K Number
K081920
Date Cleared
2008-12-19

(169 days)

Product Code
Regulation Number
866.2660
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD GeneOhm™ Cdiff Assay is a rapid in vitro diagnostic test for the direct, qualitative detection of C. difficile toxin B gene (todB) in human liquid or soft stool specimens from patients suspected of having Clostridium difficile-associated disease (CDAD). The test, based on real-time PCR, is intended for use as an aid in diagnosis of CDAD. The test is performed directly on the specimen, utilizing polymerase chain reaction (PCR) for the amplification of specific targets and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.

Device Description

The BD GeneOhm™ Cdiff Assay is a rapid in vitro diagnostic test for the direct, qualitative detection of C. difficile toxin B gene (todB) in human liquid or soft stool specimens from patients suspected of having Clostridium difficile-associated disease (CDAD). The test, based on real-time PCR, is intended for use as an aid in diagnosis of CDAD. The test is performed directly on the specimen, utilizing polymerase chain reaction (PCR) for the amplification of specific targets and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.

A liquid or soft stool specimen is collected and transported to the laboratory. A sterile dry swab is dipped into the liquid or soft stool material and processed. For testing, the swab is eluted in sample buffer and the specimen is lysed. An aliquot of the lysate is added to PCR readents which contain the tcdB specific primers used to amplify the genetic target of Clostridium difficile, if present. The assay also includes an internal control (IC) to detect PCR inhibited specimens and to confirm the integrity of assay reagents. Amplified targets are detected with hybridization probes labelled with quenched fluorophores (molecular beacons). The amplification, detection and interpretation of the signals are done automatically by the Cepheid SmartCycler® software. The entire procedure takes about 75 to 90 minutes, depending on the number of specimens processed.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the BD GeneOhm™ Cdiff Assay, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

The acceptance criteria are implied by the reported performance metrics. The device needed to demonstrate sufficient sensitivity, specificity, and reproducibility compared to a recognized reference method to be considered substantially equivalent.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Overall) - Fresh SamplesReported Device Performance (Overall) - Frozen Samples
SensitivityHigh (e.g., >90%)93.8% (95% CI: 86.2% - 98.0%)100.0% (95% CI: 89.7% - 100%)
SpecificityHigh (e.g., >95%)95.5% (95% CI: 93.8% - 96.9%)97.7% (95% CI: 94.8% - 99.3%)
Negative Predictive Value (NPV)High (e.g., >95%)99.1%99.2%
Positive Predictive Value (PPV)Reasonably High67.3%81.5%
Unresolved Rate (Overall)Low (e.g., <5%)Initial: 4.6% (39/852), After Repeat: 2.0% (17/852)Initial: 0.4% (1/256), After Repeat: 0.4% (1/256)
Invalid Run Rate (Overall)Low (e.g., <5%)0.6% (1/160)0.6% (1/160) - same overall rate as fresh samples
Analytical SpecificityNo cross-reactivityNo detectable interference with 99 organisms and 29 Clostridium strainsNot applicable (qualitative)
Analytical Sensitivity (LOD)Specific DNA copies/CFU10 DNA copies per reaction, 4 CFU per reactionNot applicable (quantitative)
Reproducibility (Low Positive Agreement)High (e.g., >95%)Site-to-Site: 96.7%, Lot-to-Lot: 100.0%Not explicitly broken down for frozen/fresh
Reproducibility (Moderate Positive Agreement)High (e.g., >95%)Site-to-Site: 100%, Lot-to-Lot: 97.8%Not explicitly broken down for frozen/fresh
Reproducibility (Negative Agreement)High (e.g., >95%)Site-to-Site: 100%, Lot-to-Lot: 100.0%Not explicitly broken down for frozen/fresh
Interfering SubstancesNo detectable interferenceMost tested substances showed no interference (NI = No Interference)Not applicable (qualitative)

