(167 days)
The BD ProbeTec Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay, when tested with the BD Viper™ System in extracted mode, uses Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in cliniciancollected female endocervical and male urethral swabs, patient-collected vaginal swab specimens (in a clinical setting), and female and male urine specimens. The assay is indicated for use with asymptomatic and symptomatic female individuals and symptomatic male individuals to aid in the diagnosis of gonococcal urogenital disease.
The BD ProbeTec GC O' Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermallycontrolled fluorescent readers. The presence or absence of N. gonorrhoeae DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.
In addition to the fluorescent probe used to detect amplified N. gonorrhoeae target DNA, a second labeled oligonucleotide is incorporated in each reaction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the N. gonorrhoeae-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and N. gonorrhoeae-specific signals to report results as positive, negative, or EC failure.
The document describes the BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay, a diagnostic device for detecting Neisseria gonorrhoeae.
Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria in terms of specific sensitivity and specificity thresholds that the device must meet. Instead, it presents the assay's performance characteristics (Sensitivity, Specificity, PPV, NPV) against a "Patient Infected Status (PIS) algorithm" as a measure of its effectiveness.
However, based on the submission being for a 510(k) and the conclusion stating "The analytical and clinical study results... support the determination of substantial equivalence in accordance with the intended use," it can be inferred that the reported performance metrics were considered acceptable for deeming the device substantially equivalent to predicate devices.
Here's a summary of the clinical performance:
| Metric | Overall Reported Performance (Across all specimen types, symptomatic statuses) |
|---|---|
| Sensitivity | 99.3% (95% CI: 98.2% - 99.8%) |
| Specificity | 99.4% (95% CI: 99.2% - 99.6%) |
| PPV | 95.5% |
| NPV | 99.9% |
Analytical Acceptance Criteria and Reported Performance:
| Criterion | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Limit of Detection (Analytical Sensitivity) | To detect N. gonorrhoeae at low concentrations. | Urine: ≤ 50 cells per mL (for neat and UPT treated urine) Swabs: ≤ 100 GC cells per mL (for expressed vaginal and endocervical swab specimens) Detected 17 GC strains with ≥ 95% proportion positive at 50 cells per mL in Q* Swab Diluent. |
| Analytical Specificity | To avoid cross-reactivity with common microorganisms. | Tested against 141 organisms. Two N. cinerea and two N. lactamica strains showed cross-reactivity. All other tested organisms did not cross-react. |
| Interfering Substances | Performance should not be significantly impacted by common substances. | No Interference Observed: Blood, seminal fluid, mucus, various vaginal products/contraceptives, hemorrhoidal cream, prescription vaginal treatments, antibiotics, analgesics, deodorant sprays/powders, hormones, leukocytes, albumin, glucose, acidic/alkaline urine, bilirubin, organisms associated with UTIs.May cause EC failures: Blood (>60%) in Swab specimens. |
| Reproducibility | Consistent results across different sites and runs. | Assessed by proportion positive/negative and coefficient of variation (SD, %CV) (details in Table 4 and 5). Demonstrated high reproducibility for both positive and negative samples, and even for samples below LOD. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Clinical Performance:
- Female Subjects: 994
- Male Subjects: 472
- Total BD ProbeTec GC Q Assay results used for sensitivity/specificity calculations:* 5387
- Data Provenance:
- Country of Origin: North America (seven geographically diverse clinical sites).
- Retrospective or Prospective: Prospective. Specimens were "collected from 1059 female subjects and 479 male subjects attending OB/GYN, sexually transmitted disease (STD) and family planning clinics."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The ground truth was not established by medical experts' consensus or interpretation. Instead, it was established by a Patient Infected Status (PIS) algorithm relying on the results of two other commercially available Nucleic Acid Amplification Tests (NAATs). Therefore, the concept of "number of experts" and "qualifications of those experts" does not directly apply in this context for the clinical performance ground truth.
4. Adjudication Method for the Test Set
The adjudication method for determining the "Patient Infected Status (PIS)" was an algorithm based on the results of the two reference NAATs:
- For females: Subjects were considered infected with GC if two of the four endocervical swab and urine specimens (from a total of two endocervical swabs and two urine specimens) tested positive in both the BD ProbeTec ET GC/AC assay AND the other commercially available NAAT (i.e., one specimen testing positive in each NAAT).
