K Number
K081895
Date Cleared
2009-01-08

(189 days)

Product Code
Regulation Number
864.7470
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term glycemic control of persons with diabetes. The A1c and total hemoglobin (tHb) values generated as part of the HbA1cN and HbA1cI results are intended for use in the calculation of the A1c / total hemoglobin ratio, and must not be used individually for diagnostic purposes.

*Note: HbA1cN reports HbA1c in % and HbA1cI reports HbA1c in mmol/mol

The ADVIA® Chemistry A1c Calibrators are for in vitro diagnostic use in the calibration of the A1c and total hemoglobin methods (Automated and Manual Pretreatment) on the ADVIA Chemistry Systems.

Device Description

The concentration of A1c and the concentration of total hemoglobin are measured and the ratio is reported (either as % or mmol/mol).

There are two different sample pretreatment methods available on the ADVIA Chemistry System. The first method is an Automated Pretreatment that uses 4 reagents: A1c Denaturant Reagent, Total Hemoglobin Reagent (tHb 2), A1c Agglutinator Reagent (R1) and A1c Antibody Reagent (R2). In this Automated Pretreatment step, the whole blood sample is mixed with the A1c Denaturant Readent. The red blood cells are lysed and the hemoglobin chains are hydrolyzed by the protease present in the reagent.

The second method is a Manual Pretreatment that uses the same reagents as the first method except that the A1c Denaturant Reagent is replaced with the Hemoglobin Denaturant Reagent. For this method, there is an off-line pretreatment that is followed bv a 10 minute incubation.

For the measurement of total hemoglobin, the Total Hemoglobin Reagent is used. The method is based on the conversion of all hemoglobin derivatives into alkaline hematin in an alkaline solution of a nonionic detergent

A latex agglutination inhibition method is used for the measurement of specific A1c. The A1c present in the sample competes with the agglutinator (synthetic latex containing multiple copies of the immunoreactive portion of A1c) for the anti-A1c antibody; thereby reducing the rate of agglutination. A concentration curve is obtained by monitoring the change in scattered light as a change of absorbance. The actual change in absorbance is inversely proportional to the concentration of A1c in the sample. The HbA1cN NGSP result (%) or the HbA1cl IFCC result (mmol/mol) is calculated using the A1c and total hemoglobin values.

The ADVIA® Chemistry A1c Calibrators are used to calibrate the methods. The calibrators consist of four (4) levels of lyophilized whole blood containing varying concentrations of HbA1c and total hemoglobin. There is a single level calibration for total hemoglobin (Cal 1) and a multi-level calibration (six levels) for A1c. Four calibrator levels (designated Cal 1 - 4) are provided in a single kit and each level is 0.5 g/vial. The other two levels consist of Saline (Cal 0) and Cal 5 (prepared by the system from Cal 4 using 1.4 times the volume used for Cal 4).

The target value of each calibrator (calibration) level is:

  • Calibration Level 1 (Cal 0): 0.00 umol/L A1c, 0.0 g/dL Total Hemoglobin ●
  • Calibration Level 2 (Cal 1): 2.30 µmol/L A1c, 11.0 g/dL Total Hemoglobin .
  • Calibration Level 3 (Cal 2): 3.65 umol/L A1c .
  • Calibration Level 4 (Cal 3): 5.15 umol/L A1c .
  • Calibration Level 5 (Cal 4): 6.80 umol/L A1c .
  • Calibration Level 6 (Cal 5): 8.20 umol/L A1c
AI/ML Overview

The Siemens Healthcare Diagnostics ADVIA® Chemistry Hemoglobin A1c Assay and Calibrators were evaluated for substantial equivalence to predicates. The study demonstrated performance through various tests, including imprecision, method comparison, interfering substances, matrix equivalency, and analytical range.

1. Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state pre-defined acceptance criteria for the performance characteristics. Instead, it presents the results of these performance studies and concludes that they are "acceptable results compared to the Predicate Device" and "support that the ADVIA Chemistry Hemoglobin A1c assay... is substantially equivalent to the Tosoh G7 HPLC Hemoglobin A1c method." However, based on the provided data, we can infer implied acceptance around certain metrics. The performance metrics are reported in comparison to the predicate device or a reference method.

