(136 days)
The LifeBed Network Patient Vigilance System is intended to measure heart rate, respiratory rate, and bed exit in adult patients for use by health care professionals.
The LifeBed Network Patient Vigilance System is identical to the LifeBed Patient Vigilance System described in the LifeBed Patient Vigilance System, K082366 with three exceptions. A Controller as been added which will not be in the patient vicinity, A wireless adapter has been added, and the Indications for Use have been reworded. The LifeBed Network integrates with the LifeBed Display to collect, store and report patient medical and status information. The Display can be configured from the Network or from the Display. The Network and Display are connected using TCP/IP protocols. To ease logistical considerations at the installation site, these two components are usually bridged across a wireless network using a LifeBed Wireless Adapter. The Network also provides an interface which may be used to integrate with a clinical information system (CIS). When used in this way, an adapter box isolates and translates protocols between the Network and the CIS. The Network centralizes information access and configuration for a group of Displays. It DOES NOT change the manner in which alerts are sent through the Nurse Call system of the hospital. Each Display is a bedside unit that collects and processes data from a sensor array on the patients' bed. It reports patient heart rate, respiration rate, alarm and if the patient is in bed or out of bed to the Controller. This information is NOT displayed real time and is NOT to be used as a method to monitor the patient. The hub of the Network is the LifeBed Controller which serves one unit (e.g., a hospital ward). From the Controller the user has full access to the information gathered from the unit and configuration of all of the Displays in the unit. The Controller also provides the interface used to integrate into a CIS or with a future product. All user access to the Network is constrained by a security system. There are two factors involved: the individual user's role and the specific access permissions assigned to the user. The user's role constrains the type of data the user has access to (e.g. medical, device status, etc.). By default, all users have specific access to information sourced by their organization. Permission to access data sourced by another organization is denied unless specifically allowed. The Network operates in a medical setting and must comply with relevant regulatory agency and legal requirements. As a medical product the Network must protect patient privacy. It must comply with all the requirements of the HIPAA Act.
This document, a 510(k) Summary for the LifeBed Network Patient Vigilance System, indicates that no new clinical testing was performed for this specific submission.
The device in question, the "LifeBed Network Patient Vigilance System" (K083534), is presented as an extension of a previously cleared device, the "LifeBed Patient Vigilance System" (K082366), with the main additions being networking capabilities (Controller, wireless adapter) and reworded Indications for Use.
The submission explicitly states:
"No changes were made to the LifeBed Patient Vigilance System which contains the algorithm and controls the patient interface (K082366). The LifeBed Network Patient Vigilance System only adds the networking capabilities. Therefore, no clinical trial tests were required to support this submission."
This means that the existing performance data and acceptance criteria would have come from the previous submission (K082366) or earlier versions of the core LifeBed system that measure heart rate, respiratory rate, and bed exit. However, the provided text does not include the acceptance criteria or device performance data for the core functions, nor does it describe the study that proved the device meets these criteria.
Therefore, based on the provided text, I cannot complete a table of acceptance criteria and reported device performance or answer most of the specific questions about the study design, sample sizes, ground truth, or expert qualifications, as no new clinical study was conducted for this particular 510(k) submission.
Summary based on the provided text:
-
Table of acceptance criteria and the reported device performance:
- N/A. The document states no new clinical trials were required for this submission because the core patient-contacting and algorithmic components were unchanged from a previously cleared device (K082366). The acceptance criteria and performance of those core functions are not provided in this document.
-
Sample size used for the test set and the data provenance:
- N/A. No new clinical study was conducted for this submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- N/A. No new clinical study was conducted for this submission.
-
Adjudication method for the test set:
- N/A. No new clinical study was conducted for this submission.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- N/A. No new clinical study was conducted for this submission. This device is described as a vigilance system reporting physiological parameters and bed exit, not an AI-assisted diagnostic tool for "human readers."
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- N/A. No new clinical study was conducted for this submission. Performance of the core algorithm would have been assessed in prior submissions.
-
The type of ground truth used:
- N/A. No new clinical study was conducted for this submission.
-
The sample size for the training set:
- N/A. No new algorithm development or training set details are provided for this submission.
-
How the ground truth for the training set was established:
- N/A. No new algorithm development or training set details are provided for this submission.
