K Number
K082220
Device Name
SOMATOM FLASH DS
Date Cleared
2008-10-10

(65 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Siemens SOMATOM P47 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

In addition the SOMATOM P47 is able to produce additional image planes and analysis results by executing optional postprocessing features, which operate on DICOM images.

The images and results delivered by the system can be used by a trained physician as an aid in diagnosis.

(*spiral planes: the axial planes resulting from the continuous rotation of detectors and xray tube, and the simultaneous translation of the patient.)

Device Description

The Siemens SOMATOM P47 is a Computed Tomography X- ray System, which features two continuously rotating tube-detector systems and functions according to the fan beam principle. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.

The SOMATOM P47 system produces CT images in DICOM format, which can be used by postprocessing applications commercially distributed by Siemens and other vendors.

The computer system delivered with the CT scanner is able to run such post processing applications optionally.

AI/ML Overview

The provided text describes a 510(k) submission for the Siemens SOMATOM P47, a computed tomography X-ray system. However, it does not contain a study that proves the device meets specific acceptance criteria in the way a diagnostic AI/ML algorithm study would.

The document is a clearance letter from the FDA, stating that the device is substantially equivalent to predicate devices. This type of clearance relies on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than providing a detailed clinical study with acceptance criteria and performance metrics for a specific diagnostic task.

Therefore, many of the requested details, such as a table of acceptance criteria, sample sizes for test/training sets, ground truth establishment, MRMC studies, or standalone performance, are not present in this type of regulatory document.

However, based on the information provided, here's what can be extracted and inferred:

1. A table of acceptance criteria and the reported device performance

This information is not available in the provided document. The 510(k) submission process for a CT scanner typically relies on demonstrating substantial equivalence to a predicate device through technical performance specifications, safety testing (electrical, mechanical, radiation), and software validation, rather than clinical performance metrics against specific diagnostic acceptance criteria.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not available in the provided document. As this is a CT scanner hardware and basic software system, clinical "test sets" for diagnostic performance, as understood for AI/ML algorithms, are not typically part of the 510(k) submission as presented here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not available in the provided document. Ground truth establishment for diagnostic performance studies is not detailed as this document concerns a CT scanner's regulatory clearance, not a diagnostic AI's performance study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not available in the provided document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not available in the provided document. MRMC studies are relevant for evaluating the impact of AI on human reader performance, which is not the focus of this CT scanner's 510(k) submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not available in the provided document. The SOMATOM P47 is a CT imaging system, not a diagnostic algorithm meant to operate standalone from human interpretation. Its output (DICOM images) are "used by a trained physician as an aid in diagnosis."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not available in the provided document. For a CT scanner, "ground truth" would more likely refer to the accuracy of the physical measurements and image reconstruction parameters rather than clinical diagnostic ground truth.

8. The sample size for the training set

This information is not available in the provided document. The SOMATOM P47 is a CT scanner, not a machine learning model that undergoes a "training" phase with a dataset in the conventional sense. Its software (SOMARIS/7) handles patient management, data management, X-ray scan control, image reconstruction, and archive/evaluation.

9. How the ground truth for the training set was established

This information is not available in the provided document.

Summary based on the provided text:

The document describes the Siemens SOMATOM P47, a Computed Tomography X-ray System, which features two continuously rotating tube-detector systems and functions according to the fan beam principle. Its software (SOMARIS/7) manages patient data, X-ray scan control, image reconstruction, and archiving.

The device's regulatory clearance (K082220) was based on Substantial Equivalence to legally marketed predicate devices:

  • Siemens SOMATOM Definition AS/AS+ (K081022 cleared 06/05/2008)
  • Siemens SOMATOM Definition (K052216 cleared 09/08/2005)

The Indications for Use for the SOMATOM P47 are to "produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles." It can also produce "additional image planes and analysis results by executing optional postprocessing features, which operate on DICOM images." The "images and results delivered by the system can be used by a trained physician as an aid in diagnosis."

