(118 days)
The HAMILTON-G5 ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients, and optionally to infant and neonatal patients. The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care, including use as a patient bedside for intra-facility transport, provided compressed gas is supplied. The device is not intended for transportation outside the hospital or for use in the home environment.
The HAMILTON-G5 is an electronically controlled pneumatic intensive care ventilator ventilation system. It uses oxygen and air or (optionally) heliox to ventilate adults, pediatrics, and optionally infants and neonates. It is powered by ac with battery backup to protect against power failure or unstable power and to facilitate intrahospital transport. The HAMILTON-G5's pneumatics deliver gas, and its electrical systems control pneumatics, monitor alarms, and distribute power. The user interface consists of a LCD-display with touch screen, keys, and a press-andturn knob. The CO2 option allows continuous mainstream monitoring of carbon dioxide.
The provided text describes a 510(k) summary for the HAMILTON-G5 continuous ventilator. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving performance against specific acceptance criteria in a clinical study. Therefore, much of the requested information regarding acceptance criteria and study details (like sample size, number of experts, adjudication methods, MRMC studies, standalone performance, and training set details) is not directly available or applicable in this document because the submission relies on non-clinical performance and equivalence.
However, I can extract the information that is present and explain why other details are missing based on the nature of a 510(k) submission.
Here's a breakdown of the available and missing information:
1. Table of Acceptance Criteria and Reported Device Performance
| Criterion/Test | "Acceptance Criteria" (Implied) | Reported Device Performance |
|---|---|---|
| Safety | Compliance with relevant IEC standards | "The test results show that the device is safe and effective for its intended use." (Based on IEC60601-1, IEC60601-1-2, IEC 60601-2-12 and other applicable standards) |
| Wave-form Performance | Substantial equivalence to a legally marketed device (as described in ASTM F1100-90) | "The data provided from these tests, were shown to be substantially equivalent to a legally marketed device." |
| Software Verification & Validation | All specified requirements implemented correctly and completely | "The results of the software verification and validation testing demonstrate that all specified requirements have been implemented correctly and completely." |
| Overall Equivalence | As safe, as effective, and performs as well as or better than legally marketed predicate devices | "The results of verification, validation, and testing activities demonstrate that the HAMILTON-G5 ventilator including the new options is as safe, as effective, and performs as well as or better than the legally marketed devices identified above." |
| Indended Use | Comparable to predicate devices | "The indication statements for the HAMILTON-G5 ventilator are comparable to those for the predicate devices." |
| Technological Characteristics & Performance Specifications | Substantially equivalent to predicate devices | "Technological characteristics and performance specifications of the HAMILTON-G5 ventilator are substantially equivalent to those of the predicate devices." |
| CO2 Option | Substantially equivalent to CO2 gas analyzer of predicate device Servo-i | "The Carbon dioxide gas analyzer is considered to be substantial equivalent to the CO2 gas analyzer of the currently marketed predicate device Servo-i." |
| Heliox Option | Comparable to option of predicate device AVEA ventilator | "The heliox option is comparable with the substantial equivalent option of the predicate device AVEA ventilator." |
| Neonate Ventilation Characteristics | Substantially equivalent to predicate devices Servo-i, Engström Carestation, and Evita XL | "The neonate ventilation characteristics are substantially equivalent to those of the predicate devices Servo-i, Engström Carestation, and Evita XL." |
2. Sample size used for the test set and the data provenance
- Sample Size: Not applicable/not provided. The submission refers to non-clinical performance tests (safety, waveform, software) rather than a clinical test set with patient data.
- Data Provenance: Not applicable. The tests appear to be laboratory-based and engineering evaluations, not involving patient data from specific countries or retrospective/prospective studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable/not provided. Ground truth establishment with experts is typical for diagnostic algorithms, not for physical medical device performance testing in this context. The "ground truth" here is adherence to engineering standards and functional specifications.
4. Adjudication method for the test set
- Not applicable/not provided. Adjudication methods are typically associated with expert consensus in clinical studies, which are not described here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done
- No, an MRMC comparative effectiveness study was not done. This type of study is common for diagnostic imaging AI, to evaluate how AI assists human readers. This submission focuses on the safety and performance of a physical ventilator.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The HAMILTON-G5 is a medical device (ventilator), not an algorithm in the AI sense. Its performance is evaluated through engineering tests and comparison to predicate devices, which can be considered "standalone" in that it performs its function without human intervention for the measurement of that function, but it's not an "algorithm-only" performance in the way an AI would be evaluated.
7. The type of ground truth used
- The "ground truth" for this ventilator device primarily constitutes:
- Compliance with International Standards: IEC60601-1, IEC60601-1-2, IEC 60601-2-12, ASTM F1100-90.
- Predicate Device Performance: The established safety and effectiveness of the legally marketed predicate devices (HAMILTON-G5 K070513, MAQUET Servo-i K041223, Viasys AVEA ventilator K062093, Datex-Ohmeda Engström Carestation K062710, Dräger EvitaXL with NeoFlow Option K051263, K983219).
- Software Requirements: The specified requirements for the device's software.
8. The sample size for the training set
- Not applicable/not provided. The document describes a medical device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
- Not applicable/not provided for the same reason as point 8.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Hamilton Medical. The logo consists of the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a circle with the letter "H" inside.