- For males: Subjects were considered infected with GC if two of the three or four urethral swab and urine specimens tested positive in both the BD ProbeTec ET GC/AC assay AND the other commercially available NAAT (i.e., one specimen testing positive in each NAAT).
- Non-infected: Subjects were considered non-infected if less than two reference NAAT results were positive.
This is a form of composite reference standard or adjudication by algorithm based on multiple reference tests.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The study evaluates the performance of the device (assay) in standalone mode against a composite reference standard, not its impact on human reader performance or diagnostic accuracy with AI assistance. This device is a diagnostic assay, not an AI-assisted interpretation tool for human readers.
6. Standalone Performance
Yes, a standalone (algorithm only without human-in-the-loop performance) study was performed. The entire clinical performance evaluation described in the document, comparing the BD ProbeTec GC Q* Assay results to the PIS algorithm, is a standalone performance study of the diagnostic assay. This assay is designed to provide a qualitative (positive/negative) result without human interpretation of raw data for diagnosis.
7. Type of Ground Truth Used
The ground truth used for clinical performance was a composite reference standard or Patient Infected Status (PIS) algorithm. This algorithm was established using the results from two other commercially available Nucleic Acid Amplification Tests (NAATs): the BD ProbeTec ET GC/AC assay and another unnamed commercially available NAAT. It is not pathology, expert consensus, or outcomes data in the traditional sense, but rather a best available evidence standard from established molecular diagnostics.
8. Sample Size for the Training Set
The document does not explicitly mention a separate "training set" or "validation set" in the context of machine learning. The clinical performance study described used a total of 994 female subjects and 472 male subjects as the test set to evaluate the device. For analytical performance (LOD, specificity), various spiked samples and clinical isolates were used. For the clinical study, it appears the entire subject cohort served as the evaluation set against the composite reference standard.
9. How the Ground Truth for the Training Set Was Established
Since a distinct "training set" for an AI or machine learning model is not explicitly mentioned, the question of how its ground truth was established is not directly applicable. If one considers the overall development of the assay, the analytical studies (LOD, specificity) would have informed the assay's design and cutoff values, where the ground truth was known by design (e.g., specific concentrations of N. gonorrhoeae for LOD, known organisms for specificity). The clinical ground truth relied on the PIS algorithm as detailed in point 4 and 7.
{0}------------------------------------------------
Korres
Image /page/0/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a graphic symbol on the left and the letters "BD" on the right. Below the letters, there is a tagline that reads "Helping all people live healthy lives." The graphic symbol appears to be a stylized representation of a person with outstretched arms under a sun-like shape.
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
| Applicant | BD Diagnostic Systems7 Loveton CircleSparks, MD 21152 |
|---|---|
| DEC 11 2008 | |
| Establishment Registration No. | 1119779 |
| Contact Person | Kathryn Babka Carr, RACtel. 410-316-4260fax. 410-316-4041Kathy.Carr@bd.com |
| Summary Date | December 1, 2008 |
| Proprietary Name | BD ProbeTec™ Neisseria gonorrhoeae (GC) Q x Amplified DNA Assay |
| Generic Name | DNA probe, nucleic acid amplification, Neisseria |
| Classification | Class II |
| Classification Name | Neisseria spp. direct serological test reagents |
| Regulation Number | 866.3390 |
| Product Code | LSL |
| Predicate Devices | BD ProbeTec ET CT/GC Amplified DNA Assay (K984631),APTIMA Combo 2 Assay (K003395) |
Device Description
The BD ProbeTec GC O' Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermallycontrolled fluorescent readers. The presence or absence of N. gonorrhoeae DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.
In addition to the fluorescent probe used to detect amplified N. gonorrhoeae target DNA, a second labeled oligonucleotide is incorporated in each reaction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the N. gonorrhoeae-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of a stylized sun-like graphic on the left and the letters "BD" in bold font on the right. Below the letters, there is a tagline that reads "Helping all people live healthy lives".
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and N. gonorrhoeae-specific signals to report results as positive, negative, or EC failure.
Intended Use
The BD ProbeTec™ Neisseria gonorrhoeae (GC) Q Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorthoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is indicated for use with asymptomatic and symptomatic females and symptomatic males to aid in the diagnosis of gonococcal urogenital disease.
