K Number
K082883
Date Cleared
2009-06-16

(260 days)

Product Code
Regulation Number
886.5925
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity..

Eyecare practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.

Device Description

EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 38% water. The hydrogel lens' material is a copolymer composed of 2-hydroxyethyl methacrylate (HEMA), which was cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photo- polymerization. The EZvue (Polymacon) Contact lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline.

AI/ML Overview

The provided 510(k) summary for the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear describes a clinical study and nonclinical tests to demonstrate safety and effectiveness. However, it does not explicitly state "acceptance criteria" in a quantitative manner for specific performance metrics. Instead, it focuses on demonstrating substantial equivalence to predicate devices.

Interpretation of Acceptance Criteria and Reported Performance:

For a contact lens, "acceptance criteria" would typically relate to factors like visual acuity correction, comfort, safety (lack of adverse events), and various optical and material properties. The document implies that the "acceptance criteria" are met if the device performs comparably to the predicate devices and demonstrates safety and effectiveness.

Here's an attempt to extract the implied acceptance criteria and reported performance based on the provided text, recognizing that these are not explicitly presented as pass/fail thresholds but rather as general findings.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Clinical Performance:
Correction of refractive ametropia (myopia and hyperopia)"Nearly a hundred percent of the participants' vision was corrected."
Patient satisfaction with wear and care"Nearly all were satisfied with the lens wearing and care of lenses."
Overall safety and good performance"In general, these products are good and safe for customers."
Biocompatibility:
Non-toxic and non-irritating"The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated."
Non-cytotoxic"The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study."
Sterilization Assurance:
Sterility Assurance Level (SAL) of $10^{-6}$"Steam sterilization process has been validated to deliver a minimum SAL of $10^{-6}$."
Shelf-life stability for sterility"There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product."
Leachability:
No detectable leachable monomers or additive residues"The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues."
Technological Characteristics (Substantial Equivalence):
Comparable to predicate lenses in:
- Diameter range- 13.90~14.30mm (Reported)
- Power range- -0.50D~-12.00D (Reported)
- Center thickness- Varies with power (0.080~0.150mm) (Reported)
- Refractive index- 1.430 (hydrated) (Reported)
- Light transmittance- >93% (Reported)
- Water content- 38 % (Reported)
- Oxygen permeability- 9×10⁻¹⁵ (cm²/sec)(mlO2/ml-mmHg)]@ 35℃ (Reported)
"These characterizations of EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear are equivalent and comparable to those of predicate lenses." (General statement of equivalence)
Overall Safety and Effectiveness"Testing performed on the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices... The risks and benefits of the subject device are the same as soft contact lenses for to the daily wear." (Conclusion of equivalence)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Study: 60 human eyes.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study data or if it was retrospective or prospective. Given the submitter information (I-SEE VISION TECHNOLOGY INC. based in Taiwan), it is likely the study was conducted in Taiwan, but this is not explicitly stated. The study type is implicit as a prospective clinical trial.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The document does not specify the number of experts used or their qualifications for establishing ground truth in the clinical study. The clinical study summary describes participants' vision being corrected and satisfaction, which are likely assessed by an optometrist or ophthalmologist involved in the study, but no details on their role as "ground truth" experts or their qualifications are provided.

4. Adjudication Method for the Test Set

  • The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical study results. Clinical studies for contact lenses typically involve direct measurements and patient feedback, rather than expert adjudication of a "ground truth" derived from images or complex diagnoses.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic tools where human readers might integrate AI outputs. This submission is for a medical device (contact lens) and does not involve AI assistance for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a physical medical device (contact lens), not an algorithm or AI product. Its performance is inherent in its physical properties and clinical effect on the wearer, not as a standalone algorithm.

7. The Type of Ground Truth Used

  • For the clinical study, the "ground truth" implicitly refers to:
    • Corrected visual acuity: Assessed by standard optometric methods.
    • Patient satisfaction: Subjective feedback from participants.
    • Safety observations: Clinical examination for adverse events.
  • For the nonclinical studies, the ground truth was based on:
    • Standardized laboratory tests and analytical measurements: For material properties, biocompatibility (e.g., irritation tests), sterilization validation, and leachability. These are objective measurements following established protocols.