Details of the Study

  1. Sample Size and Data Provenance:
    • Clinical Study (Test Set):
      • Total Specimens Tested: 1108
      • Reportable Results: 1090
      • Fresh Specimens (First Dataset): 835 specimens
      • Frozen Specimens (Second Dataset): 255 specimens (retested from aliquots of original stool specimens after an issue with the reference assay at one site)
      • Data Provenance: Multi-site prospective investigational study conducted at four medical centers (two in Canada and two in the United States).
  2. Number of Experts and Qualifications for Ground Truth (Clinical Study):
    • The document does not specify the number of experts used or their qualifications for establishing the ground truth directly. It refers to the "Reference Cytotoxicity Assay" as the comparator method.
  3. Adjudication Method (Clinical Study):
    • The document does not explicitly describe an adjudication method for discrepancies between the device and the reference assay in the clinical study. It notes that for 21 out of 34 samples that were positive by the reference assay and negative by the BD GeneOhm™ Cdiff Assay, further testing (Cytotoxicity Assay on isolated strains, standard PCR with alternative primers, and bi-directional sequencing) was performed to verify the presence of toxigenic C. difficile. This implies a form of ad-hoc adjudication for discordant results rather than a pre-defined process for all cases.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed. This device is an in vitro diagnostic (IVD) assay, not an AI-assisted diagnostic tool for human interpretation. The "interpretation of the signals are done automatically by the Cepheid SmartCycler® software."
  5. Standalone Performance:
    • Yes, the performance characteristics reported (sensitivity, specificity, NPV, PPV) are for the device (BD GeneOhm™ Cdiff Assay) in standalone mode, directly compared against the reference method. There is no human-in-the-loop performance described for the primary clinical effectiveness assessment.
  6. Type of Ground Truth Used (Clinical Study):
    • The primary ground truth for the clinical study was the Reference Cytotoxicity Assay on liquid or soft stool specimens. For discordant results, further molecular testing (standard PCR with alternative primers and bi-directional sequencing) was used to verify the presence of the tcdB gene or toxigenic C. difficile. This indicates a combination of phenotypic (cytotoxicity) and genotypic (molecular) methods to establish the most accurate ground truth.
  7. Sample Size for the Training Set:
    • The document describes performance characteristics of a completed assay rather than an algorithm trained on a dataset. Therefore, there is no explicit "training set" sample size mentioned in the context of machine learning model development. The analytical studies (specificity, sensitivity, reproducibility) served to characterize the assay's performance against known samples.
  8. How Ground Truth for the Training Set Was Established:
    • As there's no explicitly defined "training set" in the context of a machine learning algorithm, this question is not directly applicable. For the analytical studies:
      • Analytical Specificity: Strains were identified and confirmed as non-target organisms or C. difficile lacking the tcdB gene. The "neg" result indicates the absence of the target gene.
      • Analytical Sensitivity (LOD): Used a specific C. difficile strain (ATCC 43255) with known concentrations (DNA copies/CFU) and confirmed with other toxigenic strains. This involved careful preparation and quantification of known positive samples.
      • Reproducibility: Involved simulated specimens, with "low positive," "moderate positive," and "negative" categories established by inoculating known quantities of C. difficile (ATCC 43255) into simulated bowel flora.

{0}------------------------------------------------

Kori20

510(k) Summary

DEC 1 9 2008

December 17, 2008

BD Diagnostics BD GeneOhm™ Cdiff Assay

BD Diagnostics (GeneOhm Sciences Canada, Inc.) Submitted by: 2050, boul. René-Lévesque O, 4ª étage Sainte-Foy, Québec Canada G1V 2K8

Contact/U.S. Agent:Raymond Boulé
BD Diagnostics - GeneOhm
6146 Nancy Ridge Drive
San Diego, CA 92121 USA

Name of Device:

Trade Name:BD GeneOhm™ Cdiff Assay
Common Name:Clostridium difficile tcdB detection assay
Classification Name:System, Test, Genotypic Detection, Clostridium difficileToxin B
Predicate Device:Techlab C. difficile Tox-B Test (K935296)Techlab Clostridium difficile Toxin/Antitoxin Kit (K923463)

Device Description:

Intended Use:

The BD GeneOhm™ Cdiff Assay is a rapid in vitro diagnostic test for the direct, qualitative detection of C. difficile toxin B gene (todB) in human liquid or soft stool specimens from patients suspected of having Clostridium difficile-associated disease (CDAD). The test, based on real-time PCR, is intended for use as an aid in diagnosis of CDAD. The test is performed directly on the specimen, utilizing polymerase chain reaction (PCR) for the amplification of specific targets and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.

Test Description:

A liquid or soft stool specimen is collected and transported to the laboratory. A sterile dry swab is dipped into the liquid or soft stool material and processed. For testing, the swab is eluted in sample buffer and the specimen is lysed. An aliquot of the lysate is added to PCR readents which contain the tcdB specific primers used to amplify the genetic target of Clostridium difficile, if present. The assay also includes an internal control (IC) to detect PCR inhibited specimens and to confirm the integrity of assay reagents. Amplified targets are detected with hybridization probes labelled with quenched fluorophores (molecular beacons). The amplification, detection and interpretation of the signals are done automatically by the Cepheid SmartCycler® software. The entire procedure takes about 75 to 90 minutes, depending on the number of specimens processed.