Summarv and Principles of Operation
When used with the BD Viper System, the BD ProbeTec GC Q* Amplified DNA Assay (GC Q* Assay) involves automated extraction of DNA from clinical specimens through the chemical lysis of cells, followed by binding of DNA to para-magnetic particles, washing of the bound nucleic acid and elution in an amplification-compatible buffer. When present, N. gonorrhoeae DNA is then detected by Strand Displacement Amplification (SDA) of a specific target sequence in the presence of a fluorescently labeled detector probe.
Analytical Performance Characteristics
Limit of Detection (Analytical Sensitivity)
The Limits of Detection (LODs) for the GC Q* Assay with Neisseria gonorrhoeae strain ATCC 19424 in urine and swab specimens when extracted on the BD Viper System were determined to be ≤ 50 cells per mL for neat and UPT treated urine and ≤ 100 GC cells per mL for expressed vaginal and endocervical swab specimens. The GC Q' Assay on the BD Viper System in extracted mode was able to detect 17 GC strains with ≥ 95% proportion positive at a concentration of 50 cells per mL in Q* Swab Diluent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of a stylized sunburst above a figure with outstretched arms, followed by the letters "BD" in bold font. Below the logo, the text reads "Helping all people live healthy lives".
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
Analytical Specificity
The 141 organisms listed in Table 1 were tested with the BD ProbeTec GC Q Amplified DNA Assay on the BD Viger System. All potential cross-reactive species were tested at ≥ 1x10° cells/mL except where noted. Two N. cinerea and two N. lactamica strains were shown to cross-react with the GC Q* assay.
| Table 1: Potential Cross-reactant Microorganisms | |||
|---|---|---|---|
| Acinetobacter calcoaceticus | Epstein Barr Virus *** | Peptostreptococcus productus | Neisseria elongata subsp.nitroreduscens (2) |
| Acinetobacter Iwoffi | Escherichia coli | Plesiomonas shigelloides | Neisseria elongata |
| Actinomyces israelii | Flavobacterium meningosepticum | Propionibacterium acnes | Neisseria flava (4) |
| Adenovirus*** | Gardnerella vaginalis | Providencia stuartii | Neisseria flavescens (4) |
| Aeromonas hydrophilia | Gemella haemolysans | Pseudomonas aeruginosa | Neisseria gonorrhoeae |
| Alcaligenes faecalis* | Haemophilus influenzae | Salmonella minnesota | Neisseria lactamica (7) |
| Bacillus subtilis* | Herpes Simplex Virus ** | Salmonella typhimurium | Neisseria meningitidis (12) |
| Bacteroides fragilis | Human papillomavirus (16 and 18)*** | Staphylococcus aureus | Neisseria mucosa (5) |
| Candida albicans* | Kingella kingae | Staphylococcus epidermidis | Neisseria perflava (8) |
| Candida glabrata* | Klebsiella pneumoniae | Streptococcus agalactiae | Neisseria polysaccharea (2) |
| Candida tropicalis* | Lactobacillus acidophilus* | Streptococcus mitis | Neisseria sicca (5) |
| Chlamydia pneumoniae**** | Lactobacillus brevis | Streptococcus mutans | Neisseria subflava (15) |
| Chlamydia psittaci* | Lactobacillus jensenii* | Streptococcus pneumoniae* | Neisseria weaverii (3) |
| Citrobacter freundii | Listeria monocytogenes | Streptococcus pyogenes | |
| Clostridium perfringens | Mobiluncus mulieris | Streptomyces griseus** | |
| Corynebacterium renale | Moraxella lacunata* | Trichomonas vaginalis** | |
| Cryptococcus neoformans* | Moraxella osloensis | Veillonella parvula | |
| Cytomegalovirus** | Morganella morganii | Vibrio parahaemolyticus | |
| Edwardsiella turda | Mycobacterium gordonae | Yersinia enterocolitica | |
| Enterobacter cloacae | Mycobacterium smegmatis | Branhamella catarrhalis (5) | |
| Enterococcus faecalis | Peptostreptococcus anaerobius | Neisseria cinerea (2) | |
| Neisseria elongata ss |
(n) number of strains tested in the BD ProbeTec GC Q4 Assay
- Tested at > I x IO' cells on > I x 10' cells or viral particles per mL; ** Tested at ≥ 1 x l 0" genomic equivalents per mL;
*** tested at > I x IO' TCDwinL
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of a stylized sun-like graphic on the left, followed by the letters "BD" in bold, sans-serif font. Below the letters, there is a tagline that reads "Helping all people live healthy lives" in a smaller font size.
