K Number
K081712
Device Name
FINGERTIP PULSE OXIMETER, MODEL M70, AND HANDHELD PULSE OXIMETER, MODEL M700
Date Cleared
2008-09-12

(87 days)

Product Code
Regulation Number
870.2700
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment. The oximeter is not suitable to monitor patient continuously for long term. The M700 handheld pulse oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.
Device Description
The subject device of M70 Fingertip Pulse Oximeter is a fingertip device, which can display %SpO2, pulse rate value, waveform pulse amplitude bar indication. The subject device consists of detector and emitter LED, OLED display module, CPU, driving circuit, power supply circuit and battery. The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's Law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate. The subject device of M700 Handheld Pulse Oximeter is a handheld device, whose mainly function are display %SpO2, pulse rate value, pulse amplitude bar indication. The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate. The subject device consists of detachable SpO2 sensor, Main Frame and Battery. The Main Frame consist two parts, which is Control Module and SpO2 Module.
More Information

Not Found

No
The device description relies on established spectrophotometry principles (Lambert Beer's Law) and standard components (LEDs, detectors, CPU) for calculating SpO2 and pulse rate. There is no mention of AI/ML terms, training/test data, or complex algorithms beyond the fundamental physics of pulse oximetry.

No
The devices are intended to measure functional arterial oxygen saturation (SpO2) and pulse rate, which are diagnostic or monitoring functions, not therapeutic ones designed to treat or alleviate a condition.

Yes

The device measures functional arterial oxygen saturation (SpO2) and pulse rate, which are physiological parameters used to assess a patient's health status, thus aiding in the diagnosis or monitoring of conditions.

No

The device descriptions explicitly state that both the M70 and M700 devices consist of hardware components such as detectors, emitters, displays, CPUs, power supplies, batteries, and detachable sensors.

Based on the provided information, no, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information for diagnosis, monitoring, or screening.
  • Device Function: The M70 and M700 pulse oximeters measure functional arterial oxygen saturation (SpO2) and pulse rate by placing a sensor on the fingertip (M70) or using a detachable sensor (M700). This is a non-invasive measurement taken directly from the patient's body, not from a specimen taken from the body.
  • Operational Principle: The description mentions using spectrophotometry and Lambert Beer's Law to calculate oxygen saturation based on light absorption through the tissue. This is a physical measurement, not a chemical or biological analysis of a sample.

Therefore, these pulse oximeters fall under the category of non-invasive medical devices used for physiological monitoring, not In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

M70 Fingertip Pulse Oximeter: The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment. The oximeter is not suitable to monitor patient continuously for long term.

M700 Handheld Pulse Oximeter: The M700 handheld pulse oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.

Product codes (comma separated list FDA assigned to the subject device)

DQA

Device Description

M70 Fingertip Pulse Oximeter: The subject device of M70 Fingertip Pulse Oximeter is a fingertip device, which can display %SpO2, pulse rate value, waveform pulse amplitude bar indication. The subject device consists of detector and emitter LED, OLED display module, CPU, driving circuit, power supply circuit and battery. The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's Law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate. The subject device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does contain drug or biological product.

M700 Handheld Pulse Oximeter: The subject device of M700 Handheld Pulse Oximeter is a handheld device, whose mainly function are display %SpO2, pulse rate value, pulse amplitude bar indication. The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate. The subject device consists of detachable SpO2 sensor, Main Frame and Battery. The Main Frame consist two parts, which is Control Module and SpO2 Module. The subject device has 6 model sensors as the accessory, they are listed below: A0212-SA103PV (for Adult) A0212-SP103PV (for Pediatric) A0212-SW103PU (for Neonate) RSJ091DA (for Adult) RSJ091DI (for Pedicatio) RSY091DN (for Neonate)

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Fingertip (for M70 Fingertip Pulse Oximeter)

Indicated Patient Age Range

M70 Fingertip Pulse Oximeter: adult patients
M700 Handheld Pulse Oximeter: adult, pediatric and neonatal patients

Intended User / Care Setting

M70 Fingertip Pulse Oximeter: hospital type facilities as well as in the home care environment.
M700 Handheld Pulse Oximeter: hospital, hospital type facilities as well as in the home care environment.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

M70 Fingertip Pulse Oximeter: Laboratory testing was conducted to validate and verify that M70 Fingertip Pulse Oximeter met all design specifications, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of all consensus standards
M700 Handheld Pulse Oximeter: Laboratory and Clinical insting was conducted to validate and verify that M700 Handheld Pulse Oximeter met all design specifications, including electrical safety, EMC, biocompatibility, specific n. Results of these tests demonstrate compliance to the requirements of all consensus standards.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K070371, K072581

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/1 description: The image shows a sequence of handwritten digits and letters. The sequence starts with the letters 'Ko', followed by the numbers '81712'. The characters are written in a dark ink, and the image has a white background.

Premarket Notification 510(k) Chapter III 510(k) Summary Report No : A2008-002-050

Chapter III 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The Assigned 510(k) Number is:

1. Sponsor

Guangdong Biolight Meditech Co., Ltd Innovation First Road, Technology Innovation Coast Zhuhai, Guangdong, 519085, China Contact Person: Mr. Tianbao Li, Chief Engineer +86-756-3399963 Tel: Fax: +86-756-3399989 E-mail: li tb@blt.com.cn

2. Submission Correspondent

Ms. Diana Hong Mr. Lee Fu Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, China 20020 Phone: +86-21-64264467 Fax: 760-466-5084 Email: Diana.hong@mid-link.net

Subject Device Information 3.