{1}------------------------------------------------

The amplified DNA target is detected with a molecular beacon, a hairpin-forming singlestranded oligonucleotide labelled at one end with a quencher and at the other end with a fluorescent reporter dye (fluorophore). In the absence of target, the fluorescence is quenched. In the presence of target, the hairpin structure opens upon beacon/target hybridization, resulting in emission of fluorescence. For the detection of tcdB amplicons, the molecular beacon contains the fluorophore FAM at the 5' end and the nonfluorescent quencher DABCYL at the opposite 3' end of the oligonucleotide. For the detection of the IC amplicons, the molecular beacon contains the fluorophore TET at the 5' end and the quencher moiety DABCYL at the 3' end. Each beacon-target hybrid fluoresces at a wavelength characteristic of the fluorophore used in the particular molecular beacon. The amount of fluorescence at any given cycle, or following cycling, depends on the amount of specific amplicons present at that time. The SmartCycler software simultaneously monitors the fluorescence emitted by each molecular beacon, interprets all data, and provides a final result at the end of the cycling program.

Substantial Equivalence:

The BD GeneOhm™ Cdiff Assay has been found to be substantially equivalent to the Techlab C. difficile Tox-B Test (K935296) and the Techlab Clostridium difficile Toxin/Antitoxin Kit (K923463). These methods were used as the reference methods in the clinical trials.

Performance Characteristics:

Performance characteristics of the BD GeneOhm™ Cdiff Assay were determined in a multi-site prospective investigational study. Four (4) medical centers, two (2) in Canada and two (2) in the United States, participated in the study. To be enrolled in the study, specimens had to be from individuals for whom Clostridium difficile testing was indicated and/or ordered, according to institutional policies.

The Reference Cytotoxicity Assay was performed using a tissue culture Cytotoxicity assay on liguid or soft stool specimens within 48 hours of collection. The procedure was performed according to the Manufacturer's Instructions for Use.

A total of 1108 specimens were tested with both the Reference Assay described above and the BD GeneOhm™ Cdiff Assay, producing 1090 reportable results. The first dataset includes 835 fresh specimens tested at three (3) of the four (4) clinical sites (Table 1). In comparison to the Reference Assay, the BD GeneOhm™ Cdiff Assay identified 93.8% of the C. difficile positive specimens and 95.5% of the negative specimens (Table 2). For the population tested this resulted in a Negative Predictive Value (NPV) of 99.1% and a Positive Predictive Value (PPV) of 67.3%.

Testing at the fourth clinical site revealed that the Reference Cytotoxicity Assay was not reporting accurate results. Due to the high number of inaccurate reference assay results, samples were retested from aliquots of the original stool specimens which had been frozen after the initial testing. These frozen aliquots were tested with both the Reference Assay and the BD GeneOhm™ Cdiff Assay. The second dataset includes results from 255 frozen stool specimens available for analysis (Table 3).

{2}------------------------------------------------

In comparison to the Reference Assay, the BD GeneOhm™ Cdiff Assay identified 100% of the C. difficile positive specimens and 97.7% of the negative specimens in the frozen dataset: resulting in a NPV of 99.2% and PPV of 81.5% (Table 4).

Out of 852 fresh specimens tested with the BD GeneOhm™ Cdiff Assay, 39 were initially reported as unresolved (4.6%). Upon repeat testing from the frozen lysates, 22 were resolved and 17 remained unresolved (2.0%) (Table 5). Out of 256 frozen specimens tested with the BD GeneOhm™ Cdiff Assay, only one (1) specimen (0.4%) was initially reported unresolved. The specimen remained unresolved upon repeat testing from the frozen lysate (0.4%) (Table 6). One (1) run was reported invalid due to Run Control failure (0.6%). The run was reported valid upon repeat testing of the specimen lysates (Table 7).

Table 1: Fresh Stool Results Obtained with the BD GeneOhm™ Cdiff Assay in Comparison with the Reference Assay

Image /page/2/Figure/3 description: This image shows a table comparing the BD GeneOhm Cdiff Assay with a reference cytotoxicity assay. The table is a 2x2 matrix, with the rows representing the results of the BD GeneOhm Cdiff Assay (+ or -) and the columns representing the results of the reference cytotoxicity assay (+ or -). The table shows the number of samples that had each combination of results. For example, 76 samples were positive for both assays, while 720 samples were negative for both assays.

