(38 days)
The X-coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures.
The X-coated Terumo® Dual-Stage Venous Return Cannula is indicated for use for adult single tube blood drainage from the right atrium and inferior vena cava (IVC) and delivery of the blood to the extracorporeal circuit during cardiopulmonary bypass.
The X-coated Sarns™ Dual-Stage Venous Return Catheters are indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery.
The Sarns ™ Malleable Venous Return Catheters are indicated for venous drainage during cardiopulmonary bypass surgery: the 24, 28, 30, 32, 34, 36 and 40 Fr. are for dual cannulation of the superior and inferior vena cava: the 40 Fr. is for single cannulation of the right atrium.
The X-coated Sarns™ Malleable Dual-Stage Venous Return Catheters are indicated for single-tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery.
The X-coated Sarns™ High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
The X-coated Sarns™ Soft-Flow™ Extended Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
These devices are indicated for up to 6 hours of use.
The design of cannulae and catheters is such that they meet their stated intended use, and provide an acceptable level of performance and safety to the patients. Except for the X-coat application, the design and material of the X-coated cannulae will remain the same as the unmodified cannulae. The X-coat application does not affect the cannulae design.
Venous and arterial cannulae contained in this submission are structured as straight, or slightly curved, flexible tubes which are inserted into the heart to circulate blood. Depending on the location of cannulation, size, and clinicians' preference, there is a range of design variations. These features include, but are not limited, wire reinforcement, various stylets used to help guide the cannula, tip formation (conical vs. angled), number of drainage baskets (two stage), luers, and connectors.
The generic materials for these cannulae will remain the same.
The provided text describes a 510(k) summary for several X-coated cannulae. The submission focuses on demonstrating substantial equivalence to unmodified cannulae, not on new device performance claims requiring extensive clinical studies with specific acceptance criteria as might be seen for novel AI/ML devices.
Therefore, many of the requested categories (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set information) are not applicable or not explicitly detailed in this type of submission for this particular device. The performance evaluations described are primarily engineering and biocompatibility tests to show that the addition of the X-coat does not negatively impact the existing performance or safety characteristics of the predicate devices.
However, I can extract the information that is present and indicate where the requested information is absent or not relevant to this type of submission.
Acceptance Criteria and Device Performance Study for K083301
1. Table of Acceptance Criteria and Reported Device Performance
For this 510(k) submission, the "acceptance criteria" are implied by demonstrating that the X-coated devices maintain the same performance and safety profile as their uncoated predicate devices. The primary performance claim related to the X-coat itself is a reduction in platelet adhesion.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Functional Cannulae Performance | The X-coat application does not affect the functional cannulae performance. (Implied: performance remains equivalent to unmodified cannulae). |
| Reduced Platelet Adhesion | The X-coat application reduces platelet adhesion. (This is the specific enhancement claimed for the X-coat). |
| Hemolysis Equivalence | Hemolysis was evaluated to demonstrate equivalence to the uncoated cannulae. (Implied: Hemolysis rates are equivalent or within acceptable limits compared to predicate). |
| Bond Strength Equivalence | Bond strength tests (leak, pull, flex testing) were conducted to demonstrate equivalence to the uncoated cannulae. (Implied: Bond strength remains equivalent to unmodified cannulae). |
| Accelerated Aging / Shelf Life Equivalence | Accelerated aging / Shelf life tests were conducted. Materials were found to be stable over the expiry of the product through material characterization studies (physio-chemical profiles of aged and non-aged devices). (Implied: Shelf life and material stability are equivalent to unmodified cannulae). |
| Sterility Assurance Level (SAL) | Sterilization conditions validated in accordance with AAMI guidelines to provide SAL of 10⁻⁶. (Meets industry standards for sterility). |
| Residue Limits (Ethylene Oxide, etc.) | Post-sterilization release determined in consideration of maximum residue limits in accordance with EN ISO 10993-7 and AAMI TIR-19. (Meets industry standards for biocompatibility-related residues). |
| Biocompatibility | Biocompatibility studies conducted using the same materials or combination of materials as predicate devices as recommended in FDA guidance (ISO 10993). Blood contacting materials found to be biocompatible for limited exposure (≤ 24 hours) with circulating blood. (Meets industry standards for biocompatibility for its intended use). |
| No New Issues of Safety and Effectiveness | "Any noted differences between the devices do not raise new issues of safety and effectiveness." (Overall conclusion of substantial equivalence based on all evaluations). |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size:
- Not explicitly stated for the individual performance evaluations (hemolysis, bond strength, accelerated aging, biocompatibility). These are typically bench or lab tests, not clinical studies with "test sets" in the AI/ML sense.
