K Number
K081666

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-10-07

(116 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the maxillary lateral incisors and mandibular lateral and central incisors. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the Astra Tech OsseoSpeed 3.0 mm Implant.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Device Description

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.

The Atlantis™ Abutments for Astra Tech OsseoSpeed 3.0 Implant System and abutment screws are made from Titanium grade Ti-6A1-4V ELI (Meets ASTM Standard F-136). The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with Astra's 3.0 mm OsseoSpeed™ Implants.

AI/ML Overview

This document is a 510(k) summary for the Atlantis™ Abutment for Astra Tech OsseoSpeed 3.0 Implant System and does not contain information about the acceptance criteria or a study proving that the device meets such criteria. A 510(k) submission generally focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance studies against specific acceptance criteria.

The document states:

  • "The Atlantis™ Abutments for Astra OsseoSpeed 3.0 Implants are substantially equivalent in intended use, material, design and performance to the Atlantis Abutments tor Astra Implants cleared under K070833 and for the Astra Tech OsseoSpeed TM Narrow Implants cleared under K080396."

To answer the user's request, I would need a different type of document, such as a full study report or a design verification document, which is not provided here.

{0}------------------------------------------------

OCT 0 7 2008

Pre-market Notification Page - 000009


006
K081666

VII. SECTION 10 - 510(K) SUMMARY

This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

1 . Applicant's Name and Address

Astra Tech Inc. 25 First Street Cambridge, Massachusetts 02141 Telephone Number: 617-661-9799 Fax Number: 617-661-9063 Contact Person: Franklin Uyleman Manager of Quality and Regulatory Affairs

2. Name of Device

Trade Name:Atlantis™ Abutment for Astra Tech OsseoSpeed3.0 Implant System
Common Name:Endosseous dental implant abutment
Classification Name:Endosseous dental implant abutment21 CFR 872.3630 Product code NHA

3. Legally Marketed Device to which Equivalence is claimed (Predicate Device)

ManufacturerDevice510(k)Number
Astra Tech Inc.,(formerly AtlantisComponents Inc.)-Atlantis Abutment for Astra ImplantK070833
Astra Tech AB-OsseoSpeed™ NarrowK080396

4. Description of the Device

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.

{1}------------------------------------------------

K08/666

2072

Pre-marke ?age - 000010 - - - - - - - - - - - - - -

4. Description of the Device (continued)

The Atlantis™ Abutments for Astra Tech OsseoSpeed 3.0 Implant System and abutment screws are made from Titanium grade Ti-6A1-4V ELI (Meets ASTM Standard F-136). The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with Astra's 3.0 mm OsseoSpeed™ Implants.

న్. Intended Use of the Device

The devices covered by this submission are abutments which are placed into a dental implant to provide support for a prosthetic reconstruction. The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the maxillary lateral incisors and mandibular lateral and central incisors. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

6. Basis for Substantial Equivalence

The Atlantis™ Abutments for Astra OsseoSpeed 3.0 Implants are substantially equivalent in intended use, material, design and performance to the Atlantis Abutments tor Astra Implants cleared under K070833 and for the Astra Tech OsseoSpeed TM Narrow Implants cleared under K080396.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the department's name arranged in a circular fashion around a stylized symbol. The symbol is a depiction of an abstract eagle, with its wings forming three distinct, curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 0 7 2008

Astra Tech, Incorporated C/O Ms. Betsy A. Brown Consultant B.A. Brown & Associates 8994 Tamaroa Terrace Skokie, Illinois 60076

Re: K081666

Trade/Device Name: Atlantis™ Abutment for Astra Tech OsseoSpeed 3.0 Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: September 24, 2008 Received: September 26, 2008

Dear Ms. Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Brown

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal ogencies. You must comply with all the Act's requirements, including, but not limited to: registerior and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus parmits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regading postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

THAmuels Lend, ms for//

Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Pre-market Notification Page - 000227

K081666

Indications for Use

510(k) Number (if Known)

Device Name: Atlantis ™ Abutment for Astra Tech OsseoSpeed 3.0 Implant System

Indication for Use:

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the maxillary lateral incisors and mandibular lateral and central incisors. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the Astra Tech OsseoSpeed 3.0 mm Implant.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Prescription Use X (Part 21 CFR 801 SubpartD) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runrel

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)