(89 days)
No
The summary describes a reprocessed SpO2 sensor and its performance testing, with no mention of AI or ML.
No
The device is a sensor used for monitoring physiological parameters (oxygen saturation and pulse rate), not for treatment or therapy.
No
Explanation: The device monitors physiological parameters (arterial oxygen saturation and pulse rate) but does not provide a diagnosis or aid in the diagnosis of a disease or condition. It is a monitoring device.
No
The device description explicitly states it is a "sensor" and a "disposable device," indicating a physical hardware component. The performance studies also focus on testing of the "Reprocessed Pulse Oximeter Sensors," further confirming it is not software-only.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is "continuous noninvasive arterial oxygen saturation and pulse rate monitoring." This describes a device that monitors physiological parameters directly from the patient's body, not a device that tests samples taken from the body (like blood, urine, or tissue).
- Device Description: The description confirms it's a "continuous noninvasive arterial oxygen saturation and pulse rate monitoring" device. It's a sensor applied to the skin.
- Lack of IVD Characteristics: The description does not mention any of the typical characteristics of an IVD, such as:
- Analyzing biological samples (blood, urine, etc.)
- Detecting or measuring substances in biological samples
- Providing information for diagnosis, monitoring, or screening based on sample analysis.
Therefore, this device falls under the category of a non-invasive physiological monitoring device, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
Reprocessed Pulse Oximeter Sensors are indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.
Product codes
NLF
Device Description
The devices are reprocessed Low Noise Cabled Sensors (LNCS)® Series - Adult, Pediatric, and Infant SpO2 adhesive sensors. The sensors are disposable devices used for continuous noninvasive arterial oxygen saturation and pulse rate monitoring.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Adult, Pediatric, and Infant
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Pulse Oximeter Sensors. This included the following tests: Biocompatibility, Validation of reprocessing, Function test(s), Packaging Validation. Performance testing demonstrates that Reprocessed Pulse Oximeter Sensors perform as originally intended.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
JUL 2 9 2008
1081238
SECTION 5: 510(k) SUMMARY
.
| Submitter: | Ascent Healthcare Solutions
10232 South 51st Street
Phoenix, Arizona 85044 | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Contact: | Katie Bray
Regulatory Affairs Manager
(480) 763-5300 (o)
(480) 763-6089 (f)
kbray@ascenths.com | |
| Date of preparation: | April 29, 2008 | |
| Name of device: | Trade/Proprietary Name: Reprocessed Pulse Oximeter Sensors | |
| | Classification Name: Oximeter, Reprocessed | |
| Predicate Device
K060143
K041815 | 510(k) Title
LNCS and SPO2.COM Sensors
LNCS Oximetry Sensors | Manufacturer
Masimo Corporation
Masimo Corporation |
| Device description: | The devices are reprocessed Low Noise Cabled Sensors
(LNCS)® Series - Adult, Pediatric, and Infant SpO2 adhesive
sensors. The sensors are disposable devices used for
continuous noninvasive arterial oxygen saturation and pulse rate
monitoring. | |
| Indications for Use: | Reprocessed Pulse Oximeter Sensors are indicated for use in
continuous noninvasive arterial oxygen saturation and pulse rate
monitoring. | |
| Technological
characteristics: | The design, materials, and intended use of Reprocessed Pulse
Oximeter Sensors are identical to the predicate devices. The
mechanism of action of Reprocessed Pulse Oximeter Sensors is
identical to the predicate devices in that the same standard
mechanical design and size and equivalent materials are
utilized. There are no changes to the claims, intended use,
clinical applications, patient population, performance
specifications, or method of operation. In addition, Ascent
Healthcare Solutions' reprocessing of Pulse Oximeter Sensors
includes removal of adherent visible soil and decontamination.
Each individual Pulse Oximeter Sensor is tested for appropriate
function of its components prior to packaging and labeling
operations. | |
| Performance data: | Bench and laboratory testing was conducted to demonstrate
performance (safety and effectiveness) of Reprocessed Pulse
Oximeter Sensors. This included the following tests: | |
1
- Biocompatibility
- . Validation of reprocessing
- . Function test(s)
- Packaging Validation .
Performance testing demonstrates that Reprocessed Pulse Oximeter Sensors perform as originally intended.
Conclusion:
Ascent Healthcare Solutions concludes that the modified devices (Reprocessed Pulse Oximeter Sensors) are safe, effective, and substantially equivalent to the predicate devices as described herein.
2
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Public Health Service
JUL 29 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Katie Bray Regulatory Affairs Manager Ascent Healthcare Solutions 10232 South 51st Street Phoenix, Arizona 85044
Re: K081238
Trade/Device Name: Reprocessed Masimo Pulse Oximeter Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: NLF Dated: April 29, 2008 Received: May 1, 2008
Dear Ms. Bray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 - Ms. Bray
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
K. Kommelts-Lending for II
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
SECTION 4: INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Reprocessed Masimo Pulse Oximeter Sensors
Indications For Use:
This sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
14081238 510(k) Number: ___
Ascent Healthcare Solutions Reprocessed Pulse Oximeter Sensors Traditional 510(k)