  • Cytotoxicity Assay on isolated strains was positive for 21 out of the 34 samples, verifying the presence of toxigenic C. difficile. For the remaining 13 samples, standard PCR with alternative primers followed by bi-directional sequencing revealed that 11 out of the 13 samples contained the expected todB gene. + For two (2) of the five (5) false negative specimens, C. difficile was recovered by culture, and only one (1) of these two (2) was reported as toxigenic. Of the remaining three (3) false negative PCR specimens, no C. difficile was recovered by culture.
Table 2: Performance Obtained with Fresh Stools using the BD GeneOhm™ Cdiff
Assay in Comparison with the Reference Method
Clinical SitesPrevalenceSensitivity with 95% CI*Specificity with 95% CI*
Site 111.0% (40/365)90.5% (38/42) (77.4% - 97.3%)95.7% (309/323) (92.8% - 97.6%)
Site 26.7% (16/240)94.4% (17/18) (72.7% - 99.9%)96.4% (240/249) (93.2% - 98.3%)
Site 311.1% (18/162)100% (21/21) (83.9% - 100%)94.0% (171/182) (89.4% - 96.9%)
Overall9.6% (74/767)93.8% (76/81) (86.2% - 98.0%)95.5% (720/754) (93.8% - 96.9%)
  • CI: Confidence Intervals

{3}------------------------------------------------

Table 3: Frozen Stool Results Obtained with the BD GeneOhm™ Cdiff Assay in Comparison with the Reference Assay

Image /page/3/Figure/1 description: This image shows a table comparing the BD GeneOhm Cdiff Assay with a reference cytotoxicity assay. The table is a 2x2 matrix, with rows representing the BD GeneOhm Cdiff Assay results (+ and -) and columns representing the reference cytotoxicity assay results (+ and -). The table shows the following values: 34 for (+,+), 5 for (+,-), 0 for (-,+), and 216 for (-,-). The row totals are 39 and 216, and the column totals are 34 and 221, with a grand total of 255.

Table 4: Performance Obtained with Frozen Stools using the BD GeneOhm™ Cdiff Assay in Comparison with the Reference Method

Clinical SitePrevalenceSensitivity with 95% CI*Specificity with 95% CI*
Site 412.7% (34/267)100.0% (34/34) (89.7% - 100%)97.7% (216/221) (94.8% - 99.3%)
  • CI: Confidence Intervals

Table 5: Fresh Stool Unresolved Rates

Clinical SitesInitial unresolved rate with95% CI*Unresolved rate after repeat with95% CI*
Site 10.8% (3/367)(0.2% - 2.4%)0.5% (2/367)(0.1% - 2.0%)
Site 26.6% (18/273)(4.0% - 10.2%)2.2% (6/273)(0.8% - 4.7%)
Site 38.5% (18/212)(5.1% - 13.1%)4.2% (9/212)(2.0% - 7.9%)
Overall4.6% (39/852)(3.3% - 6.2%)2.0% (17/852)(1.2% - 3.2%)
  • CI: Confidence Intervals

Table 6: Frozen Stool Unresolved Rates

Clinical SiteInitial unresolved rate with95% CI*Unresolved rate after repeat with95% CI*
Site 40.4% (1/256) (0.0% - 2.2%)0.4% (1/256) (0.0% - 2.2%)
  • CI: Confidence Intervals

Table 7: Overall Invalid Run Rates

SiteInvalid Run Rates with 95% Cl*
Site 12.6% (1/38)(0.1% - 13.8%)
Site 20.0% (0/41)(0.0% - 8.6%)
Site 30.0% (0/58)(0.0% - 6.2%)
Site 40.0% (0/23)(0.0% - 14.8%)
Overall0.6% (1/160)(0.0% - 3.4%)
  • CI: Confidence Intervals

{4}------------------------------------------------

Analytical Specificity

Genomic DNA from one non toxigenic C. difficile strain, two strains of Toxinotype XI lacking tcdB gene and 29 other-Clostridium strains (including one strain of C. sordelli), along with 99 closely related organisms and other pathogenic and commensal flora found in the intestine and stools (representing 96 species) were tested. All strains were tested at a concentration of approximately 1X108 CFU/mL or 1X10° target copies/mL. None of these species tested positive with the BD GeneOhm™ Cdiff Assay (Attachment 1).

Analytical Sensitivity

Quantitated culture and purified genomic DNA diluted in BD GeneOhm™ Cdiff Assay sample buffer were tested in five (5) replicates. The LOD was defined as the lowest concentration, in DNA copy number per reaction and CFU per reaction, at which five replicates out of five were found positive.

The analytical sensitivity (limit of detection or LOD) of the BD GeneOhm™ Cdiff Assay was determined with one strain of Toxinotype 0 Clostridium difficile carrying the tcdB gene (ATCC 43255).

The BD GeneOhm™ Cdiff Assay LOD is 10 DNA copies per reaction. The LOD in Colony Forming Units (CFU) is established at 4 CFU per reaction.

