K Number
K090027
Device Name
BLOOD PRESSURE MONITOR, ZSBP-003/004/005/006/007/008 AND ZSBP-102/103/104/105/106/107
Date Cleared
2009-02-02

(28 days)

Product Code
Regulation Number
870.1130
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
ZSBP-003/004/005/006/007/008 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the Oscillometric method for the patient of the age over 18 years old. ZSBP-102/103/104/105/106/107 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old. All measurement values can be read out and keep memory on the LCD panel for home care use.
Device Description
Basically the measuring system were composite of blood pressure measuring circuit via Oscillometric method, pressure sensor, measuring cuff at wrist or upper arm, pneumatic pump, inflation and deflation system, housing, display LCD, and measuring software... and so on. The main operation for the blood pressure measurement is carried out in such a way that the measuring cuff at wrist or upper arm is inflated to the estimated pressure level, then deflated to zero automatically. During the inflation, the pressure change with respective of time were recorded as the data base of measurement. Then the following measuring results will be calculated against the measurement data base: - Blood pressure information including systolic and diastolic pressure (calculated via Oscillometric method) - Heart beat rate.
More Information

Not Found

No
The description details a standard oscillometric blood pressure monitor and its components, with no mention of AI or ML in the measurement method, software, or performance studies.

No
The device is a blood pressure monitor used for measurement and monitoring, not for treating or preventing a disease or condition.

Yes
Explanation: The device measures blood pressure and heart rate, which are physiological parameters used to assess and monitor a person's health condition. This aligns with the definition of a diagnostic device that helps in the identification or monitoring of medical conditions.

No

The device description explicitly lists hardware components such as a pressure sensor, measuring cuff, pneumatic pump, inflation and deflation system, housing, and display LCD, in addition to measuring software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device function: This device measures blood pressure and heart rate directly from the body (wrist or upper arm) using the Oscillometric method. It does not analyze samples taken from the body.
  • Intended Use: The intended use is to measure blood pressure and heart rate for home care use. This is a direct physiological measurement, not an in vitro test.

Therefore, this device falls under the category of a non-invasive physiological monitoring device, not an IVD.

N/A

Intended Use / Indications for Use

ZSBP-003/004/005/006/007/008 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the Oscillometric method for the patient of the age over 18 years old.

ZSBP-102/103/104/105/106/107 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old.

All measurement values can be read out and keep memory on the LCD panel for home care use.

Product codes

DXN

Device Description

Basically the measuring system were composite of blood pressure measuring circuit via Oscillometric method, pressure sensor, measuring cuff at wrist or upper arm, pneumatic pump, inflation and deflation system, housing, display LCD, and measuring software... and so on.

The main operation for the blood pressure measurement is carried out in such a way that the measuring cuff at wrist or upper arm is inflated to the estimated pressure level, then deflated to zero automatically. During the inflation, the pressure change with respective of time were recorded as the data base of measurement. Then the following measuring results will be calculated against the measurement data base:
Blood pressure information including systolic and diastolic pressure (calculated via Oscillometric method)
Heart beat rate.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

wrist, upper arm

Indicated Patient Age Range

over 18 years old

Intended User / Care Setting

home care use.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K070473

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).

0

Zen Strong Medical Technology Co., Ltd.

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

    1. Submitter's Identifications: Zen Strong Medical Technology Co., Ltd. 6F No. 88, Ning Jin St., Keelung, Taiwan, R.O.C. Contact: Cheng Roei-Sheng Date of Summary Preparation: December 31, 2008
    1. Name of the Device: Blood Pressure Monitor, models ZSBP-003/004/005/006/007/008 for wrist type and ZSBP-102/103/104/105/106/107 for upper arm type.
    1. Classification information: Regulation Number 870.1130 Medical Specialty Cardiovascular Product Code DXN Device Class Tier

Device Description: 4.

Basically the measuring system were composite of blood pressure measuring circuit via Oscillometric method, pressure sensor, measuring cuff at wrist or upper arm, pneumatic pump, inflation and deflation system, housing, display LCD, and measuring software... and so on.

The main operation for the blood pressure measurement is carried out in such a way that the measuring cuff at wrist or upper arm is inflated to the estimated pressure level, then deflated to zero automatically. During the inflation, the pressure change with respective of time were recorded as the data base of measurement. Then the following measuring results will be calculated against the measurement data base:

  • Blood pressure information including systolic and diastolic pressure (calculated via Oscillometric method)
  • Heart beat rate.

5. Intended Use:

ZSBP-003/004/005/006/007/008 blood pressure monitors measure automatically human being svstolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for all ages of person.

ZSBP-102/103/104/105/106/107 blood pressure monitors measure automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for all ages of person.

All measurement values can be read out and keep memory on the LCD panel for home care use.

1

Zen Strong Medical Technology Co., Ltd.

    1. Comparison to the 510(k) Cleared Device (Predicate Device): ZSBP-001, ZSBP-002, and ZSBP-101 (K070473).
    1. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:

Compliance to applicable voluntary standards includes ANSI/AAMI, SP 10, as well as EN 60601-1, and EN 60601-1-2 requirement.

In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.

    1. Conclusions
      The Zen Strong model ZSBP-003/004/005/006/007/008blood pressure monitors (measurement at wrist) have the same intended use and technical characteristics as the cleared model ZSBP-001 and ZSBP-002 (K070473), and ZSBP-102/103/104/105/106/107 blood pressure monitors (measured at upper arm) have the same intended use and technical characteristics as the cleared model ZSBP-101 (K070473).

Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.

2

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

FEB - 2 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Zen Strong Medical Technology Co., Ltd. c/o Ms. Cheng Roie-Sheng 6F, No. 88, Ning Jin Street, Nuannuan District, Keelung City, Taiwan, R.O.C.

Re: K090027

Trade/Device Name: Blood Pressure Monitor, Models ZSBP-003/004/005/006/007/008 for wrist type and ZSBP-102/103/104/105/106/107 for upper arm type Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 31, 2008 Received: January 5, 2009

Dear Ms. Roie-Sheng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations; Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

3

Page 2 - Ms. Cheng Roie-Sheng

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Dina R. Valenci

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Indications For Use

510(k) Number (if known):

`Device Name: Blood Pressure Monitor, models ZSBP-003/004/005/006/007/008 for wrist type, and ZSBP-102/103/104/105/106/107 for upper arm type.

Indications For Use:

ZSBP-003/004/005/006/007/008 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the Oscillometric method for the patient of the age over 18 years old.

ZSBP-102/103/104/105/106/107 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old.

All measurement values can be read out and keep memory on the LCD panel for home care use.

Prescription Use (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

proma & Valumes

(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K090027

Page 1 of