(243 days)
No
The device description focuses on the mechanical and electrical components of a standard diagnostic catheter and explicitly states "No new technology or circuitry is associated with the transmission of electrical signals". There is no mention of AI or ML in the provided text.
No
The device is described as a "diagnostic catheter" used for "recording of electrical potentials" and "delivering pacing stimuli from an external source" for "electrophysiology studies," which points to diagnostic and temporary pacing functions rather than long-term therapeutic treatment.
Yes
The "Intended Use / Indications for Use" section explicitly states, "The Blazer Dx-20 is a diagnostic catheter indicated for temporary use in electrophysiology studies for intracardiac stimulation (pacing) and/or recording of electrical potentials." The "Device Description" also mentions it is a "diagnostic electrophysiology catheter used to record intracardiac electrical potentials." Both statements confirm its diagnostic purpose.
No
The device description clearly details a physical catheter with electrodes, a steerable shaft, and a handle, indicating it is a hardware medical device, not software-only.
No, the Blazer Dx-20 is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices intended for use in vitro (outside the body) for the examination of specimens derived from the human body to provide information for diagnostic purposes. This includes things like blood tests, urine tests, and tissue biopsies analyzed in a lab.
- Blazer Dx-20 Function: The description clearly states the Blazer Dx-20 is a diagnostic catheter used for intracardiac stimulation (pacing) and/or recording of electrical potentials within the heart. This is an in vivo (within the body) procedure.
- Mechanism of Action: The device works by making electrical contact with the endocardium to record and deliver electrical signals directly within the heart. It does not analyze specimens outside the body.
Therefore, based on the provided information, the Blazer Dx-20 is an in vivo diagnostic device, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The Blazer Dx-20 Catheter is intended for temporary use in electrophysiology studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
The Blazer Dx-20 is a diagnostic catheter indicated for temporary use in electrophysiology studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
Product codes
DRF
Device Description
The Boston Scientific Corporation Blazer Dx-20 is a sterile, single use steerable diagnostic electrophysiology catheter used to record intracardiac electrical potentials. The Blazer Dx-20 Catheter is also used to deliver pacing stimuli from an external source. The Blazer Dx-20 Catheter is a bi-directionally steerable diagnostic catheter built on a modified SteeroCath Dx shaft platform with a molded handle. It has one tip electrode and up to 19 ring electrodes for a total of 20. The ring spacing configurations vary with each model.
The catheter shaft of the Blazer Dx-20 is equivalent to the SteeroCath Dx Catheter, which is designed to allow for a standard curve of the catheter distal section. The catheter shaft has a soft distal section in order to minimize trauma and risk of perforation.
The Blazer Dx-20 Catheters use a thumb-actuated bidirectional steering mechanism. contained within an ergonomically shaped cylindrical handle that is also utilized by the currently marketed SteeroCath Dx Diagnostic catheters. A thumb-level motion of the piston actuates the steering of the distal tip. The catheter is placed into the heart and is guided to location by steering the distal tip area of the catheter. Tip and ring electrodes come into contact with the endocardium where electrical contact is made and pacing and recording of electrograms becomes possible. No new technology or circuitry is associated with the transmission of electrical signals to or from the endocardium - the Blazer Dx-20 Catheter relies on platinum-iridium alloy, ring electrodes whose circuitry is identical to standard electrode and pacing catheters. Additionally, the electrical connections made (catheter to EP recorder) are similar to those for commercially available electrode recording and pacing v catheters.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
intracardiac
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
As the non-clinical tests demonstrated the safety and effectiveness of the device, no clinical studies were conducted for the Blazer Dx-20 Catheter.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1220 Electrode recording catheter or electrode recording probe.
(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).
