K Number
K081661
Date Cleared
2008-06-24

(11 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Applications: Ophthalmic, Fetal, Abdominal, Intra-operative (vascular/epicardial), Intra-operative (Neuro), Laparoscopic, Pediatric, Small Organ (thyroid, scrotum, breast, prostate), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-urethral, Trans-esoph. (non-Card.), Musculo-skel (conventional), Musculo-skel (superficial), Intra-luminal, Other (Gynecological), Cardiac Adult, Cardiac Pediatric, Trans-esoph. (Cardiac), Peripheral vessel, Other (Carotid, I/O).

Modes of Operation: B, M, PWD, CWD, Color Doppler, Combined (Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual), Other (Panoramic, Color Power Angio (CPA, formerly Angio), 3D/4D, Harmonics (Tissue and Contrast), Directional Angio Imaging, Tissue Doppler Imaging, Amplitude Doppler).

Device Description

The HD15 is a diagnostic ultrasound device. It consists of a system console containing the power supply and electronic circuitry required to generate the image, a display screen, and a connection to the separate transducers. It is substantially equivalent to the currently marketed HD11 Diagnostic Ultrasound System (K043535 and K062247). The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo-electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The differing acoustic properties of the tissues in the body reflect some of the transmitted energy back to the transducer, where it is converted back to electrical signals and sent back to the system. In the system, advanced signal processing technologies convert the returned signals into images of the tissues. The Doppler functions of this system process the Doppler shift frequencies from the echoes of moving targets (such as blood), to detect and graphically display the Doppler shifts of these tissues as flow.

AI/ML Overview

Here's an analysis of the provided text regarding the Philips HD15 Diagnostic Ultrasound System, focusing on acceptance criteria and supporting studies:

Assessment of the Provided Information:

The provided document is a 510(k) Premarket Notification for the Philips HD15 Diagnostic Ultrasound System. It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices (HD11 Diagnostic Ultrasound System, K043535 and K062247).

A key aspect of substantial equivalence is demonstrating that a new device is as safe and effective as a legally marketed predicate device. This K-number document is not a clinical study report that establishes performance against standalone acceptance criteria. Instead, it attests that the new device's performance is equivalent to that of existing, cleared devices.

Therefore, the requested information elements related to specific performance metrics, sample sizes, expert ground truth, adjudication methods, and comparative effectiveness studies will largely be absent from this type of regulatory submission because the device is cleared based on demonstrating equivalence, not novel performance against pre-defined metrics.


1. Table of Acceptance Criteria and Reported Device Performance

As noted above, this submission demonstrates substantial equivalence rather than establishing new performance criteria. Therefore, there are no explicit acceptance criteria for this specific device's novel performance or reported quantitative performance metrics in the format typically seen in a clinical study for a new, non-equivalent device.

The "acceptance criteria" here are effectively met by demonstrating:

  • Same Indications for Use: The HD15 is indicated for the same diagnostic ultrasonic imaging and fluid flow analysis as the predicate device.
  • Same Capabilities: Both have the same gray-scale and Doppler capabilities.
  • Same Technologies: Both use essentially the same technologies for imaging, Doppler functions, and signal processing.
  • Acoustic Output: Both have acoustic output levels below the Track 3 FDA limits.
  • Quality Systems & Biosafety: Both are manufactured under equivalent quality systems and from materials with equivalent biosafety.
  • Safety Standards: Both are designed and manufactured to the same electrical and physical safety standards.

2. Sample Size Used for the Test Set and Data Provenance

This document does not describe a "test set" in the context of validating novel performance data. The clearance is based on the device's functional equivalence to already-cleared predicate devices. Therefore, there are no patient or image data sets used for testing described in this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable for this type of submission. There is no "ground truth" establishment described in relation to a test set, as the clearance is based on substantial equivalence to existing devices, whose efficacy and safety were presumably established previously.

4. Adjudication Method for the Test Set

Not applicable. No test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not a study involving AI or human reader performance improvement. It's a regulatory submission for a diagnostic ultrasound system based on substantial equivalence.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a diagnostic ultrasound system, not an algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context is the established safety and effectiveness of the predicate devices. The HD15 is deemed safe and effective because it is substantially equivalent to a device already cleared by the FDA. There is no new primary "ground truth" established for the HD15 in this document.

8. The Sample Size for the Training Set

Not applicable. This device is an ultrasound system, not an AI algorithm requiring a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable. This device is an ultrasound system, not an AI algorithm requiring a training set and associated ground truth.


