(19 days)
The HSW Arthroscope and accessories is a tubular endoscopic device with accessory devices which attach to the Arthroscope and is intended to examine and / or perform surgery on the interior of a joint. Arthroscopic minimal invasive procedures are performed in the hip, knee, shoulder, wrist (carpal tunnel syndrome), temporal-mandibular joint, ankle, elbow and feet (plantar fascia release).
The Henke Sass Wolf of America's Arthroscope is identical in terms of materials and modes of construction, optical performance and safety to Henke Sass Wolf of America's current Arthroscope and the Smith & Nephew Arthroscope. The only difference from the current Henke Sass Wolf of America's Arthroscope is the new intended use for hip surgery. This indication for use is already in place for the Smith & Nephew Arthroscope.
The arthroscope is a long tube containing a series of lenses. At the distal end, an objective lens captures the image of the object. A series of rod lenses relay the image along the length of the tube. At the proximal end, in the case of a direct view model, an ocular lens forms an image for viewing directly with the human eye. In the case of a videoarthroscope, a proximal coupling lens relays the image to a CCD (charged couple device used as an electronic video sensor chip).
Arthroscopes generally come in two diameters, 2.7mm and 4.0mm, although other sizes are sometimes offered. Larger size arthroscopes are used for general viewing. This is due to the preferable larger and brighter image achievable with larger diameter optical components. The small sizes are used where access to the surgical site is restricted.
Arthroscopes generally come in several "directions of view", 0°, 30°, 70°, 90° and 110°, though others are sometimes offered. The center axis of the field of view of the 0° scope is along the normal axis of the Arthroscope. The other "direction of view" instruments are referenced from the scope normal axis. The various "directions of view" permit or facilitate viewing of different parts of the relevant anatomy.
All arthroscopes also contain glass fibers for illumination of the surgical site.
Henke Sass Wolf Arthroscopes are provided in Autoclavable and Non-Autoclavable models.
Here's an analysis of the provided text regarding acceptance criteria and supporting studies for the Henke Sass Wolf Arthroscope.
Important Note: The provided document is a 510(k) submission for a medical device (Arthroscope). In this context, "acceptance criteria" and "studies" are typically focused on demonstrating substantial equivalence to existing legally marketed devices, rather than establishing completely new performance benchmarks through extensive clinical trials. The criteria and studies revolve around proving that the new device is as safe and effective as its predicates. Therefore, some of the requested information (like specific performance metrics measured in a clinical study, effect sizes of human readers with AI, etc.) is generally not present in this type of submission.
Description of Acceptance Criteria and Supporting Study for Henke Sass Wolf Arthroscope
The Henke Sass Wolf Arthroscope's acceptance criteria and supporting "study" are primarily based on demonstrating substantial equivalence to predicate devices, namely the Henke Sass Wolf of America Laparoscope (K962075) and the Smith & Nephew, Inc. Arthroscope (K971253). The submission argues that the device's design, materials, manufacturing, optical performance, and safety are identical or highly similar, with the only significant change being an expanded intended use to include hip surgery, which is already an indicated use for one of the predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria (Implicit or Explicit in Submission) | Reported Device Performance (Summary from Submission) |
|---|---|---|
| Intended Use | Must align with or expand upon predicate devices in a substantially equivalent manner. | Expanded to include hip surgery; aligns with Smith & Nephew Arthroscope (K971253) for this indication. |
| Design | Must be identical or highly similar to predicate devices. | "Identical in terms of materials and modes of construction, optical performance and safety to the existing line of HSW Arthroscopes." |
| Materials | Must be medical grade and identical to predicate devices. | "Identical to the Henke Sass Wolf of America Arthroscope - K962075 and the Smith & Nephew, Inc., Arthroscope - K971253. All materials used are medical grade." (Includes specific details of stainless steel, titanium, P.E.E.K., glass, sapphire, solder, adhesive.) |
| Physical Properties | Must be identical to predicate devices. | "Identical to the Henke Sass Wolf of America Arthroscope - K962075 and the Smith & Nephew, Arthroscope - K971253." |
| Optical Performance | Must be identical to predicate devices. | "Identical... in terms of optical performance." (Specifications for diameter, direction of view, field of view, eyepiece magnification, working length are provided, consistent with arthroscope standards.) |
| Safety | Must exhibit equivalent safety profile to predicate devices. | "Identical... in terms of... safety." (Biological evaluation, electrical safety, sterilization validation cited as evidence.) |
| Sterilization | Must be capable of effective cleaning and sterilization according to recognized standards. | Non-sterile, reusable. Validated for cleanability and repeat sterilization (Sterrad 100S, Steam Sterilization). Sterilization validation conducted for predicate devices deemed applicable. Achieves SAL of 10-6. |
| Manufacturing | Must follow similar methods of construction as predicate devices. | Mentions "modes of construction" are identical. Notes that Smith & Nephew Arthroscopes are manufactured by Henke Sass Wolf, GmbH, implying consistent manufacturing processes. |
2. Sample Size Used for the Test Set and Data Provenance
- No specific "test set" in the conventional sense of a prospective clinical study with patient data is mentioned for device performance. The primary "testing" for substantial equivalence relies on comparing the device's specifications and characteristics to those of predicate devices already on the market.