0
FEB -3 2009
510(K) SUBMISSION
Blazer Dx-20 Steerable Diagnostic Catheter
5.0 510(k) Summary
Modified Device Information
Table 1: Modified Device Information
Category: | Comments: |
---|---|
Sponsor: | Boston Scientific Corporation |
150 Baytech Drive | |
San Jose, CA 95134 | |
Correspondent: | Ronald C. Allen, Ph.D. |
Director, Regulatory Affairs | |
150 Baytech Drive | |
San Jose, CA 95134 | |
Contact Information: | Email: allenr@bsci.com |
Phone: (408) 935-6310 | |
Fax: (408) 957-6202 | |
Device Common Name: | Electrode recording and pacing catheter; steerable catheter |
Device Proprietary Name: | Blazer Dx-20 Diagnostic Catheter |
Device Classification Number: | 21 CFR §870.1220 |
Device Classification | Class II |
Predicate Device Information
Table 2: Predicate Device Information
Category: | Comments: |
---|---|
Predicate Device: | SteeroCath Dx Diagnostic Catheter (K940168/ |
K913375) | |
Predicate Device Manufacturer: | Boston Scientific Corporation |
Predicate Device Common Name: | Electrode recording and pacing catheter; steerable |
catheter | |
Predicate Device Classification | |
Number: | 21 CFR §870.1220 |
Predicate Device Classification: | Class II |
Date Summary Was Prepared
June 4, 2008
Boston Scientific Corporation 510(k) Submission
CONFIDENTIAL
1
Description of the Device
The Boston Scientific Corporation Blazer Dx-20 is a sterile, single use steerable diagnostic electrophysiology catheter used to record intracardiac electrical potentials. The Blazer Dx-20 Catheter is also used to deliver pacing stimuli from an external source. The Blazer Dx-20 Catheter is a bi-directionally steerable diagnostic catheter built on a modified SteeroCath Dx shaft platform with a molded handle. It has one tip electrode and up to 19 ring electrodes for a total of 20. The ring spacing configurations vary with each model.
The catheter shaft of the Blazer Dx-20 is equivalent to the SteeroCath Dx Catheter, which is designed to allow for a standard curve of the catheter distal section. The catheter shaft has a soft distal section in order to minimize trauma and risk of perforation.
The Blazer Dx-20 Catheters use a thumb-actuated bidirectional steering mechanism. contained within an ergonomically shaped cylindrical handle that is also utilized by the currently marketed SteeroCath Dx Diagnostic catheters. A thumb-level motion of the piston actuates the steering of the distal tip. The catheter is placed into the heart and is guided to location by steering the distal tip area of the catheter. Tip and ring electrodes come into contact with the endocardium where electrical contact is made and pacing and recording of electrograms becomes possible. No new technology or circuitry is associated with the transmission of electrical signals to or from the endocardium - the Blazer Dx-20 Catheter relies on platinum-iridium alloy, ring electrodes whose circuitry is identical to standard electrode and pacing catheters. Additionally, the electrical connections made (catheter to EP recorder) are similar to those for commercially available electrode recording and pacing v catheters.
Intended Use
The Blazer Dx-20 Catheter is intended for temporary use in electrophysiology studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
Boston Scientific Corporation 510(k) Submission
CONFIDENTIAL
2
510(K) SUBMISSION
Blazer Dx-20 Steerable Diagnostic Catheter
Comparison to Predicate Device
Table 3: Comparison to Predicate Device
Predicate Device | Subject Device | |
---|---|---|
Manufacturer | Boston Scientific Corporation | Same |
Device | Electrode Recording Catheter; | |
Steerable Catheter | Same | |
510(k) Reference | K940168 | |
K913375 | Current Submission | |
Regulatory Class | II | Same |
Device | ||
Classification | 21 CFR §870.1220 | Same |
Intended Use | Record electrical potentials and | |
pacing from intracardiac locations | Same | |
Single Use? | Yes | Same |
EO Sterilized? | Yes | Same |
Summary of the Non-clinical Data
Specifically, non-clinical tests adopted by and conducted for the Blazer Dx-20 Catheter included biocompatibility, sterility, packaging, physical integrity, and electrical integrity testing that all passed and have shown substantial equivalence to the predicate device, the SteeroCath Diagnostic Catheter.
Abstract of the Clinical Data
As the non-clinical tests demonstrated the safety and effectiveness of the device, no clinical studies were conducted for the Blazer Dx-20 Catheter.
Boston Scientific Corporation 510(k) Submission
CONFIDENTIAL
3
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB -3 2009
Boston Scientific Corporation c/o Ms. April Malmborg Principal Regulatory Affairs Specialist 150 Baytech Drive San Jose, CA 95134
Re: K081576
Trade/Device Name: Blazer Dx-20 Diagnostic Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II (two) Product Code: DRF Dated: December 19, 2008 Received: December 22, 2008
Dear Ms. Malmborg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
4
Page 2 - Ms. April Malmborg
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
R. be himes
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number (if known): K081576
Device Name:
Blazer Dx-20 Diagnostic Catheter
Indications For Use:
The Blazer Dx-20 is a diagnostic catheter indicated for temporary use in electrophysiology studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
une 2. J. James
Division Sign Off
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K081576
Page 1 of __ 1
રેણ