Performance MetricAcceptance Criteria (Implied / Compared Against)Reported Device Performance (ADVIA Chemistry Hemoglobin A1c)
Imprecision (Total CV)Comparable to Predicate Device (Tosoh G7 HPLC)Automated Pretreatment: 1.9% (at 5.5% HbA1c), 1.3% (at 9.6% HbA1c)
Manual Pretreatment: 1.9% (at 5.0% HbA1c), 1.4% (at 8.8% HbA1c)
Predicate Device: 1.12% (at 5.8% HbA1c), 0.71% (at 10.9% HbA1c)
Method Comparison (Regression)Correlation with NGSP Reference Method (Tosoh G7)Automated: $y = 0.98x + 0.17$ ($r = 0.997$, Sy.x = 0.17) (Least Squares)
High correlation coefficient (r) close to 1,Automated: $y = 1.00x + 0.06$ ($r = 1.00$) (Passing Bablok / Pearson's)
slope close to 1, intercept close to 0.Manual: $y = 0.95x + 0.26$ ($r = 0.995$, Sy.x = 0.19) (Least Squares)
Manual: $y = 0.97x + 0.10$ ($r = 1.00$) (Passing Bablok / Pearson's)
Interfering Substances (Effect % change)Minimal effect (specific thresholds not stated, but results considered acceptable)Triglycerides (500mg/dL): -7.8% (Automated), -7.7% (Manual)
Bilirubin, free (60mg/dL): -1.9% (Automated), -2.0% (Manual)
Bilirubin, conjugated (60mg/dL): -3.8% (Automated), -5.9% (Manual)
Rheumatoid Factor (1100 IU/mL): 0% (Manual)
Analytical RangeStated range the device can accurately measure.2.9 - 15.4%

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Method Comparison (Test Set): 80 samples for both Automated and Manual Pretreatment methods.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The data appears to be from internal studies conducted by Siemens Healthcare Diagnostics. It is retrospective in the sense that samples were likely collected and then tested, but not explicitly labeled as prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This device is an in vitro diagnostic (IVD) assay measuring a biochemical marker (HbA1c). The "ground truth" for the test set in this context is established by a reference method or a predicate device, not by expert interpretation of images or other subjective assessments.

4. Adjudication Method for the Test Set

Not applicable. As this is an IVD assay, ground truth is established by an objective, traceable reference method (NGSP Reference Method, specifically Tosoh G7 in this case), not through expert consensus requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is an in vitro diagnostic device, not an imaging device or AI-assisted diagnostic tool that involves human readers interpreting cases. Therefore, an MRMC study and the concept of human reader improvement with AI assistance are not relevant to this submission.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, the studies presented (imprecision, method comparison, interfering substances, analytical range) demonstrate the standalone performance of the ADVIA Chemistry Hemoglobin A1c Assay. The device directly measures HbA1c concentrations and calculates the ratio, functioning independently of human interpretation of the final measurement.

7. Type of Ground Truth Used

The ground truth used for the method comparison study was established using the NGSP Reference Method (Tosoh G7 Automated HPLC Analyzer - HbA1c Variant Analysis Mode). The document states that the Tosoh G7 is certified by NGSP and used as a reference method for establishing traceable results to the Diabetes Control and Complications Trial (DCCT). This indicates a highly standardized and traceable method for establishing the true HbA1c values.

8. Sample Size for the Training Set

Not explicitly stated. The document focuses on the performance studies (validation) of the finished device. For IVD assays, "training sets" are usually involved in the initial assay development and optimization, rather than being a distinct, reported phase like in AI/ML device development. The calibrators themselves are used to "train" or calibrate the instrument on an ongoing basis.

9. How the Ground Truth for the Training Set was Established

For IVD assays like this, the "ground truth" for calibrators is typically established through a process of assigning traceable values. The ADVIA® Chemistry A1c Calibrators are stated to be "Traceable to NGSP*** and Traceable to IFCC by Master Equation." This means the assigned values for the calibrators are determined through rigorous analytical methods that link them back to internationally recognized standardization programs (NGSP and IFCC), ensuring accuracy and consistency across different laboratories and instruments. This would involve a hierarchy of reference methods and materials.