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
Interfering Substances
Potential interfering substances which may be encountered in swab and/or urine specimens were extracted from urine and vaginal swab matrix in the absence and presence of GC target (150 GC cells/mL for urine and 300 GC cells/mL for swabs) and tested with the BD ProbeTec GC Q* Amplified DNA Assay on the BD Viper System. Results are summarized in Table 2.
| Interpretation | Swab | Urine |
|---|---|---|
| No InterferenceObserved | Blood (≤ 60%) | Blood (1%) |
| Seminal Fluid | Seminal fluid | |
| Mucus | Mucus | |
| Over The Counter vaginal products andcontraceptives | Antibiotics | |
| Hemorrhoidal cream | Analgesics | |
| Prescription vaginal treatments | Over The Counter deodorant spraysand powders | |
| Leukocytes (1x106 cells/mL) | Hormones | |
| 1x106 cells/mL Chlamydia trachomatis | Leukocytes | |
| Albumin <1 mg/mL | ||
| Glucose | ||
| Acidic urine (pH 4.0) | ||
| Alkaline urine (pH 9.0) | ||
| Bilirubin | ||
| Organisms associated with UrinaryTract Infections | ||
| May causeextraction control(EC) failures | Blood (> 60%) | Not observed |
Table 2: Interfering Substances
Clinical Performance Characteristics
Clinician-collected endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female O' UPT and neat urine specimens were collected from 1059 female subjects and 479 male subjects attending OB/GYN, sexually transmitted disease (STD) and family planning clinics at seven geographically diverse clinical sites in North America. Subjects were classified as symptomatic if they reported symptoms such as dysuria, urethral discharge, coital pain/difficulty/bleeding, testicular or scrotum pain/swelling, abnormal vaginal discharge, or pelvic/uterine/adnexal pain. Subjects were classified as asymptomatic if they did not report symptoms. Sixty five female subjects and 7 male subjects were excluded from the data analysis due to age requirement violations, antibiotic treatment in the last 21 days, opting to withdraw from the study after initially consenting, failure to obtain paired swab and urine specimens, urine quantity less than 20 mL, or
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a symbol on the left and the letters "BD" on the right. The symbol appears to be a stylized representation of a person with outstretched arms, surrounded by a sun-like shape. Below the letters "BD" is the company's slogan: "Helping all people live healthy lives."
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
transport and storage errors related to specimen collection. Therefore, the final data analysis included 994 compliant female subjects and 472 compliant male subjects.
Five specimens were collected from each of the 994 eligible female subjects. A urine specimen was collected and split into Q UPT, neat urine and the two reference urine specimen collection devices followed by a vaginal swab specimen and three randomized endocervical swab specimens. Up to four specimens were collected from each of the 472 eligible male subjects. Up to three randomized urethral swab specimens were collected followed by a urine specimen that was split into Q UPT, neat urine and the two reference urine specimen collection devices. BD ProbeTec GC Q assay results were generated from the Q* UPT and neat urine specimens, the vaginal swab specimen, one endocervical swab specimen and one male urethral swab specimen. The remaining two endocervical swab specimens, up to two male urethral swab specimens, and the two reference urine specimens for each male and female subject were tested using two reference methods: the BD ProbeTec ET GC/AC assay and another commercially available NAAT (Nucleic Acid Amplification Test). Specimen testing was conducted either at the site of collection or at a designated BD Viper testing site.
All performance calculations were based on the total number of BD ProbeTec GC Q assays results for endocervical, vaginal and male urethral swab specimens, and male and female Q * UPT and neat urine specimens compared to a patient infected status (PIS) algorithm for each gender. In the algorithm, the designation of a subject as being infected with GC or not was based on endocervical swab and unne specimen results from the commercially available BD ProbeTec ET GC/AC assay and the other commercially available NAAT. Subjects were considered with GC if two of the four endocervical swab and urine specimens (or two of the three or four urethral swab and urine specimens) tested positive in the BD ProbeTec ET GC/AC assay and the other reference NAAT (one specimen testing positive in each NAAT). Subjects were considered non-infected if less than two reference NAAT results were positive. A total of 5387 BD ProbeTec GC Q* assay results was used to calculate sensitivity and specificity. Sensitivity and specificity by specimen type and symptomatic status are presented in Table 3.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for BD, a global medical technology company. The logo consists of a stylized sun-like graphic to the left of the letters "BD" in bold, sans-serif font. Below the letters, there is a tagline that reads "Helping all people live healthy lives" in a smaller font size.
