K Number
K083433

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-03-18

(118 days)

Product Code
Regulation Number
890.5290
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffiess, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.

Generally accepted indications for use:

  • Pain Relief .
  • Reduce Muscle Spasm .
  • . Decrease Joint Stiffness
  • . Contractures
  • Increase Blood Flow
  • Chronic Inflammatory Conditions .
  • . Bursitis
  • Tenosynovitis
  • Synovitis .
  • Chronic Inflammatory Pelvic Disease
Device Description

The Intelect® SWD 100/ Senior Shortwave Diathermy devices consists of a power supply that provides power to a radio frequency oscillator. This radio frequency oscillator provides stable, drift-free oscillations at the required frequency. The power amplifier generates the power required to drive the different types of electrodes. The output resonant tank tunes in the patient as part of the circuit and allows maximum power to be transferred to the patient.

The output intensity controls the percentage of maximum power transferred to the patient circuit. The tuning control adjusts the output circuit for maximum energy transfer from the radio frequency oscillator. The power output meter monitors only the current that is drawn from the power supply and not the energy being delivered to the patient.

The power output produces sufficient energy to raise the tissue temperature into a therapeutic range. The specific absorption rate (SAR) represents the rate of energy absorbed per unit area of tissue mass.

Patient sensation provides the basis for recommendations of continuous shortwave diathermy dosage and thus varies considerably with different patients. The following are recommended dosage guidelines: Dose I (lowest): No sensation of heat

Dose II (low): Mild heating sensation

Dose III (medium): Moderate (pleasant) heating sensation

Dose IV (heavy): Vigorous heating that is tolerable below the pain threshold

Therapy may be delivered to the patient via either capacitance or induction techniques. The Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy devices use the following accessories:

  • Capacitive Electrodes
  • Pair of 80mm round
  • Pair of 120mm round
  • Pair of 165mm round
  • 180mm x 120mm flexible rubber electrodes
  • 250mm x 145mm flexible rubber electrodes
  • Inductive Electrodes
  • Monode 14cm
  • Diplode 18cm x 39cm

Each of these techniques can affect different biological tissues, therefore selection of the appropriate electrodes is essential for effective treatment.

AI/ML Overview

This 510(k) premarket notification for the Intelect® SWD 100/Senior Solutions® Shortwave Diathermy device does not contain the kind of study design and results that are typical for devices that rely on AI or complex algorithms requiring performance metrics like sensitivity, specificity, or AUC.

This submission focuses on demonstrating substantial equivalence to existing legally marketed predicate devices, primarily through adherence to recognized consensus standards for safety and electrical compatibility, and by proving similar indications for use. It is a traditional medical device submission, not one for an AI/ML-driven diagnostic or therapeutic support system.

Therefore, many of the requested categories for acceptance criteria and study details (like sample sizes for test/training sets, expert ground truth, MRMC studies, standalone performance with AI, etc.) are not applicable to this specific submission.

Here's a breakdown based on the provided text, indicating where information is present and where it is not:


Acceptance Criteria and Device Performance (Not Applicable in the traditional sense for AI/ML)

The "acceptance criteria" for this type of device submission are generally met by demonstrating:

  1. Compliance with recognized safety and performance standards: The device fulfills the requirements of relevant IEC and UL standards.
  2. Substantial equivalence to predicate devices: The device operates with the same fundamental technology and has similar indications for use as previously cleared devices.
  3. Ability to produce therapeutic deep heat: Implied by the device description and its function as a shortwave diathermy unit.

Since this is a physical therapy device that generates heat, its "performance" is more related to its electrical output and energy transfer capabilities, rather than diagnostic accuracy metrics. The document describes how it works (radio frequency oscillator, power amplifier, output resonant tank, intensity controls, tuning control, power output meter) and mentions the specific absorption rate (SAR) as a measure of energy absorbed. Therapeutic dosage is described in terms of patient sensation (Dose I-IV).

Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/Standard (as implied by submission)Reported Device Performance/Evidence from StudyNotes
SafetyUL 60601-1: 2003 (Safety, 1st Ed)Declared ConformityThe device complies with this standard.
Electromagnetic Compatibility (EMC)IEC 60601-1-2: 2001 (EMC, 2nd Ed)Declared ConformityThe device complies with this standard.
Specific Safety/Therapy EquipmentIEC 60601-2-3: 1991 (Safety of Short-Wave Therapy Equipment); Amd 1 – 1998, 2nd EdDeclared ConformityThe device complies with this standard, though the FDA notes it doesn't recognize this specific standard. This is likely due to the FDA having its own guidelines or recognizing superseded versions, but the manufacturer still adhered to it.
Programmable Medical Electrical SystemsIEC 60601-1-4: 2000 (Programmable Medical Electrical Systems); Amd 1 – 1999Declared ConformityThe device complies with this standard.
Intended Use EquivalenceSimilar indications of use as legally marketed predicate devicesK042554, K030382, K022458, K973732The FDA's substantial equivalence determination implies these were met. The indications listed are identical to those of common shortwave diathermy.
Therapeutic Heat GenerationAbility to produce sufficient energy to raise tissue temperature into a therapeutic rangeDescribed in device functionThe device description states: "The power output produces sufficient energy to raise the tissue temperature into a therapeutic range." Dosages I-IV are based on patient sensation of heat. No specific numerical performance (e.g., target temperature achieved) is provided, as it's typically user-dependent and sensation-based for this type of device.
Operating FrequencyElectromagnetic energy in the radio frequency band of 27.12 megahertzDevice operates at 27.12 MHzStated in the "Intended Use" section.