{0}------------------------------------------------

K081895

510(k) Summary

JAN - 8 2009

Submitter informationContact person:Philip LiuManager, Regulatory Affairs & Compliance
Address:Siemens Healthcare Diagnostics511 Benedict AvenueTarrytown, NY 10591
Phone:914-524-2443914-524-2500 (fax)
Date summary prepared:November 12, 2008
Device Trade or Proprietary Names:ADVIA® Chemistry Hemoglobin A1c AssayADVIA® Chemistry A1c Calibrators
Device Common/Usual Name orClassification Name:Glycosylated Hemoglobin AssayCalibrators
Classification Number/Class:LCP / Class II

JIX / Class II

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) number is: _k081895

Assay Predicate Device:

Predicate Device
Device NameG7 Automated HPLC Analyzer - HbA1c Variant Analysis Mode
Common nameGlycosylated Hemoglobin Test
510(k) NumberK011434
ManufacturerTosoh Medics, Inc.

Calibrator Predicate Device:

Predicate Device
Device NameVITROS Chemistry Products Calibrator Kit 18
Common nameCalibrator
510(k) NumberK041764
ManufacturerOrtho-Clinical Diagnostics

Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary

{1}------------------------------------------------

Device Description:

The concentration of A1c and the concentration of total hemoglobin are measured and the ratio is reported (either as % or mmol/mol).

There are two different sample pretreatment methods available on the ADVIA Chemistry System. The first method is an Automated Pretreatment that uses 4 reagents: A1c Denaturant Reagent, Total Hemoglobin Reagent (tHb 2), A1c Agglutinator Reagent (R1) and A1c Antibody Reagent (R2). In this Automated Pretreatment step, the whole blood sample is mixed with the A1c Denaturant Readent. The red blood cells are lysed and the hemoglobin chains are hydrolyzed by the protease present in the reagent.

The second method is a Manual Pretreatment that uses the same reagents as the first method except that the A1c Denaturant Reagent is replaced with the Hemoglobin Denaturant Reagent. For this method, there is an off-line pretreatment that is followed bv a 10 minute incubation.

For the measurement of total hemoglobin, the Total Hemoglobin Reagent is used. The method is based on the conversion of all hemoglobin derivatives into alkaline hematin in an alkaline solution of a nonionic detergent

A latex agglutination inhibition method is used for the measurement of specific A1c. The A1c present in the sample competes with the agglutinator (synthetic latex containing multiple copies of the immunoreactive portion of A1c) for the anti-A1c antibody; thereby reducing the rate of agglutination. A concentration curve is obtained by monitoring the change in scattered light as a change of absorbance. The actual change in absorbance is inversely proportional to the concentration of A1c in the sample. The HbA1cN NGSP result (%) or the HbA1cl IFCC result (mmol/mol) is calculated using the A1c and total hemoglobin values.

The ADVIA® Chemistry A1c Calibrators are used to calibrate the methods. The calibrators consist of four (4) levels of lyophilized whole blood containing varying concentrations of HbA1c and total hemoglobin. There is a single level calibration for total hemoglobin (Cal 1) and a multi-level calibration (six levels) for A1c. Four calibrator levels (designated Cal 1 - 4) are provided in a single kit and each level is 0.5 g/vial. The other two levels consist of Saline (Cal 0) and Cal 5 (prepared by the system from Cal 4 using 1.4 times the volume used for Cal 4).

The target value of each calibrator (calibration) level is:

  • Calibration Level 1 (Cal 0): 0.00 umol/L A1c, 0.0 g/dL Total Hemoglobin ●
  • Calibration Level 2 (Cal 1): 2.30 µmol/L A1c, 11.0 g/dL Total Hemoglobin .
  • Calibration Level 3 (Cal 2): 3.65 umol/L A1c .
  • Calibration Level 4 (Cal 3): 5.15 umol/L A1c .
  • Calibration Level 5 (Cal 4): 6.80 umol/L A1c .
  • Calibration Level 6 (Cal 5): 8.20 umol/L A1c

Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary

{2}------------------------------------------------

Storage is at 2 - 8°C.

CAUTION! POTENTIAL BIOHAZARD: Contains human source material. While each human serum or plasma donor unit used in the manufacture of this product was tested by FDA-approved methods and found nonreactive for hepatitis B surface antigen (HBsAg), antibody to hepatitis C (HCV), and antibody to HIV-1/2, all products manufactured using human source material should be handled as potentially infectious. Because no test method can offer complete assurance that hepatitis B or C viruses, HIV, or other infectious agents are absent, these products should be handled according to established good laboratory practices.