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
Table 3: GC Q Assay Performance Compared to Patient Infected Status (by specimen type and symptomatic status)
| SpecimenType | Symptomatic | N | Sensitivity | 95% C.I. | Specificity | 95% C.I. | PPV% | NPV% | ErrorInitial/Final |
|---|---|---|---|---|---|---|---|---|---|
| FS | N | 450 | 96.3% (26/27) | (81.0%-99.9%) | 99.5% (421/423) | (98.3%-99.9%) | 92.9 | 99.8 | 3/0 |
| Y | 542 | 100.0% (38/38) | (90.7%-100.0%) | 99.8% (503/504) | (98.9% -100.0%) | 97.4 | 100.0 | 2/2 | |
| Total | 992 | 98.5% (64/65) | (91.7% - 100.0%) | 99.7% (924/927) | (99.1%-99.9%) | 95.5 | 99.9 | 5/2 | |
| FV | N | 449 | 100.0% (27/27) | (87.2%-100.0%) | 98.6% (416/422) | (96.9%-99.5%) | 81.8 | 100.0 | 0/0 |
| Y | 544 | 100.0% (38/38) | (90.7%-100.0%) | 99.6% (504/506) | (98.6%-100.0%) | 95.0 | 100.0 | 0/0 | |
| Total | 993 | 100.0% (65/65) | (94.5% - 100.0%) | 99.1% (920/928) | (98.3%-99.6%) | 89.0 | 100.0 | 0/0 | |
| FN | N | 450 | 96.3% (26/27) | (81.0%-99.9%) | 99.3% (420/423) | (97.9%-99.9%) | 89.7 | 99.8 | 0/0 |
| Y | 543 | 97.4% (37/38) | (86.2%-99.9%) | 99.6% (503/505) | (98.6% -100.0%) | 94.9 | 99.8 | 0/0 | |
| Total | 993 | 96.9% (63/65) | (89.3%-99.6%) | 99.5% (923/928) | (98.7%-99.8%) | 92.6 | 99.8 | 0/0 | |
| FUPT | N | 450 | 100.0% (27/27) | (87.2%-100.0%) | 99.5% (421/423) | (98.3%-99.9%) | 93.1 | 100.0 | 0/0 |
| Y | 543 | 97.4% (37/38) | (86.2% -99.9%) | 99.8% (504/505) | (98.9% -100.0%) | 97.4 | 99.8 | 0/0 | |
| Total | 993 | 98.5% (64/65) | (91.7%-100.0%) | 99.7% (925/928) | (99.1%-99.9%) | 95.5 | 99.9 | 0/0 | |
| MS | N | 215 | 100.0% (7/7) | (59.0% - 100.0%) | 100.0% (208/208) | (98.2% - 100.0%) | 100.0 | 100.0 | 0/0 |
| Y | 257 | 100.0% (100/100) | (96.4% - 100.0%) | 98.7% (155/157) | (95.5%-99.8%) | 98.0 | 100.0 | 1/0 | |
| Total | 472 | 100.0% (107/107) | (96.6% - 100.0%) | 99.5% (363/365) | (98.0%-99.9%) | 98.2 | 100.0 | 1/0 | |
| MN | N | 215 | 100.0% (7/7) | (59.0% - 100.0%) | 100.0% (208/208) | (98.2% -100.0%) | 100.0 | 100.0 | 0/0 |
| Y | 257 | 100.0% (100/100) | (96.4%-100.0%) | 98.1% (154/157) | (94.5%-99.6%) | 97.1 | 100.0 | 0/0 | |
| Total | 472 | 100.0% (107/107) | (96.6%-100.0%) | 99.2% (362/365) | (97.6%-99.8%) | 97.3 | 100.0 | 0/0 | |
| MUPT | N | 215 | 100.0% (7/7) | (59.0% - 100.0%) | 99.5% (207/208) | (97.4% - 100.0%) | 87.5 | 100.0 | 0/0 |
| Y | 257 | 100.0% (100/100) | (96.4%-100.0%) | 98.7% (155/157) | (95.5%-99.8%) | 98.0 | 100.0 | 0/0 | |
| Total | 472 | 100.0% (107/107) | (96.6% - 100.0%) | 99.2% (362/365) | (97.6%-99.8%) | 97.3 | 100.0 | 0/0 | |
| Total | 5387 | 99.3% (577/581) | (98.2% - 99.8%) | 99.4% (4779/4806) | (99.2%-99.6%) | 95.5 | 99.9 | 6/2 |
| A | Asymptomatic |
|---|---|
| CI | Confidence Interval |
| FNU | Female Neat Urine |
| FS | Female endocervical swab |
| FUPT | Female urine in Q* UPT |
| FV | Female vaginal swab |
| MNU | Male Neat Urine |
| MS | Male urethral swab |
| MUPT | Male urine in Q* UPT |
| n | number |
| S | Symptomatic |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a graphic symbol on the left and the letters "BD" on the right. Below the letters, there is a tagline that reads "Helping all people live healthy lives." The graphic symbol appears to be a stylized representation of a person with arms raised, surrounded by rays or petals, possibly symbolizing health and well-being.
