(109 days)
Not Found
No
The device description relies on bio-electrical impedance analysis and proprietary regression equations, not AI/ML algorithms, for calculating body composition.
No.
The device is used for non-invasive bio-impedance analysis to determine body composition status and does not claim to treat or diagnose any disease or condition.
No
The device is marketed for "non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status" and calculates estimated values like fat percentage and body mass index. This is for general health and wellness monitoring rather than for diagnosing a disease or condition.
No
The device description clearly describes a hardware unit ("Bodystat 1500 Body Composition Monitoring Unit") that applies an electrical signal and measures impedance. While software programs are mentioned as optional accessories, the core device performing the measurement is hardware-based.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: This device performs a non-invasive measurement of bio-impedance on the intact human body. It does not analyze samples taken from the body.
- Intended Use: The intended use is to determine body composition status in healthy adults, not to diagnose a disease or condition based on analysis of bodily fluids or tissues.
The device uses bio-electrical impedance analysis (BIA) to measure the body's resistance to an electrical current, which is a physical measurement taken directly from the subject, not an analysis of a biological sample.
N/A
Intended Use / Indications for Use
For the purposes of performing a non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status.
This device measures bio-electrical impedance when a 50 kHz signal is applied at specific locations on the subject. From this and the subject's age, gender, weight, height, waist and hip circumferences (optional), the device calculates the following estimated values; fat%, fat weight, lean weight, total body water%, total body water, dry lean weight, basal metabolic rate, basal metabolic rate/body weight ratio, average requirement (EAR), and actual body mass index and waist/hip ratio.
OPTIONAL ACCESSORIES: Bodystat (1) BODY MANAGER and (2) WELLNESS SOFTWARE PROGRAMS.
The Body Manager Program is ideal when a client returns to be tested in order to track their changes in body composition over a period of time.
The Wellness Program is ideal for first time assessments on subjects. The program's graphical presentations provide information on each of the specific measurements in an educational format.
Product codes
MNW
Device Description
The Bodystat 1500 Body Composition Monitoring Unit is an impedance plethysmograph that uses bio-electrical impedance analysis (BIA) to measure the impedance of the flow of an electrical current through the body. The impedance of tissue is proportional to the amount of fluid in the tissue; water is low in fat tissues, thus fat tissue has a high impedance, and high in lean tissues, thus lean tissue has a high impedance.
In practice, a small constant current is passed between electrodes spanning the body and the voltage drop between electrodes provides a measure of impedance. Prediction equations, previously generated by correlating impedance measures against an independent estimate of TBW (total body water), may be used subsequently to convert a measured impedance to a corresponding estimate of TB W. Lean body mass is then calculated from this estimate using an assumed hydration fraction for lean tissue; Bodystat uses a proprietary regression equation for this calculation. Fat mass is calculated as the difference between body weight and lean body mass.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
normal healthy human adults
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2770 Impedance plethysmograph.
(a)
Identification. An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.(b)
Classification. Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
0
PAGE 1 OF 2
NOV 2 5 2008
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
510(k) Number : K082253
Submitter: A.
Bodystat Ltd P O Box 50 Douglas Isle of Man IM99 1DQ BRITISH ISLES
Phone: +44-1624-629571 444-1624-611544 Fax:
Contact: IJ Meeuwsen
Date Prepared: October 318 , 2008
B. Device Names:
Classification name Impedance Plethysmograph Common/usual name Body Composition Monitor Proprietary name Bodystat 1500 Body Composition Monitoring Unit
C. Predicate Device:
The modified BODYSTAT®1500 Body Composition Monitoring Unit is substantially equivalent to previously cleared device, K951097, Bodystat 1500 Body Composition Monitoring Unit.
The BODYSTAT® 1500 Body Composition Monitoring Unit device is a hardware and software modification of the previously cleared device.