Statement of Intended Use:

The ADVIA® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term alvcemic control of persons with diabetes. The A1c and total hemoglobin (tHb) values generated as part of the HbA1cN and HbA1cl methods are intended for use in the calculation of the A1c / total hemoglobin ratio, and must not be used individually for diagnostic purposes.

*Note: HbA1cN reports HbA1c in % and HbA1cl reports HbA1c in mmol/mol

The ADVIA® Chemistry A1c Calibrators are for in vitro diagnostic use in the calibration of the A1c and total hemoglobin methods (Automated and Manual Pretreatment) on the ADVIA Chemistry Systems.

Comparisons to the Predicate Device:

Method Similarities

ADVIA ChemistryHemoglobin A1c( new device )Tosoh G7 Automated HPLC Analyzer- HbA1c Variant Analysis Mode( predicate device )
Intended UseFor in vitro diagnostic use for themeasurement of hemoglobin A1cFor in vitro diagnostic use for themeasurement of hemoglobin A1c
Specimen TypeHuman Whole Blood(EDTA and Lithium Heparin)Human Whole Blood
StandardizationTraceable to NGSP***Traceable to IFCC by Master EquationTraceable to NGSP***

{3}------------------------------------------------

Method Differences

ADVIA ChemistryHemoglobin A1c(new device)Tosoh G7 Automated HPLC Analyzer- HbA1c Variant Analysis Mode(predicate device)
Assay PrincipleFor Hemoglobin:Conversion of all hemoglobinderivatives into alkaline hematinFor HbA1c:Latex agglutination inhibitionHigh Performance LiquidChromatography(non-porous ion exchange)
Determination /Calculation of resultsRatio of HbA1c/ Total HemoglobinMeasurement of HbA1c as apercentage of total hemoglobin
CalibrationMultipoint calibration (6 levels) forHemoglobin A1cSingle point calibration for TotalHemoglobin2 point Calibration (high and low)
Analytical Range2.9 - 15.4%4.2 - 20.8%

Calibrator Similarities

ADVIA ChemistryA1c Calibrators(new device)VITROS Chemistry Products CalibratorKit 18(predicate device)
Intended UseFor in vitro diagnostic use in thecalibration of the A1c and totalhemoglobin methods (Automated andManual Pretreatment) on the ADVIAChemistry SystemsFor in vitro diagnostic useonly. VITROS Chemistry ProductsCalibrator Kit 18 is used in conjunctionwith VITROS Chemistry Products FSCalibrator 1 to calibrate VITROS 5, 1FS Chemistry Systems for thecalculation of percent glycatedhemoglobin (%A1c).
Calibrator Levels4 levels provided4 levels provided
Matrix / IngredientsProcessed human whole bloodHemolysate derived from human andovine blood, surfactants, stabilizer, andpreservatives
FormLyophilizedLyophilized
Storage2 - 8°C2 - 8°C
Shelf Life Stabilitystable until the expiration datestable until the expiration date
Reconstituted Vial /Use Stability7 days stored at 2 - 8°C2 days stored at 2 - 8ºC

Calibrator Differences

ADVIA ChemistryA1c Calibrators(new device)VITROS Chemistry Products CalibratorKit 18(predicate device)
Instructions for Use(Preparation)Reconstitute with DI water before useReconstitute with diluent before use

Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary .

{4}------------------------------------------------

Performance:

Substantial equivalence was demonstrated by testing several performance characteristics including imprecision, method comparison, interfering substances, matrix equivalency, and analytical range. The following tables summarize the precision (total), interfering substances, analytical range, and method comparison results.

All of the evaluation studies gave acceptable results compared to the Predicate Device. These studies support that the ADVIA Chemistry Hemoglobin A1c assay (with the ADVIA Chemistry A1c Calibrators) is substantially equivalent to the Tosoh G7 HPLC Hemoglobin A1c method that is currently marketed.

