(218 days)
EliA™ RNP70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to RNP70 in human serum and plasma (heparin, EDTA) as an aid in the clinical diagnosis of mixed connective tissue disease (MCTD) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA™ RNP70 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
EliA™ Scl-70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Scl-70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of scleroderma (diffuse form) in conjunction with other laboratory and clinical findings. EliA™ Sci-70 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
EliATM Jo-1 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Jo-1 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of polymyositis / dermatomyositis in conjunction with other laboratory and clinical findings. EliATM Jo-1 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
The new devices belong to a fully integrated and automated system for immunodiagnostic testing. It comprises a Fluorescence-Immunoassay test system using EliA single wells as the solid phase and is intended to be performed on the instruments ImmunoCAP 100 and ImmunoCAP 250. The conjugate for the EliA IgG method is mouse anti-human IgG beta-galactosidase, which uses 4-Methylumbellifery|-BD-Galactoside as substrate. The total IgG calibration is based on a set of six WHOstandardized IgG Calibrators derived from human serum. They are used to establish an initial calibration curve, which may be used for up to 28 days on additional assays and can be stored by the instrument. Each additional assay includes calibrator (curve) controls that have to recover in defined ranges to ensure that the stored calibration curve is still valid. The Fluorescence-Immunoassay test system includes test-, method specific and general reagents that are packaged as separate units.
This document describes the EliA™ RNP70, EliA™ Scl-70, and EliA™ Jo-1 immunoassays, which are intended for the semi-quantitative measurement of IgG antibodies to RNP70, Scl-70, and Jo-1 respectively, as an aid in diagnosing specific autoimmune diseases. The information provided is for their 510(k) premarket notification and focuses on establishing substantial equivalence to previously marketed predicate devices.
Acceptance Criteria and Reported Device Performance
The provided text does not explicitly list quantitative acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy thresholds) or the reported device performance against such criteria in a table format.
Instead, the submission states that "the comparability of predicate device and new device is supported by a data set including: results obtained within a comparison study between new and predicate device, results obtained for clinically defined sera, results obtained for samples from apparently healthy subjects (normal population)." It concludes: "In summary, all available data support that the new devices are substantially equivalent to the predicate devices."
This implies a qualitative acceptance criterion of substantial equivalence to the predicate devices (Varelisa U1RNP Antibodies K993589, Varelisa Scl-70 Antibodies K944172, Varelisa Jo-1 Antibodies K944173). The "study that proves the device meets the acceptance criteria" is broadly described as a "comparison study between new and predicate device," along with testing on "clinically defined sera" and "samples from apparently healthy subjects." However, specific quantitative results from these studies were not detailed in the provided summary.
Detailed Information as Requested:
-
Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criterion (Implicit) Reported Device Performance (Summary) Substantial equivalence to predicate devices (Varelisa U1RNP Antibodies K993589, Varelisa Scl-70 Antibodies K944172, Varelisa Jo-1 Antibodies K944173) "All available data support that the new devices are substantially equivalent to the predicate devices." (Quantitative results are not provided in this summary) -
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not specified in the provided summary. The text mentions "a data set including results obtained within a comparison study," "clinically defined sera," and "samples from apparently healthy subjects."
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be for regulatory submission, but it's not indicated if they were prospective or retrospective. Given they involve "clinically defined sera" and "normal population" samples, these would likely be retrospective selections or collections for a specific validation study.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not specified. The ground truth for "clinically defined sera" would presumably be based on established clinical diagnoses, likely determined by medical specialists, but the number and qualifications of experts involved in this determination are not mentioned.
-
Adjudication Method for the Test Set:
- Not specified. This information is typically relevant for studies where multiple readers or assessors establish ground truth.
-
If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an in vitro diagnostic (IVD) immunoassay, not an AI software intended for human reader assistance with image interpretation or complex decision-making. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant to this type of device.
-
If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, implicitly. The device is an automated immunoassay system (EliA IgG method on ImmunoCAP® 100/250 instruments) that performs measurements and provides semi-quantitative results. This is inherently a "standalone" or "algorithm only" performance, as it measures antibody levels in patient samples without direct human interpretation of the assay's output during the measurement process. The results are then used by clinicians as an "aid in clinical diagnosis."
-
The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):
- Likely clinical diagnosis based on a combination of clinical findings and possibly other laboratory tests, given the mention of "clinically defined sera" for conditions like mixed connective tissue disease (MCTD), systemic lupus erythematosus (SLE), scleroderma, and polymyositis/dermatomyositis. For the "normal population" samples, the ground truth would be the absence of these diseases.
-
The Sample Size for the Training Set:
- Not specified. The document does not provide details about a training set, which is more common for machine learning or AI-based devices. For an immunoassay, method development and optimization would occur, but it's not typically categorized as a "training set" in the same way as AI.
-
How the Ground Truth for the Training Set was Established:
- Not applicable, as a distinct "training set" with ground truth establishment for an AI/ML algorithm is not described for this immunoassay device. Quality control and assay calibration are established using standardized materials and calibrators, as mentioned (e.g., "WHO-standardized IgG Calibrators derived from human serum").
