(243 days)
The DRI Methadone Metabolite (100/300) Assay is intended for the qualitative and semi-quantitative determination of the presence of Methadone Metabolite (2-ethylidene-1,5-dighenyl-3,3-diphenylpyrrolidine or EDDP) in human urine at cutoffs of 100 and 300 ng/mL.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography / Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. Tests for methadone metabolite cannot distinguish between abused drugs and certain prescribed medications. Certain foods or medications may interfere with test for methadone metabolite and cause false positive results.
The DRI Methadone Metabolite Calibrators are intended for use in calibration of the DRI Methadone Metabolite (100/300) Assay.
The DRI Methadone Metabolite Controls are intended for use in the DRI Methadone Metabolite (100/300) Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects.
The DRI Methadone Metabolite (100/300) Assay utilizes liquid ready-to-use reagents. The Antibody/Substrate Reagent (R1) contains mouse monoclonal anti-EDDP antibody, glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as a preservative. The Conjugate Reagent (R2) contains EDDP-derivative labeled with glucose-6-phosphate Enzyme dehydrogenase (G6PDH) in Tris buffer with sodium azide as a preservative.
The assay uses specific antibodies that can detect EDDP in human urine without cross-reactivity to the parent drug methadone. The assay is based on competition between drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) and free drug from the sample for a fixed number of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of free drug from the sample, the specific antibody binds the drug labeled with G6PDH and causes a decrease in enzyme activity. This phenomenon creates a direct relationship between drug concentration in urine and enzyme activity. This enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.
Here's a detailed breakdown of the acceptance criteria and study information for the DRI® Methadone Metabolite (100/300) Assay, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Sensitivity (LOQ) | Not explicitly stated as acceptance criteria, but a performance metric. | 14 ng/mL |
| Precision (Qualitative) | All samples spiked below cutoff read negative; all samples spiked above cutoff read positive. | All samples met criteria. |
| Precision (Semi-Qualitative) | Target within-run precision: ≤ 8% CV; Target total-run precision: ≤ 10% CV. | Within-run and total-run precision at all levels of both 100 ng/mL and 300 ng/mL cutoffs met target specifications. |
| Cutoff Characterization (Qualitative) | Negative controls recover < cutoff calibrators; positive controls recover > cutoff calibrators. No overlap between cutoff levels and ± 25% levels with 95% statistical confidence. Precision of 21 replicates < 1% CV. | Controls recovered accurately; negative controls < cutoff calibrators, positive controls > cutoff calibrators. No overlap observed. Precision of 21 replicates was < 1% CV. |
| Cutoff Characterization (Semi-Qualitative) | Recovery < 15% error of nominal values; precision of 21 replicates < 3% CV. | Recovery was < 15% error of nominal values. Precision of 21 replicates was < 3% CV. |
| Linearity (Dilution Recovery) | Recovery within 10% of expected values; correlation coefficient r = 0.9990. | Recovery within 10% of expected values. Correlation coefficient r = 0.9990. |
| Interferences | No significant interference from endogenous and exogenous urine substances at tested concentrations and pH range of 4 to 11. | No significant interference observed from endogenous and exogenous urine substances at the tested concentrations and pH range of 4 to 11. |
| Specificity (Cross-reactivity) | All tested compounds negative (rate below 100 ng/mL cutoff rate; dose recovered < 100 ng/mL). | All compounds tested negative, with rates below the 100 ng/mL cutoff rate and with dose recovered less than 100 ng/mL, indicating no cross-reactivity at tested concentrations. |
| Method Comparison (Qualitative, 100 ng/mL cutoff) | Overall concordance with predicate CEDIA Assay: Not explicitly stated, but high concordance expected for substantial equivalence. Overall concordance with GC/MS: Not explicitly stated, but high concordance expected. | DRI vs. CEDIA: 95% DRI vs. GC/MS: 95% |
| Method Comparison (Qualitative, 300 ng/mL cutoff) | Overall concordance with GC/MS: Not explicitly stated, but high concordance expected. | DRI vs. GC/MS: 99% |
| Method Comparison (Semi-Qualitative, 100 ng/mL cutoff) | Overall concordance with predicate CEDIA Assay: Not explicitly stated. Overall concordance with GC/MS: Not explicitly stated. | DRI vs. CEDIA: 99% DRI vs. GC/MS: 99% |
| Method Comparison (Semi-Qualitative, 300 ng/mL cutoff) | Overall concordance with GC/MS: Not explicitly stated. | DRI vs. GC/MS: 100% |
2. Sample Size and Data Provenance
- Test Set Sample Size: For the "Method Comparison" study, "A total of 100 unaltered patient urine samples" were used.
