AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CareSens N Meter is used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. CareSens N Blood Glucose System should be used only for testing outside the body (in vitro diagnostic use only). Do not use the system for the diagnosis or screening of diabetes. Testing sites include the traditional fingertip testing along with alternate sites testing on forearm, palm, thigh and calf.

CareSens N Test Strip is used with the CareSens N Blood Glucose Meter for quantitatively measuring glucose in capillary whole blood. The CareSens N Test Strip is intended for selftesting outside the body (in vitro diagnostic use only). Do not use the system for the diagnosis of diabetes without the guidance of healthcare professional. Testing sites include the traditional fingertip testing along with alternate sites testing on forearm, palm, thigh and calf.

CareSens Control A&B Solutions are a red liquid to check that both the meters and test strips are working together properly. It contains a known range of glucose as written on the bottle.

Device Description

CareSens N Blood Glucose Monitoring System ("CareSens N BGMS") is an in vitro diagnostic device designed to measure the concentration of glucose in capillary whole blood with CareSens N Test Strips. The CareSens N model consists of the CareSens N meter (Model GM505IA, GM505IB, GM505IC: IA, IB and IC indicates the color of the devices, respectively, black, white, and light blue.), CareSens N test strips, and CareSens control solutions for two different glucose concentration ranges (called "Normal" and "Middle" ranges), lancing device, lancets, user manual, quick reference guide and a logbook.

AI/ML Overview

The provided document describes the CareSens N Blood Glucose Monitoring System, a modification of the previously cleared CareSens Blood Glucose Monitoring System. The modifications include automatic code identification, alarms, and post-meal flagging. The document presents performance evaluation summaries for both the original and modified features.

1. A table of acceptance criteria and the reported device performance

TestPre-determined Acceptance CriteriaReported Device Performance
Performance Evaluation (Original Features)
PE-01 Repeatability Test<100 mg/dL: SD<7.7 mg/dL; ≥100 mg/dL: CV<7.5%Satisfactory
PE-02 Intermediate Precision test<100 mg/dL: SD<7.7 mg/dL; ≥100 mg/dL: CV<7.5%Satisfactory
PE-03 Linearity StudyLinear response within 20~600mg/dLSatisfactory
PE-05 System Accuracy Test (CareSens N vs. YSI)<75 mg/dL: within ±15 mg/dL; ≥75mg/dL: within ± 20%Satisfactory
PE-06 Hematocrit Effect TestHct range 20-60%: Bias within ±15%Satisfactory
PE-07 Interference TestDeviations are within ±15% of the measurement for samples with no interferent.Satisfactory
PE-08 Substantial EquivalenceDistribution of test results in A-region of the Error Grid is over 95%.Satisfactory
PE-09 Consumer StudyDistribution of test results in A-region of the Error Grid is over 95%.Satisfactory
PE-10 Point-of-Care (POC) StudyDistribution of test results in A-region of the Error Grid is over 95%.Satisfactory
PE-11 Temperature StudyTemp. range 10-40°C: Bias within ±10%Satisfactory
PE-12 Humidity StudyHumidity range 10-90%: Bias within ±10%Satisfactory
PE-13 Altitude TestDistribution of test results in A-region of the Error Grid is over 95%.Satisfactory
PE-14 Alternate site StudyDistribution of test results in A-region of the Error Grid is over 95%.Satisfactory
PE-15 Drop TestHeight 0.5, 1.0, 1.5m: Acceptance within ±5 mg/dLSatisfactory
PE-16 Vibration TestRPM (50~60Hz), 300, 600, 1200: Acceptance within ±5 mg/dLSatisfactory
Validation Activities for Modified Features
Code Identification Test100% correct recognition of 15 types of code marks printed on the backside of the strip.Satisfactory
Test of self-diagnosis of code identificationThe meter should recognize any malfunction in the code-reading sensors. If the code recognition sensor is faulty, meter should detect the malfunction and display Er6 message.Satisfactory
Alarm TestThe meter should alarm at preset times within ± 1 min.Satisfactory
Time TestThe meter clock should display correct time within ± 1 min after an extended period of observation (120 hours).Satisfactory
Data Average TestThe averages (for 14-day, and before and after meal) calculated by meter software should be same as those calculated manually with spreadsheet.Satisfactory
ADC Function TestWithin 2152 ~ 2406 (12bit ADC)Satisfactory
Memory TestAll test results saved on meter memory (as many as 250 data) should exactly match the manually recorded data.Satisfactory
Electromagnetic Compatibility ReportShould meet the EMC test regulations.Satisfactory
Electrical Safety ReportShould comply with the test regulations.Satisfactory

