K Number
K082377
Manufacturer
Date Cleared
2009-02-09

(175 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TENS stands for Transcutaneous Electrical Nerve Stimulation. The itouch TENS system is indicated for:

  • Symptomatic relief and management of chronic, intractable pain. .
  • Adjunctive treatment for post-surgical and post-trauma acute pain. .
Device Description

A portable TENS device for pain relief.
Features:

  • Innovative design
  • Large LCD display
  • Dual output isolated channels
  • 2 AA Alkaline Batteries
  • Adjustable frequency, pulse width, and timing parameters
  • 9 different modes
  • Timer option
  • Doctor lock/unlock facility
  • Open circuit detectors
  • Non-volatile
AI/ML Overview

The provided 510(k) summary for the Tenscare TENS Model itouch (K082377) does not contain a study that proves the device meets specific acceptance criteria in the way typically found for novel diagnostic algorithms or devices.

Instead, this submission primarily focuses on demonstrating substantial equivalence to predicate devices (EasyMed TENS Unit TN-28C and Gemore GM300T) via technological comparison, labeling comparison, and compliance with recognized safety standards. For TENS devices, the primary "acceptance criteria" are related to safety, electrical output specifications, and efficacy established by existing, legally marketed devices.

Here's an breakdown based on the provided text, addressing your points where information exists:

1. Table of Acceptance Criteria and Reported Device Performance

As per the document, specific, quantifiable acceptance criteria (e.g., target accuracy, sensitivity, specificity) for a clinical study proving device efficacy are not presented for the Tenscare itouch. Instead, the "performance" is implicitly tied to compliance with standards and equivalence to predicate devices which are already accepted as safe and effective.

The closest to "acceptance criteria" are:

Acceptance Criterion (Implicit)Reported Device Performance/Comparison
Safety: Compliance with relevant recognized consensus standards."Design to comply with relevant safety applicable recognized consensus standards;""the output energy is well controlled in the safety and effectiveness ranges specified by relevant FDA guidance's." "Test results, Risk Analysis, and FMEA analysis show that the new unit itouch is safe with no any hazard."
Performance: Output characteristics (e.g., constant current)."Itouch was designed and added the constant current feature based on the design of TN-28C (K040253). The introduction of marketed device Gemore GM300T (K032720) is for the purpose of comparison of the feature of constant current.""The new device itouch is of the same range of parameters as those of marketed devices."
Post-Market Performance (Europe): Functioning as intended."itouch has been marketed in Europe. During this period a review of Customer Complaints, Returned Product and the results of Post Market Feedback, has demonstrated that the product has performed as Intended, to its' Specified Requirements."

2. Sample size used for the test set and the data provenance

  • No specific clinical test set is described. The submission relies on technological comparison and compliance with standards rather than a new clinical performance study with a dedicated test set of patient data.
  • The reference to "itouch has been marketed in Europe" and "review of Customer Complaints, Returned Product and the results of Post Market Feedback" suggests retrospective data analysis from real-world usage in Europe. However, no sample sizes for this feedback are provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This is not applicable as there is no described clinical study with a "test set" and associated "ground truth" derived by experts for performance evaluation.

4. Adjudication method for the test set

  • This is not applicable as there is no described clinical study with a "test set" requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not applicable. The device is a Transcutaneous Electrical Nerve Stimulator (TENS), not an AI-powered diagnostic tool, and therefore, an MRMC study involving human readers and AI assistance is outside its scope.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This is not applicable. The Tenscare itouch is a medical device for pain relief, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the European post-market review, the "ground truth" was implicitly the real-world performance as intended by the device's design and specified requirements, based on customer complaints, returned products, and feedback. This can be considered a form of "outcomes data" related to device function and patient satisfaction, rather than a diagnostic ground truth.
  • For the submission to the FDA, the "ground truth" for demonstrating equivalence is the established safety and efficacy profile of the predicate devices and compliance with recognized electrical safety and performance standards.

8. The sample size for the training set

  • This is not applicable. The device is not an AI algorithm that requires a "training set." Its design and development are based on established engineering principles for TENS devices and comparison to predicate devices.

9. How the ground truth for the training set was established

  • This is not applicable as there is no "training set" for an AI algorithm.

In summary:

The K082377 submission for the Tenscare TENS Model itouch demonstrates substantial equivalence to predicate devices through technical comparisons, adherence to safety standards, and general claims of satisfactory performance in the European market. It does not present a clinical study with specific acceptance criteria, test sets, or ground truth establishment in the manner typically seen for novel diagnostic or AI-driven medical devices. The primary "proof" relies on the device being technologically similar to and meeting the same safety standards as already cleared TENS devices.

{0}------------------------------------------------

K0r2377

FEB - 9 2009

510(k) Summary

-- 4 --

Date of Summary prepared: January 21, 2005

Submitter: Tenscare Ltd

9 Blenheim Road, Longmead Business Park, Epsom, Surrey, KT19 9BE, Address : United Kingdom

Tel +44(0)1372 723 434 .