The analytical sensitivity in CFU per reaction was confirmed with a second Toxinotype 0 (ATCC 9689) and with Toxinotypes Illa (SE844), VII (57267) and VIII (1470) Clostridium difficile toxigenic strains.

In addition to strains used for LOD determination, one hundred (100) other toxigenic C. difficile strains (including 17 other Toxinotypes), representing 21 countries, from wellcharacterized clinical isolates or public collections were evaluated using the BD GeneOhm™ Cdiff Assay. C. difficile strains were tested at a concentration of approximately 6.7 DNA copies/uL or 1 CFU/μL. The assay correctly identified all 100 C. difficile strains carrying the tcdB gene.

Reproducibility

The reproducibility panel consisted of three (3) simulated specimen categories where each tube contained 100 uL of simulated bowel flora; the two positive panel members were also inoculated with C. difficile (ATCC 43255). Additionally, two (2) Specimen Processing Controls (ATCC 9689 and ATCC 25922) and, two (2) Run Controls (Positive and Negative) were included. The specimens were tested in triplicate per panel run, on five (5) distinct days (consecutive or not), wherein each day two (2) panels were tested, one for each of two (2) technologists, at three (3) clinical sites with one (1) lot of reagents. One (1) of these clinical sites participated in the extended study where two (2) additional lots of reagents were tested.

{5}------------------------------------------------

The overall percent agreement for the low positive C. difficile specimen category is 96.7%; the moderate positive C. difficile specimen category is 100% and the negative specimen category is 100% for the Site-to-Site Reproducibility (Table 8).

The overall percent agreement for the low positive C. difficile specimen category is 100%; the moderate positive C. difficile specimen category is 97.8% and the negative specimen category is 100% for the Lot-to-Lot Reproducibility (Table 9). Cycle threshold (Ct), an internal criteria used to determine a final assay result, was selected as an additional means of assessing assay reproducibility. Overall mean Ct values with variance components (SD and %CV) are shown in Tables 8 and 9.

An additional reproducibility study was performed, in accordance with the original reproducibility study protocol, to assess high negative specimens below the BD GeneOhm™ Cdiff Assay limit of detection (LOD). A sample containing simulated bowel flora was inoculated with C. difficile (ATCC 43255) at a concentration equivalent to the assay LOD. 100-fold and 10-fold dilutions of this sample were prepared, respectively, to obtain the two (2) high negative panel members. Overall percent agreement for negative test results and overall mean Ct values with variance components (SD and %CV) are shown in Table 10. As expected, the more dilute panel member (100-fold below the LOD) containing lower levels of target, demonstrates a higher percent agreement for negative test results than the less dilute panel member (10-fold below the LOD) which contains higher levels of target. Although high negative panel members are below the analytical LOD of the assay, positive test results may still be observed due to the presence of target in these specimens.

SITEOverall PercentAgreementCt Values
CategorySite 1Site 2Site 3Overall MeanSD%CV
NEG30/30100.0%30/30100.0%30/30100.0%90/90100.0%36.2†0.3†0.8%†
LOW POS28/3093.3%29/3096.7%30/30100.0%87/9096.7%38.80.92.3%
MOD POS30/30100.0%30/30100.0%30/30100.0%90/90100.0%38.31.02.7%

Table 8: Site-To-Site Reproducibility Study Results using One Lot

  • Data represent values from the internal control.
Table 9: Lot-To-Lot Reproducibility Study Results using Three Lots
LOTCt Values
CategoryLot 1Lot 2Lot 3Overall PercentAgreement
Percent AgreementPercent AgreementPercent AgreementOverallMeanSD%CV
NEG30/30100.0%30/30100.0%30/30100.0%90/90100.0%36.1‡0.3‡0.8%‡
LOW POS30/30100.0%30/30100.0%30/30100.0%90/90100.0%38.61.02.5%
MOD POS29/3096.7%29/3096.7%30/30100.0%88/9097.8%37.81.12.8%