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
A total of 5387 GC Q Assay results was evaluated at seven geographically diverse clinical sites. A frequency distribution of the initial MaxRFU values for the GC Q' Assay with an assay cutoff of 125 MaxRFU is shown in Figure A.
Image /page/6/Figure/4 description: The image is a bar graph showing the frequency of MaxPFU. The x-axis shows the MaxPFU ranges, and the y-axis shows the frequency. The bar graph shows that the most frequent MaxPFU range is 0-49, with a frequency of 4765. The MaxPFU range of >=800 has a frequency of 594.
Image /page/6/Figure/5 description: The image shows the title of a figure. The title is "Figure A: Frequency Distribution of MaxRFU for the GC Q* Assay". The title is written in a clear, sans-serif font and is centered on the image.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a stylized graphic on the left and the letters "BD" on the right. The graphic resembles a person with arms raised towards a sun-like burst of light. Below the logo, the tagline reads "Helping all people live healthy lives" in a smaller font.
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
Reproducibility
Reproducibility of the BD Viper System using the BD ProbeTec GC Q* Assay was evaluated at three clinical sites on one BD Viper System per site. A panel of simulated specimens was tested that comprised CT and GC organisms seeded into swab diluent for the BD ProbeTec GC Q* Assay. Simulated endocervical and urethral specimens contained a clean endocervical swab whereas the simulated urine and vaginal swab specimens did not. Uninoculated swab diluent for the BD ProbeTec GC Q* Assay was used for the GC negative samples. Nine replicates of each panel member were tested every day for five days on each BD Viper System. The data are summarized in Table 4.
| Specimen Type | CT, EB/ml | GC Cells/ml | Within Run | Between Run | Within Site | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||||||||
| Endocervical/Urethral | 0 | 0 | 99.3%(134/135) | (95.9%,100.0%) | 13.8 | 151.3 | 1096.3 | 0.0 | 0.0 | 0.6 | 4.3 |
| 30 | 0 | 98.5%(133/135) | (94.8%,99.8%) | 28.1 | 220.7 | 785.3 | 0.0 | 0.0 | 33.8 | 120.3 | |
| 0 | 100 | 100.0%(135/135) | (97.3%,100.0%) | 1859.5 | 94.1 | 5.1 | 0.0 | 0.0 | 19.2 | 1.0 | |
| 30 | 250 | 100.0%(135/135) | (97.3%,100.0%) | 1847.3 | 117.6 | 6.4 | 0.0 | 0.0 | 25.9 | 1.4 | |
| 75 | 100 | 100.0%(135/135) | (97.3%,100.0%) | 1855.9 | 119.4 | 6.4 | 0.0 | 0.0 | 42.2 | 2.3 | |
| Urine/Vaginal | 0 | 0 | 99.3%(134/135) | (95.9%,100.0%) | 15.7 | 162.3 | 1031.1 | 0.0 | 0.0 | 0.0 | 0.0 |
| 30 | 0 | 100.0%(135/135) | (97.3%,100.0%) | 1.1 | 3.1 | 295.8 | 0.7 | 69.7 | 0.5 | 48.3 | |
| 0 | 100 | 100.0%(135/135) | (97.3%,100.0%) | 1899.0 | 86.1 | 4.5 | 22.8 | 1.2 | 0.0 | 0.0 | |
| 30 | 250 | 100.0%(135/135) | (97.3%,100.0%) | 1884.2 | 94.0 | 5.0 | 13.8 | 0.7 | 0.0 | 0.0 | |
| 75 | 100 | 100.0%(135/135) | (97.3%,100.0%) | 1867.2 | 87.7 | 4.7 | 0.0 | 0.0 | 19.2 | 1.0 |
Table 4: Summary of Reproducibility Data on the BD Viper System for the GC Q Assay
A sccond study was conducted internally to characterize the reproducibility of test results (i.e., proportion positive or negative) at target levels below the analytical Limit of Detection (LOD) of the BD ProbeTec GC Q Assay. A panel of simulated specimens was tested that comprised GC and CT organisms seeded into Q* swab diluent at two different levels each of which was below the respective analytical LOD for the organisms (1:10, 1:100). These levels were selected to fall within the dynamic range of the analytical L.OD curve of the assay. Fifteen replicates of each panel member were tested every day for five days across three BD Viper Systems. The data are summarized in Table 5.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a stylized graphic on the left and the letters "BD" on the right. The graphic resembles a person with arms outstretched under a sunburst. Below the letters "BD" is the tagline "Helping all people live healthy lives" in a smaller font.
BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
Table 15: Characterization of System Reproducibility at Target Levels below the Analytical Limit of Detection for the GC Q* Assay
| 2007 10:50 PM 11 | Beginningੇਰਾ (ਕਿਸ (ਸੰਮਾ) ( | 19 1985 1978 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 1 | 11/24/2019 | 11:12 2011977 - 199 | And The First of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first and | |||
|---|---|---|---|---|---|---|---|---|
| 1 1 2 2017 12 | ||||||||
| Endocervical/Urethral | 1:10 | 92.9(209/225) | (88.7, 95.9) | 1324.6 | 7.1(16/225) | (4.1, 11.3) | 41.4 | |
| Endocervical/Urethral | 1:100 | 30.7(69/225) | (24.7. 37.1) | 835.9 | 69.3(156/225) | (62.9, 75.3) | 7.2 | |
| Urine/Vaginal | 1:10 | 90.7(204/225) | (86.1, 94.1) | 1165.9 | 8.3(21/225) | (5.9, 13.9) | 34.2 | |
| Urine/Vaqinal | 1:100 | 22.7(51/225) | (17.4 28.7) | 872.7 | 77.3(174/225) | (71.3, 82.6) | 7.8 |
Conclusions
The analytical and clinical study results for the BD ProbeTec Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay support the determination of substantial equivalence in accordance with the intended use as stated in the product labeling.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized lines representing its wings or feathers.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Kathryn Babka Carr Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
DEC 1 1 2008
Re: K081825
Trade/Device Name: BD ProbeTec Neisseria gonorrhoeae (GC) Qx Amplified DNA Assay Regulation Number: 21 CFR 866.3390 Regulation Name: Neisseria spp. direct serological test reagents Regulatory Class: Class II Product Code: LSL Dated: December 5, 2008 Received: December 9, 2008
Dear Ms. Babka Carr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sally attayata
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
Indications for Use
510(k) Number: K081825
Device Name: BD ProbeTec™ Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay
Indications For Use:
The BD ProbeTec Neisseria gonorrhoeae (GC) Q* Amplified DNA Assay, when tested with the BD Viper™ System in extracted mode, uses Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in cliniciancollected female endocervical and male urethral swabs, patient-collected vaginal swab specimens (in a clinical setting), and female and male urine specimens. The assay is indicated for use with asymptomatic and symptomatic female individuals and symptomatic male individuals to aid in the diagnosis of gonococcal urogenital disease.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Uve Schuf
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K081825
Page 1 of 1
BD Diagnostic Systems Becton, Dickinson and Company
Page ix
§ 866.3390
Neisseria spp. direct serological test reagents.(a)
Identification. Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identifyNeisseria spp. from cultured isolates. Additionally, some of these reagents consist ofNeisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence ofNeisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusNeisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.(b)
Classification. Class II (performance standards).