Conclusion:

The provided document is a 510(k) Premarket Notification which demonstrates substantial equivalence of the Philips HD15 Diagnostic Ultrasound System to existing cleared devices. It does not describe a clinical study with specific acceptance criteria, test sets, ground truth establishment, or performance metrics against novel claims. Instead, its "proof" is the detailed comparison to predicate devices, showing that it functions identically, uses similar technologies, and meets the same safety standards.

{0}------------------------------------------------

510(k) Premarket Notification

JUN 2 4 2008

3.2 Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 as implemented in 21 C.F.R. 8807.92.

The submitter of this premarket notification is:

William Picatti Manager, Regulatory Affairs Philips Ultrasound, Inc. 22100 Bothell Everett Highway Bothell, WA 98021-8431 Tel: (425) 487-7151 Fax: (425) 487-8666

This summary was prepared on May 12, 2009.

The proprietary name of the device is the HD15 Diagnostic Ultrasound System. In combination with transducers - 3D9-3v, BP10-5ec, C5-2, C6-3, C8-4v, C8-5, D2cwc, L 12-3, L 12-5, L 15-7io, S5-2, S8-3, S7-2omni and V6-2 - are commonly known as a diagnostic ultrasound system and transducers.

These devices are classified as follows:

90IYNUltrasonic Pulsed Doppler Imaging System
90IYOUltrasonic Pulsed Echo Imaging System
  • Diagnostic Ultrasound Transducer 90ITX
    As stated in 21 CFR, parts 892.1550, 892.1560 and 892.1570, each of these generic types of devices have been classified as Class II.

The HD15 is a diagnostic ultrasound device. It consists of a system console containing the power supply and electronic circuitry required to generate the image, a display screen, and a connection to the separate transducers. It is substantially equivalent to the currently marketed HD11 Diagnostic Ultrasound System (K043535 and K062247).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the text "K08166", "pg. 2 of 2", and "HD15 Diagnostic Ultrasound system". The text appears to be handwritten and typed. The text "K08166" and "pg. 2 of 2" are handwritten and are located at the top of the image. The text "HD15 Diagnostic Ultrasound system" is typed and is located at the bottom of the image.

The HD15 system and transducers function in a manner identical to all Philips ultrasound systems and transducers. The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo-electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The differing acoustic properties of the tissues in the body reflect some of the transmitted energy back to the transducer, where it is converted back to electrical signals and sent back to the system. In the system, advanced signal processing technologies convert the returned signals into images of the tissues. The Doppler functions of this system process the Doppler shift frequencies from the echoes of moving targets (such as blood), to detect and graphically display the Doppler shifts of these tissues as flow.

The HD15 is intended for diagnostic ultrasound imaging and fluid flow analysis.

The HD15 is substantially equivalent in safety and effectiveness to the predicates identified above:

  • . Both the predicate device and the HD15 are indicated for the diagnostic ultrasonic imaging and fluid flow analysis.
  • . Both the predicate device and the HD15 have the same gray-scale and Doppler capabilities.
  • Both the predicate device and the HD15 use essentially the same technologies for . imaging, Doppler functions and signal processing.
  • . Both the predicate device and the HD15 have acoustic output levels below the Track 3 FDA limits.
  • . Both the predicate device and the HD15 are manufactured under equivalent quality systems.
  • . Both the predicate device and the HD15 are manufactured of materials with equivalent biosafety. The materials have been evaluated and found to be safe for this application.
  • � Both the predicate device and HD15 are designed and manufactured to the same electrical and physical safety standards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 4 2008

Philips Ultrasound, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K081661

Trade/Device Name: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: June 12, 2008 Received: June 13, 2008

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

3D9-3v 4D Endovaginal BP10-5ec Endocavity C5-2 Curved Linear Array C6-3 Curved Linear (C8-4v) Endovaginal C8-5 Curved Linear

{3}------------------------------------------------

D2cwc Non-image Pencil L12-3 Linear Array L12-5 50mm Linear Array L15-7io Linear Array S5-2 Sector Phased Array S7-2 omni TEE Phased Array S8-3 Sector Phased Array V6-2 Mechanical 3D

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set 1 forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Lauren Hefner at (240) 276-3666.