4 Data represent values from the internal control.

{6}------------------------------------------------

******************************************************************************************************************************************************************************Д 7492 УАУАУ - УАУАЧАУАУ (1998 - УАРАУАУ РАСА АВАРАНА НУУ - УАУАРА " АУ 2017 - 14 -"AYANGORIAYA PAYAYAS IN "WA CAYA YA WAYA YA "A"AVAFAASWILL A ANAFAMANAULIL A PI
SHIPAN CATATA Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y AMIR/ANARA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YACorporation Comments of Children Comments of Children CommentWAWAYAYAYACE YA YA YA YA YA YA Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T A T
Carles Control Concerners of ChildrenCATALOGAL CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT COCarles of the Catalog Colors�������������������������������������WAWAYAVAYAVAVAAPATWANY WAND
******************************************************************************************************************************************************************************AYAYAYAYAYAYARARAYAYAYAYAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
*AYATAYAYAYAYAYAYARAWAY
and and which and which and which andCarlo Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Car
NºV HOMAY164611WWW.VAT
a a managar da mana mana mana mana mana mana mana marka marka marka mara mara marka maof the before to the concent are of client of a not show their one of
AMASANAYAYAN WASHINGO AVAICAMAWAYAYAMAYAYAYAYAYAYAYAAAATAYAYAYAYAYAYAYAYAYAYAY.AN MATHINASALEAS WS SATAY A V A T A T A T A T A S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M S M SYA WAY WASHINGWAY ANARA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YA YATATATATATATATATATATATATATATATATATATAAAAAAAAAAAAAAAAAAAAATWA WENDAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAY******************************************************************************************************************************************************************************A Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Ca
AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAW N A P M A P M A P A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y A Y AAPAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAYAY WAYAYAYAYA******************************************************************************************************************************************************************************
children and stock of all of the first and one of the first of the first of the first of the productions of the production
21/20
AVATANAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
YAYAWAWARAYAYAYAYAYAYAYAYAYAAAAW
.******************************************************************************************************************************************************************************

Table 10: Additional Reproducibility Study Using a High Negative Sample Panel

*Percent agreement for a negative result.

Precision

Within-laboratory precision was evaluated for the BD GeneOhm Cdiff Assay at one (1) site. The study was performed over 12 days, with two (2) runs per day and two (2) sample replicates per run. Samples included simulated specimens representing low and moderate positive C. difficile as well as negative C. difficile. One (1) out of 24 runs was excluded due to failure of the positive control (PC). One (1) moderate positive sample produced an unresolved result. All remaining samples and controls produced reportable results for a total of 46 replicates. Precision study results for low and moderate positive samples demonstrated agreement for (46/46) and (45/46) replicates, respectively; negative sample results demonstrated agreement for (46/46) replicates.

{7}------------------------------------------------

Interfering Substances

Twenty-six (26) biological and chemical substances occasionally used or found in perianal, rectal and/or stool specimens were evaluated for interference with the BD GeneOhm™ Cdiff Assay. Potentially interfering substances include, but are not limited to, blood and mucus. The presence of excessive blood may inhibit PCR and may give unresolved results. The remaining twenty-four (24) substances illustrated in the table below showed no detectable interference with the BD GeneOhm™ Cdiff Assay.

Endogenous and Commercial Exogenous Substances Tested with the BD GeneOhm Cdiff Assay

SubstanceResult
AnusolMC Plus *NI**
Atlas Ihle's Paste *Zinc oxide 25 % w/w paste (Laboratoire Atlas Inc.)NI
Barium sulfateFresh solution from powder form (LabMat)NI
Exact™ Hydrocortisone acetate * CreamUSP 0.5 % (Taro Pharmaceuticals Inc.)NI
Exact™ stomach reliefBismuth subsalicylate liquid (Perrigo®)NI
Fecal fatNI
Fresh control® Moist towelettes pH 5,5 (Blue Skin)NI
Gyne Moistrin®Vaginal moisturizing gel (Schering)NI
Imodium AD® *Loperamide hydrochloride oral solution (McNeil)NI
Kaopectate®Oral attaplugite suspension (Pharmacia & Upjohn)NI
K-Y® Jelly (Johnson & Johnson Inc.)NI
MetronidazoleFresh solution from powder form (Acros Organics)NI
SubstanceResult
MonistatTM DermMiconazole nitrate cream USP 2% (McNeil)NI
Palmitic acid *Fresh solution from powder (LabMat)NI
Preparation H® with Bio-Dyne® *Cream (Wyeth)NI
Preparation H® with Bio-Dyne® *Ointment (Wyeth)NI
Rougier Neo-Laryngobis *Suppositories(Rougier Pharma)NI
SAB-Dimenhydrinate® * Suppositories(SABEX®)NI
Steric acid *Fresh solution from powder (LabMat)NI
Trojan® latex condoms (with nonoxynol-9)Spermicidal lubricant (Church & Dwight Co., Inc.)NI
TucksMC personal cleansing padsMoist, soft clothpads (Pfizer)NI
Vagisil® Anti-itch cream (Combe Incorporated)NI
VancomycinLiquid (MP Biomedicals, LLC)NI
VaselineTM *White petroleum jelly U.S.P. (Lever Pond's)NI
  • Substance tested with two strains of C. difficile (Tox 0 and Tox VIII) ** NI: No detectable interference with the BD GeneOhm Cdiff Assay

