K Number
K083180

Validate with FDA (Live)

Date Cleared
2009-02-12

(107 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Device Description

The Third Eye System consists of both disposable portions and facility equipment. The disposable portions include the Third Eye Retroscope and Caps. The facility equipment portions of the device include the Third Eye Video Processor and accessories. The Third Eye is designed as an auxiliary device for use during a colonoscopy procedure. After a colonoscope has been advanced to the cecurn, the Third Eye Retroscope is inserted through the instrument channel of the colonoscope. As the Third Eye Retroscope emerges from the distal tip of the instrument channel of the colonoscope, it automatically bends 180 degrees to form a "J" shape. The Third Eye then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope. The Third Eve Video Processor is a piece of electrical equipment that acts as an interface for the Third Eye Retroscope, the video output signals, and user input controls. The output video signals from the Video Processor are displayed on a monitor for viewing by the physician. Application software that runs on the Video Processor allows the user to adjust the light intensity and various picture settings of the Third Eve Retroscope image.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text for the Avantis Medical Systems' Third Eye Retroscope Auxiliary Endoscopy System:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance
Safety: Device modifications do not raise new or unresolved safety issues. Adherence to general controls provisions (registration, labeling, GMP)."Modifications made to the device and accessories did not change the fundamental scientific technology of the device. The testing described demonstrates that the differences in the device components and accessories do not raise any new or unresolved issues for safety and efficacy." "Risks to patients in terms of adverse events are comparable to those reported in the literature for traditional screening colonoscopies. Use of the Third Eye did not increase the risks to the patients..."
Effectiveness/Performance: Device performs as intended."Bench verification testing was conducted to verify that the modified device meets the design inputs and intended performance characteristics. Results demonstrate that modifications made to the Third Eye design performs as intended." "Results from a prospective, multi-center post-market study demonstrate that retrograde viewing of the colon with the Third Eye improves the detection of polyps and adenomas compared to traditional colonoscopy alone. Results were statistically significant."
Clinical Benefit: Improves detection of polyps and adenomas."Avantis Medical Systems, Inc. has determined, based on clinical data, that the Third Eye improves the detection of polyps and adenomas in patients for whom colonoscopies are indicated." "Use of the Third Eye did not increase the risks to the patients but offered a statistically significant benefit of improved detection."
Substantial Equivalence: To the predicate device (K070330) despite modifications."The Avantis Third Eye is substantially equivalent in operation and fundamental scientific technology to its previous version, cleared under K070330." "The modified device is substantially equivalent to the cleared predicate device in fundamental design materials, processes, sterile barrier packaging, and sterilization. All changes have been subjected to risk analysis and have been verified to meet current design specifications."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated in terms of a specific number of patients or cases. The document refers to a "multi-center post-market study."
  • Data Provenance:
    • Country of Origin: Not specified.
    • Retrospective or Prospective: Prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • This information is not provided in the document. The document states that the Third Eye "improves detection of polyps and adenomas," implying a comparison, but it does not detail how the ground truth for "polyps and adenomas" was established in the clinical study (e.g., standard colonoscopy findings, pathology, or a panel of experts).

4. Adjudication Method for the Test Set

  • This information is not provided in the document.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • It states that a clinical study "demonstrate(s) that retrograde viewing of the colon with the Third Eye improves the detection of polyps and adenomas compared to traditional colonoscopy alone." This indicates a comparative effectiveness study, however, it's not explicitly labeled as an MRMC study and the specific effect size quantifying "how much human readers improve with AI vs without AI assistance" is not provided. The document only states that the results were "statistically significant." The device itself is an auxiliary endoscope, not an AI, so the framing of "human readers improve with AI vs without AI assistance" isn't directly applicable in this context. It's about improved detection with the device vs. without.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • No, this was not done. The device is an auxiliary endoscopic system designed for human use during a colonoscopy to provide additional visualization. It is not an algorithm that operates independently.

7. The Type of Ground Truth Used

  • The document states the study measured "detection of polyps and adenomas." While not explicitly defined as pathology, in the context of colonoscopy clinical trials, pathology from biopsied or resected lesions is the generally accepted gold standard for confirming polyps and adenomas. However, the document does not definitively state that pathology was the sole or primary ground truth method.

8. The Sample Size for the Training Set

  • This information is not applicable/not provided. The Third Eye Retroscope is a medical device, not an AI algorithm that requires a "training set" in the machine learning sense. The clinical study described is for validation of the device's performance.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable/not provided for the same reason as point 8.

{0}------------------------------------------------

K 0831.80

Image /page/0/Picture/1 description: The image shows the logo for Avantis Medical Systems. The logo consists of the word "avantis" in lowercase letters, with the words "medical systems" in smaller letters below it. Above the word "avantis" is a stylized graphic that appears to be two curved shapes with a circle in the middle.

