(13 days)
This device is a digital x-ray system configured for use as a diagnostic x-ray angiography system. This x-ray angiography system is used for diagnostic and interventional procedures for cardiac blood vessels, cerebral blood vessels, abdominal blood vessels and lower extremities.
This device is a digital radiography/fluoroscopy system used in a diagnostic angiography configuration. This system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
These systems are single or dual plane systems that employ x-rays to obtain fluoroscopy or radiography images of the body. The system is comprised of a computer system, support arms that contain the tube and solid state detector (either one or two), x-ray generator and a patient table.
This 510(k) submission (K081624) for the Toshiba America Medical Systems, Inc. INFX-8000F (Infinix CSi and Infinix DPi) is a model number change for an electrostatic fluoroscopic x-ray system. The 510(k) summary explicitly states "Model Number Change" as the reason for submission and refers to existing predicate devices (GE Innova 4100; 510(k) Control # K052412 and Toshiba DFP-8000D/FPD; 510(k) Control # K052884).
When the reason for submission is solely a model number change, or a change that does not introduce new technological characteristics or indications for use, the FDA's typically doesn't require new performance studies. Instead, the manufacturer demonstrates substantial equivalence by showing the device employs similar materials and processes as the predicate device(s) and complies with the same or equivalent standards.
Therefore, the provided document does not contain the acceptance criteria or a study demonstrating device performance against such criteria because it is a model change submission. The focus is on demonstrating that the new model number does not alter the fundamental safety or effectiveness characteristics established by the predicate devices.
Here's a breakdown based on the information provided, explicitly noting what is not applicable (N/A) due to the nature of the submission:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| N/A | N/A |
| Explanation: This submission is a model number change; therefore, specific performance acceptance criteria and associated study results for the INFX-8000F are not presented in this document. The device relies on the established performance of its predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size (Test Set): N/A
- Data Provenance: N/A
Explanation: No new performance testing in humans was performed for this model change submission. The device's safety and effectiveness are established by its substantial equivalence to predicate devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Number of Experts: N/A
- Qualifications of Experts: N/A
Explanation: No new clinical studies requiring expert ground truth were conducted for this model change.
4. Adjudication Method for the Test Set:
- Adjudication Method: N/A
Explanation: No new clinical studies requiring adjudication were conducted for this model change.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- MRMC Study Done: No
- Effect Size of Human Reader Improvement: N/A
Explanation: This submission does not involve a multi-reader multi-case study; it's a model number change, not a new device with software requiring such evaluation.
6. Standalone (Algorithm Only) Performance Study:
- Standalone Study Done: No
Explanation: This is a hardware-based x-ray system, not an AI/algorithm-only device, and its submission is for a model number change.
7. Type of Ground Truth Used:
- Type of Ground Truth: N/A
Explanation: No new clinical studies requiring ground truth were conducted for this model change.
8. Sample Size for the Training Set:
- Sample Size (Training Set): N/A
Explanation: As a hardware system with a model number change, there's no "training set" in the context of an algorithm's development.
9. How the Ground Truth for the Training Set Was Established:
- Method for Establishing Ground Truth: N/A
Explanation: Not applicable for a hardware device model number change.
Summary of Device Rationale for Acceptance:
The acceptance of the INFX-8000F is based on its substantial equivalence to legally marketed predicate devices:
The submission states, "The INFX-8000V is of comparable type and substantially equivalent to..." and "Therefore the INFX-8000V complies with the same or equivalent standards and has the same intended use as the predicate device." This implies that the safety and effectiveness for the INFX-8000F are effectively "inherited" from the performance demonstrated by these predicate devices, which would have undergone their own rigorous testing and review processes when they were initially cleared. The change to the device is specifically noted as a "Model Number Change," indicating no fundamental alterations in the technology, intended use, or safety/effectiveness profile compared to the predicate.
{0}------------------------------------------------
K081624
pg. 1 of 2
Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) for INFX-8000F
510(k) Summary
| JUN 28 2008 | |
|---|---|
| Date: | May 29, 2008 |
| Submitter's Name: | Toshiba America Medical Systems, Inc. |
| Submitter's Address: | P.O. Box 2068, 2441 Michelle Drive,Tustin, CA 92781-2068 |
| Submitter's Contact: | Paul Biggins, Director Regulatory Affairs(714)730-5000 |
| Establishment RegistrationNumber: | 2020563 |
| Device Proprietary Name: | INFX-8000F (Infinix CSi and Infinix DPi) |
| Common Name: | Electrostatic Fluoroscopic x-ray System)[Fed. Reg. No. 892.1650, Pro. Code: 90MQB] |
| Regulatory Class: | II (per 21 CFR 892.1650) |
| Performance Standard: | 21 CFR Subchapter J,Federal Diagnostic X-ray Equipment Standard |
| Predicate Device(s): | GE Innova 4100; 510(k) Control # K052412Toshiba DFP-8000D/FPD; 510(k) Control # K052884 |
| Reason for Submission | Model Number Change |
Description of this Device:
These systems are single or dual plane systems that employ x-rays to obtain fluoroscopy or radiography images of the body. The system is comprised of a computer system, support arms that contain the tube and solid state detector (either one or two), x-ray generator and a patient table.
Summary of Intended Uses:
This device is a digital x-ray system configured for use as a diagnostic x-ray angiography system. This x-ray angiography system is used for diagnostic and interventional procedures for cardiac blood vessels, cerebral blood vessels, abdominal blood vessels and lower extremities.
{1}------------------------------------------------
Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) for INFX-8000F
Image /page/1/Picture/1 description: The image shows handwritten text on a white background. The text "K081624" is at the top, followed by "pg. 2 of 2" on the next line. The handwriting is in black ink and appears to be somewhat bold.
Technological Characteristics:
This device employs similar materials and processes as found in the predicate device. The device produces ionizing radiation that is employed to generate fluoroscopic and radiographic images of the anatomy.
Safety and Effectiveness Concerns:
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR § 1020,that apply to this device, will be met and reported via an initial report. Additionally this system is in conformance with the applicable parts of the IEC 60601-1 {applicable portions}; IEC 60601-2-32, and IEC 60601-2-28. - Medical Device Safety standards.
Substantial Equivalence:
The INFX-8000V is of comparable type and substantially equivalent to:
GE Innova 4100; 510(k) Control # K052412 Toshiba DFP-8000D/FPD; 510(k) Control # K052884
Therefore the INFX-8000V complies with the same or equivalent standards and has the same intended use as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure, with three distinct lines forming the body and wings.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
AUG 21 2013
Re: K081624
Trade/Device Name: INFX-8000F; Infinix-CSi and Infinix-DPf Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA and MOB Dated: June 9, 2008 Received: June 10, 2008
Dear Mr. Job:
This letter corrects our substantially equivalent letter of June 23, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Morri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) for INFX-8000F
Indications for Use
510(k) Number (if known): | (08) (0)
Device Name: INFX-8000F; Infinix-CSi and Infinix- DPi
Indications for Use:
This device is a digital radiography/fluoroscopy system used in a diagnostic angiography configuration. This system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arzni Ta Whang
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 1 of _
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.