{8}------------------------------------------------

Genera and SpeciesStrainBD GeneOhmTM CdiffAssay
Abiotrophia defectiveATCC 49176neg
Acinetobacter baumanniiATCC 19606neg
Acinetobacter IwoffiiCDCF 3697neg
Aeromonas hydrophilaATCC 7966/ CCRI-10071neg
Alcaligenes faecalis subsp. FaecalisATCC 15554neg
Anaerococcus tetradiusATCC 35098neg
Bacillus cereusATCC 13472neg
Bacillus cereusHER 1414neg
Bacteroides caccaeATCC 43185neg
Bacteroides merdaeATCC 43184neg
Bacteroides stercorisATCC 43183neg
Bifidobacterium adolescentisATCC 15703neg
Bifidobacterium longumATCC 15707neg
Campylobacter coliATCC 43479neg
Campylobacter jejuni subsp. jejuniATCC 33292neg
Candida albicansATCC 10231neg
Candida catenulateIDI-1729neg
Cedecea davisaeATCC 33431neg
Chlamydia trachomatisABI 08-901-000neg
Citrobacter amalonaticusATCC 25405neg
Citrobacter freundiiATCC 8090neg
Citrobacter koseriATCC 27028neg
Citrobacter sedlakiiATCC 51115 (IDI-2178)neg
Clostridium beijerinckiiATCC 8260neg
Clostridium bifermentansATCC 638neg
Clostridium bolteaeBAA-613neg
Clostridium botulinumHall Aneg
Clostridium butyricumCCRI-11128neg
Clostridium chauvoeiATCC 11957neg
Clostridium difficile non-toxigenicATCC-700057neg
Clostridium difficile Xla (A-B-tox bin+)IS58neg
Clostridium difficile XIb (A-B-tox bin+)R11402neg
Clostridium fallaxATCC 19400neg
Clostridium haemolyticumATCC 9650neg
Clostridium histolyticumATCC 19401neg
Clostridium innocuumCCRI-9927 / IDI 1986neg
Clostridium methylpentosumATCC 43829neg
Clostridium nexileATCC 27757neg
Clostridium novyiATCC 19402neg
Clostridium orbiscindensATCC 49531neg
Clostridium paraputrificumATCC 25780neg
Clostridium perfringensATCC 13124neg
Clostridium ramosumATCC 25582neg
Clostridium scindensATCC 35704neg
Clostridium septicumATCC 12464neg
Genera and SpeciesStrainBD GeneOhm™ Cdiff Assay
Clostridium sordelliiATCC 9714neg
Clostridium spCCRI-9842 / IDI 1987neg
Clostridium spCCRI-9929 / IDI-1988neg
Clostridium sphenoidesATCC 19403neg
Clostridium spiroformeATCC 29899neg
Clostridium sporogenesATCC 15579neg
Clostridium symbiosumCCRI-9928 / IDI 1989neg
Clostridium symbiosumATCC 14940neg
Clostridium tertiumATCC 14573neg
Clostridium tetaniATCC 19406neg
Collinsella aerofaciensATCC 25986neg
Corynebacterium genitaliumLSPQ 3583neg
Desulfovibrio pigerATCC 29098neg
Edwardsiella tardaATCC 15947neg
Eggerthella lentaCCRI-9926 / IDI 1990neg
Enterobacter aerogenesATCC 13048neg
Enterobacter cloacaeATCC 13047neg
Enterococcus casseliflavus (vanC2)CCRI-1566 / IDI 1981neg
Enterococcus cecorumATCC 43198neg
Enterococcus disparATCC 51266neg
Enterococcus faecalis vanBATCC 51299neg
Enterococcus faecium vanAATCC 700221neg
Enterococcus gallinarum vanCCCRI-1561 / IDI 1982neg
Enterococcus hiraeATCC 8043neg
Enterococcus raffinosusATCC 49427neg
Escherichia coliATCC 23511neg
Escherichia coliTop10 (IDI-266)neg
Escherichia fergusoniiATCC 35469neg
Escherichia hermanniiATCC 33650neg
Fusobacterium variumATCC 8501neg
Gardnerella vaginalisATCC 14019neg
Gemella morbillorumATCC 27824neg
Hafnia alveiATCC 13337neg
Helicobacter fennelliaeATCC 35683 / IDI-2180neg
Helicobacter pyloriATCC 43504neg
Homo sapiensATCC MGC-15492 / 2.16neg
Klebsiella oxytocaATCC 33496neg
Klebsiella oxytocaATCC 33497neg
Klebsiella pneumoniae subsp. PneumoniaeATCC 13883neg
Lactobacillus acidophilusATCC 4356neg
Lactobacillus reuteriATCC 23272neg
Lactococcus lactisATCC 11454neg
Leminorella grimontiiATCC 33999neg
Listeria grayiATCC 19120neg
Listeria innocuaATCC 33090neg
Listeria monocytogenesL374neg
Mitsuokella multacidaATCC 27723neg
Mobiluncus curtisii subsp. HolmesiiATCC 35242neg
Moellerella wisconsensisATCC 35017neg
Morganella morganii subsp. morganiiATCC 25830neg
Neisseria gonorrhoeaeATCC 35201neg
Genera and SpeciesStrainBD GeneOhm™ Cdiff Assay
Peptoniphilus asaccharolyticusATCC 14963neg
Peptostreptococcus anaerobiusATCC 27337neg
Plesiomonas shigelloidesATCC 14029neg
Porphyromonas asaccharolyticaATCC 25260neg
Prevotella melaninogenicaATCC 25845neg
Proteus mirabilisATCC 25933neg
Proteus penneriATCC 35198neg
Providencia alcalifaciensATCC 9886neg
Providencia rettgeriATCC 9250neg
Providencia stuartiiATCC 33672neg
Pseudomonas aeruginosaATCC 35554neg
Pseudomonas putidaLCDC D7172neg
Ruminococcus bromiiATCC 27255neg
Salmonella choleraesuis (typhimurium)ATCC 14028neg
Salmonella enterica subsp. Arizonae (formerlycholeraesuis arizonae)ATCC 13314neg
Salmonella enterica subsp. Enterica (formerlySalmonella choleraesuis subsp. choleraesuis)ATCC 7001neg
Serratia liquefaciensATCC 27592neg
Serratia marcescens2ATCC 13880neg
Shigella boydiiATCC 9207neg
Shigella dysenteriaeATCC 11835neg
Shigella sonneiATCC 29930neg
Staphylococcus aureus3ATCC 43300neg
Staphylococcus epidermidisATCC 14990neg
Stenotrophomonas maltophiliaATCC 13637neg
Streptococcus agalactiaeATCC 12973neg
Streptococcus dysgalactiaeATCC 43078neg
Streptococcus intermediusATCC 27335neg
Streptococcus uberisATCC 19436neg
Trabulsiella guamensisATCC 49490neg
Veillonella parvulaATCC 10790neg
Vibrio choleraeATCC 25870neg
Vibrio parahaemolyticusATCC 17802neg
Yersinia bercovieriATCC 43970neg
Yersinia rohdeiATCC 43380neg
Yokenella regensburgeiATCC 35313neg