Koris21
SEP 2 5 2008
510(k) Summary
| SUBMITTER: | HAMILTON MEDICAL AGVia Crusch 87402 Bonaduz, Switzerland |
|---|---|
| CONTACT PERSON: | Joerg SchneiderPhone: +41 81 660 6479Fax: +41 81 660 60 20e-mail: jschneider@hamilton-medical.ch |
| PREPARATION DATE: | May 21, 2008 |
| TRADE NAME: | HAMILTON-G5 |
| COMMON NAME: | Continuous Ventilator |
| CLASSIFICATION NAME: | Ventilator, Continuous, Facility Use(21 CFR 868.5895, Product Code: CBK) |
| LEGALLY MARKETEDDEVICES TO WHICHEQUIVALENCE ISBEING CLAIMED: | HAMILTON-G5510(k) Number: K070513 |
| MAQUET Servo-i510(k) Number: K041223 | |
| Viasys AVEA ventilator510(k) Number: K062093 | |
| Datex-Ohmeda Engström Carestation510(k) Number: K062710 | |
| Dräger EvitaXL with NeoFlow Option510(k) Numbers: K051263, K983219 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is composed of two lines of text, with "HAMILTON" on the top line and "MEDICAL" on the bottom line. To the right of the word "HAMILTON" is a circle with the letter "H" inside of it.
DEVICE DESCRIPTION
The HAMILTON-G5 is an electronically controlled pneumatic intensive care ventilator ventilation system.
It uses oxygen and air or (optionally) heliox to ventilate adults, pediatrics, and optionally infants and neonates. It is powered by ac with battery backup to protect against power failure or unstable power and to facilitate intrahospital transport.
The HAMILTON-G5's pneumatics deliver gas, and its electrical systems control pneumatics, monitor alarms, and distribute power.
The user interface consists of a LCD-display with touch screen, keys, and a press-andturn knob.
The CO2 option allows continuous mainstream monitoring of carbon dioxide.
INTENDED USE
The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients.
The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care.
The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician.
The HAMILTON-G5 ventilator may be used for transport within a hospital or hospitaltype facility provided compressed gas is supplied. The device is not to be used in the presence of flammable anesthetic agents or other ignition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment. The device is not intended for transportation outside the hospital or for use in the home environment.
In the USA, federal law restricts this device to sale by or on the order of a physician.
SUMMARY OF THE TECHNOLOGY AND PERFORMANCE SPECIFICATIONS COMPARISON WITH THE PREDICATED DEVICES
The indication statements for the HAMILTON-G5 ventilator are comparable to those for the predicate devices.
Technological characteristics and performance specifications of the HAMILTON-G5 ventilator are substantially equivalent to those of the predicate devices.
The Carbon dioxide gas analyzer is considered to be substantial equivalent to the CO2 gas analyzer of the currently marketed predicate device Servo-i.
The heliox option is comparable with the substantial equivalent option of the predicate device AVEA ventilator.
The neonate ventilation characteristics are substantially equivalent to those of the predicate devices Servo-i, Engström Carestation, and Evita XL.
HAMILTON MEDICAL has demonstrated the HAMILTON-G5 ventilator including the new options to be safe and effective.
The HAMILTON-G5 is considered to be substantial equivalent to currently marketed predicate devices which have been previously cleared by FDA.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the words "HAMILTON" and "MEDICAL" stacked on top of each other. The word "HAMILTON" is in a bold, sans-serif font, and the word "MEDICAL" is in a similar font but slightly smaller. To the right of the word "HAMILTON" is a logo that appears to be a stylized "H" inside a circle. The text and logo are all in black, and the background is white.
NON-CLINICAL PERFORMANCE TESTS
Safety testing of the HAMILTON-G5 with the new options was conducted according to IEC60601-1, IEC60601-1-2, IEC 60601-2-12 and other applicable standards. The test results show that the device is safe and effective for its intended use.
The ventilator was further subject to wave-form performance testing as described in the standard ASTM F1100-90. The data provided from these tests, were shown to be substantially equivalent to a legally marketed device.
The results of the software verification and validation testing demonstrate that all specified requirements have been implemented correctly and completely.
CONCLUSION
The results of verification, validation, and testing activities demonstrate that the HAMILTON-G5 ventilator including the new options is as safe, as effective, and performs as well as or better than the legally marketed devices identified above.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract image of an eagle.
SEP 2 5 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joerg Schneider Quality Engineer, R&D Hamilton Medical AG Via Crusch 8 CH-7402 Bonaduz SWITZERLAND
Re: K081521
Trade/Device Name: HAMILTON-G5 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: September 12, 2008 Received: September 17, 2008
Dear Mr. Schneider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
క్ష
ﻧﻪ ..
{4}------------------------------------------------
Page 2 - Mr. Schneider
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrio's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Thlmulrfled mrs frey
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE STATEMENT
| 510(k) Number: | |
|---|---|
| Device Name: | HAMILTON-G5 |
| Indication for Use: | The HAMILTON-G5 ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients, and optionally to infant and neonatal patients. The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care, including use as a patient bedside for intra-facility transport, provided compressed gas is supplied. The device is not intended for transportation outside the hospital or for use in the home environment. |
X Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
MHh
(Division Sign-Off) (Division of Anesthesiology, General Hospital Division Control, Dental Devices
510(k) Number: _
§ 868.5895 Continuous ventilator.
(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).