2. Sample sizes used for the test set and the data provenance

The document does not explicitly state the sample sizes used for each specific test in the performance evaluation section (PE-01 to PE-17). It refers to the report numbers PE-01 to PE-17 from the unmodified device (K080923) and states that the methods used for these tests can be found in that prior application. The actual reports for these tests repeated for CareSens N are not included in this submission. For the modified features, such as the Code Identification Test, the acceptance criterion mentions "15 types of code marks," but the specific sample size of strips tested is not provided.

The provenance of the data (e.g., country of origin, retrospective or prospective) is not explicitly stated for these studies. However, the submitter, i-SENS, Inc., is based in Seoul, Korea, suggesting the studies likely originated from there. The studies appear to be prospective performance evaluations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not provide details on the number or qualifications of experts used to establish ground truth for the various tests. For the System Accuracy Test (PE-05), it compares the CareSens N to a YSI (likely a YSI glucose analyzer, which is a laboratory reference method), implying the YSI results serve as the ground truth. Similarly, for the "Substantial Equivalence," "Consumer Study," and "Point-of-Care (POC) Study," the ground truth involves the "A-region of the Error Grid," which is a clinically accepted method for evaluating glucose meter accuracy, typically against a reference method (like YSI).

4. Adjudication method for the test set

The document does not describe any specific adjudication methods, such as 2+1 or 3+1, for the test sets. For accuracy-related studies like System Accuracy, the comparison is directly against a reference method (YSI), which implies no human adjudication beyond the reference measurement itself.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device under review is a blood glucose monitoring system, not an AI-assisted diagnostic imaging or interpretation device that would involve "human readers" in the context of an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This concept is not directly applicable in the same way it would be for an AI algorithm interpreting an image. The CareSens N Blood Glucose Monitoring System inherently performs its measurement automatically (algorithm only) without human interpretation of the glucose value by the device itself. Human interaction is primarily for applying the blood sample and reading the displayed result. The "System Accuracy Test" (PE-05) essentially represents a standalone performance evaluation against a reference standard. The new features like automatic code identification are also standalone functions of the device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The primary type of ground truth used for accuracy and precision studies (e.g., PE-01, PE-02, PE-05) is comparison against a laboratory reference method, specifically implied by the "CareSens N vs. YSI" test (PE-05). YSI glucose analyzers are widely considered a gold standard for glucose measurement in clinical settings. For the linearity study, known glucose concentrations would be used. For the functional tests of modified features (e.g., code identification, alarms, memory), the ground truth is established by design specifications and manual verification of expected behavior.

8. The sample size for the training set

The document does not explicitly mention a "training set" in the context of machine learning. Blood glucose monitoring systems typically do not involve machine learning models that require distinct training sets in the way AI algorithms do. Their performance is based on electrochemical principles and calibration, not learned patterns from data.

9. How the ground truth for the training set was established

As there is no explicit mention of a training set for a machine learning model, this question is not applicable. The 'training' for such a device involves factory calibration and quality control procedures using known glucose concentrations, rather than a data-driven training set.

{0}------------------------------------------------

IC083468

Image /page/0/Picture/1 description: The image shows a logo with the word "i-sens" in a stylized font. The "i" is lowercase and has a small circle above it, resembling a dot. The rest of the word, "-sens," is also lowercase. To the left of the word is a circular graphic with a complex, textured pattern inside. The overall design is simple and modern.

i-SENS, Inc. 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea

Tel. 82-2-916-6191

Special 510(k) SUMMARY

FEB 1 0 2009

(As required by 2I.CFR.807.87)