510(k) SUMMARY

FEB 1 2 2009

Applicant Information: a)

Date Summary Prepared09 February 2009
Sponsor/SubmitterAvantis Medical Systems, Inc.263 Santa Ana CourtSunnyvale, CA 94085-4511
Correspondent Contact InformationAmrita SethiDirector of Regulatory and Quality AffairsPhone: 408-733-1901 x227Fax: 408-733-1847E-mail: asethi@avantismedical.com

Device Information: b)

Device Common NameEndoscope
Device Trade & Proprietary NameThird Eye® Retroscope® AuxiliaryEndoscopy System (Third Eye)
Device Classification NameColonoscope (per 21CFR 876.1500)
Device Classification Regulation21CFR 876.1500
Device ClassificationClass II (special controls)
Device Classification & Product CodeFDF, FDS, KNW, KOG

Identification of Predicate Device: c)

The Avantis Third Eye is substantially equivalent in operation and fundamental scientific technology to its previous version, cleared under K070330.

d) Device Description Summary:

The Third Eye System consists of both disposable portions and facility equipment. The disposable portions include the Third Eye Retroscope and Caps. The facility equipment portions of the device include the Third Eye Video Processor and accessories.

The Third Eye is designed as an auxiliary device for use during a colonoscopy

{1}------------------------------------------------

Ko 83i80

Image /page/1/Picture/1 description: The image shows the logo for Avantis Medical Systems. The logo features the word "avantis" in a lowercase sans-serif font, with the words "medical systems" in a smaller font below. Above the word "avantis" is a graphic of two curved shapes facing each other with a circle in the middle.

procedure. After a colonoscope has been advanced to the cecurn, the Third Eye Retroscope is inserted through the instrument channel of the colonoscope. As the Third Eye Retroscope emerges from the distal tip of the instrument channel of the colonoscope, it automatically bends 180 degrees to form a "J" shape. The Third Eye then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope.

The Third Eve Video Processor is a piece of electrical equipment that acts as an interface for the Third Eye Retroscope, the video output signals, and user input controls. The output video signals from the Video Processor are displayed on a monitor for viewing by the physician. Application software that runs on the Video Processor allows the user to adjust the light intensity and various picture settings of the Third Eve Retroscope image.

  • Intended Use: e)
    The Third Eye Retroscope is intended for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Discussion of Substantial Equivalence to Predicate Device: f)

The Third Eye has been modified to a) accommodate additional conventional colonoscope models and brands, b) provide product enhancements, c) improve manufacturability, and d) update software. The modified device is substantially equivalent to the cleared predicate device in fundamental design materials, processes, sterile barrier packaging, and sterilization. All changes have been subjected to risk analysis and have been verified to meet current design specifications.

The IFU has been updated to a) add clinical data, b) clarify procedural directions based on the clinical experiences and c) expand the indications for use based on clinical data demonstrating that the Third Eye used with a traditional colonoscope improves detection of polyps and adenomas during colonoscopy as compared to the use of a colonoscope alone.

Differences between the modified Third Eye System and the predicate device have no impact on safety and effectiveness.

Scientific Technology g)

The Third Eye Retroscope® Auxiliary Endoscopy System includes the Retroscope, a Video Processor, Cap and accessories. Modifications made to the device and accessories did not change the fundamental scientific technology of the device. The testing described demonstrates that the differences in the device components and

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Avantis Medical Systems. The logo features the word "avantis" in lowercase letters, with the words "medical systems" underneath. Above the word "avantis" is a graphic of two curved shapes that are connected by a circle in the middle.

R08 31 80

yg 3 of 3

accessories do not raise any new or unresolved issues for safety and efficacy.

Summary of Supporting Non-Clinical Performance Data h)

Bench verification testing was conducted to verify that the modified device meets the design inputs and intended performance characteristics. Results demonstrate that modifications made to the Third Eve design performs as intended.

Summary of Supporting Clinical Data i)

Results from a prospective, multi-center post-market study demonstrate that retrograde viewing of the colon with the Third Eye improves the detection of polyps and adenomas compared to traditional colonoscopy alone. Results were statistically significant.

Conclusions j)

Avantis Medical Systems, Inc. has determined, based on clinical data, that the Third Eye improves the detection of polyps and adenomas in patients for whom colonoscopies are indicated. Risks to patients in terms of adverse events are comparable to those reported in the literature for traditional screening colonoscopies. Use of the Third Eye did not increase the risks to the patients but offered a statistically significant benefit of improved detection. Results of the clinical study are included in the Instructions for Use.

Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The present submission and statements therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing

263 Santa Ana Ct., Sunnyvale, CA 94085 TEL: (408) 733-1901, FAX: (408) 733-1847

3 of 3

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 2 2009

Ms. Amrita Sethi Director of Regulatory and Quality Affairs Avantis Medical Systems, Inc. 263 Santa Ana Court SUNNYVALE CA 94085-4511.

Re: K083180

Trade/Device Name: Third Eye® Retroscope® Auxiliary Endoscopy System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF and ODA Dated: February 4, 2009 Received: February 5, 2009

Dear Ms. Sethi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K083180

Device Name: Third Eye® Retroscope® Auxiliary Endoscopy System (Third Eye)

Indications for Use: The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Prescription Use × Over-The-Counter Use - AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Qubuthemen

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.