- Data Provenance: The data provenance is based on the existing design, materials, and validated manufacturing processes of the manufacturer's own existing arthroscope line and the Smith & Nephew arthroscope (which is also manufactured by Henke Sass Wolf, GmbH). This is retrospective in the sense that it refers to established device characteristics and prior validation reports rather than a new prospective data collection specifically for this submission's comparison.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. As this is a 510(k) submission for substantial equivalence based on design and material comparisons, rather than a clinical study evaluating diagnostic accuracy or treatment outcomes, there is no mention of external experts or a ground truth panel in the context of a test set. The "ground truth" implicitly lies with the established performance and safety of the predicate devices.
4. Adjudication Method for the Test Set
- Not Applicable. No test set requiring expert adjudication is described in the submission.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. An MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices to assess human reader performance with and without AI assistance, which is outside the scope of this arthroscope submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not Applicable. This device is an arthroscope, a physical medical instrument for direct visualization and surgical procedures, not an algorithm or software-controlled device. "Standalone" performance as described for AI algorithms is not relevant here. The submission explicitly states, "Device incorporates software? No."
7. The Type of Ground Truth Used
- The "ground truth" for this submission is the established safety and effectiveness of the legally marketed predicate devices. The new device's characteristics (design, materials, optical performance, manufacturing methods) are compared against these known predicates to demonstrate that it is equally safe and effective for its expanded intended use.
- For the sterilization validation, the ground truth is established by AAMI/ISO Standards (e.g., SAL of 10-6), with a "worst case scenario" device (a 0° 10mm Operative Laparoscope) being used to validate the Sterrad 100S system.
8. The Sample Size for the Training Set
- Not Applicable. As it's a physical device and not an AI algorithm, there is no "training set" in the context of machine learning. The design and validation are based on engineering principles and regulatory standards rather than data-driven machine learning models.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. (See point 8).
{0}------------------------------------------------
MAR 1 8 2008
<080560
510(K) Submission
Applicant's Name & Address: Henke Sass Wolf of America 44 Southbridge Rd.
John Kingston (508) 671-9300
Dudley, MA 01571
- Contact Person and Telephone Number: 2. Lynette Howard (908) 788-4580
-
- Representative / Consultant: Lynette Howard 106 East 5th Avenue Mount Dora, FL 32757 Phone: (908) 788-4580 Fax: (352) 383.8338 e-mail: Lylehoward@aol.com
-
- Device Trade or Proprietary Name: Henke Sass Wolf Arthroscope
-
- Device Common Name / Classification Name: Arthroscope
-
- Establishment Registration Number: 1222997
-
- Address of Manufacturing Site: Henke-Sass, Wolf GmbH Postfach 2459 78507 Tuttlingen Germany Registration #: 8010418
-
- Classification of Device under Section 513: Class II, HRZ Regulation #: 888.1100
{1}------------------------------------------------
9. Labels and Instructions for Use:
See Section 8 & 9
10. Drawings of device:
See Section 10
11. Marketed Device(s) to which the claim of substantial equivalence is claimed:
The Henke Sass Wolf of America Laparoscopes are substantially equivalent to the Henke Sass Wolf of America Laparoscope - K962075 and the Smith & Nephew, Inc. Arthroscope - K971253 in design, materials, methods of construction and intended use.