D. Device Description:
The Bodystat 1500 Body Composition Monitoring Unit is an impedance plethysmograph that uses bio-electrical impedance analysis (BIA) to measure the impedance of the flow of an electrical current through the body. The impedance of tissue is proportional to the amount of fluid in the tissue; water is low in fat tissues, thus fat tissue has a high impedance, and high in lean tissues, thus lean tissue has a high impedance.
In practice, a small constant current is passed between electrodes spanning the body and the voltage drop between electrodes provides a measure of impedance. Prediction
1
Image /page/1/Picture/0 description: The image shows the text "K082253 PAGE 1 OF 2" in a handwritten style. The text is arranged in two lines, with the first line containing the alphanumeric code "K082253". The second line indicates that it is page 1 of a total of 2 pages. The handwriting appears to be clear and legible.
equations, previously generated by correlating impedance measures against an independent estimate of TBW (total body water), may be used subsequently to convert a measured impedance to a corresponding estimate of TB W. Lean body mass is then calculated from this estimate using an assumed hydration fraction for lean tissue; Bodystat uses a proprietary regression equation for this calculation. Fat mass is calculated as the difference between body weight and lean body mass.
E. Intended Use:
For the purposes of performing a non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status.
This device measures bio-electrical impedance when a 50 kHz signal is applied at specific locations on the subject. From this and the subject's age, gender, weight, height, waist and hip circumferences (optional), the device calculates the following estimated values; fat%, fat weight, lean weight, total body water%, total body water, dry lean weight, basal metabolic rate, basal metabolic rate/body weight ratio, average requirement (EAR), and actual body mass index and waist/hip ratio.
OPTIONAL ACCESSORIES: Bodystat (1) BODY MANAGER and (2) WELLNESS SOFTWARE PROGRAMS.
The Body Manager Program is ideal when a client returns to be tested in order to track their changes in body composition over a period of time.
The Wellness Program is ideal for first time assessments on subjects. The program's graphical presentations provide information on each of the specific measurements in an educational format.
F. Comparison with the Predicate Device:
The modified BODYSTAT®1500 Body Composition Monitoring Unit is a hardware and software modification of the previously cleared Bodystat 1500. The modified Bodystat 1500 and the previously cleared Bodystat 1500 have the same intended use and use the same operating principle.
Based on the data and information presented here, the modified BODYSTAT® 1500 Body Composition Monitoring Unit is substantially equivalent to the previously cleared Bodystat 1500 device.
2
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves, arranged in a cascading pattern.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 25 2008
Mr. I. J. Meeuwsen President BODYSTAT® LIMITED P.O. Box 50 Douglas, Isle of Man IM99 1DQ BRITISH ISLES
K082253 Re:
Trade/Device Name: BODYSTAT® 1500 Body Composition Monitoring Unit Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: November 14, 2008 Received: November 19, 2008
Dear Mr. Meeuwsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
---|---|---|
21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
21 CFR 894.xxx | (Radiology) | 240-276-0120 |
Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Joppu M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
ATTACHMENT 4
Indications for Use
510(k) Number : K082253
Device Name: BODYSTAT®1500 Body Composition Monitoring Unit
Indications for Use:
For the purposes of performing a non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status.
This device measures bio-electrical impedance when a 50 kHz signal is applied at specific locations on the subject. From this and the subject's age, gender, weight, height, waist and hip circumferences (optional), the device calculates the following estimated values: fat%, fat weight, lean%, lean weight, total body water%, total body water, dry lean weight, basal metabolic rate, basal metabolic rate/body weight ratio, average requirement (EAR), and actual body mass index and waist/hip ratio.
OPTIONAL ACCESSORIES: Bodystat (1) BODY MANAGER and (2) WELLNESS SOFTWARE PROGRAMS.
The Body Manager Program is ideal when a client returns to be tested in order to track their changes in body composition over a period of time.
The Wellness Program is ideal for first time assessments on subjects. The program's graphical presentations provide information on each of the specific measurements in an educational format.
Prescription Use (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
Helent lemen
Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number