K Number
K080768
Date Cleared
2008-04-11

(24 days)

Product Code
Regulation Number
872.3275
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

UltraTemp Firm, Fast and Regular is a non-eugenol temporary cement indicated for interim cementation of inlays, crowns and bridges. It is methacrylate based and does not negatively affect resin bonding. Being water soluble until set, it cleans up easily.

Device Description

UltraTemp Firm, Regular and Fast are non-eugenol temporary cements. They are methacrylate based and do not negatively affect resin bonding. Being water soluble until set, they clean up easily. UltraTemp Firm is available in regular and fast set.

AI/ML Overview

This 510(k) summary (K080768) describes the device "UltraTemp Firm, Fast and Regular," which is a temporary dental cement. The submission aims to demonstrate substantial equivalence to a predicate device, UltraTemp (K994261). The testing described in this document is primarily focused on demonstrating the safety and effectiveness of the new device through product specifications and biocompatibility testing, rather than a clinical study comparing its performance against specific acceptance criteria in a real-world scenario.

Here's an breakdown of the requested information based on the provided text, with many points marked as "Not Applicable" or "Not Provided" due to the nature of this particular 510(k) submission for a dental cement:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly define acceptance criteria in terms of quantitative performance metrics for the cement (e.g., bond strength, setting time ranges) against the predicate device in a comparative table. Instead, the comparison is at a higher, qualitative level, focusing on intended use, material type, characteristics, human factors, and biocompatibility/safety.

PropertyAcceptance Criteria (Implied)Reported Device Performance
Intended UseSame as predicate (Temporary cement)Same: "Temporary cement" and "For temporary application of provisional crowns, bridges, inlays, and onlays."
CharacteristicsSame as predicate (Temporary Luting/Filling Material)Same: "Temporary Luting/Filling Material"
Biocompatibility/SafetyPass biocompatibility tests. Safe when used as directed.Same: "Cytotoxicity, Sensitization, irritation and Genotoxicity testing passed. Literature and testing to demonstrate product is safe when used as directed." The new device states, "The results of biocompatibility testing demonstrate that UltraTemp Firm, Fast and Regular are safe and effective when used according to the Instructions for Use."
Material (Type)Acceptable material for temporary dental cement (Methacrylate)Different from Predicate: Predicate is Polycarboxylate; New device is Methacrylate-based. This is noted as a technological characteristic but implied to be acceptable.
Human Factors (Delivery)Acceptable delivery systemDifferent from Predicate: Predicate is "Dual Spense Delivery System"; New device is "Double Barrel Delivery System." This is noted but implied to be acceptable.
Water Solubility (before set)Cleans up easily before settingClaimed: "Being water soluble until set, they clean up easily."
Resin Bonding EffectDoes not negatively affect resin bondingClaimed: "does not negatively affect resin bonding."

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: Not explicitly provided for any specific performance testing beyond general statements about "each lot of product." The biocompatibility testing implies biological samples were used, but the quantity of samples or subjects is not mentioned.
  • Data Provenance: The testing is internal to the manufacturer, Ultradent Products, Inc. "Each lot of product must pass internal test specifications prior to release." The document does not specify country of origin for data; it's likely internal laboratory data. It is a prospective testing of manufacturing lots and biocompatibility, not analysis of pre-existing patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. The ground truth for this device's performance would be derived from laboratory tests and established biocompatibility standards, not from expert consensus on clinical cases.

4. Adjudication Method for the Test Set

Not applicable. This type of submission for a dental cement does not involve adjudication of clinical cases or expert readings.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a dental cement, not an AI-powered diagnostic or assistive tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a dental cement, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device is based on:

  • Laboratory-measured physical and chemical properties: Ensuring the cement meets internal specifications for its intended function (though specific criteria and results are not detailed).
  • Biocompatibility standards: Confirmation that the material passes established tests for cytotoxicity, sensitization, irritation, and genotoxicity, indicating it is safe for biological contact.
  • Predicate Device Comparison: The ultimate "ground truth" for regulatory approval here is demonstrating that the new device is "substantially equivalent" to a legally marketed predicate device based on its intended use, technological characteristics, and safety/effectiveness.