Study Details (Not Applicable for AI/ML performance studies)

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • N/A. This submission is for a physical therapy device based on established technology, not an AI/ML diagnostic or predictive device with a "test set" in the computational sense. The "study" here is primarily one of engineering design verification and validation against pre-defined standards.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • N/A. Ground truth as understood in AI/ML performance evaluation (e.g., expert labels on images) is not relevant for this device. The "ground truth" here is the established scientific and engineering principles of shortwave diathermy and regulatory compliance.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • N/A. Not applicable, as there is no "test set" requiring expert adjudication in this context.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A. This is a standalone therapy device, not an AI-assisted diagnostic or decision support tool. Therefore, MRMC studies comparing human readers with and without AI are not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • N/A. While the device operates standalone in the sense that it delivers therapy without continuous intervention from an algorithm, the concept of "standalone performance" in the AI context (e.g., sensitivity/specificity of an algorithm making a diagnosis) is not relevant here. Its performance is about safely generating radiofrequency energy and heat.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • N/A. The "ground truth" for this device's acceptance is based on:
      • Compliance with recognized industry standards (UL, IEC).
      • Scientific principles of shortwave diathermy.
      • Regulatory precedent established by substantially equivalent predicate devices.
      • Engineering verification and validation to ensure the device performs as designed.
  7. The sample size for the training set:

    • N/A. This device does not use a "training set" in the AI/ML sense. Its operational parameters are based on physics and engineering, not learned from data.
  8. How the ground truth for the training set was established:

    • N/A. Since there's no training set, this is not applicable.

{0}------------------------------------------------

K083433 510(k)

Page 1 of 3

Chattanooga Group SUBMITTER: 4717 Adams Road Hixson, TN 37343

ESTABLISHMENT 1022819 REGISTRATION:

  • CONTACT: Michael Treas. Director of Regulatory Affairs
    DATE PREPARED: November 17, 2008

  • PROPRIETARY NAME: Intelect® SWD 100/ Senior Solutions®
    CLASSIFICATION: Class II

PRODUCT CODES: IMJ

21 CFR 890.5290 REGULATION NUMBER Shortwave Diathermy AND COMMON NAMES:

PANEL: Physical Medicine

Indications for Use:

Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffness, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.

Generally accepted indications for use:

  • . Pain Relief
  • . Reduce Muscle Spasm
  • Decrease Joint Stiffness ●
  • Contractures
  • Increase Blood Flow .
  • Chronic Inflammatory Conditions .
  • Bursitis .
  • Tenosynovitis .
  • Synovitis .
  • Chronic Inflammatory Pelvic Disease ●

{1}------------------------------------------------

510(k) # K083433

Intended Use:

Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy device for use in applying therapeutic deep heat for selected medical conditions by applying electromagnetic energy in the radio frequency band of 27.12 megahertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.

Description :

The Intelect® SWD 100/ Senior Shortwave Diathermy devices consists of a power supply that provides power to a radio frequency oscillator. This radio frequency oscillator provides stable, drift-free oscillations at the required frequency. The power amplifier generates the power required to drive the different types of electrodes. The output resonant tank tunes in the patient as part of the circuit and allows maximum power to be transferred to the patient.

The output intensity controls the percentage of maximum power transferred to the patient circuit. The tuning control adjusts the output circuit for maximum energy transfer from the radio frequency oscillator. The power output meter monitors only the current that is drawn from the power supply and not the energy being delivered to the patient.

The power output produces sufficient energy to raise the tissue temperature into a therapeutic range. The specific absorption rate (SAR) represents the rate of energy absorbed per unit area of tissue mass.

