(71 days)
The Master3DS is a X-ray, tomography, computed, dental that provides 3D imaging of dental and maxillofacial for diagnosis and treatment planning.
E-WOO Dental Imaging system Master3DS is a Computed Tomography X-ray System. Computed Tomography (CT) provides valuable 3-D imaging of the dental and maxillofacial structures for diagnosis and treatment planning. Uses of CT imaging include for assessment of impacted teeth, root configurations and mandibular condyle evaluation. This technology allow for 3-D imaging but at lower equipment cost, simpler image acquisition and lower patient radiation dose. Model Master3DS is diagnostic equipment which consists of panoramic dental x-ray system and computed tomography x-ray system. The panoramic dental x-ray is a system based on digital and computed tomography (CMOS CT Sensor to Capture 3D x-ray Computerized tomogram scanned image)
The provided 510(k) summary for the E-WOO Dental Imaging system Master3DS does not contain details about specific acceptance criteria, a study proving device performance against those criteria, or most of the requested information about a clinical study.
The submission focuses on establishing substantial equivalence to a predicate device (i-CAT) based on general device features and compliance with electrical, mechanical, and safety standards. It does not provide performance metrics or details of a clinical study designed to demonstrate a given level of diagnostic accuracy or effectiveness.
Therefore, many of the requested fields cannot be populated from the provided text.
Here's a breakdown of what can and cannot be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not specified in the provided text. The 510(k) focuses on substantial equivalence based on safety and general features, not on specific performance metrics or diagnostic accuracy targets. | The device provides "high quality digital image," "clear Tomography image upto minimum 0.1mm at any directions," and "drastically reduce X-Ray expose dose comparing with medical CT." However, these are descriptive claims rather than quantifiable performance metrics linked to acceptance criteria. |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified.
- Data Provenance: Not specified. The submission does not describe a clinical study with a test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not specified, as no clinical study with a ground truth assessment is described.
4. Adjudication method for the test set
- Not applicable, as no clinical study with a test set requiring adjudication is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not described. The device is a diagnostic imaging system, not an AI-assisted diagnostic tool for human readers in the context of this 510(k).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. The device is an imaging system; its performance is inherent in the image quality it produces, not in an algorithm's standalone diagnostic capability in the absence of a human interpreter.
7. The type of ground truth used
- Not specified, as no clinical study requiring a ground truth is described.
8. The sample size for the training set
- Not applicable. This device is an X-ray imaging system, not an AI/machine learning algorithm requiring a "training set" in the conventional sense for diagnostic performance.
9. How the ground truth for the training set was established
- Not applicable (see point 8).
Summary of the Study Mentioned (Non-Clinical):
The provided text mentions non-clinical testing to ensure the safety and performance of the Master3DS. This testing included:
- Electrical, mechanical, and environmental safety and performance testing: According to standards EN/IEC 60601-1, EN/IEC 60601-2-7, EN/IEC 60601-2-28, EN/IEC 60601-2-32, and EN/IEC 60601-2-44.
- EMC testing: Conducted in accordance with standard EN/IEC 60601-1-2 (2001).
The conclusion was that "All test results were satisfactory," indicating that the device met the requirements of these engineering and safety standards. This type of testing ensures the device is built safely and functions as intended from an engineering perspective, but it does not evaluate its diagnostic accuracy or clinical effectiveness in a patient population.
{0}------------------------------------------------
510(k) Summary
OCT 31 2008
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Jul. 17, 2008
-
- Company and Correspondent making the submission:
Name -- E-WOO Technology Co., Ltd.
- Company and Correspondent making the submission:
Address - 1F, 4F, Yunmin Technotown, 473-4, Bora-dong, Giheung-gu, Yongin-si,
Gyeonggi-do, 449-904, Korea
Telephone - +82-31-285-6950
Fax - +82-31-286-3007
Contact - Mr. DongTaek, Oh
Internet - http://www.e-wootech.com
-
- Device :
| Trade/proprietary name | : Master3DS |
|---|---|
| Common Name | : X-ray, tomography, computed, dental |
| Classification Name | : X-ray, tomography, computed, dental |
-
- Predicate Device :
| Manufacturer | : Imaging Sciences International, Inc. |
|---|---|
| Device | : i-CAT |
| 510(k) Number | : K061284(Decision Date - Jul. 3. 2006 |
-
- Classifications Names & Citations :
21CFR 892.1750, OAS, X-ray, tomography, computed, dental, Class2
- Classifications Names & Citations :
- Description :
5.1 General
E-WOO Dental Imaging system Master3DS is a Computed Tomography X-ray
E-WOO Technology Co., Ltd.