Additional Information from the text:
- Device Name: LifeBed Network Patient Vigilance System
- Indication for Use: "The LifeBed Network Patient Vigilance System is intended to measure heart rate, respiratory rate, and bed exit in adult patients for use by health care professionals."
- Substantial Equivalence Claim: The device is claimed to be substantially equivalent to the LifeBed Patient Vigilance System (K082366) and the wireless capabilities are compared to Verathon BVI 9400 (K071217) and Diagnostic Ultrasound BVM6500 (K030763).
- Testing Performed: Electrical, Mechanical, and EMC testing were conducted for both patient vicinity and non-patient vicinity components. Biocompatibility testing was not conducted as no changes were made to the patient-contacting surface (Coverlet).
- Core Functionality: The core functionality (algorithm for heart rate, respiratory rate, and bed exit) is stated to be identical to the predicate device K082366.
{0}------------------------------------------------
3 2009
SECTION 5 - 510(k) Summary
| Submitted By: | Hoana Medical, Inc828 Fort Street Mall, Suite 620Honolulu, HI 96813Phone: (808) 523-1225Fax: (808) 523-5480 | APR |
|---|---|---|
| Contact: | Nancy GertlarDirector of Regulatory Affairs and QualityAssurance | |
| Date Summary Prepared: | November 10, 2008 | |
| Trade Name: | Lifebed Network Patient Vigilance System | |
| Common/Classification Name: | Monitor, Cardiac74DRT, 870.2300 | |
| Substantially Equivalent Devices: | LG1 Patient Vigilance System(K052446);LifeBed Patient Vigilance System(K082366)Verathon Corp.BVI 9400 with ScanPoint(K071217)Diagnostic UltrasoundBVM6500 with ScanPoint(K030763) |
Description of the LifeBed Network Patient Vigilance System:
The LifeBed Network Patient Vigilance System is identical to the LifeBed Patient Vigilance System described in the LifeBed Patient Vigilance System, K082366 with three exceptions.
A Controller as been added which will not be in the patient vicinity, A wireless adapter has been added, and the Indications for Use have been reworded.
The LifeBed Network integrates with the LifeBed Display to collect, store and report patient medical and status information.
The Display can be configured from the Network or from the Display. The Network and Display are connected using TCP/IP protocols. To ease logistical considerations at the
Hoana Medical, Inc.
510(k) Premarket Reports
{1}------------------------------------------------
installation site, these two components are usually bridged across a wireless network using a LifeBed Wireless Adapter.
The Network also provides an interface which may be used to integrate with a clinical information system (CIS). When used in this way, an adapter box isolates and translates protocols between the Network and the CIS.
The Network centralizes information access and configuration for a group of Displays. It DOES NOT change the manner in which alerts are sent through the Nurse Call system of the hospital.
Each Display is a bedside unit that collects and processes data from a sensor array on the patients' bed. It reports patient heart rate, respiration rate, alarm and if the patient is in bed or out of bed to the Controller. This information is NOT displayed real time and is NOT to be used as a method to monitor the patient.
The hub of the Network is the LifeBed Controller which serves one unit (e.g., a hospital ward). From the Controller the user has full access to the information gathered from the unit and configuration of all of the Displays in the unit. The Controller also provides the interface used to integrate into a CIS or with a future product.
All user access to the Network is constrained by a security system. There are two factors involved: the individual user's role and the specific access permissions assigned to the user. The user's role constrains the type of data the user has access to (e.g. medical, device status, etc.). By default, all users have specific access to information sourced by their organization. Permission to access data sourced by another organization is denied unless specifically allowed.
The Network operates in a medical setting and must comply with relevant regulatory agency and legal requirements. As a medical product the Network must protect patient privacy. It must comply with all the requirements of the HIPAA Act.
Indication for Use
LifeBed Network Patient Vigilance System is intended to measure heart rate, respiratory rate, and bed exit in adult patients for use by health care professionals.
LifeBed Patient Vigilance System is intended to measure heart rate and respiratory rate in adult patients, in bed, in a general care hospital environment. The system will also monitor bed exit. K082366
Hoana Medical, Inc. 510(k) Premarket Reports
{2}------------------------------------------------
Principles of Operation
The LifeBed Network is an extension of the LifeBed Display that collects, stores and reports patient medical and device status information. It is not intended to replace nurse call or other reports systems. It does not alarm. The LifeBed Network does not provide real time data for the purposes of diagnosing or treatment.