- Data Provenance: The data provenance is not explicitly stated (e.g., country of origin). The samples were "unaltered patient urine samples," suggesting they were naturally occurring specimens. It's not specified if they were retrospective or prospective, but the phrasing implies collection from patients.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Not applicable. The ground truth was established by an objective analytical method (GC/MS) rather than expert consensus on subjective interpretations.
- Qualifications of Experts: N/A.
4. Adjudication Method
- Adjudication Method: Not applicable. The ground truth was established by Gas Chromatography / Mass Spectrometry (GC/MS), an objective laboratory method, not human interpretation that would require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done? No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in-vitro diagnostic (IVD) assay designed for laboratory analysis, not for interpretation by human readers in a clinical imaging context, for example.
- Effect Size: Not applicable.
6. Standalone Performance Study
- Standalone Study Done? Yes, the "Method Comparison" study evaluated the standalone performance of the DRI Methadone Metabolite Assay by comparing its results directly against the Gas Chromatography / Mass Spectrometry (GC/MS) reference method. Other studies like "Precision," "Cutoff Characterization," "Linearity," "Interferences," and "Specificity" also demonstrate standalone performance characteristics.
7. Type of Ground Truth Used
- Type of Ground Truth: Gas Chromatography / Mass Spectrometry (GC/MS) was used as the reference analytical method to establish the ground truth for the method comparison study. This is an objective chemical method considered the "preferred confirmatory method" for drug testing.
8. Sample Size for the Training Set
- Training Set Sample Size: The document does not specify a separate "training set" sample size. For an IVD like this, "training" typically refers to the development and optimization of the assay's reagents and parameters rather than a data-driven machine learning model. The provided data focuses on validation and performance evaluation.
9. How the Ground Truth for the Training Set was Established
- Ground Truth for Training Set: Not applicable in the context of a machine learning training set. The development of the assay reagents and conditions would be based on chemical principles and optimization to achieve the desired sensitivity and specificity to EDDP in urine matrices. This involves internal studies and experimentation, but not a distinct "training set with ground truth" in the same way as an AI algorithm.
{0}------------------------------------------------
510K SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is: K081378
COMPANY/CONTACT PERSON
Thermo Fisher Scientific Microgenics Corporation 46360 Fremont Blvd. Fremont, CA 94538
Establishment registration No: 2937369
Jack Rogers Manager of Regulatory Affairs Telephone: (317) 610-3823 Fax: (317) 610-0018
DATE PREPARED
December 3, 2008
DEVICE NAME
| Trade Name: | DRI® Methadone Metabolite (100/300) Assay |
|---|---|
| DRI® Methadone Metabolite Urine Calibrators | |
| DRI® Methadone Metabolite Urine Controls | |
| Common Name: | Methadone Test System |
| Device Classification: | 21 CFR 862.3620 Methadone Test System; Class II |
| 21 CFR 862.3200 Clinical Toxicology Calibrator; Class II | |
| 21 CFR 862.3280 Clinical Toxicology Control Material; Class II |
INTENDED USE
The DRI Methadone Metabolite (100/300) Assay is intended for the qualitative and semi-quantitative determination of the presence of Methadone Metabolite (2-ethylidene-1,5-dighenyl-3,3-diphenylpyrrolidine or EDDP) in human urine at cutoffs of 100 and 300 ng/mL.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography / Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. Tests for methadone metabolite cannot distinguish between abused drugs and certain prescribed medications. Certain foods or medications may interfere with test for methadone metabolite and cause false positive results.
The DRI Methadone Metabolite Calibrators are intended for use in calibration of the DRI Methadone Metabolite (100/300) Assay.
The DRI Methadone Metabolite Controls are intended for use in the DRI Methadone Metabolite (100/300) Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects.