M70 Fingertip Pulse Oximeter a.

(1)Classification Name: Oximeter

(2)Regulation Number:870.2700

(3)Product Code:DQA

(4)Class: II

(5)Review Panel: Anesthesiology

M700 Handheld Pulse Oximeter b.

(1)Classification Name: Oximeter

1

(2)Regulation Number:870.2700 (3)Product Code: DQA (4)Class: II (5)Review Panel: Anesthesiology

Predicate Device 4.

  • a. Fingertip Oximeter MD300C
    K-number: K070371 Product Code: DQA

Intended Use:

Fingertip Pulse Oximeter MD300C is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc). Not for continuously monitoring.

Manufactured by:

Beijing Choice Electronic Technology Co., Ltd. Room 1127-1128 Building B, Bailangyuan Fuxing Road , No. A36 Beijing, CHINA 100039

b. PM-60 Pulse Oximeter

K-number: K072581

Product Code: DQA

Intended Use:

The PM-60 handheld Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, pediatric and neonatal patients in hospitals, out-of-hospital transport and home care.

Manufactured by:

Shenzhen Mindray Bio-medical Electronics Co., LTD. Mindray Building, Keji 12th Road South, Hi-tech Industrial Park Nanshan, Shenzhen, 518057, P.R.China

2

5. M70 Fingertin Pulse Oximeter

5.1 Device Description

The subject device of M70 Fingertip Pulse Oximeter is a fingertip device, which can display %SpO2, pulse rate value, waveform pulse amplitude bar indication.

The subject device consists of detector and emitter LED, OLED display module, CPU, driving circuit, power supply circuit and battery.

The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's Law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate.

The subject device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does contain drug or biological product.

5.2 Statement of Intended Use:

The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment.

The oximeter is not suitable to monitor patient continuously for long term.

5.3 Testing

Laboratory testing was conducted to validate and verify that M70 Fingertip Pulse Oximeter met all design specifications, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of all consensus standards

5.4 Substantially Equivalence Determination

Comparison Analysis

The subject device has same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The only difference is Measurement Wavelengths, Working time, Operating temperature, PR Measuring range, Atmosphere pressure.

3

Conclusion:

The subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.

M700 Handheld Pulse Oximeter 6.

6.1 Device Description

The subject device of M700 Handheld Pulse Oximeter is a handheld device, whose mainly function are display %SpO2, pulse rate value, pulse amplitude bar indication.

The operational principle of subject device uses the method as mensurated optical density by spectrophotometer. It is base on the double wavelength Lambert Beer's law and uses the specific absorbance of oxygenation hemoglobin and deoxyhemoglobin to calculate.

The subject device consists of detachable SpO2 sensor, Main Frame and Battery. The Main Frame consist two parts, which is Control Module and SpO2 Module.

The subject device has 6 model sensors as the accessory, they are listed below:

A0212-SA103PV (for Adult) A0212-SP103PV (for Pediatric) A0212-SW103PU (for Neonate) RSJ091DA (for Adult) RSJ091DI (for Pedicatio) RSY091DN (for Neonate)

6.2 Statement of Intended Use

The M700 handheld puls oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) will pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as in the home care environment.

6.3 Testing

Laboratory and Clinical insting was conducted to validate and verify that M700 Handheld Pulse Oximeter met all design specifications, including electrical safety, EMC, biocompatibility, specific n. Results of these tests demonstrate compliance to the requirements of all consensus standards

4

6.4 Substantially Equivalence Determination

Comparison Analysis

The subject device has same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performane safety as the predicate device. The only difference is Measurement Wavelengths, Display Unit, Power, Working time, Operating temperature, PR Measuring range, Relative humidity, Atmosphere pressure.

Conclusion:

The subject device is Sul clantially Equivalent (SE) to the predicate device which is US legally market device. The subject device is determined as safe and effectiveness.

5

Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features an abstract eagle design with three stylized wing segments. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

SEP 1 2 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Guangdong Biolight Meditech Company, Limited C/O Ms. Diana Hong General Manager Shanghai Mid-Link Business Consulting Company, Limited Suite 8D, Zhongshan Zhongxin Masion No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030 CHINA

Re: K081712

Trade/Device Name: M700 Handheld Pulse Oximeter M70 Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: June 13, 2008 Received: June 17, 2008

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

6

Page 2 - Ms. Hong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of vour device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Thlmelt Kirdx>

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

7

Indication for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: M70 Fingertip Pulse Oximeter_____________________________________________________________________________________________________________________________________

Indications for Use:

The M70 fingertip pulse oximeter is intended to measure functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospital type facilities as well as in the home care environment.

The oximcter is not suitable to monitor patient continuously for long term.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

hul

Page 1 of __2

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K081712

8

Premarket Notification 510(k) Chapter II Indication for Use Statement Report No.: A2008-002-050 Comments of the country

Indication For Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: __ M700 Handheld Pulse Oximeter

Indications for Use:

The M700 handheld pulse oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.

Prescription Use __ J ________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

:

AND/OR

Over-The-Counter Use __________ (21 CFR 801 Subpart C)

and the contract and the country of the

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

and the country of the country of the country of the county of the county of

Concurrence of CDRH, Office of Device Evaluation (ODE)

e

Page 2 of _ 2_

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices

510(k) Number: K081712