K Number
K090057
Device Name
ON-CALL PLUS BLOOD GLUCOSE MONITORING SYSTEM
Date Cleared
2009-04-08

(90 days)

Product Code
Regulation Number
862.1345
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The On-Call® Plus Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in capillary whole blood from the fingertip, forearm, and/or palm by people with diabetes at home and by healthcare professionals as an aid in the monitoring the effectiveness of diabetes control programs. The On-Call® Plus Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates. The On-Call® Plus Blood Glucose control solution is for use with the On-Call® Plus Blood Glucose meter and strips as a quality control check to verify the accuracy of blood glucose test results.
Device Description
The On-Call® Plus Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip, palm, and/or forearm. The glucose, measurement is achieved by using the amperometric detection method. The test strip has a reagent system including glucose oxidase and a mediator that reacts with glucose in the whole blood sample to produce an electrical current is measured by the meter, and after calculation by the meter, the blood glucose concentration reading is displayed on the meter display, calibrated to a plasma reference.
More Information

No
The device description and performance studies focus on electrochemical detection and standard laboratory and clinical testing, with no mention of AI or ML algorithms.

No.
The device is for monitoring glucose levels as an aid in diabetes control, not for directly treating or curing a medical condition.

No

The "Intended Use / Indications for Use" section explicitly states: "The On-Call® Plus Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus..." Its purpose is to monitor the effectiveness of diabetes control programs, not to diagnose the condition itself.

No

The device description clearly outlines hardware components (meter, test strip) and their interaction to perform the measurement. The performance studies also include hardware-related testing (electromagnetic compatibility, electrical safety).

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use explicitly states it's for the "quantitative detection of glucose in capillary whole blood." This involves testing a biological sample (blood) in vitro (outside the body).
  • Device Description: The description details an "electrochemical enzymatic assay" and a "reagent system" that reacts with glucose in the blood sample. This is a classic description of an in vitro diagnostic test.
  • Purpose: The system is used "as an aid in the monitoring the effectiveness of diabetes control programs," which is a diagnostic purpose related to managing a medical condition.

The definition of an IVD generally includes devices used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or screening purposes. This device clearly fits that description.

N/A

Intended Use / Indications for Use

The On-Call® Plus Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in capillary whole blood from the fingertip, forearm, and/or palm by people with diabetes at home and by healthcare professionals as an aid in the monitoring the effectiveness of diabetes control programs.

The On-Call® Plus Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates.

The On-Call® Plus Blood Glucose control solution is for use with the On-Call® Plus Blood Glucose meter and strips as a quality control check to verify the accuracy of blood glucose test results.

Product codes (comma separated list FDA assigned to the subject device)

NBW, CGA, JJX

Device Description

The On-Call® Plus Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip, palm, and/or forearm. The glucose, measurement is achieved by using the amperometric detection method.

The test strip has a reagent system including glucose oxidase and a mediator that reacts with glucose in the whole blood sample to produce an electrical current is measured by the meter, and after calculation by the meter, the blood glucose concentration reading is displayed on the meter display, calibrated to a plasma reference.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Fingertip, palm, and/or forearm

Indicated Patient Age Range

Not Found

Intended User / Care Setting

people with diabetes at home and by healthcare professionals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Laboratory Testing:
The performance characteristics of the On-Call® Plus Blood Glucose Monitoring System were evaluated by performing the following studies: repeatability precision, intermediate precision, linearity, interfering agents, hematocrit effect, temperature effect evaluation - blood & control solution, low battery effect, altitude effect, sample volume, software validation testing, electromagnetic compatibility and electrical safety testing as part of meter and strip validation testing.

Clinical Tests:
Clinical studies were conducted with lay persons and trained laboratory technicians using the On-Call® Plus Blood Glucose Monitoring System. The study data were presented evaluating the system accuracy of the On-Call® Plus Blood Glucose Monitoring System compared to the YSI Model 2300 STAT PLUS (K913806) per the ACON Clinical Study Protocol for the Blood Glucose Monitoring System. Study results indicate that non-professional, inexperienced lay persons were able to obtain comparable blood glucose readings when using the On-Call® Plus Blood Glucose Monitoring System as compared to the results obtained by the trained technicians. In addition, the participating lay persons were questioned and responded as satisfied with the ease of operation by following the Instructions for Use in the User's Manual and the overall performance of the On-Call® Plus Blood Glucose Monitoring System.