Type of 510(k):Special
Submitter's Previously Cleared Device:Name - CareSens Blood Glucose Monitoring System510(k) number - K080923
Introduction:This 510(k) submission contains information/data on modifications madeto submitter's previously cleared device. The modifications do notaffect the intended use of the device or alter the fundamental scientifictechnology of the device. The major modifications are the addition ofan automatic code identification feature, and the user interfaceapplications: alarms and post-meal flagging.
Submitted By:i-SENS, Inc.465-6, Wolgye-dong, Nowon-gu, Seoul, 139-845, KoreaTel.) ++82-2-916-6191Fax) ++82-2-942-2514www.i-sens.com
Contact Person:Dr. Hyun Joon Oh, Phone: +82-33-903-0760, Fax +82-33-748-6191
Date Summary was Prepared:November 6, 2008
Device Name:Trade name: CareSens N Blood Glucose Monitoring SystemCommon Name: Blood Glucose Test SystemClassification Name: Class II, 862.1345 Glucose Blood Tester
Predicate Device:We claim substantial equivalence to submitter's previously clearedCareSens Blood Glucose Monitoring System ("CareSens BGMS")
The glucose measurement test principle:same as CareSens BGM
Intended Use:same as CareSens BGM
Fundamental scientific technology:same as CareSens BGM

:

:

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for i-sens. The logo consists of the word "i-sens" in a sans-serif font, with the "i" in lowercase and the rest of the word in lowercase. To the left of the word is a circular graphic made up of many small words. The words are arranged in a circular pattern, and they are all in a small font size.

i-SENS. Inc. 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191

systems by i-SENS, Inc. (K080923) and the OneTouch® Ultra® Blood Glucose Monitoring System by LifeScan, Inc. (K024194).

Device Description:

CareSens N Blood Glucose Monitoring System ("CareSens N BGMS") is an in vitro diagnostic device designed to measure the concentration of glucose in capillary whole blood with CareSens N Test Strips. The CareSens N model consists of the CareSens N meter (Model GM505IA, GM505IB, GM505IC: IA, IB and IC indicates the color of the devices, respectively, black, white, and light blue.), CareSens N test strips, and CareSens control solutions for two different glucose concentration ranges (called "Normal" and "Middle" ranges), lancing device, lancets, user manual, quick reference guide and a logbook.

Comparison to Submitter's Previously Cleared Device (K080923):

Modifications:

Automatic code identification by the meter through detection of a marked pattern on each strip, so that the user does not input code information.

For the meter, the only difference is the addition of code identification sensors in the meter, while all other major elcctronic components are the same as those of CareSens BGMS. Please refer to Page 3 of Section 10 Specifications for details.

For the strip, the only difference is printing code-specific marks on the back side of each strip with infrared absorbing ink, while all the reagent components and mcasurement process are the same as those of CareSens BGMS. Please refer to Page 11 of Section 10 Specifications for details.

Alarms (Three time set alarms, and 2-hour post meal alarm) Post-meal flagging of test results Number of Buttons on the meter is three

7-2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for i-SENS, a company that specializes in biosensor technology. The logo features a stylized lowercase "i" with a circular element above it, followed by the text "sens" in a bold, sans-serif font. The logo is simple and modern, reflecting the company's focus on innovation and technology.

.

:

Common Features
ItemCareSens N BGMSCareSens BGMS
EnzymeGlucose Oxidase
Measurement principleAmperometric method
Test principleGlucose Oxidase chemical reaction. The instrument measures the extent of current cause by presence of glucose in sample.
Intended useThe test strips work with the device to quantitatively measure glucose in capillary whole blood. The test Strips are for in vitro (i.e., outside the body) diagnostic use only.
SampleFresh capillary whole blood
ElectrodeCarbon
CalibrationPlasma-equivalent
Test Time (second)5
Sample volume (µL)0.5
Memory250
Test Range(mg/dL)20~600
Hematocrit range (%)20~60 (below 400mg/dL)
Glucose unitsEither mg/dL or mmol/L
Checking the systemControl solution
Alternate SiteYes
Capability
Operating Humidity10~90%
Differences
ItemCareSens N BGMSCareSens BGMS
CodingAutomatic code identificationManual input by button
Self-diagnosis of codeidentification functionYesNo
Three time set Alarmsand 2-hour post-mealAlarmYesNo
Post-meal flaggingYesNo
Number of buttons3 buttons2 buttons (CareSens II)1 button (CareSens POP)

:

:

:

{3}------------------------------------------------

i-SENS, Inc.