8. The Sample Size for the Training Set

  • This submission describes a medical device approval through the 510(k) pathway, which primarily focuses on demonstrating substantial equivalence to a predicate device. It does not involve machine learning or AI models that require a "training set." Therefore, there is no training set sample size.

9. How the Ground Truth for the Training Set was Established

  • As stated above, there is no training set for this type of device submission.

{0}------------------------------------------------

K082883

JUN 1 6 2009

510 (K) Summary

JUN 16 2009

SUMMARY OF SAFETY AND EFFECTIVENESS FOR EZvue (Polymacon) Soft (Hydrophilic) Contact Lens for Daily Wear

Submitter Information:
Company:I-SEE VISION TECHNOLOGY INC.
1F, No. 45, Lane 2, Section 2, Kuang-Fu Rd., Hisnchu, Taiwan,
30071
Contact Person:Mr. Tom Lin,
Special Assistant of President Office
Phone:886-3-5752822
FAX:886-3-5752823
E-Mail:tomlin@iseevision.com

Identification of Device:

Classification Name:Lens, Soft Contact, Daily Wear, per 21 CFR. 886.5925
Trade Name:EZvue (Polymacon) Soft (Hydrophilic) Contact Lens for
Daily Wear
Common or usual Name:Soft (hydrophilic) Contact lens (daily wear)
FDA Classification:Class II

Predicate Device:

  • BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted છે Contact Lens cleared via K020927 BAUSCH & LOMB®
  • SuperView (polymacon) Contact Lens for Daily Wear (K033136) made � from INNOVA VISION INC.
  • Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA 0 number N16895/S087) by BAUSCH & LOMB, INC.

Indications for Use

EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity..

Eyecare practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.

{1}------------------------------------------------

Description of Device

EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 38% water. The hydrogel lens' material is a copolymer composed of 2-hydroxyethyl methacrylate (HEMA), which was cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photo- polymerization. The EZvue (Polymacon) Contact lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline.

Summary of Clinical Study:

The EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear were tested over 60 human eyes separately within 6 months. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.

Nonclinical Studies:

A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear, and establish substantial equivalence to predicate lenses - BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted Contact Lens (K020927). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(K) Guidance Document for Class IV Contact lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:

a) Technological characteristics studies

EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear designs in the following parameter ranges:

Diameter range; 13.90~14.30mm

  • Power range: -0.50D~-12.00D
    Center thickness: varies with power (0.080~0.150mm)

Lenses have the following properties:

Refractive index: 1.430 (hydrated)

Light transmittance: >93%

Water content: 38 %

Oxygen permeability (edged corrected): 9×10~15(cm²/sec)(mlO2/ml-mmHg)]@ 35℃

There characterizations of EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear s are equivalent and comparable to those of predicate lenses.

{2}------------------------------------------------

b) Biocompatibility

In accordance with the May 1994 Guidance Document for daily wear contact lenses, toxicity studies have been conducted on the model: EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.

c) Microbiology

Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group IV. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.

d) Leachability

Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.

Substantial equivalence Statement:

Testing performed on the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted Contact Lens (K020927), SuperView (polymacon) Contact Lens for Daily Wear (K033136) made from INNOVA VISION INC. , & Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA number N16895/S087) by BAUSCH & LOMB, INC. , when used in accordance with the instructions for use. It is due to the facts that The risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.

In conclusion, it is I-SEE's conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device-- BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted Contact Lens (K020927) , SuperView (polymacon) Contact Lens for Daily Wear (K033136) made from INNOVA VISION INC. , & Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA number N16895/S087) by BAUSCH & LOMB, INC ..

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three bars representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

I-See Vision Technology Inc. c/o Ms. Jennifer Reich Senior Consultant Harvest Consulting Corp. 2904 N. Boldt Drive Flagstaff, AZ 86001

Jun 1 6 2009

Re: K082883

Trade/Device Name: EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: LPL Dated: June 9, 2009 Received: June 11, 2009

Dear Ms. Reich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Kesia Alexander for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K081883

Device Name: EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear I-SEE Vision Technnology Inc.

Indications For Use:

EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.

Eyecare practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE, IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

MD

Division Sim Division of Onthalm Nose and Throat Dev

Page 1 of 1

Nose and Throat Devices,
510(k) Number
K082883

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.