Sincerely yours,

loni M Whay

lancy C. Brogdon E Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

4.3.2 Indications for Use Tables

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

K081661

Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System Device name: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmicNNNNNNN
FetalNNNNNNN
AbdominalNNNNNNN
Intra-operative(vascular/epicardial)NNNNNNN
Intra-operative (Neuro)NNNNNNN
LaparoscopicNNNNNNN
Fetal Imaging& OtherPediatricNNNNNNN
Small Organ (thyroid,scrotum, breast)NNNNNNN
Neonatal CephalicNNNNNNN
Adult CephalicNNNNNNN
Trans-rectalNNNNNNN
Trans-vaginalNNNNNNN
Trans-urethralNNNNNNN
Trans-esoph. (non-Card.)NNNNNNN
Musculo-skel (conventional)NNNNNNN
Musculo-skel (superficial)NNNNNNN
Intra-luminalNNNNNNN
Other (Gynecological)NNNNNNN
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)NNNNNNN
Other (specify)NNNNNNN
PeripheralVesselPeripheral vesselNNNNNNN
Other (Carotid, I/O)NNNNNNN

= new indication; P= previously cleared by FDA; E=added under Appendix E

*Other modes: Panoramic, Color Power Angio (CPA, formerly Angio), 3D/4D, Harmonics (Tissue and Contrast), Directional Angio Imaging, Tissue Doppler Imaging.

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Signature Sign Off

{6}------------------------------------------------

208166 510(k) Number:

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: 3D9-3v 4D Endovaginal transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalPPPPPP
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Prostate)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPPP
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)
Cardiac Adult
Cardiac Pediatric
CardiacTrans-esoph. (Cardiac)
Other (specify)
PeripheralVesselPeripheral vessel
Other (Specify)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Panoramic, 3D/4D Imaging, Directional Angio Imaging
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color
Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Arque Arthing

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{7}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

1081661

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: BP10-5ec Endocavity transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalPPPPPP
Abdominal
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (scrotum)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPP
Trans-vaginalPPPPPP
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E*Other modes: Color Power Angio, Panoramic, Harmonics (Tissue & Contrast), 3D/4D Imaging, Directional AngioImaging, Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

lonai Th. Whang

Islon of Reproductive, Abdominal, a adiological ! 610(k) N

{8}------------------------------------------------

510(k) Number: System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: C5-2 Curved linear array transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPP
AbdominalPPPPPP
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPP
Small Organ (thyroid,scrotum, breast)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)PPPPPP
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPP
Other (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Panoramic, Harmonics (Tissue & Contrast), 3D/4D Imaging, Directional Angio

Imaging,

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Aomi McWhan

(Division Sign-Off) Division of Reproductive, Abdomina Radiological Devices 510(k) Number

{9}------------------------------------------------

Cardiac

Peripheral

Vessel

510(k) Number:19081641
System:Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System
Transducer:C6-3 Curved linear Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalPPPPPP
AbdominalPPPPPP
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricPPPPPP
Small Organ (thyroid,scrotum, breast)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM 120811000

N= new indication; P= previously cleared by FDA; E= added under Appendix E *Other modes: Color Power Angio, Panoramic, Harmonics (Tissue & Contrast), 3D/4D Imaging, Directional Angio

P

P

P

Imaging, Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

P

P

P

P

P

P

P

P

P

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEBDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Intra-luminal

Other (Gynecological) Cardiac Adult

Cardiac Pediatric Trans-esoph. (Cardiac) Other (Specify)

Peripheral vessel

Other (Carotid, VO)

Aoru Tn Whan

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{10}------------------------------------------------

1081661 510(k) Number:

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: (C8-4v) Endovaginal transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPP
Abdominal
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (thyroid,scrotum, breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPPP
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Panoramic, 3D/4D Imaging, Directional Angio Imaging
Combined modes: Duplex = 2D + Doppler, Triplex = 2D + Doppler + Color
Previous submission: First use on this ultrasound system (See 4 4 1 4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Arzu M. Khan

(Division Sign-Off) ision of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{11}------------------------------------------------

K081661 510(k) Number:

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: C8-5 Curved linear Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPP
AbdominalPPPPPP
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPP
Small Organ (thyroid,scrotum, breast)
Neonatal CephalicPPPPPP
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)PPPPPP
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (specify)
PeripheralVesselPeripheral vesselPPPPPP
Other (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E*Other modes: Color Power Angio, Panoramic, 3D/4D Imaging, Harmonics (Tissue), Directional Angio Imaging,Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler, Triplex = 2D + Doppler + Color, Dual

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Arzne M. Whtz

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{12}------------------------------------------------

510(k) Number:

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: D2cwc Non-Imaging Pencil

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
(Track I Only)(Tracks I & III)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (thyroid,scrotum, breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E*Other modes include: NoneCombined modes: None.Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Sidn-Off Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{13}------------------------------------------------