{0}------------------------------------------------
510(k) Summary
MAY 28 2009
This summary information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
Assigned 510(k) Number:
| Date of Summary Preparation: | October 20, 2008 |
|---|---|
| Manufacturer: | Phadia ABRapsgatan 7SE-751 37 Uppsala, Sweden |
| 510 (k) Contact Person: | Martin MannRegulatory Affairs ManagerPhadia US Inc.4169 Commercial AvenuePortage, Mi 49002, USA+1 (-269-492) -1957 (Phone)+1 (-269-492) -7541 (Fax)martin.mann@phadia.com |
| Device Name: | EliA™ RNP70 ImmunoassayEliA™ Scl-70 ImmunoassayEliA™ Jo-1 Immunoassay |
| Common Name: | Antinuclear antibody immunological test system andControl |
Classification
| Product Name | Product Code | Class | CFR |
|---|---|---|---|
| EliATM RNP70 | LJM | II | 866.5100 |
| EliATM Scl-70 | LJM | II | 866.5100 |
| EliATM Jo-1 | LJM | II | 866.5100 |
Substantial Equivalence to
| Varelisa U1RNP Antibodies | K993589 |
|---|---|
| Varelisa Scl-70 Antibodies | K944172 |
| Varelisa Jo-1 Antibodies | K944173 |
{1}------------------------------------------------
Intended Use Statements of the New Devices
-
EliATM RNP70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to RNP70 in human serum and plasma (heparin, EDTA) as an aid in the clinical diagnosis of mixed connective tissue disease (MCTD) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA™ RNP70 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
-
EliA™ Scl-70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Scl-70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of scleroderma (diffuse form) in conjunction with other laboratory and clinical findings. EliA™ Sci-70 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
-
EliATM Jo-1 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Jo-1 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of polymyositis / dermatomyositis in conjunction with other laboratory and clinical findings. EliATM Jo-1 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
Special condition for use statement
The device is for prescription use only.
Special instrument requirements
ImmunoCAP® 100/ImmunoCAP® 250 are fully automated immunoassay analyzers, which include software for evaluation of test results.
General Description of the New Devices
The new devices belong to a fully integrated and automated system for immunodiagnostic testing. It comprises a Fluorescence-Immunoassay test system using EliA single wells as the solid phase and is intended to be performed on the instruments ImmunoCAP 100 and ImmunoCAP 250. The conjugate for the EliA IgG method is mouse anti-human IgG beta-galactosidase, which uses 4-Methylumbellifery|-BD-Galactoside as substrate. The total IgG calibration is based on a set of six WHOstandardized IgG Calibrators derived from human serum. They are used to establish an initial calibration curve, which may be used for up to 28 days on additional assays and can be stored by the instrument. Each additional assay includes calibrator (curve) controls that have to recover in defined ranges to ensure that the stored calibration curve is still valid. The Fluorescence-Immunoassay test system includes test-, method specific and general reagents that are packaged as separate units.
{2}------------------------------------------------
Test Principle of the New Devices
The EliA Wells are coated with the following antigens:
| Test | Antigen coated to the wells: |
|---|---|
| EliA RNP70 | human recombinant RNP (70 kDa) protein |
| EliA Scl-70 | human recombinant Scl-70 protein |
| EliA Jo-1 | human recombinant Jo-1 protein |
If present in the patient's specimen, antibodies to the antigens mentioned above bind to their specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the more specific IgG is present in the specimen. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.
Device Comparison
The new and the predicate devices both represent non-competitive solid phase ELISAs. Both IVDs are used as an aid in the diagnosis of the following diseases:
| Disease | Detection of antibodies to |
|---|---|
| mixed connective tissue disease (MCTD) | RNP70 |
| systemic lupus erythematosus (SLE) | RNP70 |
| scleroderma (diffuse form) | Scl-70 |
| polymyositis / dermatomyositis | Jo-l |
Laboratory equivalence
The comparability of predicate device and new device is supported by a data set including
- · results obtained within a comparison study between new and predicate device
- results obtained for clinically defined sera
- results obtained for samples from apparently healthy subjects (normal population).
In summary, all available data support that the new devices are substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
MAY 28 2009
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Phadia US Inc. c/o Mr. Martin Mann Regulatory Affairs Manager 4169 Commercial Avenue Portage, MI 49002
Re: K083117
Trade/Device Name: EliA™ RNP70, ELIA™ Scl-70 and EliA™ Jo-1 Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LJM, LKO Dated: April 03, 2009 Received: April 06, 2009
Dear Mr. Mann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced-above and-have determined-the device is substantially-equivalent-(for-the-indicationsfor use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice
{4}------------------------------------------------
Page 2 - Mr. Martin Mann
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
ia m. Chen
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE
510(k) Number:
K083117 -
EliA™ Jo-1
Device Name:
Indications For Use:
EliA™ Jo-1 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Jo-1 in human serum and plasma (heparin, EDTA, citrate) as . an aid in the clinical diagnosis of polymyositis / dermatomyositis in conjunction with other laboratory and clinical findings. EliA™ Jo-1 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
Prescription Use AND/OR V (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
the Thu
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
51000 K083117
Concurrence of CDRH, Office of Device Evaluation (ODE)
{6}------------------------------------------------
INDICATIONS FOR USE
510(k) Number:________________________________________________________________________________________________________________________________________________________________ K083117
Device Name: EliA™ RNP70
Indications For Use:
EliA™ RNP70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to RNP70 in human serum and plasma (heparin, EDTA) as an aid in the clinical diagnosis of mixed connective tissue disease (MCTD) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA™ RNP70 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
AND/OR Over-The-Counter Use Prescription Use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
w Nh
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K083117
Concurrence of CDRH, Office of Device Evaluation (ODE)
{7}------------------------------------------------
INDICATIONS FOR USE
510(k) Number:
Device Name:
EliATM Scl-70
Indications For Use:
EliA™ Scl-70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Scl-70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of scleroderma (diffuse form) in conjunction with other laboratory and clinical findings. EliA™ Scl-70 uses the EliA IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Tur
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K083117
Concurrence of CDRH, Office of Device Evaluation (ODE)
§ 866.5100 Antinuclear antibody immunological test system.
(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).