Attachment 1: BD GeneOhm™ Cdiff Assay Reactivity Study using DNA and Lysates from Various Species

{9}------------------------------------------------

{10}------------------------------------------------

A SC curve with a strong background was obtained at the first testing leading to a positive status.

Retest in triplicate generated a final negative result.

in a con of the artistic and and and and and and of the one of the one of the one of the one of the one of the fin
t Two lysates were prepared because the first one gave an

*Tested with two lots of isolated DNA

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in black.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Raymond J Boule Senior Director, Regulatory Affairs, Quality Assurance BD Diagnostics (GeneOhm Sciences Inc.) 6146 Nancy Ridge Drive San Diego, CA 92121

DEC 1 9 2008

Re: K081920

Trade/Device Name: BD GeneOhm™ Cdiff Assay Regulation Number: 21 CFR § 866.2660 Regulation Name: Clostridium difficile toxin Regulatory Class: I Product Code: LLH Dated: July 2, 2008 Received: July 3, 2008

Dear Mr. Boule:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally

{12}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric''s (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attayna

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications For Use Statement

510(k) Number (if known): _ K081920

Device Name: BD GeneOhm™ Cdiff Assay

Indications For Use

Intended Use:

The BD GeneOhm™ Cdiff Assay is a rapid in vitro diagnostic test for the direct, qualitative detection of C. difficile toxin B gene (tcdB) in human liguid or soft stool specimens from patients suspected of having Clostridium difficile-associated disease (CDAD). The test, based on real-time PCR, is intended for use as an aid in diagnosis of CDAD. The test is performed directly on the specimen, utilizing polymerase chain reaction (PCR) for the amplification of specific targets and fluorogenic target-specific hybridization probes for the detection of the amplified DNA.

Prescription UseXXX
(Per 21 CFR 801.109)

OR

Over-The-Counter Use
(Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices Evaluation and Safety (OIVD)

Sally att
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K081920

§ 866.2660 Microorganism differentiation and identification device.

(a)
Identification. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.