JAN 1 4 2009
{1}------------------------------------------------
LEGALLY MARKETED DEVICE TO WHICH EQUIVALENCY IS CLAIMED
The DRI Methadone Metabolite (100/300) Assay is substantially equivalent to the previously cleared CEDIA DAU EDDP Assay (K980746, Microgenics Corporation).
The DRI Methadone Metabolite Calibrators are substantially equivalent to the previously cleared CEDIA DAU Multi-Drug Calibrators (K980853, Microgenics Corporation).
The DRI Methadone Metabolite Controls are substantially equivalent to the previously cleared MGC DAU Control Sets: Primary, Clinical, Select (K040758, Microgenics Corporation).
DESCRIPTION OF DEVICE
The DRI Methadone Metabolite (100/300) Assay utilizes liquid ready-to-use reagents. The Antibody/Substrate Reagent (R1) contains mouse monoclonal anti-EDDP antibody, glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as a preservative. The Conjugate Reagent (R2) contains EDDP-derivative labeled with glucose-6-phosphate Enzyme dehydrogenase (G6PDH) in Tris buffer with sodium azide as a preservative.
The assay uses specific antibodies that can detect EDDP in human urine without cross-reactivity to the The assay is based on competition between drug labeled with glucose-6parent drug methadone. phosphate dehydrogenase (G6PDH) and free drug from the sample for a fixed number of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of free drug from the sample, the specific antibody binds the drug labeled with G6PDH and causes a decrease in enzyme activity. This phenomenon creates a direct relationship between drug concentration in urine and enzyme activity. This enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.
{2}------------------------------------------------
2015
:
| Comparison | DRI Methadone Metabolite (100/300)Assay | Predicate Device - CEDIA DAU EDDPAssay |
|---|---|---|
| Intended Use | Qualitative and semi-quantitativedetermination of the presence ofMethadone Metabolite (2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP)in human urine at cutoffs of 100 and 300ng/mL. | Qualitative and semi-quantitative assay ofEDDP (2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP) in humanurine. |
| Test Principle | Homogeneous Enzyme Immunoassaybased on competition between a druglabeled with glucose-6-phosphatedehydrogenase (G6PDH) and free drugfrom the urine sample for a fixed amountof specific antibody binding sites, | Homogeneous Enzyme Immunoassaybased on competition between a druglabeled with ß-galactosidase, and freedrug from the urine sample for a fixedamount of specific antibody binding sites. |
| Direct relationship between drugconcentration in urine and enzymeactivity. | Direct relationship between drugconcentration in urine and enzymeactivity. | |
| Enzyme activity is determinedspectrophotometrically at 340 nm bymeasuring its ability to convertnicotinamide adenine dinucleotide (NAD)to NADH. | Enzyme activity is determinedspectrophotometrically at 570 nm bymeasuring its ability to convert CPRG toCPR. | |
| Cutoff | 100 and 300 ng/mL | 100 ng/mL |
| Matrix | Human urine | Human urine |
| Reagents | Liquid Ready-to-UseTwo reagent assay (R1 and R2) | Lyophilized (reconstitution required)Two reagent assay (R1 and R2) |
| Calibrators | Liquid ready-to-use(0, 100, 300, 500, 1000 ng/mL) | Liquid ready-to-use(0, 100, 500, 2000 ng/mL) |
| Controls | Liquid ready-to-use(± 25% from cutoffs) | Liquid ready-to-use(± 25% from cutoffs) |
:
:
:
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
10.000
{3}------------------------------------------------
SUMMARY OF CLINICAL TESTING
Sensitivity
Sensitivity measured semi-quantitatively as limit of quantitation (LOQ) was 14 ng/mL.
Precision
A precision study was performed using the CLSI guideline EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition
For qualitative mode, acceptance criteria required that all samples spiked at levels below the cutoff read as negative and all samples spiked at levels above the cutoff read as positive.
For semi-quantitative mode, target within-run precision was less than or equal to 8% CV and target total-run precision was less than or equal to 10% CV.
The within-run and total-run precision at all levels of the 300 cutoff and the 300 cutoff met target specifications in both qualitative and semi-quantitative modes.
