(178 days)
The EVOX Electro Diagnostic Device is intended for the testing of sacral reflexes, pudendal nerve terminal motor latency (PNTML) studies. electromyography (EMG), and cerebral pudendal somatosensory evoked potential (SEP).
The EVOX Electro-Diagnostic Device utilizes Laborie Urodynamic equipment 510(k) exempt under section § 876.1620 or with another microprocessor that is compatible with the required capability.
The Laborie EVOX Electro Diagnostic Device will help diagnose and perform evaluation tests for urinary and fecal incontinence as well as other neurological and physiological assessment, including:
Pudendal Nerve Terminal Motor Latency with St Marks Electrodes Anal Sphincter CN EMG
Cerebral Pudendal Somatosensory Evoked Potential
Sacral Reflexes
Calculations will include: latency, peak, duration, conduction velocity and area.
The equipment includes the Evox Electro-diagnostic Software and the hardware required to perform the stated tests which include:
- Laborie 94-based Hardware with Integrated Electrodiagnostic Stimulation .
- . Ring/Bar Stimulation Electrodes
- Needle/Cup/Patch EMG/ECG Electrodes .
- St. Mark's Electrode and Cable .
- EMG/ECG Cabling .
The provided 510(k) summary for the KOR2614 Laborie EVOX Electro Diagnostic Device does not contain the detailed study results, acceptance criteria, or performance data typically found in a comprehensive clinical or performance study report.
Instead, this document focuses on establishing substantial equivalence to predicate devices through technical characteristics and intended use. The "Testing" section explicitly states that "Bench studies have confirmed the efficacy of the EVOX Electro Diagnostic Device. Safety testing included electrical safety testing and electromagnetic compatibility testing to recognized standards." This indicates that the device's efficacy was confirmed via bench testing, not through a clinical study with human subjects, ground truth, or expert review in the way a diagnostic AI/ML device would be evaluated.
Therefore, many of the requested categories cannot be populated from the provided text.
Here's a breakdown based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Efficacy (via bench testing) | Confirmed by bench studies (details not provided) |
| Electrical Safety | Complies with IEC 60601-1 and IEC 60601-2-10 standards. |
| Electromagnetic Compatibility | Complies with recognized standards (details not provided). |
| Technical Characteristics | Detailed in the "Summary of the Technical Characteristics" table, showing the Laborie EVO's specifications against predicate devices. The implicit acceptance criterion here is that the technical specifications are comparable to, or better than, the predicate devices for achieving the intended use. (See full technical table in the original document for specific values like Common Mode Rejection, Noise, etc.) |
2. Sample size used for the test set and the data provenance
Not applicable. The document states "Bench studies have confirmed the efficacy," implying a technical verification rather than a clinical test set from patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. No clinical test set or human ground truth described.
4. Adjudication method for the test set
Not applicable. No clinical test set or human ground truth described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done. This device is an electro-diagnostic tool and not an AI-assisted diagnostic software in the context of this 510(k) summary.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is a standalone electro-diagnostic device. Its performance is inherent in its technical capabilities for signal acquisition and processing. The "bench studies" mentioned would represent this standalone performance, although specific metrics are not detailed in this summary.
7. The type of ground truth used
For efficacy, the ground truth would have been established by comparing the device's output signals and measurements (e.g., latency, peak, duration, conduction velocity, area) against known, accurate electrical stimuli and reference measurements in a laboratory setting (bench testing).
For safety, the ground truth is established by the specified electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-2-10).
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a training set of data.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device that requires a training set of data.
{0}------------------------------------------------
KOR2614
510(k) Summary (per 21 CFR 807.92)
MAR = $ 2009
I. Applicant
Laborie Medical Technologies, Inc. 6415 Northwest Drive, Unit 10 Mississauga Ontario Canada L4V 1X1
Contact Person: Barbara Mornet. Regulatory Affairs Deputy Tel: (802) 857-1300 Fax: (802) 878-1122 Email: bmornet@laborie.com
July 30, 2008 Date Prepared:
II. Device Name
Laborie EVOX Electro Diagnostic Device Proprietary Name: Common/ Usual Name: Classification Name: Stimulator, Electrical, Evoked Response Regulation Number: 882.1870 and 882.1550 GWF, JXE Product Codes: Classification: 2 Classification Panel: Neurology
III. Predicate Device
The Laborie Evo Electro Diagnostic Device is substantially equivalent to the Dantec Kevpoint K944547 and the Cadwell K962455 in its intended use and in its technical characteristics as well as the safety and effectiveness of the device as a diagnostic tool.
IV. Intended Use of the Device
The EVOX Electro Diagnostic Device is intended for the testing of sacral reflexes, pudendal nerve terminal motor latency (PNTML) studies. electromyography (EMG), and cerebral pudendal somatosensory evoked potential (SEP).
V. Description of the Device
The EVOX Electro-Diagnostic Device utilizes Laborie Urodynamic equipment 510(k) exempt under section § 876.1620 or with another microprocessor that is compatible with the required capability.
The Laborie EVOX Electro Diagnostic Device will help diagnose and perform evaluation tests for urinary and fecal incontinence as well as other neurological and physiological assessment, including:
Pudendal Nerve Terminal Motor Latency with St Marks Electrodes Anal Sphincter CN EMG
Cerebral Pudendal Somatosensory Evoked Potential
{1}------------------------------------------------
510(k) Summary
(per 21 CFR 807.92)
Sacral Reflexes
Calculations will include: latency, peak, duration, conduction velocity and area.