Image /page/3/Picture/1 description: The image shows a logo for "i-sens". The logo is composed of a stylized letter "i" followed by the word "sens". The "i" is surrounded by a circular pattern of small dots, giving it a technological or digital feel. The word "sens" is written in a bold, sans-serif font, and the overall design is simple and modern.

465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191

Since for CareSens N, the glucose measurement test principle, intended use, and fundamental scientific technology are the same as the submitter's unmodified device (K080923), there is no reason to expect that the performance would be different, but most of the performance evaluation tests done for the unmodified device were repeated to confirm that performance still satisfies the acceptance criteria. For performance evaluation, the following summary table shows the tests that were repeated.

Additionally, the substantial equivalence report for comparison to OneTouch® Ultra® is provided in Section 11.

Performance evaluation summary of modified device, CareSens N BGMS

The report numbers PE-01 to PE-17 in this table indicate those of the unmodified device (K080923), and the methods used for the tests can be found in that prior application. Please note that the actual reports for tests PE-01 to PE-17 that were repeated for CareSens N, are not included in this submission. .

ReportNo.TitlePre-determined Acceptance CriteriaResults
PE-01Repeatability TestConcentrationPrecisionSatisfactory
<100 mg/dLSD<7.7 mg/dL
≥100 mg/dLCV<7.5%
PE-02Intermediate Precision testConcentrationPrecisionSatisfactory
<100 mg/dLSD<7.7 mg/dL
≥100 mg/dLCV<7.5%
PE-03Linearity Studylinear response within 20~600mg/dLSatisfactory
PE-04-1Shelf Lifetime Test for TestStrips (Before Opening)N/AThese tests not repeated becauseall aspects of the strip chemistryare the same, and the markingon the back should not affectshelf life.
PE-04-2Shelf Lifetime Test for TestStrips (After Opening)N/A
PE-05System Accuracy Test(CareSens N vs. YSI)ConcentrationAccuracy
<75 mg/dLwithin ±15mg/dLSatisfactory
≥75mg/dLwithin ± 20%
PE-06Hematocrit Effect TestHct rangeBias (%)Satisfactory
20-60%within ±15 %
PE-07Interference TestDeviations are within ±15 % of the measurementfor the samples with no interferent.Satisfactory
PE-08Substantial EquivalenceThe distribution of test results in A-region of the ErrorGrid is over 95%.Satisfactory
PE-09Consumer StudyThe distribution of test results in A-region of the ErrorGrid is over 95%.Satisfactory
PE-10Point-of-Care(POC) StudyThe distribution of test results in A-region of the ErrorGrid is over 95%.Satisfactory
PE-11Temperature StudyTemp. rangeBias (%)Satisfactory
10-40°Cwithin ±10 %
PE-12Humidity StudyHumidity rangeBias (%)Satisfactory
10-90%within ±10 %
PE-13Altitude TestThe distribution of test results in A-region of the ErrorGrid is over 95%.Satisfactory
PE-14Alternate site StudyThe distribution of test results in A-region of the ErrorGrid is over 95%.Satisfactory
PE-15Drop TestHeightAcceptanceSatisfactory
0.5, 1.0, 1.5mwithin ±5 mg/dL
PE-16Vibration TestRPMAcceptance
(50~60Hz) ,Satisfactory
300, 600, 1200within ±5 mg/dL
PE-17Control solutionN/ATest not repeated because thecontrol solutions for bothsystems are the same.

7-4

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for i-sens. The logo features a stylized globe with a small plus sign in the center, followed by the word "i-sens" in a bold, sans-serif font. The globe is made up of small circles, giving it a textured appearance. The overall design is simple and modern.

i-SENS, Inc. 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191

.

۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔۔

11 11:11 PM 11 11

,

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for a company called i-sens. The logo features a stylized letter 'i' followed by the word 'sens' in a simple, sans-serif font. To the left of the 'i' is a circular graphic composed of many small, interconnected dots, creating a textured, almost globe-like effect. The overall design is clean and modern, suggesting a technology or innovation-focused company.

Based on risk analysis provided in Section 14, the following validation activities were carried out and submitted in this application under the Scction Numbers specified.

Validation Activities for Modified Features

The test reports in this table are included in this submission, and can be found under the Section Numbers below.