See Section 11
12. The intended use of the devices to which we claim substantial equivalence:
The Henke Sass Wolf of America Arthroscopes (K962075) are intended to be used by surgeons in diagnostic and therapeutic procedures. Arthroscopic minimal invasive procedures are performed in the knee, shoulder, wrist (carpal tunnel syndrome), temporal-mandibular joint, ankle, elbow and feet (plantar fascia release). The benefits realized through the use of arthroscopic procedures include: the minimal trauma experienced from the small punctures necessary to gain access for the Arthroscope and compared to a complete open procedure, the ability to gain visual access to areas of the anatomy which would be extremely difficult and require substantially more complicated procedures to approach by other methods.
Diagnostic procedures are used when it is difficult to make an exact diagnosis without direct visual observation. Therapeutic procedure provide a means for repair of cartilage. detachment of lesions and removal of debris. Procedural names include subacromial decompression, notchplasty, meniscectomy, synovectomy, small ioint arthroscopy, examination of the menisco-synovial junction, access and release of carpal tunnel ligament and plantar fascia release.
The Smith & Nephew, Inc. Rigid Arthroscope & Accessories - K971253, are indicated for use in diagnostic and operative arthroscopic procedures to provide illumination and visualization of the shoulder, knee, elbow, ankle, wrist and jaw, also to provide illumination and visualization during arthroscopic diagnostic procedures and removal of loose bodies and soft tissue within the hip joint as size/length appropriate.
{2}------------------------------------------------
The Smith & Nephew, Inc. Endoscopy Division Sterile Disposable Endoscopic Surgery Blades - K971253 are indicated for use during arthroscopic resection of soft and osseous tissue in various large and small articular cavities including the hip joint. Blades which are sizeappropriate are indicated for use in the following joints: shoulder, knee, elbow, ankle, wrist, jaw and the hip joint.
13. Description of the new device:
The Henke Sass Wolf of America's Arthroscope is identical in terms of materials and modes of construction, optical performance and safety to Henke Sass Wolf of America's current Arthroscope and the Smith & Nephew Arthroscope. The only difference from the current Henke Sass Wolf of America's Arthroscope is the new intended use for hip surgery. This indication for use is already in place for the Smith & Nephew Arthroscope.
The arthroscope is a long tube containing a series of lenses. At the distal end, an objective lens captures the image of the object. A series of rod lenses relay the image along the length of the tube. At the proximal end, in the case of a direct view model, an ocular lens forms an image for viewing directly with the human eye. In the case of a videoarthroscope, a proximal coupling lens relays the image to a CCD (charged couple device used as an electronic video sensor chip).
Arthroscopes generally come in two diameters, 2.7mm and 4.0mm, although other sizes are sometimes offered. Larger size arthroscopes are used for general viewing. This is due to the preferable larger and brighter image achievable with larger diameter optical components. The small sizes are used where access to the surgical site is restricted.
Arthroscopes generally come in several "directions of view", 0°, 30°, 70°, 90° and 110°, though others are sometimes offered. The center axis of the field of view of the 0° scope is along the normal axis of the Arthroscope. The other "direction of view" instruments are referenced from the scope normal axis. The various "directions of view" permit or facilitate viewing of different parts of the relevant anatomy.
All arthroscopes also contain glass fibers for illumination of the surgical site.
Henke Sass Wolf Arthroscopes are provided in Autoclavable and Non-Autoclavable models.
| Characteristics | Non-Autoclavable | Autoclavable |
|---|---|---|
| Material of externaloptical elements | Conventional optical qualityglass | Sapphire |
| Mounting Methodsfor external opticalelements | Epoxied or mechanicallyrestrained | Soldered |
| Bonding methodsmechanicalcomponents | Epoxied | Welded |
The characteristics which distinguish the two models are:
{3}------------------------------------------------
The imaging systems incorporated consist of:
- An objective lens which performs the image acquisition. .
- A series of rod lenses relay the image from the distal end to the proximal end of . the Arthroscope.
- A final optical element at the proximal end of the Arthroscope. In the case of a . direct view model, an ocular lens forms and image for viewing directly with the human eye. In the case of a videoarthroscope, a proximal coupling lens relays the image to a CCD (charged couple device used as an electronic video sensor chip).
