conventional scanning techniques.

The PET scanner's software generates quantitative images through Attenuation Correction, which corrects for loss of information due to absorption and scatter by overlying soft tissue, and through theoretical, measured, or hybrid (segmented attenuation correction) Image Reconstruction, which generates transaxial images from acquired data using a filtered back projection or iterative reconstruction process.

{1}------------------------------------------------

The PET scanner consists of Detector System, Gantry, Front-end Electronics, Patient Table, Transmission Source, Operator's Control Workstation, Image Processing Workstation and optional Remote Display Workstation.

Intended Use:

The PET scanner is intended to be used for whole body, multi-slice, Positron Emission Tomography diagnostic imaging.

Environment of Use / Patient Population:

The PET scanner is intended to be used for imaging individual patients in a controlled medical facility, such as a hospital, medical clinic, or mobile medical unit environment.

Comparison of Major Technological Characteristics:

The main design differences between the predicate device, the mPower™ PET scanner, and the candidate device, the NPMS PET scanner, are listed below:

  • The system hardware dead time of the candidate device is reduced relative to the 1. predicate device, e.g., the predicate device has a Discriminator dead time period of 520 nanoseconds, whereas the candidate device Discriminator dead time period is 400 nanoseconds; the predicate device has a coincidence dead time period of 160 nanoseconds and encoder dead time period of 42 nanoseconds, whereas the candidate device has a coincidence dead time period of 60 nanoseconds and encoder dead time period of 30 nanoseconds.
  • The predicate device Front End Electronics include 64 Discriminator boards, 8 2. Encoder boards, 4 LAFD boards, 1 Coincidence board, and 1 Arbitration and Transfer board, whereas the candidate device integrates the same functionality into 8 Bank Encoder Boards (BEB) and 1 Data Assembly Board (DAB) .
    1. The predicate device uses 34 fans and blowers to pull cooling air through the Gantry and Front End Electronics enclosures. TThe candidate device uses 4 fans located in the top cover of the Gantry to pull cooling air through the Gantry.
  • The predicate device rod source loading mechanism requires manual intervention to 4. switch between the high activity transmission rod source and low activity emission calibration rod source whereas the candidate device provides an automated rod source loading mechanism that loads either the high activity transmission rod source or the low activity emission calibration source without manual intervention.

{2}------------------------------------------------

  • The predicate device has a single LED display on the front cover of the Gantry along 5. with two pushbutton control panels. The candidate device has two LCD displays combined with pushbutton control panels on the front cover of the Gantry, along with two pushbutton control panels on the sides of the Gantry (accessible from behind the Gantry), and an LED display on the rear cover of the Gantry.
  • The predicate device provides a stop switch which makes the device slow down and 6. stop gradually, whereas the candidate device provides emergency stop switches which shut off the power of table, septa and wobble, and so the moving components of the device stop quickly.
    1. The predicate device patient table weight limit is 159 kg. The candidate device patient table weight limit is 200 kg.
    1. The predicate device patient table is a fixed height design. The candidate device patient table can move over a 24cm vertical range from a loading level to a working level.
  • റ്റ. The predicate device utilizes one Ge-68 4mCi transmission rod source. The candidate device utilizes one Ge-68 5mCi transmission source.
    1. The candidate device includes an updated control workstation and image processing workstation.
    1. The predicate device data acquisition system is a rack mounted VMEbus-based real time computer running the VxWorks real-time operating system on a MC68030-based single board computer. The data acquisition and processing functions are executed on a set of proprietary VME/VSBbus compatible boards. The candidate device provides a Gigabit Ethernet interface on the Data Assembly Board that transfers the event data and system control data to the control workstation where the data are processed by software.
    1. The candidate device provides the same 30 party OSEM image reconstruction software used on the predicate device. In addition, the candidate device uses the following 30 party software: Linux Operating System, RTAI (Real Time Application Interface). These 30 party software packages were not used on the predicate device.
    1. The performance test data of the candidate device have been obtained using NEMA NU2 phantoms, under the test procedure of NEMA NU2 standards and detailed in section 18 of the submission.

{3}------------------------------------------------

NEUSOFT POSITRON MEDICAL SYSTEMS CO., LTD.

Testing

The following tests can demonstrate the candidate device is as safe, as effective, and performs as well as or better than the predicate device:

  • Radiation measurement ●
  • NEMA NU2 Performance test
  • IEC 60601-1 Safety tests .
  • IEC 60601-1-1 Safety tests .
  • IEC 60601-1-2 Electromagnetic Compatibility test
  • Software validation, US FDA Software Used in Medical Devices Guidance

Conclusions:

All the above differences have been verified and determined to not affect safety and effectiveness of the device. We submit that the candidate device and the predicate device are substantially equivalent.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.

Public Health Service

APR 2 4 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Sun Yan Chief Quality Engineer, Quality Management Department Neusoft Positron Medical Systems Co., Ltd. No., 16, Shiji Road, Hunnan New District Shenyang, Liaoning. 110179 P. R. China

Re: K090178

Trade/Device Name: Attrius™, Attrius LTM, Truesight PET scanner · Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system

Regulatory Class: II Product Code: KPS

Dated: January 26, 2009 Received: January 26, 2009

Dear Mr. Yan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx (Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 884.xxx (Obstetrics/Gynecology) (240) 276-0115 21 CFR 892.xxx (Radiology) (240) 276-0120 Other (240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely wours,

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for use statement

510(k) Number:

K090178

Attrius™, Attrius L™, Truesight PET scanner Device Name:

Intended Use / Indications for Use:

The PET scanner is intended to be used for whole body, multislice, positron emission tomography diagnostic imaging.

Environment of Use / Patient Population:

The PET scanner is intended to be used to image individual patients in a controlled medical facility, such as a hospital, medical center, or mobile medical unit environment.

Prescription Use YES

Over-The-Counter Use NO

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices
510(k) NumberK090178

11

.

73

4-2

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.