- Data Provenance: Not specified. These are likely internal laboratory tests conducted by the manufacturer.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- Not Applicable. This type of information (experts for ground truth) is not relevant for this device submission, which involves physical device performance (mechanical, chemical, biological) rather than interpretation of diagnostic data where expert consensus creates ground truth.
4. Adjudication Method for the Test Set
- Not Applicable. As there's no diagnostic interpretation or subjective assessment by multiple readers, an adjudication method is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. MRMC studies are used for evaluating diagnostic devices, particularly those involving human interpretation, sometimes with AI assistance. This device is a surgical cannula, not a diagnostic AI tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not Applicable. This device does not involve an algorithm.
7. The Type of Ground Truth Used
- For functional performance (Hemolysis, Bond Strength, Accelerated Aging): Ground truth is established by standardized test methods and comparisons to the performance characteristics of the unmodified predicate devices.
- For reduced platelet adhesion: The "ground truth" for this claim would be based on specific in vitro or ex vivo blood compatibility tests designed to quantify platelet adhesion, compared to a control (e.g., the uncoated device). The specific methods and results are not detailed in this summary.
- For Sterility, Residue, Biocompatibility: Ground truth is established by adherence to recognized international standards (AAMI, ISO 10993, EN ISO 10993-7, AAMI TIR-19).
8. The Sample Size for the Training Set
- Not Applicable. This device does not use an AI model that requires a "training set."
9. How the Ground Truth for the Training Set was Established
- Not Applicable. As there is no training set, there is no ground truth to establish for it.
{0}------------------------------------------------
6. 510(k) Summary
:
DEC 1 8 2008
| 6. 510(k) Summary | 1 | |
|---|---|---|
| 6.1. | Submitter Information | 2 |
| 6.2. | Device Names, Predicate Device Names, and Predicate Device 510(k) Numbers: | 2 |
| 6.3. | Statement of Identical Intended Use(s)/Indications: | 3 |
| 6.4. | Principles of Operation and Technology: | 3 |
| 6.5. | Design and Materials: | 4 |
| 6.6. | Performance Evaluations: | 4 |
| 6.7. | Substantial Equivalence Comparison: | 4 |
| 6.7.1. Intended Use: | 4 | |
| 6.7.2. Principles of Operation and Technology: | 4 | |
| 6.7.3. Design and Materials: | 5 | |
| 6.7.4. Substantial Equivalence Summary: | 5 | |
| 6.8. | Additional Safety Information | 5 |
| 6.9. | Conclusion | 5 |
.
:
:
{1}------------------------------------------------
6.1.
.
| Submitter Information | |
|---|---|
| 510(k) Submitter: · | Terumo Cardiovascular Systems Corporation6200 Jackson Road,Ann Arbor, MI 48103 |
| Primary Contact Name: | Junko Kurosawa,Senior Associate, Regulatory ManagementTel: 1(800)262-3304, Extension 6563Fax: (734)741-6030 |
| Secondary Contact Name: | Christina Thomas,Manager, Regulatory ManagementTel: 1(800)262-3304, Extension 6278Fax: (734)741-6030 |
| Date Prepared: | November 7, 2008 |
:
6.2. Device Namcs, Predicate Device Names, and Predicate Device 510(k) Numbers:
| Name of theUnmodified Device | 510(k) Number ofSponsor'sUnmodified Device | Clearance date | Name of the ModifiedDevice |
|---|---|---|---|
| Terumo® Dual-StageVenous ReturnCannula | X-coated Terumo®Dual-Stage VenousReturn Cannula | ||
| Sarns™ Dual StageVenous ReturnCatheters | X-coated Sarns™Dual Stage VenousReturn Catheters | ||
| Sarns™ MalleableDual Stage VenousReturn Catheters | X-coated Sarns™Malleable Dual StageVenous ReturnCatheters | ||
| Sarns™ MalleableVenous ReturnCatheters | K905224 | 1991-02-05 | X-coated Sarns™Malleable VenousReturn Catheters |
| Sarns™ High FlowAortic Arch Cannula | K770429 | 1977-03-16 | X-coated Sarns™High Flow AorticArch Cannula |
| Sarns™ Soft-Flow™Extended AorticCannula | K934127 | 1994-02-22 | X-coated Sarns™Soft-Flow™Extended AorticCannula |
·
{2}------------------------------------------------
These devices are substantially equivalent in intended use, materials, design, technology, principles of operation, and performance.