Illumination of the surgical site is provided by fiber optics internal to the Arthroscope. A mechanical coupling located at the proximal end of the device provides a method of connecting flexible light guides to external lightsources.
| ArthroscopeDiameter(mm) | Direction ofView(degrees) | Field ofView(degrees) | *EyepieceMagnification | WorkingLength(mm) |
|---|---|---|---|---|
| 2.3-2.9 | 0, 30, 70 | ~ 85 | 24 | ~ 195 |
| 4.0 | 0, 30, 45,70,110 | ~ 105 | 21 | ~185 |
| ShortArthroscopeDiameter(mm) | Direction ofView(degrees) | Field ofView(degrees) | *EyepieceMagnification | WorkingLength(mm) |
| 1.7-1.9 | 0, 30 | ~ 85 | 21 | ~60 |
| 2.3-2.9 | 0, 30, 70 | ~ 85 | 24 | ~70/140 |
Specifications for the proposed device:
- Eyepiece magnification pertains to direct view arthroscopes only.
Notes:
- Locking Device can be either Storz Type, Stryker Speedlock, J-lock or . Linvatec Quick Latch locking mechanism
- Adapters for connection to all conventional light-guides are supplied as . standard.
| Life supporting or Life Sustaining? | No |
|---|---|
| Implanted (Short term or Long term)? | No |
| Device incorporates software? | No |
{4}------------------------------------------------
Summary of:
Device Design
The Henke Sass Wolf Arthroscope is identical in terms of materials and modes of construction, optical performance and safety to the existing line of HSW Arthroscopes, differing only in an additional intended use.
Note: The Smith & Nephew Arthroscopes are manufactured by Henke Sass Wolf, GmbH facility.
Henke Sass Wolf manufactures autoclavable and non-autoclavable models. The autoclavable models are identified by the word "Autoclavable" on the instrument. Both models, autoclavable and non-autoclavable, are sold as non-sterile, reusable instruments.
| Component | Material # | GermanStandard | USAUNS | ASTM |
|---|---|---|---|---|
| Shaft | StainlessSteel | QQ-S-763QQ-W-423QQ-S-764 | ||
| Body | StainlessSteel | QQ-S-763QQ-W-423QQ-S-764 | ||
| Sidearm | Titanium | B367 | ||
| StainlessSteel | QQ-S-763QQ-W-423QQ-S-764 | |||
| Eyepiece | TitaniumP.E.E.K. | MIL-P-46183 –type 1 | B367 | |
| Lenses | Glass | BK-7 (SchottGlass No.517642) | ||
| Sapphire | OpticalGrade PureSapphire, |
Materials
Additional Materials:
The solder used to seal the objective lens in the autoclavable modes of the Solder: Arthroscope is the lead free alloy "CASTIN".
The adhesive used to seal the objective lens in the non-autoclavable Adhesive:
{5}------------------------------------------------
models of the Arthroscope is Loctite product # 3321.
The external exposure of either of the above materials is limited to an area of 0.95 millimeters2. This is equivalent to a circle with a diameter of 1.1 millimeters.
The materials of construction are identical to the Henke Sass Wolf of America Arthroscope - K962075 and the Smith & Nephew, Inc., Arthroscope -- K971253. All materials used are medical grade. These devices are not implantable device.
Physical properties
The physical properties are identical to the Henke Sass Wolf of America Arthroscope -K962075 and the Smith & Nephew, Arthroscope -- K971253.
Is the Device Sterile?
No
The device is intended to be sterilized prior to use. The Arthroscopes Endoscopes are designed to facilitate cleaning and disinfection, and are validated for clean ability and repeat sterilization.
Instructions for cleaning and sterilization are included in the instructions for use. Sterilization and disinfection validation includes Sterrad 100S and is performed to assure an SAL of 10-6 per AAMI / ISO Standards for non-autoclavable scopes.
Steam Sterilization as per ANSI/AAMI standards is used for the autoclavable scopes.
All sterilization methods are defined in the instructions for use.
| Is the Device for single use? | No |
|---|---|
| Is the Device for home use or prescription use? | Yes,Prescription Use |
| Does the Device contain drug or biologicalproduct as a component? | No |
| Is the Device a kit? | No |
| Is this a software-controlled device? | No |
14. The intended use of the new device:
The HSW Arthroscope and accessories is a tubular endoscopic device with accessory devices which attach to the Arthroscope and is intended to examine and / or perform surgery on the interior of a joint. Arthroscopic minimal invasive procedures are performed in the hip, knee, shoulder, wrist (carpal tunnel syndrome), temporal-mandibular joint, ankle, elbow and feet (plantar fascia release).