Fax +44(0)1372 745 434

E-Mail sales@tenscare.co.uk :

Contact person: Anthony Jankel

Address of the manufacturing facility:

EasyMed Instrument Ltd 3/F, No.2 Bei Hai Da Road, Lun Jiao, Shun De, Guang Dong, China. Zip: 528308

Address of American Representative: DEBORAH IAMPIETRO QRC CONSULTING ASSOCIATES 130 gotzens rd

conway, NH 03818 Phone: 603 4476086 ext Fax: 734 4236086

Email: grcassoc@aol.com

{1}------------------------------------------------

Submitted Device:

Generic name: Transcutaneous Electrical Nerve Stimulator (TENS) Tenscare itouch Trade name: Common name: TENS device Classification name: Stimulator, Nerve, Transcutaneous, for Pain Relief - GZJ; 21 CFR 882.5890. Device Classification: Class II Predicate Devices: EasyMed TENS Unit TN-28C (K040253) Gemore GM300T (K032720) (http://www.gemore.com.tw/) The class of the predicate Devices: Class II Device Description: A portable TENS device for pain relief.

Features:

  • · Innovative design
  • · Large LCD display
  • · Dual output isolated channels
  • · 2 AA Alkaline Batteries
  • · Adjustable frequency, pulse width, and timing parameters
  • 9 different modes
  • · Timer option
  • · Doctor lock/unlock facility
  • · Open circuit detectors
  • · Non-volatile

The intended use of the device:

TENS stands for Transcutaneous Electrical Nerve Stimulation. The itouch TENS system is indicated for:

  • Symptomatic relief and management of chronic, intractable pain. .
  • Adjunctive treatment for post-surgical and post-trauma acute pain. .

The intended use and indications for use of the new device are very similar to that of the predicate devices.

{2}------------------------------------------------

Technological Comparison:

    1. The marketed device EasyMed TN-28C (K040253) is a digital TENS unit which has been well exploited the digital technology, supplying the user with full ranges of parameters, traditional TENS functions, and non-traditional new functions to be selected.
    1. The designed circuitry of new device itouch is very similar to the marketed device TN-28C, they are of similar circuit diagrams and similar working principle; The main difference between the two devices is that TN-28C is not a constant current TENS device while itouch is a constant current device.
    1. Both itouch and TN-28C (K040253) are passed the same tests of applicable recognized international consensus standards.
    1. The marketed device Gemore GM300T (K032720) is a TENS device with constant current output. The specification of GM300T is very similar with that of itouch.
    1. The new device itouch is of the same range of parameters as those of marketed devices.
    1. Itouch was designed and added the constant current feature based on the design of TN-28C (K040253). The introduction of marketed device Gemore GM300T (K032720) is for the purpose of comparison of the feature of constant current.
    1. The accessories of the new device itouch are similar to those of the marketed devices.

Labelling Comparison: The Labelling is substantially equivalent to that of the predicate devices.

Safety information:

Design to comply with relevant safety applicable recognized consensus standards; the output energy is well controlled in the safety and effectiveness ranges specified by relevant FDA guidance's. Testing has been carried out in very detailed and strictly. Test results, Risk Analysis, and FMEA analysis show that the new unit itouch is safe with no any hazard.

itouch has been marketed in Europe. During this period a review of Customer Complaints, Returned Product and the results of Post Market Feedback, has demonstrated that the product has performed

-- ( --

{3}------------------------------------------------

as Intended, to its' Specified Requirements.

It is the first submission to FDA for this new device

The new device itouch is of the feature of constant current. But this feature is only applied in the range of load from 500Ω to 1500Ω, the current is being get down and tends to zero when the load is being get far from this range, this character makes the device safe reliably.

Submitted times:

Conclusions:

Important remark:

The itouch TENS unit is substantially equivalent to the predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.

There are two models for Tenscare itouch TENS devices: "Easy" and "Plus". Because all functions of "Easy" can be covered by "Plus" (see instruction manual in Section V), we do not discuss "Easy" model in this submission and consider it as no significant change.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its back, representing the department's services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Tenscare Ltd % Mr. Anthony Jankel 9 Blenheim Road Longmead Business Park, Epsom Surrey, KT19 9BE United Kingdom

Re: K082377

Trade Name: Tenscare TENS Model itouch Regulation Number: 21 CFR 882.5890 Regulation Names: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ Dated: December 12, 2008 Received: December 12, 2008

FEB - 9 2009

Dear Mr. Jankel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Anthony Jankel

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark N Millerson

Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known):

Device Name

Tenscare TENS Model itouch

Indications for Use

TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is indicated for:

  • Symptomatic relief and management of chronic, intractable pain. .
  • Adjunctive treatment for post-surgical and post-trauma acute pain.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division of General, Restorative, and Neurological Devices

-- 9 --

510(k) Number

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).