Cutoff Characterization - Spike Recovery
In qualitative mode, the control levels were recovered accurately, with the negative controls recovering less than the cutoff calibrators and the positive controls recovering greater than the cutoff calibrators. No overlap was observed between cutoff levels and their ± 25% levels, with 95% statistical confidence. The precision of the 21 replicates was < 1% CV.
In semi-quantitative mode, recovery was less than 15% error of the nominal values and the precision of the 21 replicates was below 3% CV.
Linearity - Dilution Recovery
A high sample containing around 1000 ng/mL methadone metabolite was serially diluted in 10% increments with analyte-free urine. Recovery was within 10% of expected values, with a correlation coefficient of r = 0.9990.
Interferences
Results demonstrate that no significant interference was observed from endogenous and exogenous unne substances at the tested concentrations and the pH range of 4 to 11.
Specificity
Cross-reactivity to metabolites and structurally related compounds was tested on spiked samples in both the qualitative and semi-quantitative modes. All compounds tested negative, with rate below the 100 ng/mL cutoff rate and with dose recovered less than 100 ng/mL, indicating that the assay does not cross react with compounds at the tested concentrations.
Method Comparison
A total of 100 unaltered patient urine samples containing various concentrations of methadone metabolite were tested in both qualitative and semi-quantitative modes. Approximately 10% of the samples included in the study had concentrations between cutoff and 50% above the cutoff, and approximately 10% of the samples had concentrations between cutoff and 50% below the cutoff.
Samples were tested by:
- . DRI Methadone Metabolite Assay (on-test method)
- o CEDIA DAU EDDP Assay (predicate device)
- . GC/MS (reference analytical method)
{4}------------------------------------------------
Qualitative Results
At the 100 ng/mL cutoff, the overall concordance between the DRI Assay and the predicate CEDIA Assay was 95%. The overall concordance between DRI and GC/MS was also 95%.
At the 300 cutoff, the overall concordance between the DRI and GC/MS was 99%.
Semi-quantitative Results
At the 100 cutoff, the overall concordance between the DRI Assay and the predicate CEDIA Assay was 99%. The overall concordance to GC/MS was also 99%.
At the 300 cutoff, the overall concordance between the DRI Assay and GC/MS was 100%.
CONCLUSION
As summarized, the DRI Methadone Metabolite (100/300) Assay is substantially equivalent to the CEDIA DAU EDDP Assay. Substantial equivalence has been demonstrated through performance testing to verify that the device functions as intended and that design specifications have been satisfied.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Thermo Fisher Scientific Microgenics Corporation c/o Mr. Jack Rogers Manager of Regulatory Affairs 46360 Fremont Blvd. Fremont, CA 94538
JAN 1 4 2009
K081378 Re:
Trade Name: DRI® Methadone Metabolite (100/300) Assay, DRI® Methadone Metabolite Urine Calibrators, DRI® Methadone Metabolite Urine Control Regulation Number: 21 CFR §862.3620 Regulation Name: Methadone Test System Regulatory Class: Class II Product Code: DJR, DKB, DIF Dated: January 6, 2009 Received: January 7, 2009
Dear Mr. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
Page - 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Coy C. He
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K081378
Device Name: DRI Methadone Metabolite (100/300) Assay
Indications for Use:
The DRI Methadone Metabolite (100/300) Assay is intended for the qualitative and semi-quantitative determination of the presence of Methadone Metabolite (2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP) in human urine at cutoffs of 100 and 300 ng/mL.
The DRI Methadone Metabolite Calibrators are intended for use in calibration of the DRI Methadone Metabolite (100/300) Assay.
The DRI Methadone Metabolite Controls are intended for use in the DRI Methadone Metabolite (100/300) Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography / Mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional iudoment should be applied to any drug of abuse test result. particularly when preliminary results are used. Tests for methadone metabolite cannot distinguish between abused drugs and certain prescribed medications. Certain foods or medications may interfere with tests for methadone metabolite and cause false positive results.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division/Sign-Off
Office of in Vitro Diagnostic Device Evaluation and Safety 51 Offic
Page 1 of
§ 862.3620 Methadone test system.
(a)
Identification. A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.(b)
Classification. Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).