Conclusion:
The laboratory testing and clinical study results demonstrate that the On-Call® Plus Blood Glucose Monitoring System is safe, effective and easy-to-use. It also demonstrates that the On-Call® Plus Blood Glucose Monitoring System meets the accuracy requirements per ISO 15197 and as such is substantially equivalent to the One Touch Ultra Blood Glucose Monitoring System, currently sold on the U.S. market (K002134).

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K002134

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K913806

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

11

5. 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The Assigned 510(k) number isk0910057
-----------------------------------------

Submitter's Identification:

ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121

Tel.: 858-875-8019 Fax: 858-875-8099

Date Prepared: January 7, 2009

Contact Person:

Richard Lenart Regulatory Affairs Manager

Proprietary Name of the Device:

On-Call® Plus Blood Glucose Monitoring System

Common Name:

Glucose Test System

Classification Name:

Class II §862.1345 Glucose Test System (To be manufactured and marketed for consumer home and professional use)

Predicate Device:

One Touch Ultra Blood Glucose Monitoring System Lifescan, Inc., Iocated at 1000 Gibraltar Dr., Milpitas, CA 95035, USA. 510(k) Number: K002134

1

Proprietary NameClassificationProCodeDescriptionCommon Name
On-Call® Plus Blood
Glucose Monitoring
System862.1345
Class II75 NBWSystem, Test,
Blood Glucose,
Over The CounterGlucose Test
System
On-Call® Plus Blood
Glucose Meter and
On-Call® Plus Blood
Glucose Test Strips862.1345
Class II75 CGAGlucose MonitorGlucose Meter &
Test Strips
On-Call® Plus Glucose
Control Solution862.1660
Class I75 JJXSingle Analyte
ControlControl Solution
On-Call® Plus Lancets and
On-Call® Plus Lancing
Device878.4800
Class I79 FMKLancet, BloodLancets

Device Name: On-Call® Plus Blood Glucose Monitoring System

Description:

The On-Call® Plus Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip, palm, and/or forearm. The glucose, measurement is achieved by using the amperometric detection method.

The test strip has a reagent system including glucose oxidase and a mediator that reacts with glucose in the whole blood sample to produce an electrical current is measured by the meter, and after calculation by the meter, the blood glucose concentration reading is displayed on the meter display, calibrated to a plasma reference.

Intended Use:

The On-Call® Plus Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in capillary whole blood from the fingertip, forearm, and/or palm by people with diabetes at home and by healthcare professionals as an aid in the monitoring the effectiveness of diabetes control programs.

2

Technological Characteristics:

FeatureSpecification
Measurement Range1.1-33.3 mmol/L (20 to 600 mg/dL)
Result CalibrationPlasma-equivalent
SampleFresh capillary whole blood
Minimum Sample Size1 μL
Test Time10 seconds
Power SourceOne (1) CR 2032 3.0V coin cell battery
Battery Life12 months or approximately 1,000 tests
Glucose Units of MeasureThe meter is pre-set at time of manufacturing to either millimoles per
liter (mmol/L) or milligrams per deciliter (mg/dL) depending on the
country of use standard
MemoryUp to 300 records with time and date
Meter Size85 mm x 54 mm x 20.5 mm
Display Size35 mm x 32.5 mm
WeightApproximately 49.5 g (with battery installed)
Operating Temperature5-45°C (41 - 113°F)
Operating Relative Humidity20-90% (non-condensing)
Hematocrit Range30-55%
Data Port9600 baud, 8 data bits, 1 stop bit, no parity
Sample SiteFingertip, palm and forearm

Specification of Blood Glucose Meter:

3

Comparison to Predicate Devices:

The On-Call® Plus Blood Glucose Monitoring System is substantially equivalent to the One Touch Ultra Blood Glucose Monitoring System, K002134.