The LifeBed Network extends the LifeBed Display to collect, store and report patient medical and status information. The Network and Display are connected using TCP/IP protocols. To ease logistical considerations at the installation site, these two components (The Network and the Display) are usually bridged across a wireless network using a LifeBed Wireless Adapter.
The Network centralizes information access for a group of Displays. Each Display is a bedside unit that collects and processes data from a sensor array on the patient's bed. It reports patient heart rate, respiratory rate, alarm, and status information to the Controller.
The hub of the Network is the LifeBed Controller which serves one unit (e.g. a hospital ward. From the Controller the user has full access to the information gathered from the unit and configuration of all the Displays in its unit. The Network also allows for reports to be configured and sent by email, using an email to SMS gateway to mobile phones. These reports are NOT alarms and are not real time. Just as when reading from the Controller, the data in the reports is historical in perspective, only.
Users access information via the user interface, which runs on the Controller. All user access to the Network is constrained by a security system. Each user is assigned a role which constrains what type of data the user can access.
The Network operates in a medical setting and complies with the relevant regulatory agency and legal requirements. Wireless components are used to connect Displays to the Controller. The Controller will not be in the patient vicinity. As a medical product the Network must protect patient privacy. It must comply with all the requirements of the HIPAA Act.
All components of the Network interface via TCP/IP over Ethernet. There may be intermediate hardware to transport TCP/IP. Commonly, this will be a wireless network and the LifeBed Wireless Adapter will accommodate the configuration.
There are two software interfaces in the Network, the connection to the Display. The user interface and the proprietary protocol. Beacon.
Hoana Medical, Inc. 510(k) Premarket Reports
{3}------------------------------------------------
Electrical, Mechanical and EMC Testing for the LifeBed Network Patient Vigilance System, patient vicinity components:
Test reports are found in section 17.
Electrical, Mechanical and EMC Testing for the LifeBed Network Patient Vigilance System, non-patient vicinity components:
- o The LifeBed Network Controller is located in the IT department equipment areas within the hospital and classified as non-patient vicinity or contacting equipment. This equipment falls within the Information Technology Equipment category and is tested to the appropriate standards for safety and EMI compliance.
- The Wireless Adapter is attached to the LifeBed Display in the patient room and, 0 therefore, is test to the appropriate standards for patient monitoring equipment.
Test reports are found in section 17.
Biocompatibility Testing
No changes were made to the Coverlet, which is the patient contacting surface. No biocompatibility tests were conducted as part of this submission.
Clinical Testing
No changes were made to the LifeBed Patient Vigilance System which contains the algorithm and controls the patient interface (K082366). The LifeBed Network Patient Vigilance System only adds the networking capabilities. Therefore, no clinical trial tests were required to support this submission.
Conclusion
All verification and validation testing conducted demonstrate that the LifeBed Network Patient Vigilance System is substantially equivalent to the LifeBed Patient Vigilance System, K082366.
It has been further demonstrated that the wireless addition to the LifeBed Patient Vigilance System is substantially equivalent to the wireless Verathon BVI 9400, Diagnostic Ultrasound BVM 6500, and ScanPoint which transmits data, wirelessly, to an office or facility for later viewing, diagnosing, calibrating or archiving, K071217.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 3 2009
Hoana Medical, Inc. c//o Ms. Nancy Gertlar Director, Regulatory Affairs and Quality Assurance 828 Fort Street Mall, Suite 620 Honolulu, HI 96813
Re: K083534
Trade/Device Name: LifeBed Network Patient Vigilance System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor Regulatory Class: Class II (two) Product Code: DRT Dated: April 1, 2009 Received: April 3, 2009
Dear Ms. Gertlar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Nancy Gertlar
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): K083534
Device Name: LIFEBED NETWORK PATIENT VIGILANCE SYSTEM
Indications for Use:
The LifeBed Network Patient Vigilance System is intended to measure heart rate, respiratory rate, and bed exit in adult patients for use by health care professionals.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) 4/8/09
Division of Cardiovascular Devices
510(k) Number K0803534
Page 1 of 1
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).