Acceptance Criteria and Studies:
The document does not describe specific acceptance criteria and a study to prove the device meets these criteria in the context of diagnostic performance metrics (e.g., sensitivity, specificity, accuracy against a clinical ground truth). Instead, the focus of this 510(k) summary is on demonstrating safety and effectiveness via substantial equivalence to previously cleared CT scanners. This typically involves:

  • Technical Performance: Demonstrating that the device meets established performance standards for CT scanners (e.g., image quality metrics, dose reduction features, reconstruction speed, resolution). These would be derived from internal engineering specifications, but are not explicitly detailed as "acceptance criteria" with performance results in this public summary.
  • Safety: Adherence to recognized and established industry practice and standards to minimize electrical, mechanical, and radiation hazards, confirmed through hazard analysis, verification, and validation testing.
  • Software Validation: The software (SOMARIS/7) for patient/data management, scan control, image reconstruction, and archiving is validated to ensure it performs as intended and meets specifications.

The FDA's clearance states, "We have reviewed your Section 510(k) premarket notification of intent to market the device... and have determined the device is substantially equivalent... You may, therefore, market the device, subject to the general controls provisions of the Act." This indicates that the submitted information was sufficient to demonstrate equivalence, thereby implying adequate safety and effectiveness without requiring a new detailed clinical study to prove diagnostic performance against specific acceptance criteria.

{0}------------------------------------------------

K082220

OCT 1 0 2008

SIEMENS

510(K) SUMMARY FOR SOMATOM P47

Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355

September 4, 2007

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

1. Contact Person:

Mrs. Corrine McLeod Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA

51 Valley Stream Parkway E-50 Malvern, PA 19355-1406 Phone:(601) 448-1772 Fax: (610) 448-1778

2. Device Name and Classification

Product Name:SOMATOM P47
Propriety Trade Name:SOMATOM Flash DS
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750
Device Class:Class II
Product Code:90 JAK

{1}------------------------------------------------

SIEMENS

3. Substantial Equivalence:

Siemens SOMATOM P47 Computed Tomography X-ray system, configured with software SOMARIS/7 is substantially equivalent to the following medical device in commercial distribution:

Predicate Device NameFDA Clearance NumberFDA Clearance Date
Siemens SOMATOM Definition AS/AS+K08102206/05/2008
Siemens SOMATOM DefinitionK05221609/08/2005

4. Device Description:

The Siemens SOMATOM P47 is a Computed Tomography X- ray System, which features two continuously rotating tube-detector systems and functions according to the fan beam principle. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.

The SOMATOM P47 system produces CT images in DICOM format, which can be used by postprocessing applications commercially distributed by Siemens and other vendors.

The computer system delivered with the CT scanner is able to run such post processing applications optionally.

5. Indications for Use:

The Siemens SOMATOM P47 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

In addition the SOMATOM P47 is able to produce additional image planes and analysis results by executing optional postprocessing features, which operate on DICOM images.

The images and results delivered by the system can be used by a trained physician as an aid in diagnosis.

(*spiral planes: the axial planes resulting from the continuous rotation of detectors and xray tube, and the simultaneous translation of the patient.)

{2}------------------------------------------------

SIEMENS

6. General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)", which is arranged in a circular fashion around the logo.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 0 2008

Ms. Corrine McLeod Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway E50 MALVERN PA 19355-1406

Re: K082220

Trade/Device Name: Somatom P47 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: August 1, 2008 Received: August 11, 2008

Dear Ms. McLeod:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Arqui M. Zhang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 4: Indication for Use Statement

SIEMENS

510(k) Number (if known): NO (22220

Product Name:

Somatom P47

Indications for Use:

The Siemens SOMATOM P47 system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

In addition the SOMATOM P47 is able to produce additional image planes and analysis results by executing optional postprocessing features, which operate on DICOM images.

The images and results delivered by the system can be used by a trained physician as an aid in diagnosis.

(*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Arun M Wang

Division of Reproductive. and Radiological De

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.