| Features | On-Call® Plus Blood Glucose
Monitoring System | One Touch Ultra Blood
Glucose Monitoring
System (K002134) |
|-----------------------------|--------------------------------------------------|-----------------------------------------------------------------|
| Similarities | | |
| Measurement Range | 20 to 600 mg/dL (1.1-33.3 mmol/L) | Same |
| Result Calibration | Plasma-equivalent | Same |
| Sample | Fresh capillary whole blood | Same |
| Minimum Sample Size | 1 µL | Same |
| Assay Method | Glucose oxidase biosensor | Same |
| Power Source | One (1) CR 2032 3.0V coin cell battery | Same |
| Battery Life | 12 months or approximately 1,000 tests | Same |
| Glucose Units of Measure | mg/dL | Same |
| Hematocrit Range | 30-55% | Same |
| Automatic Shutoff | Two minutes after last user action | Same |
| Data Port | One data port | Same |
| Differences | | |
| Test Time | 10 seconds | 5 seconds |
| Memory | Up to 300 records with time and date | 150 blood glucose and control
solution tests |
| Meter Size | 85 mm x 54 mm x 20.5 mm | 3.12" x 2.25" x 0.85" |
| Weight | Approximately 49.5 g (with battery installed) | 1.5 ounces with battery
(Approximately 42 g) |
| Operating Temperature | 5-45°C (41 - 113°F) | 6-44°C (43 - 111°F) |
| Operating Relative Humidity | 20-90% (non-condensing) | 10-90% |
| Sample Site | Fingertip, palm and forearm | Fingertip and forearm |

Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Guidance documents included the "FDA Guidance for Industry In Vitro Diagnostic Glucose Test System" and "FDA Guidance for Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems" as well as "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

Compliance to applicable voluntary standards includes EN ISO 15197:2003 "In vitro diagnostic test systems - Requirements for in vitro whole blood glucose monitoring systems intended for use by patients for self testing in management of diabetes mellitus".

4

Laboratory Testing:

The performance characteristics of the On-Call® Plus Blood Glucose Monitoring System were evaluated by performing the following studies: repeatability precision, intermediate precision, linearity, interfering agents, hematocrit effect, temperature effect evaluation - blood & control solution, low battery effect, altitude effect, sample volume, software validation testing, electromagnetic compatibility and electrical safety testing as part of meter and strip validation testing.

Discussion of Clinical Tests Performed:

Clinical studies were conducted with lay persons and trained laboratory technicians using the On-Call® Plus Blood Glucose Monitoring System. The study data were presented evaluating the system accuracy of the On-Call® Plus Blood Glucose Monitoring System compared to the YSI Model 2300 STAT PLUS (K913806) per the ACON Clinical Study Protocol for the Blood Glucose Monitoring System. Study results indicate that non-professional, inexperienced lay persons were able to obtain comparable blood glucose readings when using the On-Call® Plus Blood Glucose Monitoring System as compared to the results obtained by the trained technicians. In addition, the participating lay persons were questioned and responded as satisfied with the ease of operation by following the Instructions for Use in the User's Manual and the overall performance of the On-Call® Plus Blood Glucose Monitoring System.

Conclusion:

The laboratory testing and clinical study results demonstrate that the On-Call® Plus Blood Glucose Monitoring System is safe, effective and easy-to-use. It also demonstrates that the On-Call® Plus Blood Glucose Monitoring System meets the accuracy requirements per ISO 15197 and as such is substantially equivalent to the One Touch Ultra Blood Glucose Monitoring System, currently sold on the U.S. market (K002134).

5

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Acon Laboratories Co. c/o Mr. Richard Lenart 4108 Sorrento Valley Blvd. San Diego, CA 92121

APR - 8 2009

Re: K090057 Trade/Device Name: On-Call Plus Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA, JJX Dated: January 07, 2009 Received: January 08, 2009

Dear Mr. Lenart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

6

Page - 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Corg C. He

Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

7

Indication for Use

510(k) Number (if known): K090057

Device Name: On-Call® Plus Blood Glucose Monitoring System

Indication For Use:

The On-Call® Plus Blood Glucose Monitoring System is an electrochemical enzymatic assay for the quantitative detection of glucose in capillary whole blood from the fingertip, forearm, and/or palm by people with diabetes at home and by healthcare professionals as an aid in the monitoring the effectiveness of diabetes control programs.

The On-Call® Plus Blood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus, nor intended for use on neonates.

The On-Call® Plus Blood Glucose control solution is for use with the On-Call® Plus Blood Glucose meter and strips as a quality control check to verify the accuracy of blood glucose test results.

Prescription Use (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Rute Chelin

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

K090057 510(k)