Imprecision

ADVIA ChemistryHemoglobin A1cADVIA 1800 - AutomatedADVIA ChemistryHemoglobin A1cADVIA 1800 - ManualTosoh G7 HPLCHbA1c
LevelHbA1c (%)Total CV(%)LevelHbA1c (%)Total CV(%)LevelHbA1c (%)Total CV(%)
5.51.95.01.95.81.12
9.61.38.81.410.90.71

Correlation

(y = ADVIA Chemistry Hemoglobin A1c, x = comparison method/system)

System (y)NRegressionEquationrSy.xSample Range(%)
ADVIA 1800 Automated –Least Squares80$y = 0.98x + 0.17$0.9970.174.40 - 12.60
ADVIA 1800 Automated –Passing Bablok / Pearson's80$y = 1.00x + 0.06$1.00--4.40 - 12.60
ADVIA 1800 Manual –Least Squares80$y = 0.95x + 0.26$0.9950.194.40 - 12.60
ADVIA 1800 Manual –Passing Bablok / Pearson's80$y = 0.97x + 0.10$1.00--4.40 - 12.60

x = NGSP Reference Method (Tosoh G7), y = ADVIA Chemistry System

Interfering Substances (ADVIA Chemistry Hemoglobin A1c on ADVIA 1800)

Automated Sample Pretreatment

Interfering SubstanceInterferent Conc.(mg/dL)HbA1c conc.(%)Effect(% change)
Triglycerides (Intralipid)500 mg/dL5.2- 7.8
Bilirubin, free60 mg/dL5.3- 1.9
Bilirubin, conjugated60 mg/dL5.2- 3.8

Siemens Healthcare Diagnostics (formerly Siemens Medical Solutions Diagnostics) ADVIA Chemistry Hemoglobin A1c Assay and Calibrators Premarket Notification - 510(k) 510(k) Summary

Page 5 of 6

{5}------------------------------------------------

Manual Sample Pretreatment

Interfering SubstanceInterferent Conc.(mg/dL)HbA1c conc.(%)Effect(% change)
Triglycerides (Intralipid)500 mg/dL5.2- 7.7
Bilirubin, free60 mg/dL5.1- 2.0
Bilirubin, conjugated60 mg/dL5.1- 5.9
Rheumatoid Factor1100 IU/mL4.80

Analytical Range

PlatformAnalytical Range
All ADVIA Chemistry ' '- Galibrator Leval 6-(12-3-18.6%)1 -1 -1 -1 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 -- 2 --2.4 /0 -10

Conclusions:

The ADVIA Chemistry Hemoglobin A1c method and Calibrators are substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Tosoh G7 Automated HPLC Analyzer- HbA1c Variant Analysis Mode. The Tosoh G7 Automated HPLC Analyzer is certified by the National Glycohemoglobin Standardization Program (NGSP) and it is used as a reference method for establishing traceable results of other methods to the Diabetes Control and Complications Trial (DCCT).

The ADVIA Chemistry A1c Calibrators are substantially equivalent to the Predicate Device VITROS Chemistry Products Calibrator Kit 18.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Siemens Healthcare Diagnostic c/o Dr. Philip Liu Manager, Regulatory Affairs and Compliance 511 Benedict Avenue Tarrytown, NY 10591

JAN - 8 2009

K081895 Re:

Trade/Device Name: ADVIA Chemistry Hemoglobin A1c Assay, ADVIA Chemistry

Alc Calibrators

Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylate hemoglobin assay. Regulatory Class: Class II Product Code: LCP, JIX Dated: November 18, 2008 Received: November 19, 2008

Dear Dr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).

{7}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Gregory C. Hammett, Ph.D.

Courthey C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indication for Use

510(k) Number (if known): K081895

Device Name(s):

ADVIA® Chemistry Hemoglobin A1c ADVIA® Chemistry A1c Calibrators

Indications For Use:

The ADVIA® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term glycemic control of persons with diabetes. The A1c and total hemoglobin (tHb) values generated as part of the HbA1cN and HbA1cI results are intended for use in the calculation of the A1c / total hemoglobin ratio, and must not be used individually for diagnostic purposes.

*Note: HbA1cN reports HbA1c in % and HbA1cI reports HbA1c in mmol/mol

The ADVIA® Chemistry A1c Calibrators are for in vitro diagnostic use in the calibration of the A1c and total hemoglobin methods (Automated and Manual Pretreatment) on the ADVIA Chemistry Systems.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C),

(PLEASE DO NOT WRITE BELOW THIS LINE: CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K081895

§ 864.7470 Glycosylated hemoglobin assay.

(a)
Identification. A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.(b)
Classification. Class II (performance standards).