K Number
K082508
Date Cleared
2008-11-21

(84 days)

Product Code
Regulation Number
862.3650
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices are rapid immunochromatographic immunoassays for the qualitative detection of Buprenorphine in human urine at a cut-off concentration of 10 ng/ml. The devices are intended for professional use only. They are not intended for over-the-counter sale to non-professionals.

The Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard tests are used to obtain visual, qualitative results. A more specific alternate method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS are the preferred confirmatory methods.

Device Description

The Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard are based on the principle of highly specific immunochemical reactions between antigens and antibodies. Both devices utilize a competitive immunoassay procedure in which an immobilized drug conjugate competes with the drug present in urine for limited antibody binding sites. If Buprenorphine is present in the urine, it competes with the immobilized drug conjugate for the limited binding sites on the colored antibody colloidal gold conjugate. When a sufficient amount of drug is present, the drug will saturate the antibodies, and the colored colloidal gold conjugate cannot bind to the drug conjugate immobilized on the membrane. Thus, the absence of the purple-red band at the test region indicates a preliminary positive result. However, if there is no drug present to compete for the binding sites of the colored colloidal gold conjugate, it binds to the immobilized drug conjugate to form a visible purple-red band at the test region of the membrane. Thus, the presence of a purple-red band at the test region indicates a negative result. The Fastect II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices are standardized to detect Buprenorphine in human urine at a cut-off concentration of 10 ng/ml. These tests can be performed without the use of any additional instruments.

A control band with a different antigen/antibody reaction is added to the immunochromatographic membrane strip and should always appear regardless of the presence of drug or metabolite. The appearance of the control band during testing indicates that the test has completed and the test is valid.

AI/ML Overview

The provided text describes the Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices, which are immunochromatographic tests for the qualitative detection of Buprenorphine in human urine. The acceptance criteria are based on their accuracy in detecting Buprenorphine at a cut-off concentration of 10 ng/ml compared to GC/MS analysis and a predicate device.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are implied by the desired agreement with GC/MS and the predicate device. The performance is reported as percent agreement.

Fastect® II BUP Drug Screen Dipstick Test

CategoryAcceptance Criteria (Implied)Reported Device Performance (% Agreement with GC/MS)
Positive (+)High Agreement100%
Negative (-)High Agreement94%

QuickTox® Drug Screen Dipcard Test

CategoryAcceptance Criteria (Implied)Reported Device Performance (% Agreement with GC/MS)
Positive (+)High Agreement100%
Negative (-)High Agreement94%

Note: The acceptance criteria are not explicitly stated as numerical targets (e.g., "95% agreement"). Instead, the study reports the observed agreement to demonstrate substantial equivalence.

2. Sample size used for the test set and the data provenance

  • Sample size: A total of 90 urine specimens were evaluated.
  • Data provenance: 50 urine samples were clinical specimens, meaning they were obtained from patients. The remaining samples (40) were likely laboratory-prepared or spiked samples to ensure representation across different concentration ranges. These were previously analyzed by GC/MS.
    • Country of origin: Not specified.
    • Retrospective or prospective: The description states "clinical specimens previously analyzed by GC/MS," which indicates a retrospective collection and analysis for the clinical samples. The preparation of "near-negative" and "near-positive" samples suggests some prospective manipulation of existing samples or creation of new ones.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not provided in the text. The ground truth was established by GC/MS analysis, which is an analytical laboratory method, not human expert consensus.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • None in the context of human adjudication. The ground truth was established by GC/MS analysis, an objective laboratory method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC study was not done. This device is a rapid immunochromatographic test, and its interpretation is typically straightforward (presence or absence of a band). There is no mention of human readers assisting an AI, or vice versa. The comparison was between the device's results, a predicate device, and GC/MS.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, a standalone study was done. The "accuracy" study directly compared the Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices' results against GC/MS analysis, without human interpretation influencing the device's output. The devices themselves produce a visual result (presence/absence of a band) which is then read. While a human reads the band, the "performance" described refers to the device's ability to produce the correct band pattern.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The ground truth was established using GC/MS (Gas Chromatography/Mass Spectrometry), which is a gold standard analytical method for confirming the presence and concentration of substances like Buprenorphine in urine.

8. The sample size for the training set

  • The text does not explicitly mention a training set sample size. This document describes performance validation for a medical device (screening test), not the development and training of a machine learning model. The devices are based on immunochemical reactions, not algorithms that require training data in the typical sense.

9. How the ground truth for the training set was established

  • As the device is based on immunochemical principles rather than a machine learning algorithm, the concept of a "training set" and its "ground truth establishment" in the AI/ML context is not applicable. The underlying chemical reactions are inherently designed to detect the target analyte. The performance studies mentioned are for validation and verification, not for training a model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for BRANAN BMC MEDICAL CORP. The logo features the letters "BMC" in large, bold font, with the word "BRANAN" arched above it and "MEDICAL CORP" written below. The logo has a geometric shape surrounding the letters, giving it a structured and corporate appearance. The text and shapes are all in black and white.

510(k) SUMMARY

NOV 21 2008

This summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

The Assigned 510(k) number is: K082508

Submitter

Branan Medical Corporation 140 Technology Dr., Suite 400 Irvine, CA 92618

Tel: (949) 598-7166 Fax: (949) 598-7167

Date Prepared

August 25, 2008

Contact Person

Raphacl Wong President

Product Names

Fastect® II BUP Drug Screen Dipstick QuickTox® Drug Screen Dipcard

Common Name

Immunochromatographic test for the qualitative detection of Buprenorphine in urine.