6.3. Statement of Identical Intended Use(s)/Indications:
TCVS asserts that the intended use(s)/indications for the X-coated Terumo® Dual-Stage Venous Cannula are the same intended use(s)/indications that currently exist for the unmodified Terumo® Dual-Stage Venous Cannula.
TCVS asserts that the intended use(s)/indications for the X-coated Sarns™ Dual Stage Venous Return Catheters are the same intended use(s)/indications that currently exist for the unmodified Sarns™ Dual Stage Venous Return Catheters.
TCVS asserts that the intended use(s)/indications for the X-coated Sarns™ Malleable Venous Return Catheters are the same intended use(s)/indications that currently exist for the unmodified Sarns™ Malleable Venous Return Catheters.
TCVS asserts that the intended use(s)/indications for the X-coated Sarns™ Malleable Dual Stage Venous Return Catheters are the same intended use(s)/indications that currently exist for the unmodified Sarns™ Malleable Dual Stage Venous Keturn Catheters.
TCVS asserts that the intended usc(s)/indications for the X-coated Sarns™ High How Aortic Arch Cannula are the same intended use(s)/indications that currently exist for the unmodified Sarns™ High Flow Aortic Arch Cannula.
TCVS asserts that the intended use(s)/indications for the X-coated Sarns™ Soft-Flow™ Extended Aortic Cannula are the same intended use(s)/indications that currently exist for the unmodificd Sarns™ Soft-Flow™ Extended Aortic Cannula.
These devices are indicated for up to 6 hours of use.
6.4. Principles of Operation and Technology:
The X-coated venous and arterial cannulae discussed in this submission are used in open heart surgery.
During open heart surgery blood is drained into a venous cannula just upstream of the heart, at the superior / inferior vena cava and right atrium. The cannula is connected to tubing that routes the blood to a heart / lung machine where the blood is pumped and oxygenated. The blood then continues through this perfusion circuit back to the outlet side of the heart (the paticnt's aorta), where the blood re-enters the patient's circulatory system via the ascending aorta through an arterial cannulae. Platelets adhesion is reduced while blood travels through this circulatory path on the surface covered with X-coat polymer.
{3}------------------------------------------------
6.5. Design and Materials:
The design of cannulae and catheters is such that they meet their stated intended use, and provide an acceptable level of performance and safety to the patients. Except for the X-coat application, the design and material of the X-coated cannulae will remain the same as the unmodified cannulae. The X-coat application does not affect the cannulae design.
Venous and arterial cannulae contained in this submission are structured as straight, or slightly curved, flexible tubes which are inserted into the heart to circulate blood. Depending on the location of cannulation, size, and clinicians' preference, there is a range of design variations. These features include, but are not limited, wire reinforcement, various stylets used to help guide the cannula, tip formation (conical vs. angled), number of drainage baskets (two stage), luers, and connectors.
The generic materials for these cannulae will remain the same.
6.6. Performance Evaluations:
The performance of the X-coated Cannulac will rcmain the same as the unmodified cannulae. The X-coat application does not affect the functional cannulae performance. The X-coat application reduces the platelet adhesion.
The following performance evaluations of the cannulae were conducted to demonstrate its equivalence to the uncoated cannulae.
- . Hemolysis
- . Bond Strength
- o Leak testing
- o Pull strength testing
- o Flex testing
- Accelcrated aging / Shelf life test .
6.7. Substantial Equivalence Comparison:
6.7.1. Intended Use:
The intended use of the modified device will remain the same as the predicate device.
6.7.2. Principles of Operation and Technology:
The principle of operation & technology used in the modified device will remain the same as the predicate device.