The equipment includes the Evox Electro-diagnostic Software and the hardware required to perform the stated tests which include:
- Laborie 94-based Hardware with Integrated Electrodiagnostic Stimulation .
- . Ring/Bar Stimulation Electrodes
- Needle/Cup/Patch EMG/ECG Electrodes .
- St. Mark's Electrode and Cable .
- EMG/ECG Cabling .
{2}------------------------------------------------
510(k) Summary
(per 21 CFR 807.92)
,
Summary of the Technical Characteristics
| Componentsand or Features | Laborie EVO | Predicate Device- DantecKeypointK944547 | Cadwell Sierra/Cadwell6200A K962455 |
|---|---|---|---|
| Safety | Complies withIEC 60601-1standards andIEC 60601-2-10standards. | Designed to comply withrequirements of UL 544;Isolated patient connectionIEC 601-1: Type BF | |
| Electrodeinputs | 2 ElectrodeInputs: Bar,Ring, Needle,Cup or St. MarkElectrodes. | 5-pin Din connectors | 4 Buffered electrodeinputs with separate activeand reference 1.5 mmtouch proof connectors or5 pin DIN connector |
| IsolatedGroundConnections | There is mainpower supplydouble/reinforcedisolation betweenthe UDS systemandLive/Neutral/Earth, which actas anotherisolation barrier. | 2 Connections | |
| Isolation ModeRejection | No isolationamplifier is used.For IMRRinstrumentationamplifier is used. | > 150dB | |
| Common ModeRejection | 100dB | > 100 dB | 90dB |
| Sensitivities | 2,100MicroV/div | .5μV to +/- 20,00µV/div | 2,5,10,20,50,100,200,500Micro V/div; 1,2,5,10,20m V/div. |
| Noise | 6 microV peak topeak | 2 micro V peak to peak(10Hz to 10kHz) | |
| InputImpedance | > 1,000 Mohms(common mode) | > 1,000 Mohms (common mode) | |
| Notch Filter | 50-60Hz | 50-60 Hz | 50-60Hz |
| Low-cut Filters | FIR filterSelectable at 10,50, 100, 300, 500,1000, 2000, 3000,4000 Hz | Filter selectable at20, 50,100,200,300,500,1000,2000,3000,5000,10000 Hz | 1-2 pole filterSelectable at 0.04, 0.1, 1,3, 10, 30, 100, 500 Hz |
| High-cut filters | N/A | Filter selectable at 0,0.1,0.2,0.5,1,2,5,10,20,50,100,200,500,1000,2000,3000 Hz | 2-pole filter Selectable at30, 50,100,200,300,500 Hz;1, 1.5,2,3,5,10,15kHz |
| Commonrecordingreference input | 1 input | 1 input | 1 input |
| Temperatureprobe input | N/A | 20 to 45°C | |
| StimulationSignal | MonophasicCurrent controlledfor a maximum ofup to 100mA and270 V | Biphasic and monophasic | |
| Signal Capture-EMG | +/- 1mV range | +/-5µV to +/- 20,000 µV range available +/-1to 50mV used for test | |
| Signal Capture-EEG | +/- 20µV range | +/-5µV to +/- 20,000 µV range available +/-5µV to 50 µV used for test | |
| MultipleSampleAcquisition forAccuracy | Yes | Yes | Yes |
| Event Marking | Yes | Yes | Yes |
| Software DataAnalysis | Yes | Yes | Yes |
| ReportGeneration &Data Storage | Yes | Yes | Yes |
| Patient Info | Yes | Yes | Yes |
{3}------------------------------------------------
510(k) Summary
(per 21 CFR 807.92) ,
{4}------------------------------------------------
510(k) Summary (per 21 CFR 807.92)
VI. Testing
Bench studies have confirmed the efficacy of the EVOX Electro Diagnostic Device. Safety testing included electrical safety testing and electromagnetic compatibility testing to recognized standards.
VII. Safety & Effectiveness
The Laborie EVOX Electro-Diagnostic Device is substantially equivalent to the Dantec Keypoint and the Cadwell Sierra.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the bird figure.
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Laborie Medical Technologies, Inc. % Ms. Barbara Mornet Regulatory Affairs Deputy 400 Avenue D, Suite 10 Williston, VT 05495-7828
MAR - 6 2009
Re: K082614
Trade/Device Name: Laborie EVOX Electro Diagnostic Device Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked response electrical stimulator Regulatory Class: II Product Code: GWF, JXE Dated: January 23, 2009 Received: January 26, 2009
Dear Ms. Mornet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I va may, annols provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device to such additional controls. Existing major regulations affecting your device can may or cally of the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r tease oc acrised that I be roundevice complies with other requirements of the Act that I Drives Intacted and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Ms. Barbara Mornet
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark McMillan
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K082614
Device Name: Laborie EVOX Electro Diagnostic Device
Indications For Use:
The EVOX Electro Diagnostic Device is intended for the testing of sacral reflexes, pudendal nerve terminal motor latency (PNTML) studies, electromyography (EMG), and cerebral pudendal somatosensory evoked potential (SEP).
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K08 2614 |
|---|---|
| --------------- | ---------- |
Page 1 of ____
§ 882.1870 Evoked response electrical stimulator.
(a)
Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.(b)
Classification. Class II (performance standards).