Patient sensation provides the basis for recommendations of continuous shortwave diathermy dosage and thus varies considerably with different patients. The following are recommended dosage guidelines: Dose I (lowest): No sensation of heat

Dose II (low): Mild heating sensation

Dose III (medium): Moderate (pleasant) heating sensation

Dose IV (heavy): Vigorous heating that is tolerable below the pain threshold

Therapy may be delivered to the patient via either capacitance or induction techniques. The Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy devices use the following accessories:

  • Capacitive Electrodes
  • Pair of 80mm round
  • Pair of 120mm round
  • Pair of 165mm round
  • 180mm x 120mm flexible rubber electrodes
  • 250mm x 145mm flexible rubber electrodes
  • Inductive Electrodes
  • Monode 14cm
  • Diplode 18cm x 39cm

Each of these techniques can affect different biological tissues, therefore selection of the appropriate electrodes is essential for effective treatment.

{2}------------------------------------------------

510(k) Summary per 21 CFR 807.92(c)

510(k) # K083433

Page 3 of 3

510(k) #ProprietaryNamePredicate DeviceManufacture'sDeviceClassificationNamesRegulationNumbersClassification(s)and ProductCodes
K042554Auto Therm 390Mettler Electroinics,CorporationShortwaveDiathermy890.5290Class II, IMJ
K030382SeliThermSelicor, Inc.ShortwaveDiathermy890.5290Class II, IMJ
K022458Auto Therm 395Mettler Electronics,CorporationShortwaveDiathermy890.5290Class II, IMJ
K973732Megapulse IIPTI, Inc.ShortwaveDiathermy890.5290Class II, IMJ

Substantially Equivalent Predicate Devices

Declarations of Conformity

The Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy devices comply with the following FDA recognized Consensus Standards:

UL 60601-1: 2003, Standards for Medical Equipment Part 1: General Requirements for Safety, 1* Edition

IEC 60601-1-2: 2001, Medical Electrical Equipment Part 1 - 2: General requirements for Safety -Collateral Standard, Electromagnetic Compatibility - Requirements and Tests, 2nd Edition

IEC 60601-2-3: 1991, Medical Electrical Equipment – Part 2: Particular Requirements for the Safety of Short-Wave Therapy Equipment; Amendment 1 – 1998, 2nd Edition (The United States FDA does not recognize IEC 60601-2-3)

IEC 60601-1-4: 2000, Medical Electrical Equipment -- Part 1 -- 4: General Requircments for Safety -Collateral Standard: Programmable Medical Electrical Systems Consolidated with Amendment 1, 1999

Truthful and Accurate Statement

A statement attesting to the truthfulness and accuracy of the information was included in the premarket submission.

Further Information

Contact:

Chattanooga Group 4717 Adams Road Hixson, TN 37343 U.S.A.

Michael Treas, Director of Regulatory Affairs Telephone: (423) 870-2281 ext. 7207 Fax: (423) 870-7404 Email: michael.treas@djoglobal.com

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is facing left and has a sharp beak and talons.

Public Health Service

MAR 1 8 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Chattanooga Group % Mr. Michael Treas Director of Regulatory Affairs 4717 Adams Road Hixson, Tennessee 37343

Re: K083433

Trade/Device Name: Intelect® SWD 100/Senior Solutions® Shortwave Diathermy Regulation Number: 21 CFR 890.5290 Regulation Name: Shortwave diathermy Regulatory Class: II Product Code: IMJ Dated: February 17, 2009 Received: February 18, 2009

Dear Mr. Treas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices markcted in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Michael Treas

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device. to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number:

Device Name: Intelect® SWD 100/ Senior Solutions® Shortwave Diathermy

Indications for Use:

Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffiess, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.

Generally accepted indications for use:

  • Pain Relief .
  • Reduce Muscle Spasm .
  • . Decrease Joint Stiffness
  • . Contractures
  • Increase Blood Flow �
  • Chronic Inflammatory Conditions .
  • . Bursitis
  • Tenosynovitis �
  • Synovitis .
  • Chronic Inflammatory Pelvic Disease
Prescription Use(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number14083433

Submission Page 12 of 905 Page 1 of 1

§ 890.5290 Shortwave diathermy.

(a)
Shortwave diathermy for use in applying therapeutic deep heat for selected medical conditions —(1)Identification. A shortwave diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body electromagnetic energy in the radiofrequency (RF) bands of 13.56 megahertz (MHz) or 27.12 MHz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.(2)
Classification. Class II (performance standards).(b)
Nonthermal shortwave therapy —(1)Identification. A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.(2)
Classification: Class II (special controls). The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible.
(ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined:
(A) Peak output power;
(B) Pulse width;
(C) Pulse frequency;
(D) Duty cycle;
(E) Characteristics of other types of modulation that may be used;
(F) Average measured output powered into the RF antenna/applicator;
(G) Specific absorption rates in saline gel test load or other appropriate model;
(H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and
(I) Characterization of the deposited energy density in saline gel test load or other appropriate model.
(iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use.
(v) Labeling must include the following:
(A) Output characteristics of the device;
(B) Recommended treatment regimes, including duration of use; and
(C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications.
(vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.