{6}------------------------------------------------
Diagnostic procedures are used when it is difficult to make an exact diagnosis without direct visual observation. Therapeutic procedures provide a means for repair of cartilage, detachment of lesions and removal of debris. Procedural names include subacromial decompression, notchplasty, meniscectomy, synovectomy, small & large joint arthroscopy, examination of the menisco-synovial junction, access and release of carpal tunnel ligament and plantar fascia release.
15. Comparison of the new device to the marketed device(s):
Refer to Section 11
-
- Summary of Risk Analysis:
See Section 14
- Summary of Risk Analysis:
-
- Summary of safety and effectiveness of the new device:
See Section 15
- Summary of safety and effectiveness of the new device:
18. Testing conducted to assure safety and effectiveness includes but is not limited to:
Biological Evaluation of Medical Devices -- ISO 10993-1, ISO 10993-5, ISO 10993-12 Electrical Safety Requirements as per IEC 601-2-18 as a type BF or Type CF applied as part of medical electrical equipment. AAMI / ISO Standards for Sterilization of Medical Devices
For Sterile Device: Devices are intended to be sterilized prior to use as described below:
-
- Name, Address, telephone number, contact person of contract sterilizer:
N/A
- Name, Address, telephone number, contact person of contract sterilizer:
-
- Sterilization Method:
Steris, Sterrad or Autoclaving
- Sterilization Method:
Non-Autoclavable Scopes
Sterilize by one of the following methods:
Steris System 1 Processor
Sterilize the device using the Steris System 1 sterilization process. Follow all recommended directions and warnings from STERIS Corporation in the use of the Steris System 1 Processor.
{7}------------------------------------------------
Sterrad 100s Sterilization System
Sterilize the device using the Sterrad 100s Sterilization System according to ANSI/AAMI/ISO 14937 (2000) Sterilization of Medical Devices. Please follow your Sterrad 100s System Operator's Manual instructions for use when processing these devices.
Autoclavable Scopes
Autoclave Wrapped
Follow standard hospital procedure for:
Pre-Vacuum methods at 270° F - 275° F (132° C - 135° C) for 4 minutes; or
Gravity methods at 270° F - 275° F (132° C - 135° C) for 10 minutes.
In addition to autoclave wrapped methods, you may also use one of the following sterilization methods:
Steris System 1 Processor - Refer to the sterilization parameters under "Non-Autoclavable Scopes".
Prior to all methods of sterilization or disinfection the scope must be cleaned thoroughly. The scopes should be sterilized or disinfected in a container, which secures the instrument in place.
Description of method that is used to validate the sterilization cycle: 3.
Sterilization validation was conducted for predicate devices. There are no significant differences between the proposed device and HSW predicate devices that would compromise the sterilization process and/or results. The instructions for sterilization are identical as those for the Smith & Nephew predicate devices.
In addition. Sterilization Validation was completed for the Henke-Sass Wolf GmbH 0° 10mm Operative Laparoscope (Part Number 83 000 4877) for the Sterrad 100S Sterilization System. This scope was chosen as a "worst case" scenario and is applicable to the proposed device.
See Section 16
4. The sterility assurance level (SAL):
Validation conducted to SAL of 10-6
5. Description of packaging to maintain the devices sterility:
.. .
{8}------------------------------------------------
N/A
6. The radiation dose, for radiation sterilized devices:
N/A
·
:
・
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an abstract human form, represented by three curved lines that resemble a person with outstretched arms. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Henke Sass Wolf of America % Ms. Lynette Howard Submission Correspondent 106 East 5th Avenue Mount Dora, Florida 32757
MAR 1 8 2008
Re: K080560
Trade/Device Name: Henke Sass Wolf Arthroscope Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRZ Dated: February 26, 2008 Received: February 29, 2008
Dear Ms. Howard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2 - Ms. Lynette Howard
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Mulkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
Indications for Use
510(k) Number (if known): _ 人 ○ { ロ ▽ レ ○ ▽ レ ○
Device Name:
Indications for Use:
The HSW Arthroscope and accessories is a tubular endoscopic device with accessory devices which attach to the Arthroscope and is intended to examine and / or perform surgery on the interior of a joint. Arthroscopic minimal invasive procedures are performed in the hip, knee, shoulder, wrist (carpal tunnel syndrome), temporal-mandibular joint, ankle, elbow and feet (plantar fascia release).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-OC) Division of General, Restorative, and Neurologica. Devices
10805660 510(k) Number
§ 888.1100 Arthroscope.
(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.