{1}------------------------------------------------
System. Computed Tomography (CT) provides valuable 3-D imaging of the dental and maxillofacial structures for diagnosis and treatment planning. Uses of CT imaging include for assessment of impacted teeth, root configurations and mandibular condyle evaluation. This technology allow for 3-D imaging but at lower equipment cost, simpler image acquisition and lower patient radiation dose.
Model Master3DS is diagnostic equipment which consists of panoramic dental x-ray system and computed tomography x-ray system. The panoramic dental x-ray is a system based on digital and computed tomography (CMOS CT Sensor to Capture 3D x-ray Computerized tomogram scanned image)
5.2 Product features
- · Master3DS, the dental CT system with digital panoramic unit, provides the high quality digital image.
- · It is not necessary to attach / detach the panoramic sensor when you capture CT images by Auto-Switching system.
- · It is possible to capture Maxillary, mandible and whole arch by wider FOV (12 * 7).
- · With Picasso-Trio it is not required to purchase a CT and a panoramic system each separately.
- · It is helpful for treatments by viewing the invisible part with a 3-D CT images.
- · The disk space for installation is no bigger than that of a general panoramic system.
- · A clear Tomography image upto minimum 0.1mm at any directions
- · Drastically reduce X-Ray expose dose comparing with medical CT in consideration of patient's safety
- · You can set and control the Examination Program Mode with a console PC
- · Support more accurate diagnosis imaging than LCD as well as voice announcement function for patients and staffs
- · Master3DS supports the DICOM Format
- ഗ Indication for use :
The Master3DS is a X-ray, tomography, computed, dental that provides 3D imaging of dental and maxillofacial for diagnosis and treatment planning.
- Comparison with predicate device : ଦ
Image /page/1/Picture/19 description: The image is a black and white abstract design. The image is mostly white, with black shapes scattered throughout. The black shapes are irregular and vary in size.
{2}------------------------------------------------
E-WOO Technology Co., Ltd., believes that the Master3DS is substantially equivalent to the i-CAT of Imaging Sciences International, Inc.,
- 7 Safety, EMC and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1, EN/IEC 60601-2-7, EN/IEC 60601-2-28, EN/IEC 60601-2-32 and EN/IEC 60601-2-44 was performed, and EMC testing was conducted in accordance with standard EN/IEC 60601-1-2(2001). All test results were satisfactory.
9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification E-WOO Technology Co., Ltd. concludes that The Master3DS is safe and effective and substantially equivalent to predicate devices as described herein.
-
- E-WOO Technology Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA,
END
- E-WOO Technology Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA,
E-WOO Technology Co., Ltd.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 31 2008
E-WOO Technology Co., Ltd. % Mr. Vincent Lee President E-WOO Technology USA, Inc. 256 North Sam Houston Pkwy E. #115 HOUSTON TX 77060
Rc: K082410
Trade/Device Name: X-ray, tomography, computed, dental/Master3DS Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: August 17, 2008 Received: August 21, 2008
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
lope M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number(if known):
KO82410
Device Name: X-ray, tomography, computed, dental /Master3DS
Indications for Use:
The Master3DS is a X-ray, tomography, computed, dental that provides 3D imaging of dental and maxillofacial for diagnosis and treatment planning.
Over-The-Counter Use_ Prescription Use AND/OR (Part 21 CFR 801 Subpart D) (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Helus Remen
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices 82410 ් (1)(k) Number
Page 1 of _1
Image /page/5/Picture/12 description: The image shows the number 28 in a simple, bold font. The numbers are slightly separated and have a classic, almost typewriter-like appearance. The background is plain and light, which makes the numbers stand out clearly.
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.