SectionNumbersTitlePre-determined Acceptance CriteriaResults
12-11CodeIdentificationTest100% correct recognition of 15 types of code marks printed onthe backside of the strip.Satisfactory
12-22Test of self-diagnosis ofcodeidentificationThe meter should recognize any malfunction in the code-reading sensors. If the code recognition sensor is faulty by anycauses, meter should detect the malfunction and display Er6message to prevent the user from reading inaccurate results.Satisfactory
13(TR-EI-15)3Alarm TestThe meter should alarm at preset times within $\pm$ 1 min.Satisfactory
13(TR-EI-16)Time TestThe meter clock should display correct time within $\pm$ 1 minafter an extended period of observation (120 hours).Satisfactory
13(TR-EI-17)Data AverageTestThe averages (for 14-day, and before and after meal)calculated by meter software should be same as thosecalculated manually with spreadsheet.Satisfactory
13(TR-EI-18)ADC FunctionTestWithin 2152 $\sim$ 2406 (12bit ADC)Satisfactory
13(TR-EI-19)4Memory TestAll test results saved on meter memory (as many as 250 data)should exactly match the manually recorded data.Satisfactory

1 Page 19 of Section 14 Risk analysis: H11-C11.2 Error of code recognition function

2 Page 19 of Section 14 Risk analysis: H11-C11.1 Error (shorted or disconnected circuit) in the infraced photoreflectors

for auto coding function

3 Page 20 of Section 14 Risk analysis: H16 software

4 Page 20 of Section 14 Risk analysis: H16 software-C16.3 Functional error of memory recall

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with the word "isens" in a stylized font. The "i" in "isens" is designed to look like an eye, with a circular shape above the letter. The rest of the letters in "isens" are in lowercase and have a bold, sans-serif font. The logo appears to be for a company or organization named "isens."

i-SENS, Inc. Tel. 82-2-916-6191 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea

15-15Electromagnetic Compatibility ReportIt should meet the EMC test regulations given in p3 of the title report (Electronic Compatibility Report).Satisfactory
15-2Electrical Safety ReportIt should comply with the test regulations given in p1 of the title report.Satisfactory

Conclusion:

:

All predetermined acceptance criteria were satisfied. The data also demonstrates that the CareSens N BGMS is substantially equivalent to the following predicate device systems:

K080923 - i-SENS, Inc. CareSens Blood Glucose Monitoring System K024194 - LifeScan, Inc. OneTouch® Ultra® Blood Glucose Monitoring System

్ Page 16 of Section 14 Risk analysis: H4 electromagnetic field compatibility

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign, indicating an association with a public health organization.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

I-Sens, Inc. c/o Hyun Joon Oh Division Manager, Quality Assurance 465-6 Wolgye-Dong, Nowon-Gu Seoul, Republic of Korea 139-845

FEB 1 0 2009

Re: K083468

Trade Name: CareSens N Blood Glucose Monitoring System Regulation Number: 21 CFR 862, 1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW. CGA. JJX Dated: January 26, 2009 Received: January 26, 2009

Dear Hyun Joon Oh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

Image /page/7/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three bodies, representing health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

{8}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

signature

Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostics Device Evaluation and Safety Center for Devices and Radiological Health

{9}------------------------------------------------

Indication for Use

510(k) Number (if known): 长۵૪ 3468

Device Name: CareSens N Blood Glucose Monitoring System

Indication For Use: The CareSens N Meter is used for the quantitative measurement of glucose level in capillary whole blood as an aid in monitoring the effectiveness of diabetes management at home or in clinical settings. CareSens N Blood Glucose System should be used only for testing outside the body (in vitro diagnostic use only). Do not use the system for the diagnosis or screening of diabetes. Testing sites include the traditional fingertip testing along with alternate sites testing on forearm, palm, thigh and calf.

CareSens N Test Strip is used with the CareSens N Blood Glucose Meter for quantitatively measuring glucose in capillary whole blood. The CareSens N Test Strip is intended for selftesting outside the body (in vitro diagnostic use only). Do not use the system for the diagnosis of diabetes without the guidance of healthcare professional. Testing sites include the traditional fingertip testing along with alternate sites testing on forearm, palm, thigh and calf.

CareSens Control A&B Solutions are a red liquid to check that both the meters and test strips are working together properly. It contains a known range of glucose as written on the bottle.

Prescription Use (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Yici May acting

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K083468

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.