510(k) Number: System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: L12-3 Linear array transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPP
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricPPPPPP
Small Organ (thyroid,scrotum, breast)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)PPPPPP
Musculo-skel (superficial)PPPPPP
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPP
Other (Carotid, I/O)PPPPPP

previo usiy cle ared by FD. ﺔ, ﻟ *Other modes: Color Power Angio, Panoramics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging

Combined modes: Duplex = 2D + Doppler, Triplex = 2D + Doppler + Color Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Tori In Whang

(Division ar-Off Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number

{14}------------------------------------------------

K081661 510(k) Number:

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: L12-5 50mm Linear array transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
(Track I Only)(Tracks I & III)
OphthalmicOphthalmic
Fetal
AbdominalPPPPPP
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Fetal ImagingPediatricPPPPPP
& OtherSmall Organ (thyroid,scrotum, breast)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)PPPPPP
Musculo-skel (superficial)PPPPPP
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralPeripheral vesselPPPPPP
VesselOther (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Imaging, Directional AngioImaging, Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Aoyce M. Wher

Division of Reproductive, Abdomina Radiolo 510(k) Numb

{15}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

K08166

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

L15-7io Linear array transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColor DopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative(vascular/epicardial)PPPPPP
Intra-operative (Neuro)PPPPPP
Laparoscopic
Fetal Imaging& OtherPediatricPPPPPP
Small Organ (thyroid,scrotum, breast)PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)PPPPPP
Musculo-skel (superficial)PPPPPP
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPP
Other (Carotid, I/O)PPPPPP
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Amplitude Doppler, Panoramic, Directional Angio Imaging
Combined modes: Duplex = 2D + Doppler, Triplex = 2D + Doppler + Color
Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Lorne M. Why

(Division Sign (Division of Sign-Ull)
Division of Reproductive, Abdominal, and Radiological De 510(k) Number

Page 24

{16}------------------------------------------------

510(k) Number:

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: S5-2 Sector phased array transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmicPPPPPPP
FetalPPPPPPP
AbdominalPPPPPPP
Intra-operative(vascular/epicardial)PPPPPPP
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricPPPPPPP
Small Organ (thyroid,scrotum, breast)
Neonatal Cephalic
Adult CephalicPPPPPPP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)PPPPPPP
Cardiac AdultPPPPPPP
CardiacCardiac PediatricPPPPPPP
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselPPPPPPP
Other (Carotid, I/O)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: Color Power Angio, Panoramic, Harmonics (Tissue & Contrast), 3D/4D Imaging, Directional Angio Imaging, Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

tons mhm

(Division S Division of Reproductive, Abdominal, and Radiological Dev 510(k) Number

{17}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

Koflidel

System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: S7-2 omni TEE phased array transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (thyroid,scrotum, breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)
CardiacCardiac AdultPPPPPPP
Cardiac PediatricPPPPPPP
Trans-esoph. (Cardiac)PPPPPPP
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Tissue Doppler Imaging, Directional Angio Imaging,
Combined modes: Duplex = 2D + Doppler

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Hose M. Whing
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{18}------------------------------------------------

510(k) Number: System: Philips Ultrasound, Inc. HD15 Diagnostic Ultrasound System

Transducer: S8-3 Sector phased array transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalPPPPPP
AbdominalPPPPPPP
Intra-operative(vascular/epicardial)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricPPPPPPP
Small Organ (thyroid,scrotum, breast)
Neonatal CephalicPPPPPPP
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)PPPPPPP
Cardiac AdultPPPPPPP
CardiacCardiac PediatricPPPPPPP
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralPeripheral vesselPPPPPPP
VesselOther (Carotid, I/O)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
* Other modes: Color Power Angio, Panoramic, 3-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color

Previous submission: First use on this ultrasound system (See 4.4.1.4)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

Horge M. Whan

(Division Sidn-Off) Division of Reproductive, Abdominal, a Radiological Devices 510(k) Number

{19}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound, Inc. Pathfinder Diagnostic Ultrasound System Transducer: V6-2 Mechanical 3D Transducer

Clinical ApplicationIntended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:Mode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalPPPPPP
AbdominalPPPPPP
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricPPPPPP
Small Organ (Scrotum,
Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel (conventional)
Musculo-skel (superficial)
Intra-luminal
Other (Gynecological)PPPPPP
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (specify)
PeripheralVesselPeripheral vessel
Other (Carotid)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

VPrescription Use (Per 21 CFR 801.109)

Aonny M. Whing

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.