8. The Sample Size for the Training Set

Not applicable. This device is a physical product (dental cement), not a machine learning model, so there is no training set in the AI context.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for a machine learning model, this question does not apply.

{0}------------------------------------------------

0g0768

SPECIAL 510(K) SUMMARY

APR 1 1 2008

UltraTemp Firm, Fast and Regular

This summary of the Traditional 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807 for UltraTemp Firm, Fast and Regular.

Applicant's Name and Address

Ultradent Products, Inc. 505 West 10200 South South Jordan, UT 84095

Contact Person:Diane Rogers
Title:Regulatory Affairs Product Specialist
Telephone:800-552-5512 x4491, 801-553-4491
FAX:801-553-4609
Date Summary Prepared:January 23, 2008

Name of the Device

Trade Name:
Common Name:
Device Classification:
Classification Product Code:

UltraTemp Firm, Fast and Regular Dental Cement ll EMA

Legally Marketed Predicate Devices to Which Equivalence is Claimed

The predicate device is UltraTemp (K994261) This device is manufactured and distributed by Ultradent Products, Inc., 505 West, 10200 South, South Jordan, Utah 84095.

Product Description: UltraTemp Firm, Regular and Fast are non-eugenol temporary cements. They are methacrylate based and do not negatively affect resin bonding. Being water soluble until set, they clean up easily. UltraTemp Firm is available in regular and fast set.

Indications for Use: For temporary application of provisional crowns, bridges, inlays, and onlays.

{1}------------------------------------------------

Table 1: Product Comparison

PropertyPredicate:UltraTemp (K994261)UltraTemp FirmFast and Regular
Intended UseTemporary cementSame
Type of materialPolycarboxylateMethacrylate
CharacteristicsTemporary Luting/FillingMaterialSame
Human factorsDual Spense Delivery SystemDouble BarrelDelivery System
Biocompatibility/SafetyCytotoxicity, Sensitization,irritation and Genotoxicitytesting passed. Literature andtesting to demonstrateproduct is safe when used asdirectedSame

Technological Characteristics

UltraTemp Firm, Fast and Regular are non-eugenol temporary cements. They are methacrylate based and do not negatively affect resin bonding. Being water soluble until set, they clean up easily. UltraTemp Firm is available in regular and fast set.

Brief Description of Testing Performed

Each lot of product must pass internal test specifications prior to release. The results of biocompatibility testing demonstrate that UltraTemp Firm, Fast and Regular are safe and effective when used according to the Instructions for Use.

Conclusion and Substantial Equivalence

In conclusion, UltraTemp Firm, Fast and Regular, are to be manufactured and marketed by Ultradent Products, Inc., 505 West 10200 South, South Jordan, UT 84095, is substantially equivalent to UltraTemp (K994261), also manufactured by Ultradent Products, Inc. The two products are composed of similar materials, have the same intended use and technological characteristics, and both are safe and effective when used for the indications described.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

APR 1 1 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Diane Rogers Regulatory Affairs Product Specialist Ultradent Products, Incorporated 505 West 10200 South South Jordan, Utah 84095

Re: K080768

Trade/Device Name: UltraTemp Firm, Fast and Regular Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: March 18, 2008 Received: March 18, 2008

Dear Ms. Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general ' controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Rogers

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Clues

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Statement of Indications for Use

510(k) Number (if known):

Device Name:

UltraTemp Firm, Fast and Regular is a non-eugenol temporary Indications for Use: cement indicated for interim cementation of inlays, crowns and bridges. It is methacrylate based and does not negatively affect resin bonding. Being water soluble until set, it cleans up easily.

Prescription Use _X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page_1 of

(Posted November 13, 2003)

Susan Runne

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

§ 872.3275 Dental cement.

(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.