{4}------------------------------------------------
6.7.3. Design and Materials:
The design and materials, except for the X-coat application, will remain the same.
6.7.4. Substantial Equivalence Summary:
In summary, the X-coated cannulae and the uncoated cannulae are substantially equivalent in intended use, principles of operation and technology, design, and performance. Any noted differences between the devices do not raise new issues of safety and effectiveness.
6.8. Additional Safety Information
- . Sterilization conditions have been validated in accordance with AAMI guidelines to provide a Sterility Assurance Lovel (SAL) of 10".
- Post-stcrilization release for use will be determined in consideration of maximum . Ethylene Oxide, Ethylene Chlorohydrin and Ethylene Glycol (as appropriate) residue limits and maximum levels of patient exposure in accordance with EN ISO 10993-7 and AAMI TIR-19.
- . Biocompatibility studies were conducted using the same matcrials or combination of materials for predicate devices as recommended in the FDA General Program Mcmorandum #G95-1 (5/1/95): Use of International Standard ISO 10993: 2003, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing," [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible.
- . Terumo has conducted material characterization studies - including physio-chemical profiles of aged and non-aged devices to demonstrate stability of the materials, and found the materials to be stable over the expiry of the product.
6.9. Conclusion
In summary, the X-coated cannulac are substantially equivalent in intended use, principles of operation and technology, design and performance to the uncoated cannulae.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the symbol.
ood and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Terumo Cardiovascular Systems Corp. c/o Mr. Junka Kurosawa Senior Associate, Regulatory Management 6200 Jackson Rd. Ann Arbor, MI 48103
DEC 1 8 2008
Re: K083301
X-coated Terumo® Dual-Stage Venous Return Cannula, X-coated Sarns™ Dual-Stage Venous Return Catheters X-coated Sams™ Malleable Venous Return Catheters, X-coated Sarns™ Malleable Dual-Stage Venous Return Catheters, X-coated Sarns™ High Flow Aortic Arch Cannula, X-coated Sarns™ Soft-Flow™ Extended Aortic Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing Regulatory Class: Class II (two) Product Code: DWF Dated: November 7, 2008 Received: November 18, 2008
Dear Mr. Kurosawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device
{6}------------------------------------------------
Page 2 - Mr. Junka Kurosawa
can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
R. Vahner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K083301
Device Name: X-coated Terumo® Dual-Stage Venous Return Cannula, X-coated Sarns™ Dual-Stage Venous Return Catheters X-coated Sarns™ Malleable Venous Return Catheters, X-coated Sarns™ Malleable Dual-Stage Venous Return Catheters, X-coated Sarns™ High Flow Aortic Arch Cannula, X-coated Sarns ™ Soft-Flow TM Extended Aortic Cannula.
Indications For Use:
The X-coating is intended to reduce the adhesion of platelets to the blood contacting surfaces of the device during cardiopulmonary bypass procedures.
The X-coated Terumo® Dual-Stage Venous Return Cannula is indicated for use for adult single tube blood drainage from the right atrium and inferior vena cava (IVC) and delivery of the blood to the extracorporeal circuit during cardiopulmonary bypass.
The X-coated Sarns™ Dual-Stage Venous Return Catheters are indicated for single tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery.
The Sarns ™ Malleable Venous Return Catheters are indicated for venous drainage during cardiopulmonary bypass surgery: the 24, 28, 30, 32, 34, 36 and 40 Fr. are for dual cannulation of the superior and inferior vena cava: the 40 Fr. is for single cannulation of the right atrium.
The X-coated Sarns™ Malleable Dual-Stage Venous Return Catheters are indicated for single-tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery.
The X-coated Sarns™ High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
The X-coated Sarns™ Soft-Flow™ Extended Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
These devices are indicated for up to 6 hours of use.
12-12-0
-- I Page
CONFIDENTIAL
. . . . . 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
{8}------------------------------------------------
Prescription Use X OR
Over-The-Counter Use
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
. .
Concurrence of CDRH, Office of Device Evaluation (ODE)
une R.v.hnes
(Division Sign-Off) (Division of Cardiovascular Devices
510(k) Number_Ko8 330 |
§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.
(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).