Classification Panel

Buprenorphine Test Systems

Product Code

DJG

Classification Number

21 CFR, 862.6350

{1}------------------------------------------------

Device Classification

The Buprenorphine test systems are classified as Class II devices with moderate complexity. These tests are used to provide only a preliminary result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS or LC/MS confirmatory analysis.

Intended Use

The Fastect® II BUP Drug Screen Dipstick and the QuickTox® Drug Screen Dipcard devices are chromatographic immunoassays for the qualitative detection of Buprenorphine in human urine at a cut-off concentration of 10 ng/ml. These tests are screening devices and provide only a preliminary result. All positive test results obtained should be confirmed by another test method, preferably GC/MS or LC/MS analysis. The Fastect® II BUP Drug Screen Dipstick and the QuickTox® Drug Screen Dipcard deviccs are intended for professional use only. They are not intended for over-the-counter sale to non-professionals.

Description

The Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard are based on the principle of highly specific immunochemical reactions between antigens and antibodies. Both devices utilize a competitive immunoassay procedure in which an immobilized drug conjugate competes with the drug present in urine for limited antibody binding sites. If Buprenorphine is present in the urine, it competes with the immobilized drug conjugate for the limited binding sites on the colored antibody colloidal gold conjugate. When a sufficient amount of drug is present, the drug will saturate the antibodies, and the colored colloidal gold conjugate cannot bind to the drug conjugate immobilized on the membrane. Thus, the absence of the purple-red band at the test region indicates a prcliminary positive result. However, if there is no drug present to compete for the binding sites of the colored colloidal gold conjugate, it binds to the immobilized drug conjugate to form a visible purple-red band at the test region of the membrane. Thus, the presence of a purple-red band at the test region indicates a negative result. The Fastect II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices are standardized to detect Buprenorphine in human urine at a cut-off concentration of 10 ng/ml. These tests can be performed without the use of any additional instruments.

A control band with a different antigen/antibody reaction is added to the immunochromatographic membrane strip and should always appear regardless of the presence of drug or metabolite. The appearance of the control band during testing indicates that the test has completed and the test is valid.

{2}------------------------------------------------

Safety and Effectiveness Data

Accuracy

A total of 90 urine specimens were evaluated with the Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices. Of the 90 urine specimens, 50 urine samples were clinical specimens previously analyzed by GC/MS with known Buprenorphine concentrations. Approximately 10% were near-negative samples and at least 10% of the clinical urine samples were near-positive. When near-negative and near-positive clinical urine samples could not be easily obtained, more concentrated clinical urine samples were diluted to make these samples. Both the Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices were compared with the ACON® BUP One Step Buprenorphine Test Device as well as against the values obtained through GC/MS analysis. The following shows the results obtained from this study:

Fastcct® II BUP Drug Screen Dipstick Test

Negative(<50%)GC/MSNear Negative(50% to C/O)GC/MSNear Positive(C/O to +50%)GC/MSHigh Positive(>+50%)%Agreement
Positive (+)03633100%
Negative (-)4350094%

QuickTox® Drug Screen Dipcard Test

Negative(<50%)GC/MSNear Negative(50% to C/O)GC/MSNear Positive(C/O to +50%)GC/MSHigh Positive(>+50%)% Agreement
Positive (+)03633100%
Negative (-)4350094%

Performance Characteristics and Other Information

The performance characteristics of the Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices were verified by sensitivity, precision, specificity, cross-reactivity, interference (pH and Specific Gravity), stability, and optimal read time studies. Results from these studies show that the above-mentioned devices are reliable and meet performance claims when used according to the instructions provided in the Package Insert.

Conclusion

The performance characteristics studies performed demonstrate substantial equivalency between the Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices and ACON® BUP One Step Buprenorphine Test Device with the same cut-off concentration of 10 ng/ml. We have demonstrated that these tests are safe and effective in qualitatively detecting Burrenorphine at a cut-off concentration of 10 ng/ml. Therefore these tests are suitable for professional use only.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Branan Medical Corporation c/o Mr. Raphacl Wong President 140 Technology Drive, Suite 400 Irvine, CA 92618

NOV 21 2008

Re: K082508

Trade Name: Fastect® II BUP Drug Screen Dipstick QuickTox® Drug Screen Dipcard Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Codes: DJG Dated: August 25, 2008 Received: September 02, 2008

Dear Mr. Wong:

We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Jean M. Cooper, M.S., D.V.M.

Jéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

K082508

Fastect® II BUP Drug Screen Dipstick Device Name: QuickTox® Drug Screen Dipcard

Indications for Use:

The Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard devices are rapid immunochromatographic immunoassays for the qualitative detection of Buprenorphine in human urine at a cut-off concentration of 10 ng/ml. The devices are intended for professional use only. They are not intended for over-the-counter sale to non-professionals.

The Fastect® II BUP Drug Screen Dipstick and QuickTox® Drug Screen Dipcard tests are used to obtain visual, qualitative results. A more specific alternate method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS are the preferred confirmatory methods.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In,Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of 1

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).